K193275 · Shenzhen Xft Medical Limited · IPF · Apr 11, 2020 · Physical Medicine
Device Facts
Record ID
K193275
Device Name
Nerve and Muscle Stimulator
Applicant
Shenzhen Xft Medical Limited
Product Code
IPF · Physical Medicine
Decision Date
Apr 11, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Over-The-Counter Use: TENS[ (Program 1~9, Program 15 (Frequency < 5Hz or > 15Hz)]: XFT-2000 is used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities. XFT-2000 is also intended for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. NMES[ (Program 10~14, Program 15(5Hz≤Frequency≤50Hz)]: XFT-2000 is used to stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases. XFT-2000 is also intended to temporarily increase local blood circulation in health muscles of lower extremities. Prescription Use: XFT-2000 is intended for the following use: - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy - Increasing local blood circulation - Muscle re-education - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis - Maintaining or increasing range of motion - Adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
Device Story
XFT-2000 is a dual-channel nerve and muscle stimulator utilizing NMES and TENS principles. Device inputs include user-selected programs (15 total) via interface; outputs are electrical pulses delivered through self-adhesive electrodes. Powered by DC adapter or 3 AAA batteries. Used in home or clinical settings by patients or clinicians. Device transforms inputs into biphasic, rectangular pulsed waveforms to stimulate nerves or muscles. Output affects clinical decision-making by providing symptomatic pain relief, muscle strengthening, or post-surgical rehabilitation. Benefits include pain management, improved muscle performance, and prevention of venous thrombosis. Device is neither life-supporting nor implantable.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance requirements for nerve/muscle stimulators (IEC 60601-2-10), and biocompatibility (ISO 10993-5/10).
Technological Characteristics
Materials: Silicone and ABS. Sensing/Actuation: NMES/TENS electrical stimulation. Energy: DC adapter or 3 AAA batteries. Dimensions: 114x66x20mm. Connectivity: Standalone. Software: Microprocessor-controlled with 15 programs. Safety: Automatic overload/no-load trip, automatic shut-off, patient pause control. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993.
Indications for Use
Indicated for OTC use for temporary relief of pain from sore/aching muscles (strain/exercise/household), chronic intractable pain, and arthritis pain. Indicated for NMES to improve muscle tone/firmness/strength in arms, abdomen, legs, buttocks, and increase local blood circulation in healthy lower extremity muscles. Indicated for prescription use for muscle spasm relaxation, disuse atrophy prevention, muscle re-education, post-surgical calf stimulation to prevent venous thrombosis, range of motion maintenance, and adjunctive management of post-surgical/post-traumatic acute pain.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 11, 2020
Shenzhen XFT Medical Limited % Field Fu Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. Room 1122#, International Mayors Communication Centre NO. 55 Shizhou middl Shenzhen, Guangdong, China Re: K193275 Trade/Device Name: Nerve and Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: January 6, 2020 Received: January 13, 2020
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193275
Device Name Nerve and Muscle Stimulator
Indications for Use (Describe) Over-The-Counter Use: TENS[ (Program 1~9, Program 15 (Frequency < 5Hz or > 15Hz)]: XFT-2000 is used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities.
XFT-2000 is also intended for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
NMES[ (Program 10~14, Program 15(5Hz≤Frequency≤50Hz)]:
XFT-2000 is used to stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
XFT-2000 is also intended to temporarily increase local blood circulation in health muscles of lower extremities.
