K203701 · Ivory Super Holdco Inc. / Zest Anchors, LLC · DZE · Apr 15, 2021 · Dental
Device Facts
Record ID
K203701
Device Name
Locator Overdenture Implant System
Applicant
Ivory Super Holdco Inc. / Zest Anchors, LLC
Product Code
DZE · Dental
Decision Date
Apr 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LOCATOR® Overdenture Implant System is designed to retain overdentures or partial dentures in the mandible or maxilla. Immediate loading is indicated when good primary stability has been achieved and with appropriate occlusal loading.
Device Story
The LOCATOR Overdenture Implant (LODI) System is an endosseous dental implant system used to retain overdentures or partial dentures in the mandible or maxilla. The device consists of a threaded and tapered titanium alloy implant. It is surgically placed by a dental professional. The implant features a roughened surface created via Resorbable Blast Media (RBM) to facilitate integration. The system is designed for immediate loading when primary stability is achieved. The device provides a mechanical anchor for dental prosthetics, improving denture stability and patient function. It is supplied sterile for single-use.
Clinical Evidence
No clinical studies were provided. Evidence is based on bench testing, including mechanical property evaluation, fatigue testing, biocompatibility (ISO 10993-1), sterilization validation (ISO 11137-2:2015), cleanability, and endotoxin testing (USP <85>, USP <161>, AAMI ST72).
Technological Characteristics
Material: Ti 6Al-4V ELI (ASTM F136). Surface: Resorbable Blast Media (RBM). Design: Threaded and tapered endosseous implant with dual-lead threads and blunted tip. Sterilization: Gamma irradiation (25 kGy) to SAL 10^-6. Packaging: Sterile plastic tray with Tyvek lid.
Indications for Use
Indicated for patients requiring retention of overdentures or partial dentures in the mandible or maxilla. Immediate loading is indicated given good primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
Zest Anchors LLC Marysa Loustalot Sr. Regulatory Affairs Specialist 2875 Loker Avenue East Carlsbad, California 92010
Re: K203701
Trade/Device Name: Locator® Overdenture Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: March 19, 2021 Received: March 22, 2021
Dear Marysa Loustalot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203701
Device Name
The LOCATOR® Overdenture Implant System
Indications for Use (Describe)
The LOCATOR® Overdenture Implant Ssystem is designed to retain overdentures in the mandible or maxilla. Immediate loading is indicated when good primary stability has been achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☒</span> Prescription Use (Part 21 CFR 201 Subpart D) | ☐ Over-The-Counter Use (21 CFR 201 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Zest Dental Solutions. The logo has the letters "ZD" in a bold, blue font on the left side. To the right of the letters is the text "ZEST DENTAL" in a bold, blue font, with the word "SOLUTIONS" underneath in a lighter blue font.
## April 08. 2021
#### SUBMITTER I.
Zest Anchors, LLC 2875 Loker Ave. East. Carlsbad. CA 92010 Phone: +1 (760) 743-7744 Contact: Marysa Loustalot ext. 596 Date Prepared: April 8, 2021
#### II. DEVICE INFORMATION
Device / Trade Name: LOCATOR® Overdenture Implant System Common Name: Dental Implant System Classification Name: Endosseous dental implant Regulatory Classification: Class II, 21 CFR 872.3640 Product Code: DZE
#### III. PREDICATE DEVICE(s)
Primary Predicate: LOCATOR® Overdenture Implant System (K120198)
Reference Device(s): Legacy2, Legacy3, Legacy4, SimplyLegacy2, SimplyLegacy3 Dental Implants (K192221)
#### IV. DEVICE DESCRIPTION
The LOCATOR Overdenture Implant (LODI) was originally cleared under K120198. This submission is to add additional implant sizes to the LODI family that for distinction purposes, will hereinafter be called the Standard Ridge LOCATOR Overdenture Implant (LODI) System. The added implant sizes are listed in Table 1.0 below:
| | LENGTH | | | |
|----------|--------|------|------|------|
| DIAMETER | 8mm | 10mm | 12mm | 14mm |
| 3.5mm | x | x | x | x |
| 3.9mm | x | x | x | x |
| 4.4mm | x | x | x | x |
| 4.9mm | x | x | x | x |
Table 1.0 Subiect Device Sizes
Identical to K120198, the subject LODI Standard Ridge System is designed to retain overdentures of partial dentures in the mandible or maxilla. The System includes a threaded and tapered endosseous dental implant that is made from
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" in a dark blue color on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a dark blue color, and the word "SOLUTIONS" is written in a lighter blue color.
the identical material, 6Al-4V ELI Titanium, and maintains its conformance to ASTM F136. The implant surface is roughened by Resorbable Blast Media (RBM) up to the abutment seating platform by the same manufacturing processes with the same manufacturing equipment as the existing LODI implant.
Additional wording was added to the Indications for Use statement for clarification purposes only. The Subject device's Indications for Use statement remains within scope of that cleared in K120198.
