Electronic Pulse Stimulator Model: PL-029K12 and PL-029K13
Applicant
Jkh Health Co., Ltd.
Product Code
NUH · Neurology
Decision Date
Apr 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TENS (Modes 1, 2, 4, 5, 6, 8) To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities. It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. PMS (also called EMS, Modes 1, 3, 7) To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases. It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities. Heating Mode Temporary relief of minor aches and pains.
Device Story
Electronic Pulse Stimulator delivers electrical pulses to body areas via electrodes for TENS and PMS (EMS) therapy; includes optional heating mode. Device features multiple preset modes with adjustable pulse frequencies, intensities, and timers. Operated via physical buttons or optional wireless remote/Bluetooth app. Heating mode provides automatic thermal heat to skin (max 43 °C). Used in home/OTC settings by patients for pain management and muscle performance improvement. Output affects muscle stimulation and local blood circulation; provides symptomatic pain relief. Device is portable and compact; powered by rechargeable battery.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 verified. Skin temperature testing confirmed max temperature of 41 °C during operation.
Technological Characteristics
Portable electrical stimulator; rechargeable battery powered. Provides TENS/PMS via electrodes and thermal heat. Max output temperature 43 °C. Connectivity via Bluetooth/wireless remote. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10.
Indications for Use
Indicated for adults seeking temporary relief of pain from sore/aching muscles (shoulder, waist, back, arm, leg) due to strain, chronic intractable pain, or arthritis. Also indicated for healthy individuals to improve muscle tone, firmness, and performance in arms, abdomen, legs, and buttocks, and to increase local blood circulation in lower extremities. Heating mode indicated for minor aches and pains.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
JKH Health Co., Ltd. % Bill Quangin Dai 1142 S. Diamond Bar Blvd. #861 Diamond Bar, CA 91765
Re: K162517
Trade/Device Name: Electronic Pulse Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX, NYN, IRT Dated: March 14, 2017 Received: March 21, 2017
Dear Dr. Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Carlos L. Pena -S) A
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162517
Device Name Electronic Pulse Stimulator
#### Indications for Use (Describe)
TENS (Modes 1, 2, 4, 5, 6, 8)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
PMS (also called EMS, Modes 1, 3, 7)
To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.
Heating Mode Temporary relief of minor aches and pains.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: JKH Health Co., Ltd. (Previous name: Shenzhen Jingkehui Electronic Co., Ltd.) Address: 4-5F, Building 12, Hengmingzhu Industrial Park, Xinqiao Tongfuyu Industrial Area, Shajing, Baoan, Shenzhen, China Contact Person: Pu Jiang Tel: +86-755-27926589 Fax: +86-755-29970323 Email: bill@JKHhealth.com Date of Preparation: 08/30/2016
## 2. Subiect Device
Trade/Device Name: Electronic Pulse Stimulator Common Name: Transcutaneous Electrical Nerve Stimulation (TENS) unit and Powered Muscle Stimulation (PMS) unit Regulation Medical Specialty: Neurology Review Panel: Neurology Product Code: NUH, NGX, NYN, IRT Regulation Number: 21 CFR 882.5890 Device Class: II Use: Over-The-Counter (OTC)
## 3. Predicate device
Predicate Device: Electronic Pulse Stimulator 510(k) Number: K153520 Clearance Date: Mav 05, 2016 Submitter: JKH Health Co., Ltd. (Previous name: Shenzhen Jingkehui Electronic Co., Ltd.)
Predicate Device: ezFit Digital Heating TENS Clearance Date: July 16, 2007 510(k) Number: K070299 Submitter: ezFit Technology, Inc.
## 4. Description of Subject Device
Electronic Pulse Stimulator delivers electric pulses generated to the user's body areas such as the back neck and foot through the electrodes. The portable and compact device has multiple modes of different pulse frequencies, covering TENS and PMS that is also called Electrical Muscle Stimulation (EMS). It includes operating elements of ON/OFF button, intensity increase button, intensity decrease button, mode selection button, and/or timer selection button, and could be attached to electrodes. In addition, the device may also provide heat/temperature. While used in the heating mode, the device is coupled with electronically controlled electrodes to provide automatic thermal heat to the skin with the maximum temperature of 43 °C.
The device could be easily operated through its buttons to manually realize its functions, such as
{4}------------------------------------------------
turning on/off, increasing/decreasing intensity, changing mode/timer, and providing heat/temperature if needed. The optional wireless control via a remote or Bluetooth APP could provide a secondary operation way to the user, who could be able to wirelessly realize the functions mentioned above.
The electrodes cleared include the electrode pads and electrode garments, which could be packaged separately and/or together with the subject device.
## 5. Indications for Use
#### TENS (Modes 1, 2, 4, 5, 6, 8)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
## PMS (also called EMS, Modes 1, 3, 7)
To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.
