K202966 · Ge Healthcare (Ge Medical Systems, LLC) · LNH · Nov 13, 2020 · Radiology
Device Facts
Record ID
K202966
Device Name
SIGNA Architect
Applicant
Ge Healthcare (Ge Medical Systems, LLC)
Product Code
LNH · Radiology
Decision Date
Nov 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The SIGNA Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Architect system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
SIGNA Architect is a whole-body superconducting MR scanner; acquires MR signals via gradient and RF transmissions; produces axial, sagittal, coronal, oblique images, spectroscopic images, parametric maps, and spectra. Operated by trained clinicians in clinical settings. New features: HyperSense 2.0 (sparse data compressibility for scan time reduction) and Star (motion-robust, free-breathing imaging). Output interpreted by physicians to assist in diagnosis. Benefits include high SNR, high resolution, and reduced scan times. No hardware changes; software modifications only.
Clinical Evidence
External reader evaluation for HyperSense 2.0 compared image quality of 3 reconstruction techniques against predicate; results showed acceptable image quality and uniformity. External clinical testing for Star feature confirmed images were of sufficient quality for diagnostic use by a U.S. Board Certified radiologist.
Technological Characteristics
Superconducting magnet MR system; 128 independent receive channels; supports axial, sagittal, coronal, oblique, and double oblique planes. Software-based acceleration (HyperSense 2.0) and motion-robust imaging (Star). No changes to magnet homogeneity, stability, gradient strength, or slew rate compared to predicate.
Indications for Use
Indicated for diagnostic imaging of the entire body (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring MR imaging. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
November 13, 2020
GE Healthcare (GE Medical Systems, LLC) % Mr. Joseph Beach Regulatory Affairs Leader, MR 3200 N Grandview Blvd. WAUKESHA WI 53188
Re: K202966
Trade/Device Name: SIGNA Architect Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: September 28, 2020 Received: September 30, 2020
Dear Mr. Beach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202966
Device Name
SIGNA Architect
#### Indications for Use (Describe)
The SIGNA Architect system is a whole body magnetic resonance scanner designed to support high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breat, abdomen, pelvis, joints, prostate, blood vessels, and musculoskelatal regions of the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA Architect system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
#### Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
ng Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. There are three stylized swirls or arrows equally spaced around the outside of the circle.
Section 005
510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The letters are also blue, matching the color of the circle that surrounds them.
# GE Healthcare
510(k) Premarket Notification Submission SIGNA Architect
# 510(k) Summary
## K202966
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | September 29, 2020 |
|-----------------------|----------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contacts: | Joseph Beach<br>Regulatory Affairs Leader<br>Phone: 262 200-1270<br>Email: joseph.beach@ge.com |
| | Sandra Westphal<br>Regulatory Affairs Leader<br>Phone: 262 720-8872<br>Email: sandra.westphal@ge.com |
| Secondary Contact: | Andrew Menden<br>Senior Regulatory Affairs Manager<br>Phone: 262 308-5719<br>Email: andrew.menden@ge.com |
| Device Trade Name: | SIGNA Architect |
| Common / Usual Name: | MR System |
| Classification Name: | Magnetic Resonance Diagnostic Device<br>21 CFR 892.1000 |
| Primary Product Code: | LNH<br>LNI, MOS |
| Predicate Device: | K163331<br>SIGNA Architect<br>GE Medical Systems, LLC |
| Reference Devices: | K162722<br>HyperSense<br>GE Medical Systems, LLC |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The logo is colored in a light blue hue, and the circular frame has a slightly darker shade of blue.
## Device Description:
SIGNA Architect is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The system features a superconducting magnet. The data acquisition system accommodates up to 128 independent receive channels in various increments and multiple independent coil elements per channel during a single acquisition series. Each system uses a combination of time varying magnetic fields (gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. Each system can imaqe in the saqittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences and reconstruction algorithms.
This submission is prompted by the introduction of two new software features called HyperSense 2.0 and Star onto SIGNA Architect. HyperSense 2.0 is an acceleration technique based on sparse data compressibility allowing scan time reduction while maintaining SNR efficiency. Star is a motion-robust, free-breathing imaging technique. HyperSense 2.0 is a modification to the previously cleared HyperSense, while Star is a technique that can be used with the previously cleared DISCO feature. Both HyperSense and DISCO are listed above as reference devices along with their associated 510(k) submission numbers.
The addition of both the HyperSense 2.0 and Star features involved modifications to the SIGNA Architect system software. There were no changes from either of these features that were related to the system's hardware components.
### Indications for Use:
The Indications for Use statement for the proposed device is identical to that of the predicate device:
The SIGNA Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal reqions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA Architect system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
The addition of HyperSense 2.0 and Star does not impact the intended use of the SIGNA Architect system.
## Comparison of Technological Characteristics:
Many of the technological characteristics of the proposed SIGNA Architect system are unchanged from the predicate device. There are no changes to the maqnet, gradient, and RF subsystems compared to the predicate K163331. Key performance specifications (such as magnet homogeneity and stability, maximum gradient strength, and slew rate, etc.) for the system are also unchanged.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The color of the logo is a light blue.
# GE Healthcare 510(k) Premarket Notification Submission SIGNA Architect
The software used on the proposed SIGNA Architect system has been modified to include the HyperSense 2.0 and Star features. The user interface provides operators of the system with new options for selecting these features and adjusting any parameters associated with them. However, the proposed SIGNA Architect system with HyperSense 2.0 and Star employs the same fundamental technology as the predicate device.
## Summary of Nonclinical Testing:
The following quality assurance measures were applied to the development of the device:
- Risk Analysis .
- Requirements Reviews .
- . Design Reviews
- . Testing on a Unit Level (Module Verification)
- Integration Testing (System Verification) .
- Performance Testing (Verification) .
- . Safety Testing (Verification)
- . Simulated Use Testing (Validation)
Exploratory testing was also performed to ensure that the proposed SIGNA Architect is at least as safe and effective as the predicate device. Verification documents, validation documents, and test reports have been provided for more details. No new questions of safety and effectiveness were raised during nonclinical testing.
### Summary of Clinical Testing:
Studies were performed for both proposed HyperSense 2.0 and Star features.
An external reader evaluation study was performed for HyperSense 2.0. The images involved were generated using 3 different reconstruction techniques across different anatomies. Radiologists were asked to evaluate side-by-side image quality of the HyperSense 2.0 images compared to the predicate. Overall image quality and uniformity was acceptable.
External clinical testing was performed for the Star feature. Images produced by Star were judged to be of sufficient quality for diagnostic use by a U.S. Board Certified radiologist. Images from the assessment are provided.
Clinical testing confirms that both HyperSense 2.0 and Star can be used safely and effectively in a clinical setting.
## Conclusions Drawn from Performance Testing:
Nonclinical and clinical testing demonstrates that the proposed SIGNA Architect with HyperSense 2.0 and Star is at least as safe and effective as the predicate device and does not raise different questions of safety and effectiveness.
The proposed SIGNA Architect system was developed under GE Healthcare's quality system. The performance testing did not identify any new hazards, adverse effects, or safety or performance concerns that are significantly different from those associated with MR imaging in general.
Therefore, GE Healthcare believes that the proposed SIGNA Architect with HyperSense 2.0 and Star is substantially equivalent to the predicate device, and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.