Prescription Use:
XFT-2000 is intended for the following use:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
- Adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
### 5.1 Administrative Information
| Date of Summary<br>prepared | Oct. 23, 2019 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Company title: Shenzhen XFT Medical Limited. |
| | Company address:<br>Room 203, Building 1, Biomedicine Innovations<br>Industrial Park, #14 Jinhui Road, Pingshan New<br>District, Shenzhen, China.<br>Contact person: Cindy Peng<br>Phone: +86 755 29888818 |
| | Fax: +86-0755-28312625 |
| | E-mail: xftrs2@xft.cn |
| Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: 1713A, 17th Floor, Block A,<br>Zhongguan Times Square, Liuxian Avenue, Xili |
| | Town, Nanshan District, Shenzhen, Guangdong<br>Province, China. |
| | Contact person: Mr. Field Fu |
| Image: logo | E-Mail: field@cefda.com; |
| Establishment<br>registration number | |
## 5.2 Device Information
| Type of 510(k)<br>submission: | Traditional |
|-------------------------------|----------------------------------|
| Trade Name: | Nerve and Muscle Stimulator |
| Model: | XFT-2000 |
| | Stimulator, Muscle, Powered; |
| Classification<br>name: | Stimulator, Nerve, |
| | Transcutaneous, For Pain Relief. |
{5}------------------------------------------------
| Review Panel: | Physical Medicine; Neurology. |
|--------------------|-------------------------------|
| Product Code: | IPF; GZJ. |
| Device Class: | II |
| Regulation Number: | 890.5850; 882.5890. |
### 5.3 Predicate Device Information
| Sponsor: | Rio-Medical Research Ltd. |
|----------------|-----------------------------|
| Device: | MediStim XP, Type 281; |
| 510(K) Number: | K082011 |
| Sponsor: | JKH Health Co., Ltd. |
| Device: | Electronic Pulse Stimulator |
| 510(K) Number: | K162517 |
### 5.4 Device Description
Nerve and Muscle Stimulator (model: XFT-2000) is based on the principles of neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS).The instrument has two independent channels, easy to operate and use. It can relieve muscle spasm, prevent or delay disuse muscle atrophy, promote local blood circulation, promote muscle function recovery, stimulate calf muscle after surgery to prevent venous thrombosis and maintain or increase the range of activity, and assist in the management of post-operative and post-traumatic acute pain problems.
Nerve and Muscle Stimulator (model: XFT-2000) is powered by DC adapter or 3 dry AAA batteries, and used together with self-adhesive Electrode.
The device is for both prescription and over the counter. It is neither for lifesupporting nor for implanting. It does not contain any drug or biological product and it does not need to be sterilized.
## 5.5 Intended Use/ Indications for Use
### Over-The-Counter Use:
TENS[ (Program 1~9, Program 15 (Frequency<5Hz or >15Hz)]: XFT-2000 is used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities.
{6}------------------------------------------------
XFT-2000 is also intended for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
NMES[ (Program 10~14, Program 15(5Hz≤Frequency≤50Hz)]: XFT-2000 is used to stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
XFT-2000 is also intended to temporarily increase local blood circulation in health muscles of lower extremities.
#### Prescription Use:
XFT-2000 is intended for the following use:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
- Adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
## 5.6 Technological characteristics of the subject device compared to the predicate device
See next pages.
{7}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Predicate Device | Predicate Device | Remarks | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|------|
| 510(k) Number | K193275 | K082011 | K162517 | | |
| Manufacturer | Shenzhen XFT<br>Medical Limited | Rio-Medical<br>Research Ltd | JKH USA, LLC | | |
| Device<br>type/model | XFT-2000 | Medistim XP Type 281; | PL-029K12; | / | |