Heating Mode
Temporary relief of minor aches and pains.
## 6. Summary of Substantial Equivalence
The following comparison Table 1 summarizes the comparison between the subject device and predicate device, indicating the intended use and technical characteristics of the subject device are substantially equivalent to those of the predicate device.
| | Subject Device | Subject Device | Predicate Device | Predicate Device |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K162517 | K162517 | K153520 | K070299 |
| Submitter/Manufacturer | JKH Health Co., Ltd. | JKH Health Co., Ltd. | JKH Health Co., Ltd. | ezFit Technology, Inc. |
| Device Name/Model | PL-029K12 | PL-029K13 | PL-029K6 | ezFit Digital Heating TENS |
| Intended Use | TENS Mode<br>To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist, back,<br>arm, and leg, due to strain<br>from exercise or normal<br>household and work<br>activities.<br><br>It is also intended for<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief of<br>pain associated with arthritis. | TENS Mode<br>To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist, back,<br>arm, and leg, due to strain<br>from exercise or normal<br>household and work<br>activities.<br><br>It is also intended for<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief of<br>pain associated with arthritis. | TENS<br>To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist, back,<br>arm, and leg, due to strain<br>from exercise or normal<br>household and work<br>activities.<br><br>It is also intended for<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief of<br>pain associated with arthritis. | For Transcutaneous<br>Electrical Nerve<br>Stimulation, ezFit Digital<br>Heating TENS (Model No.:<br>HR-661/UC-101) is<br>intended for<br>Symptomatic relief and<br>management of chronic<br>intractable pain.<br><br>For powered heating<br>therapy, ezFit Digital<br>Heating TENS (Model No.:<br>HR-661/UC-101) is<br>intended for |
Table 1. Device comparison
{5}------------------------------------------------
| | PMS Mode<br>To stimulate healthy muscles<br>in order to improve and<br>facilitate muscle<br>performance. To be used for<br>the improvement of muscle<br>tone and firmness, and for<br>strengthening muscles in the<br>arms, abdomen, legs, and<br>buttocks. Not intended for<br>use in any therapy or for the<br>treatment of any medical<br>conditions or diseases.<br>It is also intended to<br>temporarily increase local<br>blood circulation in the<br>healthy muscles of lower<br>extremities. | PMS Mode<br>To stimulate healthy muscles<br>in order to improve and<br>facilitate muscle<br>performance. To be used for<br>the improvement of muscle<br>tone and firmness, and for<br>strengthening muscles in the<br>arms, abdomen, legs, and<br>buttocks. Not intended for<br>use in any therapy or for the<br>treatment of any medical<br>conditions or diseases.<br>It is also intended to<br>temporarily increase local<br>blood circulation in the<br>healthy muscles of lower<br>extremities.<br><br>Heating Mode<br>Temporary relief of minor<br>aches and pains | PMS<br>To stimulate healthy muscles<br>in order to improve and<br>facilitate muscle<br>performance. To be used for<br>the improvement of muscle<br>tone and firmness, and for<br>strengthening muscles in the<br>arms, abdomen, legs, and<br>buttocks. Not intended for<br>use in any therapy or for the<br>treatment of any medical<br>conditions or diseases.<br>It is also intended to<br>temporarily increase local<br>blood circulation in the<br>healthy muscles of lower<br>extremities. | Temporary relief of minor<br>aches and pains and muscle<br>spasms |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Prescription or OTC | OTC | OTC | OTC | Prescription |
| Power Source(s) | Rechargeable battery | Rechargeable battery | Rechargeable battery | Rechargeable battery |
| Functions and design | Electrical stimulation | Electrical stimulation and<br>heat | Electrical stimulation | Electrical stimulation and<br>heat |
| Heating Setting | N/A | Low and high | N/A | Adjustable (36-42 °C) |
| Maximum temperature setting | N/A | 43 °C | N/A | 42 °C |
| Maximum output voltage<br>(Volts +/- 20%) at 500Ω | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 36.4<br>Mode 3: 47.6<br>Mode 4: 57.6<br>Mode 5: 29.6<br>Mode 6: 29.6<br>Mode 7: 40.8<br>Mode 8: 24.0 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 31.2<br>Mode 3: 46.0<br>Mode 4: 42.0<br>Mode 5: 27.6<br>Mode 6: 27.6<br>Mode 7: 40.8<br>Mode 8: 23.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 39.2<br>Mode 3: 64<br>Mode 4: 56.8<br>Mode 5: 33.2<br>Mode 6: 31.6 | Undisclosed |
