Both Meditens XP and Medistim XP are intended for prescription use only. Meditens XP delivers stimulation based on the principles of transcutaneous electrical nerve stimulation (TENS). It is indicated for the symptomatic relief and management of chronic intractable pain. It is also an adjunctive treatment in the management of post-surgical and post-traumatic pain. The device has no curative value and should only be used in conjunction with medical supervision. Medistim XP delivers stimulation based on the principles of Neuromuscular Electrical Nerve Stimulation (NMES) and also Transcutaneous Electrical Nerve Stimulation (TENS). The indications for use are as follows: Neuromuscular Electrical Stimulation for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis and maintaining or increasing range of motion. Transcutaneous Electrical Nerve Stimulation (TENS) for an adjunctive treatment in the management of post-surgical and posttraumatic acute pain problems.
Device Story
Portable, two-channel, battery-operated stimulators (Meditens XP/Medistim XP); deliver TENS and/or NMES via adhesive electrodes. Units are identical hardware with different bezels; pre-configured software programs (5 for Meditens, 9 for Medistim). Operated by patient under medical supervision. Device provides electrical stimulation to nerves/muscles to manage pain, prevent atrophy, or improve circulation. Safety features included to prevent misuse. No user/clinician access to software.
Clinical Evidence
No clinical studies submitted. Bench testing only. Compliance with EN 60601-1-2:2001 (EMC), EN 60601-1 (General Safety), and IEC 60601-2-10 (Nerve/Muscle Stimulator Safety) demonstrated.
Technological Characteristics
Portable, two-channel, 9V battery-powered stimulator. Delivers TENS/NMES. Includes lead-wires and adhesive electrodes. Software is pre-configured and locked. Complies with EN 60601-1, IEC 60601-2-10, and EN 60601-1-2.
Indications for Use
Indicated for patients requiring symptomatic relief of chronic intractable pain, post-surgical/post-traumatic pain (TENS), or NMES for muscle spasm relaxation, disuse atrophy prevention, blood circulation improvement, muscle re-education, venous thrombosis prevention, and range of motion maintenance. Prescription use only.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for BMR. The logo consists of the letters "BMR" in a bold, sans-serif font. To the left of the letters is a stylized graphic that resembles a gear or a stylized letter "C". The logo is black and white.
# Research Ltd.
Parkmore Business Park West, Galway, Ircland Tel: +353 (0)91 774300 - Fax: +353 (0)91 774301
#### 510(k) Summary
K082011
NOV 28 2008
This summary of 510(k) is being NOV 2 8 2008 807.92.
## Contact Details:
| Name: | Anne-Marie Keenan |
|------------|-------------------------------------------------------------------------------|
| Address: | Bio-Medical Research Ltd.,<br>Parkmore Business Park, West<br>Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774301 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | 11th July 2008 |
## Device Information:
| Trade Name: | Meditens XP, Type 458 (Device 1) |
|----------------------|-----------------------------------------------------------------------------------|
| Common Name: | Muscle Stimulator |
| Classification Name: | Stimulator, Nerve, Transcutaneous, For Pain Relief |
| Regulation Number: | 21 CFR 882.5890 |
| Product Code: | GZJ |
| Equivalent Device: | Meditens Plus, Type 290 |
| Manufacturer: | Bio-Medical Research Ltd |
| 510(k) No: | K014020, July 2002 |
| Trade Name: | Medistim XP, Type 281 (Device 2) |
| Common Name: | Muscle Stimulator |
| Classification Name: | Stimulator, Muscle, Powered<br>Stimulator, Nerve, Transcutaneous, For Pain Relief |
| Regulation Number: | 21 CFR 890.5850, 21 CFR 882.5890 |
| Product Code: | IPF, GZJ |
| Equivalent Device: | Medistim Plus, Type 291 |
| Manufacturer: | Bio-Medical Research Ltd |
| 510(k) No: | K014019, July 2002 |
{1}------------------------------------------------
#### Description Of Devices:
Meditens XP and Medistim XP are portable two-channel battery operated stimulators. They consist of a Meditens XP or Medistim XP unit, instructions for use, 9-volt battery, two connecting lead-wires for connecting the stimulator unit to the electrodes, adhesive electrodes and a device box. The units for Meditens XP and Medistim XP are identical and can be identified by the use of different bezels. The software for each device is pre-configured during manufacturing and may not be accessed by either the user or clinician. Meditens XP offers a total of five treatment programs and Medistim XP offers nine treatment programs. Safety features are incorporated into the device to reduce the possibility of misuse.
