Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T; SIGNA Architect and SIGNA Artist
K163331 · Ge Medical Systems, LLC · LNH · Mar 17, 2017 · Radiology
Device Facts
Record ID
K163331
Device Name
Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T; SIGNA Architect and SIGNA Artist
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Mar 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The SIGNA Architect, SIGNA Artist, Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems are whole body magnetic resonance scanners designed to support high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal. coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Architect, SIGNA Artist, Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
Device Story
Whole-body MR scanners using superconducting magnets, time-varying magnetic field gradients, and RF transmissions to acquire MR signals. Data acquisition system supports up to 128 independent receive channels. Systems generate axial, sagittal, coronal, oblique, and double-oblique images, spectroscopic images, and parametric maps. Operated by trained professionals in clinical settings. Software features (MAGIC DWI, CardioMaps) process raw MR data to produce diagnostic images/maps. Output interpreted by physicians to assist in diagnosis of anatomical structures and functions. Benefits include high signal-to-noise ratio and reduced scan times.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing, including phantom testing for Synthetic DWI and CardioMaps, and submission of sample clinical images. Compliance with AAMI/ANSI 62304, AAMI/ANSI ES60601-1, and IEC 60601-2-33 standards confirmed.
Technological Characteristics
Superconducting magnet MR system; 128-channel data acquisition; NEMA DICOM standard compliance. Software-based image reconstruction and processing (MAGIC DWI, CardioMaps). Complies with IEC 60601-2-33, AAMI/ANSI ES60601-1, and AAMI/ANSI 62304.
Indications for Use
Indicated for whole-body diagnostic MR imaging (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, vasculature, musculoskeletal) to produce images, spectra, and parametric maps for physician interpretation. Applicable to all patient populations requiring MR imaging; contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2017
GE Medical Systems, LLC % Mary A Mayka, Ph.D. Regulatory Affairs Manager 3200 Grandview Blvd. WAUKESHA WI 53188
Re: K163331
Trade/Device Name: SIGNA Architect. SIGNA Artist. Discovery MR750 3.0T. Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: February 17, 2017 Received: February 21, 2017
Dear Dr. Mayka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K163331
.
Device Name
SIGNA Architect, SIGNA Artist, Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T
#### Indications for Use (Describe)
The SIGNA Architect, SIGNA Artist, Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems are whole body magnetic resonance scanners designed to support high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal. coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA Architect, SIGNA Artist, Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable)
| | <span></span> X Prescription Use (Part 21 CFR 801 Subpart D) |
|--|------------------------------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a decorative border of curved lines. The logo is simple and recognizable, and it is associated with a well-known and established company.
Section 05: 510(k) Summary Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima MR450w 1.5T, Discovery MR450 1.5T, SIGNA Architect, SIGNA Artist
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. There are also white droplet-like shapes surrounding the circle, giving it a dynamic and fluid appearance. The logo is simple, recognizable, and represents the multinational conglomerate corporation.
| In accordance with 21 CFR 807.92 the following summary of information is provided: | | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date: | November 22, 2016 | |
| Primary Contact Person: | Mary A. Mayka, Ph.D.<br>Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 262-527-3148<br>Fax: 262-364-2785 | |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director, MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848<br>Fax: 414-908-9585 | |
| Device Trade Name: | Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima<br>MR450w 1.5T, Discovery MR450 1.5T, SIGNA Architect,<br>SIGNA Artist | |
| Common/Usual Name: | Magnetic Resonance Diagnostic Device | |
| Classification Names: | Magnetic Resonance Diagnostic Device<br>(21 CFR 892.1000) | |
| Product Code: | LNH | |
| Predicate Devices: | K160618: Discovery MR750 3.0T, Discovery MR450 1.5T,<br>Discovery MR750w 3.0T, Optima MR450w 1.5T, SIGNA<br>Architect, SIGNA Artist | |
| Device Description: | The Discovery MR750 3.0T, Discovery MR450 1.5T,<br>Discovery MR750w 3.0T, Optima MR450w 1.5T, SIGNA<br>Architect and SIGNA Artist systems are whole body<br>magnetic resonance scanners designed to support high<br>resolution, high signal-to-noise ratio, and short scan times.<br>The systems each feature a superconducting magnet. The<br>data acquisition system accommodates up to 128<br>independent receive channels in various increments and<br>multiple independent coil elements per channel during a<br>single acquisition series. Each system uses a combination of<br>time varying magnetic fields (gradients) and RF<br>transmissions to obtain information regarding the density<br>and position of elements exhibiting magnetic resonance.<br>Each system can image in the sagittal, coronal, axial,<br>oblique, and double oblique planes, using various pulse<br>sequences and reconstruction algorithms.<br>The Discovery MR750 3.0T, Discovery MR450 1.5T,<br>Discovery MR750w 3.0T, Optima MR450w 1.5T, SIGNA<br>Architect. SIGNA Artist systems are designed to conform to | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The space between the letters and the border is filled with a light blue color. The logo is simple and recognizable, representing the company's brand identity.