| Maximum output voltage<br>(Volts +/- 20%) at 2kΩ | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 80.8<br>Mode 3: 96.0<br>Mode 4: 93.6<br>Mode 5: 66.4<br>Mode 6: 66.4<br>Mode 7: 86.4<br>Mode 8: 53.6 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 68.0<br>Mode 3: 90.4<br>Mode 4: 68.8<br>Mode 5: 60.0<br>Mode 6: 60.0<br>Mode 7: 84.0<br>Mode 8: 50.4 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 82.4<br>Mode 3: 84<br>Mode 4: 79.2<br>Mode 5: 70.4<br>Mode 6: 67.2 | Undisclosed |
| Maximum output voltage<br>(Volts +/- 20%) at 10kΩ | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 134<br>Mode 3: 132<br>Mode 4: 108<br>Mode 5: 126<br>Mode 6: 126<br>Mode 7: 129<br>Mode 8: 105 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 118<br>Mode 3: 124<br>Mode 4: 78.4<br>Mode 5: 115<br>Mode 6: 115<br>Mode 7: 124<br>Mode 8: 99.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 129<br>Mode 3: 120<br>Mode 4: 84.8<br>Mode 5: 121<br>Mode 6: 124 | Undisclosed |
| Maximum output current<br>(mA +/- 20%) at 500Ω | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 72.8<br>Mode 3: 95.2<br>Mode 4: 115.2<br>Mode 5: 59.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 62.4<br>Mode 3: 92.0<br>Mode 4: 84.0<br>Mode 5: 55.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 78.4<br>Mode 3: 128<br>Mode 4: 113.6<br>Mode 5: 64.4 | Undisclosed |
{6}------------------------------------------------
| | Mode 6: 59.2<br>Mode 7: 81.6<br>Mode 8: 48.0 | Mode 6: 55.2<br>Mode 7: 81.6<br>Mode 8: 46.4 | Mode 6: 63.2 | | | | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------|--------------------------------------------------------------------------------------|-------------|
| Maximum output current<br>(mA +/- 20%) at 2KΩ | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 40.4<br>Mode 3: 48.0<br>Mode 4: 46.8<br>Mode 5: 33.2<br>Mode 6: 33.2<br>Mode 7: 43.2<br>Mode 8: 26.8 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 34.0<br>Mode 3: 45.2<br>Mode 4: 34.4<br>Mode 5: 30.0<br>Mode 6: 30.0<br>Mode 7: 42.0<br>Mode 8: 25.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 41.2<br>Mode 3: 42<br>Mode 4: 39.6<br>Mode 5: 35.2<br>Mode 6: 33.6 | Undisclosed | | | |
| Maximum output current<br>(mA +/- 20%) at 10KΩ | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 13.4<br>Mode 3: 13.2<br>Mode 4: 10.8<br>Mode 5: 12.6<br>Mode 6: 12.6<br>Mode 7: 12.9<br>Mode 8: 10.5 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 11.8<br>Mode 3: 12.4<br>Mode 4: 7.84<br>Mode 5: 11.5<br>Mode 6: 11.5<br>Mode 7: 12.4<br>Mode 8: 9.92 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 12.9<br>Mode 3: 12<br>Mode 4: 8.5<br>Mode 5: 12.1<br>Mode 6: 12.4 | Undisclosed | | | |
| Pulse period (mSec) | 6.4~840 | 5.6~806 | 10~833 | Undisclosed | | | |
| Frequency (Hz) | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 62.5<br>Mode 3: 12.8~54.3<br>Mode 4: 1.19<br>Mode 5: 104.1<br>Mode 6: 104.1<br>Mode 7: 19.8<br>Mode 8: 156.2 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 73.5<br>Mode 3: 13.7~59.5<br>Mode 4: 1.24<br>Mode 5: 104.1<br>Mode 6: 104.1<br>Mode 7: 20.8<br>Mode 8: 178.5 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 69.4<br>Mode 3: 13.0~52.1<br>Mode 4: 1.2<br>Mode 5: 96.2<br>Mode 6: 96.2 | Undisclosed | | | |
| Maximum Phase charge<br>(µC) at 500Ω | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 14.6<br>Mode 3: 19.0<br>Mode 4: 23.0<br>Mode 5: 11.8<br>Mode 6: 11.8<br>Mode 7: 16.3<br>Mode 8: 9.6 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 11.5<br>Mode 3: 16.9<br>Mode 4: 15.5<br>Mode 5: 10.2<br>Mode 6: 10.2<br>Mode 7: 15.0<br>Mode 8: 8.54 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 15.1<br>Mode 3: 25.6<br>Mode 4: 18.2<br>Mode 5: 12.8<br>Mode 6: 10.1 | Undisclosed | | | |
| Maximum current density<br>(mA/cm²) at 500Ω | Mode 1:This mode cycles<br>the following modes<br>Mode 2: 2.02<br>Mode 3: 2.64<br>Mode 4: 3.20<br>Mode 5: 1.64<br>Mode 6: 1.64<br>Mode 7: 3.26<br>Mode 8: 1.92 | Mode 1:This mode cycles<br>the following modes<br>Mode 2: 2.23<br>Mode 3: 3.29<br>Mode 4: 3.00<br>Mode 5: 1.97<br>Mode 6: 1.97<br>Mode 7: 2.91<br>Mode 8: 1.66 | Mode 1:This mode cycles<br>the following modes<br>Mode 2: 2.18<br>Mode 3: 3.56<br>Mode 4: 3.16<br>Mode 5: 1.84<br>Mode 6: 1.76 | Undisclosed | | | |
| Maximum average power<br>density (mW/cm²) at 500Ω | Mode 4: 0.04 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 0.92<br>Mode 3: 0.32~1.37 | Mode 4: 0.03 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 0.94<br>Mode 3: 0.38~1.65 | Mode 4: 0.03 | Mode 1: This mode cycles<br>the following modes<br>Mode 2: 1.14<br>Mode 3: 0.64~2.56 | Undisclosed |
