K110430 · Patterson Dental Supply, Inc. · LLZ · May 11, 2011 · Radiology
Device Facts
Record ID
K110430
Device Name
DOLPHIN IMAGING
Applicant
Patterson Dental Supply, Inc.
Product Code
LLZ · Radiology
Decision Date
May 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Dolphin Imaging software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
Device Story
Dolphin Imaging software suite for dental practices; captures, stores, presents patient images; assists in treatment planning/diagnosis. Inputs: 2D digital radiography (x-ray), TWAIN-compatible scanners, standard image files (JPEG, Bitmap). Modules: ImagingPlus (capture/organize/edit), Ceph Tracing (landmark digitization/measurement/superimposition), Treatment Simulation (VTO for orthodontic/surgical planning), Arnett/Gunson FAB (face/airway/bite analysis), McLaughlin Dental VTO (tooth position/treatment planning), Implanner (implant placement), Dolphin 3D (volumetric data rendering/clipping/implant planning), Dolphin Letter System (reporting). Operation: Clinician-operated; images manipulated/traced/simulated; results serve as guidelines for professional decision-making. Output: Enhanced images, cephalometric measurements, simulated treatment scenarios, reports. Benefits: Facilitates case presentation, treatment planning, and clinical decision-making.
Clinical Evidence
Bench testing only. Software underwent integration testing, verification testing, and Beta validation. Potential hazards evaluated and controlled.
Technological Characteristics
Software-based imaging system. TWAIN and DICOM compatible. Supports 2D and 3D image processing. Proprietary XML-based data format for storage. Features include landmark digitization, volumetric rendering, clipping operations, and simulated treatment morphing. Standalone application for dental practice environments.
Indications for Use
Indicated for use by specialized dental practices for capturing, storing, and presenting patient images, and assisting in treatment planning and case diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
VistDent at Complete
VitaDent 3D
Submission Summary (Full Text)
{0}------------------------------------------------
### 5.0 Executive Summary
Dolphin Imaging software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are dependent on the interpretation of trained and licensed practitioners.
The software contains eight major components (modules) for achieving its basic functionalities. They are ImagingPlus™, Ceph Tracing, Treatment Simulation, Arnett/Gunson FAB, McLaughlin Dental VTO, Impanner™, Dolphin 3D, and Dolphin Letter System.
ImagingPlus™ is the foundation of the Dolphin product suites. It allows the user to capture, organize, edit, print, store and present patient image records. Acquisition of 2D image data can be performed from multiple types of devices, including digital radiography devices such as x-ray machines, generic imaging devices that support the TWIN interface standard, such as flatbed scanners and standard image file formats (Jpeg, Bitmap, etc.) stored on a computer, camera memory card or similar. Once acquired, images can be cropped, rotated, enhanced and otherwise manipulated before being stored in the corresponding patient record in the Dolphin Data Storage.
Ceph Tracing allows the user to digitize landmarks on a patient's radiograph, trace cephlometric structures, view cephalometric measurements, and superimpose images for analysis. If a suitable image (lateral x-ray, frontal x-ray, or arch view) has been acquired, the cephalometric digitizing module can be utilized to define a number of landmarks to establish the locations of specific anatomical features. The inter-relational positions of these landmarks are used to render tracing lines and calculate cephalometric measurements used in diagnosing patients and planning orthodontic and/or surgical procedures. If a patient record contains multiple x-ray images for which cephalometric data has been digitized, the resulting tracings can be overlaid to indicate changes and/or differences in the anatomy.
Treatment Simulation (VTO) Module provides a tool to simulate orthodontic and surgical treatment results using Visual Treatment Objective (VTO). If a patient's record contains a lateral x-ray head film for which a set of cephalometric landmarks has been defined using the Cephalometric Digitizing Module, this image can be imported into the Treatment Simulation Module to create a simulated post-treatment view for a specific treatment scenario. The Treatment Simulation Module presents the user with a number of operations (orthodontic and surgical) which can be performed on the tracing to bring the patient's anatomy closer to the
MAY 1 1 2011
{1}------------------------------------------------
established normal values. The result of these operations is a morphed x-ray or photograph of the simulated post-treatment patient. This result is to be viewed as a guideline for the medical professional when making his or her treatment decisions, not as advice or a guaranteed outcome. The resulting image, with its altered cephalometric data, can be stored in the Dolphin Data Storage as a potential treatment scenario. A multiple of these treatment simulation images can be created and stored to serve as a decision-making tool for the medical professional when deciding on a course of treatment.
