K200572 · Planmeca Oy · LLZ · Dec 2, 2020 · Radiology
Device Facts
Record ID
K200572
Device Name
Planmeca Romexis
Applicant
Planmeca Oy
Product Code
LLZ · Radiology
Decision Date
Dec 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
Planmeca Romexis is a medical imaging software intented for use in dental and medical care as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, diagnose, review, store, print, and distribute images of both adult and pediatric patients. Planmeca Romexis is also a preoperative software used for dental implant planning. Based on the planned implant position a model of a surgical guide for a guided implant surgery can be designed. The designed objects can be exported to manufacture a separate physical product. Planmeca Romexis is also a preoperative software for simulating surgical treatment options. Planmeca Romexis is also intended to be used for monitoring, storing and displaying mandibular jaw positions and movements relative to the maxilla. Additionally, Planmeca Romexis includes monitoring features for Planmeca devices for maintenance purposes. The software is designed to work as a stand-alone or as an accessory to Planmeca imaging and Planmeca dental unit products in standard PC.
Device Story
Planmeca Romexis is modular dental/medical imaging software; operates on standard PCs; functions as standalone or accessory to Planmeca imaging/dental units. Inputs include 2D/3D image files (e.g., projection radiography, CBCT), optical impressions, and jaw motion data. Software processes, renders, and displays images; provides tools for enhancement, linear/angular/volumetric measurements, implant planning, surgical simulation, and mandibular movement monitoring. Includes maintenance monitoring for Planmeca hardware. Output consists of visualized diagnostic images, surgical guide designs for export/manufacturing, and treatment simulations. Used by radiologists, clinicians, and surgeons in clinical settings to assist in diagnosis and preoperative planning. Benefits include improved visualization, precise implant positioning, and surgical workflow efficiency.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation included risk analysis, requirements/design reviews, and performance/safety testing. Bench testing compared images rendered by Planmeca Romexis against predicate software versions, confirming equal effectiveness in essential functions and substantially equivalent clinical data output.
Technological Characteristics
Software-based imaging system; operates on standard PC hardware. Features multi-planar reconstruction (MPR), 3D volume rendering, and surface rendering. Supports DICOM format. Includes modules for implant planning, surgical simulation, and jaw motion tracking. Connectivity includes integration with Practice Management Systems and hospital PACS. No specific materials or energy sources as it is a software-only device.
Indications for Use
Indicated for adult and pediatric patients requiring dental and medical 2D/3D image visualization, dental implant planning, surgical treatment simulation, and monitoring of mandibular jaw positions relative to the maxilla. Intended for use by authorized healthcare professionals including radiologists, clinicians, and referring physicians. Implant planning requires training in implantology/surgical dentistry; surgical planning requires training in maxillofacial surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
December 2, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Planmeca Oy % Lars Moring Regulatory Affairs Manager Asentajankatu 6 Helsinki. 00880 FINLAND
Re: K200572
Trade/Device Name: Planmeca Romexis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 19, 2020 Received: October 23, 2020
Dear Lars Moring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200572
Device Name Planmeca Romexis
#### Indications for Use (Describe)
Planmeca Romexis is a medical imaging software intental and medical care as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, diagnose, review, store, print, and distribute images of both adult and pediatric patients.
Planmeca Romexis is also a preoperative software used for dental implant planning. Based on the plant position a model of a surgical guide for a guided implant surgery can be designed objects can be exported to manufacture a separate physical product.
Planmeca Romexis is also a preoperative software for simulating surgical treatment options.
Planmeca Romexis is also intended to be used for monitoring, storing and displaying mandibular jaw positions and movements relative to the maxilla.
Additionally, Planmeca Romexis includes monitoring features for Planmeca devices for maintenance purposes. The software is designed to work as a stand-alone or as an accessory to Planmeca dental unit products in standard PC.
The software is for use by authorized healthcare professionals. Use of the software for implant planning requires that the user has the necessary medical training in implantology and surgical dentistry. Use of the software for surgical treatment planning requires that the user has the necessary medical training in maxillofacial surgery.
Indications of the dental implants do not change with guided surgery compared to conventional surgery.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PLANMECA" in large, bold, blue letters. The letters are all capitalized and evenly spaced. The font appears to be sans-serif.