| Mode 5: 1.01 | Mode 5: 1.04 | Mode 5: 1.13 | | | | | |
| Mode 6: 1.01 | Mode 6: 1.04 | Mode 6: 0.85 | | | | | |
| Mode 7: 0.53 | Mode 7: 0.46 | | | | | | |
| Mode 8: 1.44 | Mode 8: 1.26 | | | | | | |
## 7. Substantial Equivalence
As shown in the above comparison Table 1, the six modes of the predicate PL-029K6 (K153520) are the first six preset modes in both of the subject devices PL-029K12 and PL-029K13. The seventh and eighth preset modes of PL-029K12 and PL-029K13 have the technical characteristics (such as
{7}------------------------------------------------
maximum output voltage, maximum output current, maximum current density, and maximum average power density) almost the same as the six modes of the predicate device. Also, the changes to the display on the subject device do not affect the safety or effectiveness. Accordingly, the subject devices PL-029K12 and PL-029K13 are substantially equivalent to the predicate device PL-029K6, and the same intended use of PL-029K6 as TENS and PMS could apply to PL-029K12 and PL-029K13.
In addition to the TENS and PMS modes, the subject device PL-029K13 also has a heating mode, and is compared to the predicate device that could deliver electrical stimulation and heat separately and simultaneously in K070299. Similar to the predicate device, the subject device is used for the electrical stimulation or heat; consult with your physician before using electrical stimulation and heat simultaneously. The subject device delivers heat through the same technology as the predicate device, and is viewed as substantially equivalent to the predicate device of the exact same technology and substantially equivalent indications for use. The differences that exist between the subject and predicate devices are insignificant in the terms of safety or effectiveness.
Design and Technology - The basic design and technology of providing electrical stimulation and heat is the same or similar.
Performance and Specifications - The subject device has similar electrical stimulation and heat performance and specifications to the predicate device.
Indications - The indications are similar.
Prescriptive - The predicate K070299 is for prescription, while the remaining predicate and subject devices are for OTC.
As demonstrated, the differences between the subject and predicate devices do not affect the intended use or alter the fundamental technology of the device. There are no new safety or effectiveness issues concerning the subject device, which offers substantially equivalent technical specifications, features, intended use, safety, and effectiveness as the predicate device.
The same electrode pads and garments cleared from the predicate K153520 could apply to the subject device. Also, other-shape electrode pads and garments with the same components, such as conductive shoes, socks, belts, wraps, , will not change the biocompatibility or safety. Some electrodes used have an additional heating layer inside, compared to the K153520 cleared electrode. This heating layer isolated inside the electrode could not be touched by the user and does not raise any biocompatibility issue. Concerns of the safe and proper use of the electrodes and other accessories have been fully addressed through details in the Labeling.
# 8. Non-Clinical Tests Performed
The subject device does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.
- (a) IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance".
- (b) IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
- (c) IEC 60601-2-10 "Medical electrical equipment Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators".
In addition to the compliance of voluntary standards, the verification of software used in the subject
{8}------------------------------------------------
device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The electrode pads, electrode garments, and other accessories of the subject device also meet the requirement of safety.
The skin temperatures were measured at Mode 8 with the largest frequency of 160 Hz among the eight preset modes, the highest Intensity Level 20, and the highest Heat Level 2 for 60 min. The results showed the subject device reached the highest skin temperature of about 41 °C in 20 min, and then kept stable.
## 9. Conclusion
The tests performed and the comparison of technical characteristics and intended use demonstrate the subject device is substantially equivalent to the predicate device. Therefore, the subject device is as safe and effective as the predicate device that has been legally marketed in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.