K133320 · Sicat GmbH & Co. KG · LLZ · Feb 25, 2014 · Radiology
Device Facts
Record ID
K133320
Device Name
SICAT FUNCTION
Applicant
Sicat GmbH & Co. KG
Product Code
LLZ · Radiology
Decision Date
Feb 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
SICAT Function is a software application for the visualization and segmentation of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. It is also used as a software system to aid qualified dental professionals with the evaluation of dental treatment options. The dental professionals' planning data may be exported from SICAT Function and used as input data for CAD or Rapid Prototyping Systems.
Device Story
SICAT Function is a software application for dental professionals to visualize and segment oral-maxillofacial imaging data (CT/CBCT). It integrates optical impression data and jaw tracking data to visualize tooth geometry and mandibular movement alongside radiological data. The software aids in evaluating dental treatment options; planning data can be exported for use in CAD or rapid prototyping systems for therapeutic device manufacturing. Used in clinical settings by dental professionals, the software provides tools for 3D volume/surface rendering, measurements (length, angle), and occlusion evaluation. It benefits patients by enabling precise, data-driven dental treatment planning.
Clinical Evidence
Bench testing only. No clinical data presented. Verification and validation activities included design reviews, unit tests, code reviews, usability tests, integration tests, and system verification tests. Special bench testing verified the import, registration, and visualization of jaw motion data and the safety/effectiveness of image segmentation features.
Technological Characteristics
Pure software device; runs on Windows OS. Programming languages: C# and C++. Inputs: DICOM volume data, STL/proprietary optical impression data, and jaw motion tracking data. Features: 3D volume/surface rendering, 2D slice views (axial, coronal, sagittal, panorama, tangential, cross-sectional), and measurement tools (length, angle, grey values). Connectivity: Data export for CAD/Rapid Prototyping. Software verification follows IEC 62304 and ISO 14971 standards.
Indications for Use
Indicated for use by qualified dental professionals for the visualization and segmentation of oral-maxillofacial imaging data (CT/CBCT) to aid in the evaluation of dental treatment options and planning for CAD/Rapid Prototyping systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K133320
Page 1 of 6
# 510(k) Summary for SICAT Function
FEB 2 5 2014
# Content and format as required by section 21 CFR 807.92 ( http://www.fda.gov/ MedicalDevices/ DeviceRequlationandGuidance/ HowtoMar ketYourDevice/PremarketSubmissions/PremarketNotification510k/ucm142651.
# htm)
## 1. SUBMITTER/510(K) HOLDER
SICAT GmbH & Co. KG Brunnenallee 6 53177 Bonn Germany
Establishment Registration Number:
3006098230
| Telephone: | +49 (228) 854697-82 |
|--------------------|------------------------|
| Facsimile: | +49 (228) 854697-99 |
| Primary Contact: | Mr. Dr. Manfred Breuer |
| Secondary Contact: | Mr. Frederik Kunze |
October 17th, 2013 Date Prepared:
#### 2. DEVICE NAME AND DEVICE CLASSIFICATION
| Proprietary Name: | SICAT Function |
|-------------------------|------------------------------------------------------------------------------------|
| Common/Usual Name: | Radiological Visualization Software for Diagnosis and Dental<br>Treatment Planning |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Description: | Picture archiving and communications system |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Classification Class: | Class II Product |
#### 3. PREDICATE DEVICES
- . SICAT Implant (K103723)
- Materialise SimPlant V 12.0 with OMS Module (K033849, K053592, K081402) .
#### 4. DEVICE DESCRIPTION
SICAT Function is a pure software device.
SICAT Function is a software application for the visualization and segmentation of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or Cone Beam - CT scanners.
This information can be complemented by the imaging information from optical
{1}------------------------------------------------
5 510(k) Summary for SICAT Function
> impression systems and jaw tracking devices. The additional information about the exact geometry of the tooth surfaces and the mandibular movement can be visualized together with the radiological data.
> SICAT Function is also used as a software system to aid qualified dental professionals with the evaluation and planning of dental treatment options.
The dental professionals' treatment planning information may be exported from SICAT Function to be used as input data for the manufacturing of therapeutic devices such as oral appliances.
# 5. Intended Use
SICAT Function is a software application for the visualization and segmentation of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. It is also used as a software system to . aid qualified dental professionals with the evaluation of dental treatment options. The dental professionals' planning data may be exported from SICAT Function and used as input data for CAD or Rapid Prototyping Systems.
