LOCATOR OVERDENTURE IMPLANT SYSTEM
Device Facts
| Record ID | K120198 |
|---|---|
| Device Name | LOCATOR OVERDENTURE IMPLANT SYSTEM |
| Applicant | Zest Anchors, LLC |
| Product Code | DZE · Dental |
| Decision Date | Jun 6, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
Indications for Use
The LOCATOR® Overdenture Implant System is designed to retain overdentures or partial dentures in the mandible or maxilla.
Device Story
The LOCATOR® Overdenture Implant (LODI) System is a narrow-diameter endosseous dental implant system used to retain overdentures or partial dentures. The system consists of a narrow-diameter dental implant (2.4mm or 2.9mm diameter; 10-16mm length) and a screw-retained abutment. The implant features a flared collar to accommodate the abutment. The device is surgically placed in the mandible or maxilla by a dental professional. It functions as a mechanical anchor for prosthetic dentures, providing stability for the patient. No electronic or software components are involved.
Clinical Evidence
Bench testing only. Mechanical testing performed in accordance with ISO 14801, including static compression and dynamic compression (fatigue) tests, demonstrated substantial equivalence to the predicate device.
Technological Characteristics
Materials: Ti 6Al-4V ELI (ASTM F136 for implants, ASTM B348 for abutments). Surface: Resorbable Blast Media. Dimensions: 2.4mm or 2.9mm diameter implants; 10, 12, 14, or 16 mm lengths; 2.9mm platform abutments with 2mm or 4mm cuff heights. Mechanical fixation principle. No software, electronics, or connectivity.
Indications for Use
Indicated for patients requiring retention of overdentures or partial dentures in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- Ace Surgical Secure-Mini™ LOCATOR® Implant System (K092594)