K192864 · AngioDynamics, Inc. · QEY · Jun 1, 2020 · Cardiovascular
Device Facts
Record ID
K192864
Device Name
UNI*FUSE Infusion System with Cooper Wire
Applicant
AngioDynamics, Inc.
Product Code
QEY · Cardiovascular
Decision Date
Jun 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K192864 · Jun 1, 2020
UNI*FUSE Infusion System with Cooper Wire
AngioDynamics, Inc.
Published clinical literature (systematic reviews and meta-analyses); Retrospective clinical study cohorts; Prospective clinical study cohorts
The sponsor utilized a literature review of clinical studies to assess the safety of catheter-directed therapy in the pulmonary artery, specifically to evaluate the risk of vessel damage or injury compared to other vascular applications.
Pulmonary Embolism; Catheter-Directed Therapy; Literature Review; Safety Assessment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Kuo, W. et al. – Catheter-Directed Therapy for the Treatment of Massive Pulmonary Embolism: Systematic Review and Meta-Analysis of Modern Techniques (2009); Systematic review and meta-analysis (including 6 prospective and 29 retrospective studies)
Patients with massive pulmonary embolism; Sample Size: 594 patients; Number of Sites: 35 studies
Not applicable for this study
Clinical success rate, minor and major procedural complications
Indications for Use
The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.
Device Story
The UNI*FUSE™ Infusion System with Cooper Wire is a single-lumen, nylon infusion catheter designed for targeted fluid delivery. It features longitudinal slits at the distal end for fluid administration and an occluding ball wire (occlusion guidewire) to provide end-hole occlusion during use. The device is available in 4F and 5F diameters and various lengths. It is used by physicians in clinical settings to deliver thrombolytics or contrast media into the peripheral or pulmonary artery vasculature. The active infusion segment is identified under fluoroscopy via radiopaque markers. By enabling precise, localized delivery of therapeutic agents, the device assists in treating conditions like pulmonary embolism, potentially improving patient outcomes while minimizing systemic exposure to thrombolytics. The device is navigated through the vasculature under imaging guidance.
Clinical Evidence
No clinical trials were conducted on the subject device. Evidence consists of a human factors study on a simulated-use vascular model and a literature review of catheter-directed therapy (CDT) in the pulmonary artery. The human factors study confirmed device navigability, imaging visibility, and usability of the Directions for Use. The literature review (including meta-analyses and clinical studies like SEATTLE II, PERFECT, ULTIMA, and PEITHO) demonstrated high clinical success rates for CDT (86.5%) and no evidence of increased risk of vessel damage in the pulmonary artery compared to other vascular applications.
Technological Characteristics
Single-lumen nylon catheter; 4F/5F diameters; 90cm/135cm lengths; 2cm/5cm infusion segments. Features longitudinal slits and an occluding ball wire. Radiopaque markers for imaging. Biocompatible materials per ISO 10993-1. Mechanical operation; no energy source or software.
Indications for Use
Indicated for administration of fluids, including thrombolytic agents and contrast media, into peripheral and pulmonary artery vasculature in patients requiring catheter-directed therapy.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Bashir™ and Bashir™ N-X Endovascular Catheters (K183290)
Reference Devices
PULSE*SPRAY™ and UNI*FUSE™ Infusion Catheters (K163356)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
November 22, 2021
Angiodynamics, Inc. Brandon Brackett Manager, Global Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752
Re: K192864
Trade/Device Name: UNI*FUSE Infusion System with Cooper Wire Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA
Dear Brandon Brackett:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 1, 2020. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.
Sincerely,
Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2021.11.22 13:42:55 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 1, 2020
Angiodynamics, Inc. Brandon Brackett Manager, Global Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752
Re: K192864
Trade/Device Name: UNI*FUSE™ Infusion System with Cooper Wire Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: April 2, 2020 Received: April 3, 2020
Dear Mr. Brackett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/5 description: The image shows the name "Gregory W. O'connell -S" in a large, bold font. The text is black and appears to be part of a document or publication. The background is a light blue color.
