Echotip Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore Endobronchial High Definition Ultrasound Biopsy Needle
Applicant
Cook Ireland, Ltd.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Mar 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
Echotip® Ultra Endobronchial High Definition Ultrasound Needle for use with Olympus EBUS scopes: This device is used to sample targeted submucal lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA). Echotip® Ultra Endobronchial High Definition Ultrasound Needle for use with Pentax EBUS scopes: This device is used to sample targeted submucal lesions within or adjacent to the tracheobronchial tree through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA). Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Olympus EBUS scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract. Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Pentax EBUS scopes: This device is used with an ultrasound endoscope for fine needle biopsy. (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree.
Device Story
Device consists of stainless steel needle cannula, syringe, and handle; used with endobronchial ultrasound (EBUS) endoscopes for tissue sampling. Needle features distal dimpling for ultrasound visualization; bevelled tip; optional cutting surface for biopsy. Preloaded stylet remains during advancement; sheath covers needle when retracted. Operated by trained healthcare professionals in clinical settings. Device inserted through endoscope accessory channel to target submucosal/extramural lesions. Physician uses ultrasound guidance to position needle; performs aspiration or biopsy to retrieve tissue samples. Enables minimally invasive tissue acquisition for diagnostic evaluation; benefits patient by providing definitive diagnosis of lesions within or adjacent to tracheobronchial/gastrointestinal tracts.
Clinical Evidence
Bench testing only. Includes dimensional/visual inspections, stylet removal force, simulated use, aspiration testing, endoscope evaluation, leak testing, and joint strength testing. Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Stainless steel needle cannula with distal dimpling; 22 and 25 Ga sizes. Manual operation via endoscope accessory channel. Sterile, single-use. Sterilization via Ethylene Oxide (EO). No electronic components or software.
Indications for Use
Indicated for patients requiring fine needle aspiration (FNA) or fine needle biopsy (FNB) of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract. Used by trained healthcare professionals via ultrasound endoscope accessory channels.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Echotip® Ultra Endobronchial High Definition Ultrasound Needle (K093195)
Echotip Procore® High Definition Ultrasound Biopsy Needle (K142688)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2016
Cook Ireland Ltd. Lynsey Shine Regulatory Affairs Specialist O' Halloran Road, National Technology Park Limerick, Ireland
Re: K160229 Trade/Device Name: Echotip Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore Endobronchial High Definition Ultrasound Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: FCG Dated: January 28, 2016 Received: February 1, 2016
Dear Lynsey Shine.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160229
### Device Name
Echotip® Ultra Endobronchial High Definition Ultrasound Needle, Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle
### Indications for Use (Describe)
Echotip® Ultra Endobronchial High Definition Ultrasound Needle for use with Olympus EBUS scopes: This device is used to sample targeted submucal lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
Echotip® Ultra Endobronchial High Definition Ultrasound Needle for use with Pentax EBUS scopes: This device is used to sample targeted submucal lesions within or adjacent to the tracheobronchial tree through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Olympus EBUS scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract.
Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Pentax EBUS scopes: This device is used with an ultrasound endoscope for fine needle biopsy. (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(k) Summary
#### I. SUBMITTER
Cook Ireland Ltd. O' Halloran Road National Technology Park Limerick Ireland
Phone: +353 61 334440 Fax: +353 61 239293
Contact Persons: Lynsey Shine, Regulatory Affairs Specialist Jacinta Kilmartin, Regulatory Affairs Manager
Phone: +353 61 334440 Fax: +353 61 239293
Date Prepared: March 16, 2016
#### II. DEVICE
Trade Name of Device: Echotip® Ultra Endobronchial High Definition Ultrasound Needle and Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle.
The model numbers are ECHO-HD-XX-EBUS-X(-C), where XX denotes the needle gauge size, X denotes the type of endobronchial ultrasound endoscope and -C denotes if it is the Procore version of the device.
| Model numbers: | ECHO-HD-22-EBUS-O |
|----------------|---------------------|
| | ECHO-HD-22-EBUS-P |
| | ECHO-HD-25-EBUS-O |
| | ECHO-HD-25-EBUS-P |
| | ECHO-HD-22-EBUS-O-C |
| | ECHO-HD-22-EBUS-P-C |
| | ECHO-HD-25-EBUS-O-C |
| | ECHO-HD-25-EBUS-P-C |
{4}------------------------------------------------
Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit)
Classification Name: Gastroenterology-urology biopsy instrument (21 CFR 876.1075)
Regulatory Class: II
Product Code: FCG
#### III. PREDICATE DEVICE
Echotip® Ultra Endobronchial High Definition Ultrasound Needle, K093195 cleared January 21, 2010.