## 6. Device Comparison Table
The following table shows a summary of the intended use, technological characteristics, design and function of SICAT Function and the predicate devices.
| SICAT Function<br>Proposed | SICAT Implant<br>K103723 | Materialise SimPlant<br>V 12.0 with OMS<br>Module (K033849,<br>K053592, K081402) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| <b>Intended Use</b> | | | |
| SICAT Function is a<br>software application<br>for the visualization<br>and segmentation of<br>imaging information of<br>the oral-maxillofacial<br>region. The imaging<br>data originates from<br>medical scanners such<br>as CT or CBCT<br>scanners. It is also<br>used as a software<br>system to aid qualified<br>dental professionals<br>with the evaluation of<br>dental treatment<br>options. The dental<br>professionals' planning<br>data may be exported<br>from SICAT Function<br>and used as input data<br>for CAD or Rapid<br>Prototyping Systems. | SICAT Implant is a<br>software application<br>for the visualization of<br>imaging information of<br>the oral-maxillofacial<br>region. The imaging<br>data originates from<br>medical scanners such<br>as CT or DVT<br>scanners. SICAT<br>Implant is intended for<br>use as planning and<br>simulation software to<br>aid qualified dental<br>professionals in the<br>placement of dental<br>implants and the<br>planning of surgical<br>treatments. The dental<br>professionals planning<br>data may be exported<br>from SICAT Implant<br>and used as input data<br>for CAD or Rapid<br>Prototyping Systems. | The SimPlant System is<br>intended for use as a<br>software interface and<br>image segmentation<br>system for the transfer of<br>imaging information from<br>a medical scanner such as<br>a CT scanner or a<br>Magnetic Resonance<br>scanner. It is also<br>intended as pre-planning<br>software for dental<br>implant placement and<br>surgical treatment.<br>Materialise Dental's<br>SimPlant Ortho 3D<br>software is indicated for<br>use as a medical front-<br>end software that can be<br>used by medically trained<br>people for the purpose of<br>visualizing gray value<br>images. It is intended for<br>use as a software<br>interface and image<br> | |
| SICAT Function<br>Proposed | SICAT Implant<br>K103723 | Materialise SimPlant<br>V 12.0 with OMS<br>Module (K033849,<br>K053592, K081402) | |
| segmentation<br>system for the transfer of<br>imaging information from<br>a medical scanner such as<br>a CT scanner or Magnetic<br>Resonance scanner. It is<br>also used as a software<br>system for<br>simulating/evaluating<br>orthodontic treatment i.e.<br>dental bite options. | | | |
| <b>Technological Characteristics and Function</b> | | | |
| Material | pure software device | pure software device | |
| Operating<br>system | Windows | Windows | |
| Program-<br>ming<br>language | C# and C++ | C++ | |
| <b>Medical Data I/O</b> | | | |
| Volume data<br>import | DICOM | DICOM | |
| Optical<br>surface data/<br>optical<br>impression<br>import | Standard STL format<br>and proprietary SSI<br>container Format. | Standard STL format and<br>proprietary OrthoPlex<br>format. | |
| Jaw motion<br>tracking data | No | No | |
| Export of<br>data for CAD<br>and rapid<br>proto-typing | Yes | Yes | |
| <b>Medical Data Viewing</b> | | | |
| Data types<br>visualized | 3D volume data,<br>optical impressions | 3D volume data, optical<br>impressions | |
| 2D Slice<br>Views | axial, coronal, sagittal,<br>dental panorama,<br>dental tangential and<br>cross-sectional slice<br>views | axial, coronal, sagittal,<br>dental panorama, dental<br>cross-sectional slice view,<br>cephalometric | |
| 3D Volume<br>rendering | Yes | Yes | |
| 3D Surface<br>rendering | Yes | Yes | |
| View<br>manipu-<br>lating tools | Scroll, Zoom, Pan,<br>Change of orientation,<br>Brightness, Contrast | Scroll, Zoom, Pan,<br>Change of orientation,<br>Brightness, Contrast | |
| Measure-<br>ments | Length, Angle, Grey<br>values | Length, Angle, Grey<br>values | |
| Segmen-<br>tation (i.e. of<br>mandible) | No | Yes, using a segmentation<br>wizard | |
| SICAT Function<br>Proposed | SICAT Implant<br>K103723 | Materialise SimPlant<br>V 12.0 with OMS<br>Module (K033849,<br>K053592, K081402) | |
| Registration<br>of optical<br>impression<br>data to<br>volume data | Yes | Yes | Yes |
| Evaluation of<br>occlusion<br>based on<br>optical<br>impression<br>data | Only visually | No | Yes, color coded |
| Misc. functions | | | |
| Simulation of<br>orthodontic<br>procedures,<br>osteotomies<br>and<br>distractions. | No | No | Yes |
| Cephalo-<br>metric<br>analysis | No | No | Cephalometric view with<br>length and angular<br>measurements and special<br>tools for different<br>cephalometric schools. |
| Implant<br>planning | No | Yes | Yes |
| Soft tissue<br>simulation<br>and photo<br>mapping | No | No | Yes |
{2}------------------------------------------------
.