Digitally signed by Gregory W. O'connell -S Date: 2020.06.01 13:44:17 -04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192864
Device Name
UNI*FUSE™ Infusion System with Cooper Wire
Indications for Use (Describe)
The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary For the Angiodynamics, Inc. UNI*FUSE™ Infusion System with Cooper Wire
Date Prepared: June 1, 2020
### Sponsor
Angiodynamics, Inc. 26 Forest Street Marlborough, MA 01752 USA
### Regulatory Contact
### Brandon M. Brackett, RAC
Senior Manager, Global Regulatory Affairs AngioDynamics, Inc. Telephone: 1-508-658-7984 Facsimile: 1-508-658-7976 Email: bbrackett@angiodynamics.com
#### Subject Device
- Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
#### Predicate Device
- 510(k) Reference Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
### Reference Device
- 510(k) Reference Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel:
Angiodynamics, Inc. UNI*FUSE™ Infusion System with Cooper Wire Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices
#### K183290
Thrombolex, Inc. Bashir™ and Bashir™ N-X Endovascular Catheters Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices
#### K163356
Angiodynamics, Inc. PULSE*SPRAY™ and UNI*FUSE™ Infusion Catheters Infusion Catheter 21CFR870.1210 Catheter, Continuous Flush Class 2 KRA Cardiovascular Devices
{5}------------------------------------------------
# Purpose
The intent of this Traditional 510(k) is to propose a modification to the UNI*FUSE™ Infusion System with Cooper Wire Indications for Use statement, incorporative of a clarification being made in order to specify the pulmonary artery as a vessel indicated for device access. The UNI*FUSE™ Infusion System with Cooper Wire overall design is not changing as a result of this clarification is being made in response to the clinical use of Infusion Catheters in the pulmonary artery vasculature, identified as a usage primarily via available medical literature.
### Device Description
The proposed UNI*FUSE™ Infusion System with Cooper Wire devices are single-lumen, nylon catheters with longitudinal slits located at 90° intervals around the distal end of the catheter for fluid delivery. An occluding ball wire (or occlusion guidewire) provides end-hole occlusion during fluid delivery. The catheters are available in 4F and 5F diameters and overall lengths of 90cm and 135cm. Additionally, they are available in multiple infusion segment lengths, including 2cm and 5cm lengths. The active infusion area can be identified under imaging by means of radiopaque markers on the catheter shaft at the distal and proximal ends of the infusion segment. As shown in Table 4 below, all of these characteristics are substantially equivalent to those of the predicate devices. Additionally, all of these characteristics are identical when comparing the proposed to the reference devices, as shown in Table 5.
The proposed UNI*FUSE™ Infusion System with Cooper Wire devices are intended for administration of fluids such as thrombolytics and contrast media into vessels that are impacted by thrombus, including the peripheral and pulmonary artery vasculature. Given the minimal physical differences of the vessels for which the device is intended to access (e.g. diameter, structure, tortuousness), the operating principle mechanism of action, and intended use is the same independent of anatomical location, use in the pulmonary artery is equivalent to other areas of the vasculature that the device is currently indicated for.
# Indications for Use/Intended Use
The UNI*FUSE™ Infusion System with Cooper Wire is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.
# Comparison of Similarities and Differences in Technological Characteristics and Performance
The proposed UNI*FUSE™ Infusion System with Cooper Wire and the predicate Bashir™ N-X Endovascular Catheter Model 7200 are substantially equivalent to one another in terms of design, materials of manufacture, specifications, dimensions, Indications for Use, and sizes and/or configurations, as depicted in the comparison via Table 4 below:
| Table 4: Comparison of Similarities and Differences in Technological Characteristics and Performance | | | |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------|------------|
| Proposed UNI*FUSE™ Infusion System with Cooper Wire vs. | | | |
| Predicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290) | | | |
| | Proposed UNI*FUSE™ Infusion System with Cooper Wire | Predicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290) | Comparison |