Echotip Procore® High Definition Ultrasound Biopsy Needle, K142688 cleared December 19, 2014.
None of the predicate devices detailed above have been subject to a design related recall.
#### IV. DEVICE DESCRIPTION
The Echotip Ultra/ Procore Endobronchial High Definition Ultrasound Needle is used in conjunction with an endobronchial ultrasound endoscope and is available with needle gauge sizes of 22 and 25 Ga. The device is composed of a needle assembly and a syringe. An adapter can also be supplied for use with endobronchial ultrasound endoscopes with metal non-Luer hubs. The stainless steel needle cannula has a dimpling pattern on the distal end to allow visualization of the needle tip under endoscopic ultrasound. The needle cannula has a bevelled tip design and comes either with or without an additional cutting surface on the distal end of the needle cannula. The purpose of the needle cannula is for puncturing/ sampling of the target site. The needle is provided with a preloaded stylet which remains in place during advancement of the needle. The sheath covers the needle when the needle is retracted and not in use. The device handle allows for needle and sheath length adjustment. The stylet/ syringe can aid in specimen retrieval. The device is supplied sterile, intended for single use only and is available for prescription use only. Use of this device is restricted to a trained healthcare professional.
{5}------------------------------------------------
#### V. INDICATIONS FOR USE
# Echotipe Ultra Endobronchial High Definition Ultrasound Needle for use with Olympus EBUS scopes:
This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
Echotipe Ultra Endobronchial High Definition Ultrasound Needle for use with Pentax EBUS scopes:
This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
## Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Olympus EBUS scopes:
This device is used with an ultrasound endoscope for fine needle biopsy. (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract.
# Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for use with Pentax EBUS scopes:
This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree.
The predicate Echotip Procore® High Definition Ultrasound Biopsy Needle is indicated for use with an ultrasound endoscope for fine needle biopsy. (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract. The Indications for Use statement for the predicate Echotip Procore® High Definition Ultrasound Biopsy Needle device is not identical to the subject device. However, the differences do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for Fine Needle Aspiration/ Biopsy within or adjacent to the gastrointestinal tract.
{6}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH A PREDICATE DEVICE
The subject device is substantially equivalent to the currently marketed devices, the Echotip® Ultra Endobronchial High Definition Ultrasound Needle, cleared under K093195 on January 21, 2010 and the Echotip Procore® High Definition Ultrasound Biopsy Needle, cleared under K142688 December 19, 2014. The subject and predicate devices operate in the same manner to obtain a specimen using an ultrasound endoscope.
In brief, the subject device is identical to/ within the range of the Echotip® Ultra Endobronchial High Definition Ultrasound Needle, K093195, primary predicate device, and the Echotip Procore High Definition Ultrasound Biopsy Needle, K142688, with respect to the following:
- -Needle gauge size, material, additional cutting surface on the needle cannula,
- । Needle length extension and sheath length extension range of adjustment, method of adjustment for needle and sheath,
- Sheath material and colour, —
- -Stylet wire material, diameter, distal tip options,
- -Compatibility with an endobronchial ultrasound endoscope,
- -Supplied with a syringe,
- Principle of operation, -
- For professional use, -
- -For single use,
- -Sterility (Ethylene oxide, EO),
- Compatibility with endobronchial ultrasound endoscopes with metal non-Luer hubs (with an adapter),
- -Compatibility with ultrasound endoscopes with metal Luer hubs,
- Same endoscope minimum accessory channel size. -
The following technological differences exist between the subject device and the Echotip® Ultra Endobronchial High Definition Ultrasound Needle, K093195, primary predicate device, and the Echotip Procore® High Definition Ultrasound Biopsy Needle, K142688, with respect to the following:
- -Sheath diameter,
- | Detachable sheath adjuster for scopes with a metal Luer hub,
- Compatibility with endobronchial ultrasound endoscopes with metal Luer hubs (no | adapter required).
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
The biocompatibility evaluation for the Echotip Ultra/ Procore Endobronchial High Definition Ultrasound Needle was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" and FDA's biocompatibility guidance, G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (May 1, 1995).
The device specific guidance document was consulted in preparing this premarket submission, Guidance for the Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology. Performance testing such as dimensional and visual inspections, stylet removal force testing, simulated use testing, device aspiration testing, endoscope evaluation and leak testing, and joint strength testing were performed as per Cook's design control system.
#### VIII. CONCLUSIONS
The non-clinical data supports the safety of the subject device and demonstrates that the Echotip Ultra/ Procore Endobronchial High Definition Ultrasound Needle is safe and effective and should perform as intended in the specified use conditions. This non-clinical data supports the substantial equivalence of the Echotip Ultra/ Procore Endobronchial High Definition Ultrasound Needle to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.