K162722 · Ge Medical Systems, LLC · LNH · Apr 20, 2017 · Radiology
Device Facts
Record ID
K162722
Device Name
HyperSense
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Apr 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162722 · Apr 20, 2017
HyperSense
Ge Medical Systems, LLC
Peer-reviewed clinical literature
A peer-reviewed journal article was provided as supporting evidence to demonstrate that the HyperSense technique maintains comparable imaging performance to the predicate device.
Peer reviewed journal article describing a study of the HyperSense technique
—
Not applicable for this study
Imaging performance/image quality
Indications for Use
HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen.
Device Story
HyperSense is a software-only acceleration feature for GE 1.5T and 3.0T MR systems; utilizes sparse data sampling and iterative reconstruction to enable faster imaging or higher spatial resolution; operates as an add-on to existing volumetric acquisition workflows; can be combined with other acceleration methods like ARC. Used by radiologists and technologists in clinical settings to optimize scan times and image quality. Output is reconstructed MR images; assists clinicians in diagnostic assessment by maintaining image quality while reducing acquisition time. Benefits patients through shorter exam durations and potentially improved image resolution.
Clinical Evidence
Clinical study evaluated image quality via legibility of morphological features; supported by a peer-reviewed journal article. Results demonstrated comparable imaging performance to the predicate device (K142085). Bench testing included phantom studies evaluating Signal to Noise Ratio (SNR) and spatial resolution, with all tests meeting pass/fail criteria.
Technological Characteristics
Software-only feature for GE 1.5T/3.0T MR systems. Employs compressed sensing (sparse data sampling) and iterative reconstruction. Complies with AAMI/ANSI 62304, AAMI/ANSI ES60601-1, IEC 60601-2-33, and NEMA PS3.1-3.18 (DICOM).
Indications for Use
Indicated for patients undergoing non-contrast enhanced MRI of the head, neck, spine, extremities, pelvis, and abdomen on GE 1.5T and 3.0T systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2017
Ge Medical Systems, LLC Glen Sabin Regulatory Affairs Director 3200 Grandview Blvd Waukesha. Wisconsin 53188
Re: k162722
Trade/Device Name: HyperSense Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: March 15, 2017 Received: March 16, 2017
Dear Glen Sabin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162722
Device Name HyperSense
#### Indications for Use (Describe)
HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue. The background is white.
| In accordance with 21 CFR 807.92 the following summary of information is provided: | | |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date: | 19 April 2017 | |
| Primary Contact Person: | Glen Sabin<br>Regulatory Affairs Director, MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848<br>Fax: 414-908-9585 | |
| Secondary Contact Person: | James McMahon<br>Senior Director, Regulatory Affairs - MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 508-382-2858 | |
| Device Trade Name: | HyperSense | |
| Common/Usual Name: | Magnetic Resonance Diagnostic Device | |
| Classification Names: | 892.1000 | |
| Product Code: | LNH | |
| Predicate Devices: | Predicate Device: ARC (K142085) | |
| Device Description: | HyperSense is a software feature used on GE 1.5T and<br>3.0T MR systems.<br>HyperSense is an acceleration technique based on<br>sparse data sampling and iterative reconstruction,<br>enabling faster imaging without the penalties<br>commonly found with conventional parallel imaging.<br>HyperSense is intended to be used with volumetric<br>acquisitions, and can be combined with other methods<br>of acceleration (ARC) for achieving high signal to noise<br>ratio with shorter acquisition times. HyperSense can<br>deliver higher spatial resolution images or reduced<br>scan times. | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplet shapes around the letters. The logo is simple and recognizable.
| Indications for Use: | HyperSense is a software feature intended for use on<br>GE MR 1.5T and 3.0T Systems. HyperSense is an<br>acceleration technique based on sparse data sampling<br>and iterative reconstruction that allows users to reduce<br>scan times or increase scan resolution. HyperSense<br>can be used for non-contrast enhanced imaging of the<br>head, neck, spine, extremities, pelvis, and abdomen. |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of Intended<br>Use | Both HyperSense and the predicate device are<br>magnetic resonance imaging devices intended for<br>diagnostic use.<br>Both indications for use statements are functional in<br>nature, and do not list specific diseases or conditions.<br>HyperSense and the predicate device are indicated for<br>the same patient population, and for the same clinical<br>setting. Both indications for use statements include a<br>range of intended anatomies.<br>Therefore, GE Healthcare believes that HyperSense has<br>the same intended use as the predicate device in<br>accordance with the FDA's guidance document "The<br>510(k) Program: Evaluating Substantial Equivalence in<br>Premarket Notifications [510(k)]", dated 28 July 2014. |
| Comparison of<br>Technological<br>Characteristics: | The most notable technological difference between<br>HyperSense and the predicate device is that HyperSense<br>adopts compressed sensing acquisition method based<br>on sparse data compressibility allowing scan time<br>reduction while maintaining appropriate image quality.<br>Per bench and clinical data collected to validate the<br>indications for use, this technological difference does<br>not raise any different questions of safety and<br>effectiveness. Both devices must address questions of<br>whether they provide an adequate level of image<br>quality appropriate for diagnostic use. The performance<br>data described in this submission include results of<br>both bench testing and clinical testing that show the<br>image quality performance of HyperSense compared to<br>the predicate device |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are also some water droplet-like shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity.
| Performance Data | Summary of Non-Clinical Tests: |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | HyperSense is a software only feature and complies with the following voluntary standards:<br>AAMI/ANSI 62304 AAMI/ANSI ES60601-1 IEC 60601-2-33 |
| | In addition, HyperSense complies with NEMA PS3.1-3.18 for DICOM conformance. |
| | HyperSense complies with the same applicable standards as the software used in the predicate device. |
| | As with the predicate device, the following quality assurance measures were applied to the development of HyperSense:<br>Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) |
| | Phantom testing has been conducted to evaluate the impact of HyperSense on Signal to Noise Ratio (SNR), and spatial resolution. |
| | The non-clinical testing was completed with passing results per the pass/fail criteria defined in the test cases. |
| | Summary of Clinical Tests: |
| | A clinical study has been performed to evaluate the impact of HyperSense on image quality, as measured by the legibility of morphological features. Additionally, a peer reviewed journal article describing a study of the HyperSense technique has been provided in this submission as supporting evidence. The clinical results demonstrated that HyperSense maintains comparable imaging performance results as its predicate devices (K142085). Sample clinical images are included in this submission. |
| Conclusion: | GE Healthcare believes that the HyperSense software<br>feature has substantially the same intended use as the<br>predicate ARC. This 510(k) submission includes<br>information on the technological characteristics of the<br>HyperSense feature, as well as performance data<br>demonstrating that HyperSense is as safe and effective<br>as the predicate, and does not raise different questions<br>of safety and effectiveness.<br>In conclusion, GE Healthcare believes that the<br>HyperSense feature is substantially equivalent to the<br>predicate device. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the letters, creating a dynamic and recognizable design. The logo is simple, clean, and easily identifiable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.