| | NEMA DICOM standards (Digital Imaging and<br>Communications in Medicine). |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The original description hasn't changed from predicate<br>devices (K160618), other than reflecting the additional<br>receive channels available.<br><br>The modifications to these systems include the MAGIC DWI<br>and CardioMaps software features, delivered via the DV26<br>program. The proposed software features will be ported<br>over to other GE Healthcare MR systems based on<br>appropriate design controls and evaluation of the change in<br>accordance with FDA's Guidance—Deciding When to<br>Submit a 510(k) for a Change to an Existing Device. |
| Indications for Use: | The SIGNA Architect, SIGNA Artist, Discovery MR750<br>3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and<br>the Optima MR450w 1.5T systems are whole body magnetic<br>resonance scanners designed to support high resolution, high<br>signal-to-noise ratio, and short scan times. It is indicated for<br>use as a diagnostic imaging device to produce axial, sagittal,<br>coronal, and oblique images, spectroscopic images,<br>parametric maps, and/or spectra, dynamic images of the<br>structures and/or functions of the entire body, including, but<br>not limited to, head, neck, TMJ, spine, breast, heart,<br>abdomen, pelvis, joints, prostate, blood vessels, and<br>musculoskeletal regions of the body. Depending on the<br>region of interest being imaged, contrast agents may be used.<br>The images produced by the SIGNA Architect, SIGNA<br>Artist, Discovery MR750 3.0T, Discovery MR450 1.5T,<br>Discovery MR750w 3.0T and the Optima MR450w 1.5T<br>systems reflect the spatial distribution or molecular<br>environment of nuclei exhibiting magnetic resonance. These<br>images and/or spectra when interpreted by a trained<br>physician yield information that may assist in diagnosis. |
| Comparison of Intended<br>Use | The addition of the proposed software features does not<br>impact the indications for use. The indications for use have<br>not changed, other than the addition of the SIGNA Architect<br>and SIGNA Artist systems (K160618/001 and<br>K160618/002).<br><br>Therefore, the intended use is the same as the predicate<br>device in accordance with the FDA's guidance document<br>"The 510(k) Program: Evaluating Substantial Equivalence in |
| Comparison of<br>Technological<br>Characteristics: | The MR systems with the proposed software features<br>employ the same fundamental technology as the predicate<br>devices. |
| Performance Data: | Summary of Non-Clinical Tests:<br>The changes are software only features and comply with the<br>following voluntary standards:<br>• AAMI/ANSI 62304<br>• AAMI/ANSI ES60601-1<br>• IEC 60601-2-33 |
| | The following quality assurance measures were applied to<br>the development of the device:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Testing on unit level (Module verification)<br>• Integration testing (System verification)<br>• Performance testing (Verification)<br>• Safety testing (Verification)<br>• Simulated use testing (Validation) |
| | The non-clinical tests have been summarized in this 510(k).<br>Phantom testing for both Synthetic DWI and CardioMaps<br>was completed to demonstrate acceptable performance.<br>Testing was completed with passing results per the pass/fail<br>criteria defined in the test cases. This supports substantial<br>equivalence to its predicates because it was also developed<br>under quality assurance Design Controls.<br>Summary of Clinical Tests: |
| | Sample clinical images are included in this submission in<br>accordance with the MR guidance on premarket notification<br>submissions. |
| Conclusion: | The GE MR systems with the modified software features<br>have the same intended use as the predicate. This 510(k)<br>submission includes information on the technological<br>characteristics of the proposed software features, as well as<br>performance data demonstrating that the features are as safe<br>and effective as the predicate, and does not raise different<br>questions of safety and effectiveness. |
| | In conclusion, GE Healthcare believes that the MR systems<br>with the modified software features are substantially<br>equivalent to the predicate devices. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a series of curved lines that resemble water droplets or splashes. The logo is simple and recognizable, and it is often used to represent the company's brand.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.