| ProCode | KRA | KRA | Identical |
| Regulation Number | 21CFR870.1210 | 21CFR870.1210 | Identical |
| Regulation Name | Catheter, Continuous Flush | Catheter, Continuous Flush | Identical |
| Regulatory Class | Class 2 | Class 2 | Identical |
{6}------------------------------------------------
| Table 4: Comparison of Similarities and Differences in Technological Characteristics and Performance | | | |
|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Proposed UNI*FUSE™ Infusion System with Cooper Wire vs.<br>Predicate Bashir™ N-X Endovascular Catheter Model 7200 (K183290) | | | |
| | Proposed UNI*FUSE™ Infusion<br>System with Cooper Wire | Predicate Bashir™ N-X<br>Endovascular Catheter Model<br>7200 (K183290) | Comparison |
| Indications for<br>Use | The UNI*FUSE™ Infusion<br>System with Cooper Wire is<br>intended for the administration<br>of fluids, including thrombolytic<br>agents and contrast media, into<br>the peripheral and pulmonary<br>artery vasculature. | The Bashir™ N-X Endovascular<br>Catheter is intended for the<br>controlled and selective<br>infusion of physician-specified<br>fluids into the peripheral and<br>pulmonary artery vasculature. | Substantially Equivalent |
| Catheter<br>Diameter (F) | 4F, 5F | 7F | Substantially Equivalent |
| Catheter Length<br>(cm) | 90cm, 135cm | 92.5cm | Substantially Equivalent |
| Catheter Infusion<br>Segment Length<br>(cm) | 2cm, 5cm | 12.5cm | Substantially Equivalent |
| Materials | All materials are commonly<br>used for this type of medical<br>device and are biocompatible in<br>accordance with ISO 10993-1. | All materials are commonly<br>used for this type of medical<br>device and are biocompatible in<br>accordance with ISO 10993-1. | Substantially Equivalent |
Additionally, these attributes are identical when comparing the proposed UNI*FUSE™ Infusion System with Cooper Wire to the reference UNI*FUSE™ Infusion Catheter, as shown in Table 5 below:
| Table 5: Comparison of Similarities and Differences in Technological Characteristics and Performance | | | |
|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| | Proposed UNI*FUSE™ Infusion<br>System with Cooper Wire | Reference UNI*FUSE™ Infusion<br>Catheter (K163356) | Comparison |
| ProCode | KRA | KRA | Identical |
| Regulation<br>Number | 21CFR870.1210 | 21CFR870.1210 | Identical |
| Regulation Name | Catheter, Continuous Flush | Catheter, Continuous Flush | Identical |
| Regulatory Class | Class 2 | Class 2 | Identical |
| Indications for<br>Use | The UNI*FUSE™ Infusion<br>System with Cooper Wire is<br>intended for the administration<br>of fluids, including thrombolytic<br>agents and contrast media, into<br>the peripheral and pulmonary<br>artery vasculature. | The UNI*FUSE™ Infusion<br>System is intended for the<br>administration of fluids,<br>including thrombolytic agents<br>and contrast media, into the<br>peripheral vasculature. | Identical<br>(Aside from proposed<br>modification) |
| Catheter<br>Diameter (F) | 4F, 5F | 4F, 5F | Identical |
| Catheter Length<br>(cm) | 90cm, 135cm | 45cm, 90cm, 135cm | Substantially Equivalent |
| Catheter Infusion<br>Segment Length<br>(cm) | 2cm, 5cm | 2cm, 5cm, 10cm, 15cm, 20cm,<br>30cm, 40cm, 50cm | Substantially Equivalent |
{7}------------------------------------------------
| | Proposed UNI*FUSE™ Infusion<br>System with Cooper Wire | Reference UNI*FUSE™ Infusion<br>Catheter (K163356) | Comparison |
|-----------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------|
| Materials | All materials are commonly<br>used for this type of medical<br>device and are biocompatible in<br>accordance with ISO 10993-1. | All materials are commonly<br>used for this type of medical<br>device and are biocompatible in<br>accordance with ISO 10993-1. | Identical |
Furthermore, the proposed UNI*FUSE™ Infusion System with Cooper Wire devices and predicate Bashir™ N-X Endovascular Catheter Model 7200's are incorporative of the same operating principle, mechanism of action, and intended use as one another (in addition to exhibiting substantial equivalence in terms of the overall design, materials of manufacture, sizes/configurations, and Indications for Use as shown above). Lastly, there are no changes being made to the UNI*FUSE™ Infusion System with Cooper Wire devices as compared to that cleared via reference 510(k) K163356. As a result, all previous biocompatibility, shelflife, performance, and other testing identified within K163356 remains applicable to the proposed version.