{3}------------------------------------------------
#### 5 510(k) Summary for SICAT Function
Missing features of SICAT Function compared to the predicate devices are connected to implant planning, i.e. the planning feature itself and the measurement of grey values, and to the planning of orthognatic surgery, i.e. the simulation of orthodontic procedures, osteotomies and distraction, soft tissue simulation and the cephalometric analysis. This does not impact the safety and effectiveness of SICAT Function concerning the visualization and segmentation of imaging information and the evaluation of other dental treatment options.
Additional features of SICAT Function compared to the predicate devices are the import and visualization of jaw motion data. Performance testing has been used to validate the safety and effectiveness of SICAT Function related to the import and visualization of jaw motion data.
The predicate device SICAT Implant does not provide features for the segmentation of imaging information and for jaw motion data. Otherwise the software algorithms used in SICAT Function are identical to the predicate device SICAT Implant. Performance testing has been used to validate the safety and effectiveness of the SJCAT Function segmentation features in comparison to the predicate device Simplant.
## 7. Non-Clinical Performance Testing and Verification and Validation Activities
For SICAT Function, software verification and validation activities were performed, in accordance with the following Guidances and Standards:
- NEMA PS 3.1 3.20 (2011), Digital Imaging and Communications in Medicine . (DICOM) Set. ·
{4}------------------------------------------------
- 510(k) Summary for SICAT Function 5
- . ISO 14971 Second edition 2007-03-01, Medical devices - Application of risk management to medical devices.
- IEC 62304 First edition 2006-05, Medical device software Software life cycle . processes.
- . ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD).
- ◆ IEC 62366:2007, Medical devices - Application of usability engineering to medical devices
- SMPTE Recommended Practice RP 133-1991: Specifications for Medical . Diagnostic Imaging Test Pattern for Television Monitors and Hard-copy Recording Cameras
- ◆ HIPAA 45 CFR Part 160 - General Administrative Requirements
- . HIPAA 45 CFR Part 164 - Security and Privacy
- ◆ Guidance for the Submission of Premarket Notifications for Medical Image Management Devices. Document issued on: July 27, 2000
- . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Document issued on: May 11, 2005
- Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf .
- . Software Use in Medical Devices, Document issued on: September 9, 1999
- . Device Labeling Guidance, March 8, 1991 (G91-1)
- . General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002
- . Guidance for Industry, Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, Document issued on: January 14, 2005
- . Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management: July 18, 2000
Among others the following verification and validation activities were performed:
- Design Validation/Reviews .
- Unit Tests �
- Code Reviews .
- � Usability Tests
- Integration Tests ◆
- System Verification Tests .
- User Site Tests .
Special bench testing has been performed with non-clinical data:
- to verify the correct import, registration and visualization of jaw motion data . and
- . to verify the safety and effectiveness of image segmentation features.
October 17th, 2013
{5}------------------------------------------------
5 510(k) Summary for SICAT Function
K133320
Page 6 of 6
Test reports for integration testing, system verification testing and bench testing are included with this premarket notification.
A verification and validation activities summary report provided with this premarket notification concludes that SICAT Function passed all verification and validation activities and that safety and effectiveness of the product has been demonstrated in the context of its intended use.
#### 8. Conclusion
Based on the information and supporting documentation provided in the premarket notification, SICAT Function is considered to be substantially equivalent in design, material and function to the predicate devices. It is believed to perform as well as the predicate devices for the visualization and segmentation of imaging information and the evaluation of dental treatment options. Accordingly we respectfully request the Agency to find this traditional 510(k) premarket notification to be Substantially Equivalent.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three wave-like lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2014
SICAT GMBH & CO. KG % OLAF TEICHERT THIRD PARTY REVIEWER TUV SUD AMERICA INC. 1775 OLD HWY 8 NW, STE 104 NEW BRIGHTON MN 25112
Re: K133320
Trade/Device Name: Sicat Function Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 16, 2014 Received: January 22, 2014
Dear Mr. Teichert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Mr. Teichert
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K133320
#### Device Name Sicat Function
#### Indications for Use (Describe)
SICAT Function is a software application for the visualization and segmentation of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. It is also used as a software system to aid qualified dental professionals with the evaluation of dental treatment options. The dental professionals' planning data may be exported from SICAT Function and used as input data for CAD or Rapid Prototyping Systems.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Sm.h.7)
FORM FDA 3881 (1/14)
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.