#### INDICATIONS FOR USE V.
The LOCATOR® Overdenture Implant System is designed to retain overdentures or partial dentures in the mandible or maxilla. Immediate loading is indicated when good primary stability has been achieved and with appropriate occlusal loading.
| FEATURES | SUBJECT | PREDICATE | REFERENCE DEVICE |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K203701 | K120198 | K192221 |
| Device Name | LOCATOR Overdenture<br>Implant (LODI) System | LOCATOR Overdenture<br>Implant (LODI) System | Legacy2, Legacy3,<br>Legacy4, simplyLegacy2,<br>simplyLegacy3 |
| Manufacturer | Zest Anchors, LLC | Zest Anchors, LLC | Implant Direct Sybron<br>Manufacturing, LLC |
| Indications<br>for Use | The LOCATOR®<br>Overdenture Implant<br>System is designed to<br>retain overdentures or<br>partial dentures in the<br>mandible or maxilla.<br>Immediate loading is<br>indicated when good<br>primary stability has been<br>achieved and with<br>appropriate occlusal<br>loading. | The LOCATOR®<br>Overdenture Implant<br>System is designed to<br>retain overdentures or<br>partial dentures in the<br>mandible or maxilla | These implants are<br>intended for use in partially<br>and fully edentulous upper<br>and lower jaws in support<br>of single or multiple-unit<br>restorations and terminal or<br>intermediate abutment<br>support for fixed<br>bridgework.<br>Implants can be indicated<br>for immediate loading<br>when good primary<br>stability has been achieved<br>and with appropriate<br>occlusal loading. |
| Product Code | DZE | DZE | DZE |
| Classification | Class II | Class II | Class II |
| Regulation | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 |
| RX / OTC | RX | RX | RX |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI | Ti 6Al-4V ELI |
#### SUBSTANTIAL EQUIVALENCY COMPARISON VI.
LODI Standard Ridge
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, teal font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a dark teal font, with the word "SOLUTIONS" written below in a lighter teal font.
| FEATURES | SUBJECT | PREDICATE | REFERENCE DEVICE |
|----------------------------------|-------------------------------|---------------------------|-------------------------------------------------------|
| Material<br>Surface<br>Treatment | Resorbable Blast Media | Resorbable Blast Media | Soluble Blast Media |
| Diameter (Ø) | 3.5mm, 3.9mm, 4.4mm,<br>4.9mm | 2.4mm and 2.9mm | 3.2mm, 3.7mm, 4.2mm,<br>4.7mm, 5.2mm, 5.7mm,<br>7.0mm |
| Length | 8mm, 10mm, 12mm,<br>14mm | 10mm, 12mm, 14mm,<br>16mm | 6mm, 8mm, 10mm,<br>11.5mm, 13mm 16mm |
#### VII. PERFORMANCE TESTING
| ATTRIBUTE TEST | DISCUSSION / RESULTS |
|---------------------------|--------------------------------------------------------------------------------------------------------------|
| Mechanical Properties | |
| - Body Type | Identical Threaded and Tapered Body Type as Predicate |
| - Thread Type | Thread type is a Dual Lead compared to the variable single lead of the |
| - Tip Type | Predicate. A dual lead thread does not create a new worst case regarding<br>implant performance or strength. |
| | Blunted (Predicate was sharp): The blunted tip of the Subject stops the |
| | implant from advancing past the drilled depth. |
| Corrosion Testing | N/A - Implant and implant attachment are made from the same type of material |
| | (titanium alloy); therefore, the implant will see no effects from galvanic corrosion. |
| | Titanium alloy is inherently corrosion resistant; therefore, the implant will see no |
| | effects of corrosion of the base metal. |
| Clinical Studies | No clinical studies are provided as part of this 510(k) submission. |
| Biocompatibility | The category, contact, and contact duration per ISO 10993-1 Table A.1 are the same |
| | for the Subject and Predicate device. The increased surface are of the device does not |
| | alter the ISO 10993-1 medical device categorization and therefore does not impact the |
| | biocompatibility profile of the LODI implant cleared under K120198. The |
| | biocompatibility of Ti-6AL-4V ELI remains unchanged from that of the Predicate. |
| Sterilization | Validation testing per ISO 11137-2:2015 was conducted to ensure that the Standard |
| | Ridge LODI Implant can also be sterilized in the same manner as the predicate implant |
| | by 25 kGy to achieve a sterility assurance level (SAL) of 10-6. |
| Cleanability | The Total Organic Carbon and cytotoxicity test result demonstrated that the new |
| | worst-case implant, the Standard Ridge LODI 4.9mm implant, meets the acceptance |
| | criteria for HFE cleaning, HNO3 cleaning, and IPA cleaning. |
| Fatigue Testing | Due to the larger cross-sectional area of the subject device, the identical finishing, |
| | material and manufacturing processes, it can be concluded that the subject device does |
| | not introduce a new worst case in respect to fatigue strength. Therefore, the fatigue |
| | testing reported in 510(k) K120198 remains applicable to the subject device. |
| Shelf-Life | The Subject device is supplied sterile in standard plastic tray with Tyvek™ lid using |
| | the same validated processes and materials as the Predicate. Accelerated aging |
| | conducted on the Predicate device as the worst case (2.9mm X 10mm with 4mm Cuff |
| | Height LOCATOR Abutment) validated for a period to to 5 years. The Subject device |
| | does not introduce a new worst case and therefore falls within the scope of the |
| | Predicate Shelf-Life testing. |
| Surface Treatment | The Resorbable Blast Media used for the implant surface treatment is the identical |
| | process and material used on the predicate implant cleared under K120198. |
| Endotoxin Testing | The method of detection for bacterial endotoxin uses the kinetic turbidimetric and |
| | chromogenic techniques per USP <85>, USP <161>, and AAMI ST72, with a limit of |
| | 20 EU/Device. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" in a dark blue color on the left side. To the right of the letters, the words "ZEST DENTAL" are written in dark blue, with the word "SOLUTIONS" written below in a lighter blue color.
# VII. CONCLUSION
The Standard Ridge LODI System is identical to the LODI System cleared under K120198 in terms of intended use, system components, materials, and packaging. The equivalency of the Standard Ridge LODI System to the original LODI System has been demonstrated through verification and validation activities.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.