### Comparison of Performance Data
Angiodynamics, Inc. has also compared the performance testing that the Bashir™ N-X Endovascular Catheter Model 7200 was subjected to in support of its clearance, to that which the proposed UNI*FUSE™ Infusion System with Cooper Wire has been previously subjected to. While each battery of testing is not identical in a 1:1 nature, they are substantially equivalent to one another in that they fully test the functions of each device. Please refer to Table 6, below:
| Table 6: Comparison of Performance Testing | |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Proposed UNI*FUSE™ Infusion System with Cooper Wire vs. Bashir™ N-X Endovascular Catheter Model 7200 (K183290) | |
| Proposed UNI*FUSE™ Infusion System with Cooper Wire | Bashir™ N-X Endovascular Catheter Model 7200 (K183290) |
| -Dimensional Verification | -Kink Radius |
| -Length Sufficiency | -Trackability |
| -Catheter Hub-to-Catheter Shaft Connection | -Advancement Force |
| Compatibility | -Slider Actuator |
| -Catheter-to-Guidewire Compatibility | -Catheter Retraction |
| -Catheter-to-Introducer Sheath Compatibility | -Radial Force |
| -Catheter Tip Radius | -Delivery Flow-Rate |
| -Catheter Infusion | -Infusion Pressure at Various Flow Rates |
| -Slit Pattern Radiopacity | -Infusion Pressure with Pulse Spray |
| -Catheter Degradation | -Pressure Measurement Through Central Lumen |
| -Catheter Pressure | -Guidewire Compatibility |
| -Catheter-to-Occlusion Wire Configuration (Slow Infusion | -Dimensional Verification |
| Compatibility) | -Compliance of Injection Hubs |
| -Catheter/Accessory Compatibility | -Air Leakage |
| -Catheter/Fluid Compatibility | -Fluid Leakage |
| -Catheter Hub-to-Shaft Joint Kink Resistance | -Stress Cracking |
| -Occlusion Wire Flexibility | -Resistance to Separation |
| -Occlusion Wire Flow | -Torque Strength |
| -Occlusion Wire Seal | -Corrosion Resistance |
| -Hub-to-Wire Bond/Connection | -Joint Tensile Strength |
| -Distal Spring Tip-to-Mandrel Connection | -Particulate Generation |
| -Occlusion Wire Withdrawal | |
{8}------------------------------------------------
In addition to the performance testing summarized above, Angiodynamics, Inc. has also conducted a human factors study on the proposed UNI*FUSE™ Infusion System with Cooper Wire. Specifically, Angiodynamics, Inc. solicited the participation of multiple practicing physicians experienced in the use and placement of infusion catheters to evaluate the proposed UNI*FUSE™ Infusion System with Cooper Wire on a simulated-use vascular model. The results of this testing demonstrated the following:
- . The UNI*FUSE™ Infusion System with Cooper Wire configurations proposed via this submission are able to be navigated-to and used-within the pulmonary artery vasculature;
- . The infusion segments of the UNI*FUSE™ Infusion System with Cooper Wire configurations proposed via this submission are able to be clearly imaged under fluoroscopy (and thus confirmed to reside completely within the pulmonary artery vasculature);
- . The Directions for Use proposed for the UNI*FUSE™ Infusion System with Cooper Wire provides adequately detailed instructions in order to enable users to accurately and reliably place and use the device(s) within the pulmonary artery vasculature.
As a means of further validating the conclusions of the human factors study, Angiodynamics, Inc. also conducted an in-house "bench test" version of the study. The in-house testing corroborated those of the human factors study summarized in the dialogue above, and the cumulative results demonstrate that the proposed UNM*FUSE™ Infusion System with Cooper Wire configurations are able to navigate the vasculature, be placed and used within the pulmonary artery, and confirm their location under imaging (e.g. fluoroscopy). These determinations further justify the Indications for Use modification proposed via this submission.
# Clinical Literature Evaluation and Determinations
AngioDynamics, Inc. has assessed a variety of publicly available articles and other literature to identify instances of vasculature damage/endothelial cell destruction resulting from the use of various catheters (including infusion catheters) in the pulmonary artery vasculature. While the articles have varying endpoints, each study does document the other possible effects the device may pose during use (i.e. risks, complications, and other consequences) such that the risks and complications related to given devices and/or therapies are also known. None of the risks and complications identified related to pulmonary artery (or other vessel) damage and/or injury. This leads AngioDynamics, Inc. to conclude that there is a lack of evidence indicating the pulmonary artery vasculature to be more markedly prone to damage than other vessels for this application, and that the proposed Indications for Use does not increate existing risks OR introduce new risks. Secondarily, while the articles vary in the specific types of devices being used, infusion catheters (including the proposed UNI*FUSE™ Infusion System with Cooper Wire) are typically much smaller in diameter than other catheters being used, and therefore the other larger catheter types would represent a worst-case scenario in terms of hypothetical intrusiveness and potential for vessel damage. Please note: the clinical literature being cited and discussed is general in nature, and it is important to acknowledge that the subject UNI*FUSE™ Infusion Catheter with Cooper Wire device was not itself evaluated in the referenced studies. Additionally, the discussion developed from the clinical literature is intended to relate only to the UNI*FUSE™ Infusion Catheter with Cooper Wire devices, and only those with infusion segment lengths that have been confirmed to be completely contained within non-peripheral pulmonary arteries in a straight configuration (i.e. 2cm and 5cm infusion segment lengths).