| Intended use/<br>Indication<br>for<br>use | Neuromuscular<br>Electrical Stimulation<br>(NMES) for relaxation<br>of muscle spasms,<br>prevention or<br>retardation of disuse<br>atrophy, increasing<br>local blood circulation,<br>muscle re-education,<br>immediate post-<br>surgical stimulation of<br>calf muscles to prevent<br>venous thrombosis<br>and maintaining or<br>increasing range of<br>motion.<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS) for<br>an adjunctive<br>treatment in the<br>management of post-<br>surgical and post-<br>traumatic acute pain<br>problems. | Medistim XP delivers<br>stimulation based on the<br>principles of<br>Neuromuscular Electrical<br>NerveStimulation<br>(NMIES) and also<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS).<br>Theindications for use<br>are as follows:<br>Neuromuscular Electrical<br>Stimulation for relaxation<br>ofmuscle spasms,<br>prevention or retardation<br>of disuse atrophy,<br>increasing local blood<br>circulation,muscle re-<br>education, immediate<br>post-surgical stimulation<br>of calf muscles to<br>prevent venousthrombosis and<br>maintaining or increasing<br>range of motion.<br>Transcutaneous<br>Electrical<br>NerveStimulation<br>(TENS) for an adjunctive<br>treatment in the<br>management of post-<br>surgical and post-<br>traumatic acute pain<br>problems. | TENS (Modes 1, 2, 4,<br>5, 6, 8):<br>To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back, arm, and leg,<br>due to strain from<br>exercise or normal<br>household and work<br>activities.<br>It is also intended for<br>symptomatic relief and<br>management of<br>chronic, intractable<br>pain and relief of pain<br>associated with<br>arthritis.<br>PMS (also called EMS,<br>Modes 1, 3, 7):<br>To stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance.<br>To be used for the<br>improvement of muscle<br>tone and firmness, and<br>for strengthening<br>muscles in the arms,<br>abdomen, legs, and<br>buttocks. Not intended<br>for use in any therapy<br>or for the treatment of | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device | Predicate Device | Remarks | |
| | | | any medical conditions<br>or diseases.<br>t is also intended to<br>temporarily increase<br>local blood circulation<br>in the healthy muscles<br>of lower extremities. | | |
| Prescription<br>or<br>OTC | Prescription and OTC | Prescription | OTC | SE | |
| Electrode | 5cm round;K132588 | 5×5cm;7cm round;<br>5cm×9cm etc.<br>K970426,<br>K874469,K965194 | unknown | SE | |
| Waveform | Pulsed,Symmetrical<br>balanced biphasic<br>wave (rectangular) | Pulsed, Symmetrical<br>Biphasic, Rectangular<br>with interphase interval | Biphasic, Rectangular | SE | |
| Performance | Compliance with IEC<br>60601-2-10 | Compliance with IEC<br>60601-2-10 | Compliance with IEC<br>60601-2-10 | Same | |
| Biocompatibility | All the patient<br>contacting materials<br>are compliance with<br>ISO 10993-1/-5/-10 | All the patient contacting<br>materials are<br>compliance with ISO<br>10993-1/-5/-10 | All the patient<br>contacting materials<br>are compliance with<br>ISO 10993-1/-5/-10 | Same | |
| Electrical Safety | Compliance with IEC<br>60601-1 | Compliance with IEC<br>60601-1 | Compliance with IEC<br>60601-1 | Same | |
| EMC | Compliance with IEC<br>60601-1-2 | Compliance with IEC<br>60601-1-2 | Compliance with IEC<br>60601-1-2 | Same | |
| Connecting<br>Safety | 21CFR 898 | 21CFR 898 | 21CFR 898 | Same | |
| Parameter | Subject Device | Predicate Device | Predicate Device | Remarks | |
| 510(k) Number | K193275 | K082011 | K162517 | / | |
| Device Name and<br>Model | Nerve and Muscle<br>Stimulator, XFT-<br>2000 | MediStimi XP, Type<br>281 | JKH Stimulator:<br>PL-029K12, | / | |
| Manufacturer | Shenzhen XFT<br>Medical Limited | Rio-Medical<br>Research Ltd | JKH USA, LLC | / | |
| Power Source(s) | DC4.5V, 3 × AAA<br>batteries;<br>DC Adapter. | 9V Battery (type<br>6LR61) | Rechargeable or<br>non-rechargeable<br>battery | Note 01 | |
| ---Method of Line Current<br>Isolation | Battery Supply, or<br>DC Adapter | Battery Supply | Battery Supply | Note 01 | |
| ---Patient Leakage Current:<br>Normal Condition (µA) | N/A | N/A | N/A | / | |
| ---Patient Leakage Current:<br>Single Fault Condition (µA) | N/A | N/A | N/A | / | |
| Number of Output<br>Modes(programs) | 15 | 9 | 8 | Note 02 | |
| Number of Output Channels: | 2 | 2 | 1 | SE | |
| ---Synchronous or Alternating? | Both | Synchronous(Multip<br>lexed) | N/A | SE | |
| ---Method of Channel Isolation | Transistor | Transistor | N/A | SE | |
| Regulated Current or<br>Regulated Voltage? | Regulated Current | Regulated Current | Regulated Voltage | SE | |
| Software/Firmware/Microproce<br>ssor Control? | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.