# 510(k) SUMMARY
K200572
# DATE
October 19, 2020
# PRODUCT, CLASSIFICATION NAME
| Device name: | Planmeca Romexis |
|----------------------|------------------------------------------------|
| Common name: | System, Image Processing, Radiological |
| Classification: | Class II |
| Classification name: | Imaging Processing System, LLZ, 21 CFR 892.205 |
## SUBMITTED BY
Planmeca Oy Asentajankatu 6 00880 Helsinki, Finland Phone: +358 20 7795 500 Contact person: Mr. Lars Moring
# U.S DESIGNATED AGENT
Planmeca USA Inc. 2600 Forbs Ave., Hoffman Estates, IL 60192 Phone: (630) 529 2300 Contact person: Ed McDonough
# DEVICE DESCRIPTION
Planmeca Romexis is a modular imaging software for dental and medical use. It is divided into modules to provide user access to different workflow steps involving different diagnostic views of images. Patient management screen with search capabilities lets users to find patients and identify correct patient file before starting work with a patient. After creating or selecting patient, new images can be acquired using select Planmeca X-ray units.
Planmeca Romexis is capable of processing and displaying 2D images in different formats and 3D CBCT images in DICOM format. 3D CBCT images can be viewed in near real-time multi projection reconstruction (MPR) views. 2D and 3D image browsers are provided to allow user access to relevant images. Typical image enhancement filters and tools are available to assist the user in making diagnosis, but original exposure is always kept in the database for reference.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PLANMECA" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The word is the main focus of the image and takes up most of the space.
lmages can be exported to files or writable media, printed to paper or DICOM media or transferred securely to other users using Planmeca online services. Interfaces to select external software are provided to facilitate exchange of patient information and images or data between Software and 3rd party applications.
## INDICATIONS FOR USE
Planmeca Romexis is a medical imaging software intented for use in dental and medical care as a tool for displaying and visualizing dental and medical 2D and 3D image files from imaging devices, such as projection radiography and CBCT. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, diagnose, review, store, print, and distribute images of both adult and pediatric patients.
Planmeca Romexis is also a preoperative software used for dental implant planning. Based on the planned implant position a model of a surgical guide for a guided implant surgery can be designed. The designed objects can be exported to manufacture a separate physical product.
Planmeca Romexis is also a preoperative software for simulating surgical treatment options.
Planmeca Romexis is also intended to be used for monitoring, storing and displaying mandibular jaw positions and movements relative to the maxilla.
Additionally, Planmeca Romexis includes monitoring features for Planmeca devices for maintenance purposes. The software is designed to work as a stand-alone or as an accessory to Planmeca imaging and Planmeca dental unit products in standard PC.
The software is for use by authorized healthcare professionals. Use of the software for implant planning requires that the user has the necessary medical training in implantology and surgical dentistry. Use of the software for surgical treatment planning requires that the necessary medical training in maxillofacial surgery.
Indications of the dental implants do not change with quided surgery compared to conventional surgery.
## EQUIVALENT DEVICES
| | Primary predicate device: | |
|---------|---------------------------|--------------|
| K123519 | Anatomage | InVivoDental |
#### Reference devices:
Televere Systems K061035 K141570 3Shape Medical K133320 SICAT
TigerView Implant Studio SICAT Function
| Newly added reference device: | |
|-------------------------------|------------------|
| K110430 | Patterson Dental |
Dolphin Imaging
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PLANMECA" in large, bold, blue letters. The letters are all capitalized and evenly spaced. The font is sans-serif and the overall impression is one of strength and stability. The background is plain white.
# INDICATIONS FOR USE COMPARISON
The indications for use of Planmeca Romexis is substantially equivalent to that of the predicate devices, with few exceptions that do not raise any new or potential safety risks to the user or patient. The most evident is that Planmeca Romexis includes service managing features for Planmeca devices for maintenance purposes, which are not intended for clinical use.
# TECHNOLOGICAL CHARACTERISTICS
A comparison of key software characteristics, including operating environment, functionalities, image files, and other major features was performed. Some general comparison data:
## Primary predicate device:
| | Planmeca Romexis | InVivoDental (K123519) |
|-----------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Operating environment | standard PC hardware | standard PC hardware |
| Functionalities | Viewing, enhancing,<br>processing, archiving, printing<br>and sharing images | Retrieve, process, render,<br>review, store, print, assist in<br>diagnosis, and distribute<br>images |
| Image files | 2D, 3D | 3D |
| Major features | Multiplanar views | Section and Multislice View<br>Operations |
| | 3D volume rendering view | Volume Rendering of scan<br>data |
| | Linear, angular, area and<br>volumetric measurements | Linear, Angular,<br>Circumferential, Area and<br>Volumetric Measurements |
| | Implant Planning module | Implant, Abutment and<br>Restoration Treatment<br>Planning |
| | Airway volume measurement | Quick Airway Volume<br>Measurement and Evaluation |
| | TMJ module | TMJ Applications |
| | Virtual Cephalometric image | SuperCeph |
| | Virtual Panoramic image | SuperPano |
| | Software Viewer | InVivo Viewer |
## Reference devices:
| | Planmeca Romexis | TigerView Professional<br>(K061035) |
|-----------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Operating environment | standard PC hardware | standard PC hardware |
| Functionalities | Viewing, enhancing,<br>processing, archiving, printing<br>and sharing images | acquire, display, edit, review,<br>store, print, and distribute<br>images |
| Image files | 2D, 3D | 2D |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PLANMECA" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and is centered.