Amett/Gunson FAB Analyses performs face, ainway, bite analysis, and simulate treatment for orthodontic and surgical cases based on the methodologies of Dr. William Arnett.
The McLaughlin Dental VTO is an interactive treatment-planning and case presentation software program based on the theories of Dr. Richard P. McLaughlin, a renowned clinician, author and lecturer. It analyzes and evaluates tooth positions and dental treatment options, which assists clinicians in planning precise, quantifiable movement of dentition using clinical examination and treatment planning values.
Dolphin Implanner Module is for planning implant procedures. Using this module, simulated dental implants can be placed on a patient's lateral or panoramic x-ray images. Implant data gets written to a section in the image footer for storage.
The Dolphin 3D module contains features for generating a multitude of views of the volumetric data, including simulated x-ray views and 3D-rendered views of the volume. A number of clipping operations can be performed on the volume in order to produce a clear view of the anatomy of interest. Any images generated by the Dolphin 3D module can be stored as snapshots in the Dolphin Data Storage, or sent to the 3D report engine for generated printed output. The Dolphin 3D module contains functions for isolating specific parts of the patient's anatomy for diagnosis, and for placing simulated dental implants. Any view generated in the Dolphin 3D module can be sent to the 3D report engine for formatting and output to printer devices or storage in the Dolphin Data Storage. Reports are designed in a WYSIWYG (What You See Is What You Get) interface and allows free placement of images and text to generate the desired outcome. Images and data items (such as the patient's name, address, and other information) can be automatically imported from the Dolphin Data Storage and included on the reports can be stored in the Dolphin Data Storage using a proprietary XML-based data format and opened from the 2D Image layouts interface at a later date for further manipulation and/or output. Reports can be output
{2}------------------------------------------------
to any installed printer drive device, or as report page snapshots stored as bitmap images on the Windows clipboard or to a hard drive or other storage device.
Dolphin Letter System generates letters that include the images and diagnostic questionnaire data entered by the user can choose a pre-defined letter template or create a custom template.
Dolphin Imaging has similar intended uses as the predicate devices and has very similar technological characteristics. They all allow for the analysis, communication and media interchange of digital images acquired from a variety of acquisition devices. Dolphin Imaging is TWAIN compatible and has the similar 2D functionalities with VistDenat at Complete; DICOM compatible and has the similar 3D functionalities with VitaDent 3D for communication of images with other medical imaging devices.
Minor technological differences do not raise any new questions regarding safety or effectiveness of the device. Thus, Dolphin Imaging is substantially equivalent to the predicate devices. Please refer to the Substantial Equivalence Discussion section for detailed comparison.
The software was designed, developed, tested, and validated according to written procedures. These procedures specify individuals within the organization responsible for developing and approving product specifications, coding, testing, validating, and maintenance. Dolphin Imaging has successfully completed integration testing/verification testing and Beta validation. In addition, potential hazards have been evaluated and controlled to an acceptable level.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings and body. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the upper half of the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MAY 1 1 2011
Ms. Dawn Li, RAC Regulatory Compliance Specialist Patterson Dental Supply 1031 Mendota Heights Rd. SAINT PAUL MN 55120
Re: K110430
Trade/Device Name: Dolphin Imaging Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 11, 2011 Received: February 14, 2011
#### Dear Ms. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
バ
#### 3.0 Intended Use Statement
# 510(k) Number (if known): ¥11 0430
## Device Name: Dolphin Imaging
Dolphin Imaging software is designed for use by specialized dental practices for capturing, storing and presenting patient images and assisting in treatment planning and case diagnosis. Results produced by the software's diagnostic and treatment planning tools are diagnotibly on the interpretation of trained and licensed practitioners.
Over-The-Counter Use Prescription Use__ × AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of of
Mary's Patt
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnos
Division of Radiological Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
510K.
6/30
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.