#### Intended Use/Indications for Use:
Both Meditens XP and Medistim XP are intended for prescription use only.
Meditens XP delivers stimulation based on the principles of transcutaneous electrical nerve stimulation (TENS). It is indicated for the symptomatic relief and management of chronic intractable pain. It is also an adjunctive treatment in the management of post-surgical and post-traumatic pain. The device has no curative value and should only be used in conjunction with medical supervision.
Medistim XP delivers stimulation based on the principles of Neuromuscular Electrical Nerve Stimulation (NMES) and also Transcutaneous Electrical Nerve Stimulation (TENS). The indications for use are as follows: Neuromuscular Electrical Stimulation for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis and maintaining or increasing range of motion. Transcutaneous Electrical Nerve Stimulation (TENS) for an adjunctive treatment in the management of post-surgical and posttraumatic acute pain problems.
#### Technological Characteristics:
There are no new technological characteristics that could affect safety or effectiveness of the Meditens XP or Medistim XP devices. The new devices have technological features as the predicate devices.
{2}------------------------------------------------
### Clinical and Non-Clinical Tests:
Bio-Medical Research Ltd ("BMR") has over 30 years experience in the research, design, manufacture and marketing of medical grade products for both muscle strengthening and pain relief. Bio-Medical Research I.td. complies with 21 CFR 820 and is registered to I.S. EN ISO 13485:2003, Mcdical Device Quality Management System for the design, manufacture and distribution of electro-medical devices. No clinical studies have been submitted as part of this premarket notification.
Meditens XP and Medistim XP comply with the following international safety standards:
- C EN 60601-1-2:2001 Medical clectrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and Tests.
- ロ EN 60601-1 (1990) + A1 (1993) + A2 (1995) +A13 (1996) + Corrigenda (July 1994) -Medical Electrical Equipment - Part 1: General Requirements for Safety
- O IEC 60601-2-10, 1st ed., 1987 Medical Electrical Equipment Part 2: Particular Requirements For The Safety Of Nerve And Muscle Stimulators.
A risk management plan has been carried out to I.S. EN ISO 14971 2001 AMD 1 2003.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a sans-serif font. The text is centered on the image and is the only element present.
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### NOV 2 8 2008
Bio-Medical Research, Ltd. % Ms. Anne-Marie Keenan Parkmore Business Park West Galway, Ireland
Re: K082011 Trade/Device Name: MediStim XP, Type 281 and MediTens XP, Type 458 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ, IPF Dated: October 9, 2008 Received: October 21, 2008
Dear Ms. Kecnan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally
{4}------------------------------------------------
Page 2 - Ms. Anne-Marie Keenan
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:
MediTens XP, Type 458
Indications for Use:
The symptomatic relief and management of chronic intractable pain. It is also an adjunctive treatment in the management of post-surgical and post-traumatic pain. The device has no curative value and should only be used in conjunction with medical supervision.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
tence of CDRH, Office of Device Evaluation (ODE) (Division Sigm Division of General, Restorative and Neurological Devices
510(k) Number
{6}------------------------------------------------
# Indications for Use
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known):
Device Name:
MediStim XP, Type 281
Indications for Use:
- к Neuromuscular Electrical Stimulation (NMES) for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis and maintaining or increasing range of motion.
- Transcutaneous Electrical Nerve Stimulation (TENS) for an adjunctive 제 treatment in the management of post-surgical and post-traumatic acute pain problems.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.