| Integration with Practice<br>Management Systems | YES | YES |
|-------------------------------------------------|---------------------------------------------------------------|--------------------------------------------|
| Integration with hospital PACS<br>systems | YES | YES |
| Features | image processing, brightness &<br>contrast, crop, annotations | resize, adjust contrast, crop,<br>annotate |
| | Planmeca Romexis | Implant Studio<br>(K141570) |
|-----------------------|---------------------------------------------------------------------------|-----------------------------------------------------------------|
| Operating environment | standard PC hardware | standard PC hardware |
| Functionalities | Viewing, 3D data fitting, implant<br>planning, implant guide<br>designing | prosthetic driven implant<br>planning, surgical guide<br>design |
| Image files | 2D, 3D | 3D |
| Major features | Implant guide designing,<br>implant and sleeve libraries | Implant guide designing,<br>implant and sleeve libraries |
| | Planmeca Romexis | SICAT Function<br>(K133320) |
|-----------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Operating environment | standard PC hardware | standard PC hardware |
| Functionalities | Viewing, enhancing,<br>processing, archiving, printing<br>and sharing images | Visualization and<br>segmentation of<br>imaging information |
| Visualized data types | 2D, 3D, optical impressions,<br>jaw motion data | 3D volume data, optical<br>impressions, jaw motion data |
| Major features | Visualization of 3D images<br>using multi-planar 2D views, 3D<br>volume rendering, 3D surface<br>rendering | Visualization of 3D images<br>using 2D slice views, 3D<br>volume rendering, 3D surface<br>rendering |
| | Fitting of optical impressions to<br>3D image | Registration of optical scans to<br>volume data |
| | Segmentation of the jaws | Segmentation of jaws using<br>wizard |
| | Visualization of jaw movement<br>data | Visualization of jaw movement<br>data |
Comparison table between the subject and newly added reference device:
| | Planmeca Romexis | Dolphin Imaging<br>(K110430) |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Operating environment | Standard PC hardware | Windows OS |
| Visualized data types | 2D, 3D, optical impressions, 3D<br>face photos | 2D, digital study models, 3D,<br>3D face camera photos |
| Major features | Visualization of 3D images<br>using multi-planar 2D views, 3D<br>volume rendering, 3D surface<br>rendering, airway rendering | Multi-planar visualization,<br>rendering of 3D volume, 3D<br>photo, digital impression,<br>airway |
| | Linear, angular, area and | 3D-2D measurements: |
| | volumetric measurements | distance, angle, area |
| | Segmentation of jaws | Segmentation of image data |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PLANMECA" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word appears to be a logo or brand name.
| Definition of osteotomy planes | Definition of osteotomy planes |
|--------------------------------------------------|-------------------------------------------------------|
| Free movement and rotation of<br>bone fragments. | Treatment simulation by<br>moving skeletal structures |
Based on the presented technical comparison of the Planmeca Romexis and the predicate devices it was concluded that these devices are technically substantially equivalent.
The differences found are not significant as they do not raise any new or potential safety risks to the user or patient, or questions related to safety or effectiveness.
# NON-CLINICAL TEST RESULTS
The following quality assurance measures were applied to the development of the Software:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- . Performance testing (Verification)
- . Safety testing (Verification)
- . Final acceptance testing (Validation)
- Bench testing to compare with predicate software ●
Testing confirmed that Planmeca Romexis is stable and operating as designed. Testing also confirmed that Planmeca Romexis has been evaluated for hazards and that the risk has been reduced to acceptable levels.
The non-clinical bench-testing of Planmeca Romexis with predicate software versions were performed by comparison of images rendered by Planmeca Romexis and the predicate software versions. The results confirm that the software applications are equally effective in performing the essential functions and provide substantially equivalent clinical data.
## SUMMARY
Based on the intended use, product, performance, and testing information provided in this notification, the subject device has been shown to be substantially equivalent in technology, functionality, and indicated use to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.