Summary: AngioDynamics, Inc. has identified and reviewed relevant articles that discuss catheterdirected interventions and/or therapy for pulmonary embolism, and each summary includes discussion pertaining to related complications identified. The summary begins on the following page.
{9}------------------------------------------------
"Catheter-Directed Therapy in Acute Pulmonary Embolism with Right Ventricular Dysfunction: A Promising Modality to Provide Early Hemodynamic Recovery" (Dilektasli, A.G.; et al. - 2016)} evaluates the use of catheter-directed therapy (CDT) as an alternative to systemic thrombolysis (ST) in patients. Catheter-directed therapy is a percutaneous procedure used to dissolve thrombus by administering a lytic directly into said thrombus. The primary outcomes were mortality, clinical success, and complications. The study included 15 consecutive patients who underwent immediate catheter-directed therapy. An essential conclusion of the study was related to the complications of catheter-directed therapy, specifically, "there were no technications, such as perforation of a cardiac/vascular structure, tamponade, or procedure-related death in our study." Instead, it continues by asserting that "interventionalist experience is known to influence the technical success." In this article, that conclusion is contrasted against a statement made within a referenced article titled Goldhaber's "Percutaneous Mechanical Thrombectomy for Acute Pulmonary Embolism" (Goldhaber, S. 2007)2 related to "percutaneous mechanical thrombectomy (PMT)" catheters", which states: "the percutaneous mechanical thrombectomy catheter can perforate the pulmonary artery, cause massive distal embolization, or cause hemolysis." The significance of these statements and their comparison to one another supports the following conclusion:
- . The literature suggests that catheters that employ some type of mechanical component (or component that is in-addition-to a "typical" catheter design) may have a greater risk for the potential of vessel damage compared to catheters that do not.
"Catheter-Directed Therapy for the Treatment of Massive Pulmonary Embolism: Systematic Review and Meta-Analysis of Modern Techniques" (Kuo, W. et al. – 2009)3 summarizes the authors' systematic review of modern techniques related to catheter-directed therapy for the treatment of massive pulmonary embolism; specifically, to evaluate the safety and effectiveness of modern catheter-directed therapy (CDT) as an alternative treatment for massive pulmonary embolism. 594 patients from 35 studies (6 prospective, 29 retrospective) were analyzed both for the clinical success rate associated with catheterdirected therapy, as well as the minor and major complications encountered during catheter-directed therapy treatment (along with the rates associated with each complication). The pooled clinical success rate from catheter-directed therapy was 86.5%, and the pooled risks of minor and major procedural complications were 7.9% and 2.4% respectively. Minor and major complications were listed by specific type of complication, and none were related to injury or damage to the pulmonary artery itself. The significance of this data is twofold:
- 1. It shows that catheter-directed treatment in the pulmonary artery vasculature is a treatment that has a high success rate attributed to it; and
- 2. It shows that the complications associated with catheter-based interventions in the pulmonary artery vasculature do not include – in either "minor" or "major" complication categories – damage and/or injury to the vessel itself.
"Catheter-Directed Interventions for Pulmonary Embolism" (Zarghouni, M; et al. – 2016)" is an analysis of the information and conclusions of numerous key studies related to pulmonary artery vasculature interventions, most notably (see next paqe):
<sup>4</sup> Dilektasli, A.G.; et al. - "Catheter-Directed Therapy in Acute Pulmonary Embolism with Right Ventricular Dysfunction: A Promising Modality to Provide Early Hemodynamic Recovery" (2016)
<sup>2</sup> Goldhaber, S. – "Percutaneous Mechanical Thrombectomy for Acute Pulmonary Embolism" (2007)
<sup>3</sup> Kuo, W.; et al. – "Catheter-Directed Therapy for the Treatment of Massive Pulmonary Embolism: Systematic Review and Meta-Analysis of Modern Techniques" (2009)
<sup>4</sup> Zarghouni, M.; et al. – "Catheter-Directed Interventions for Pulmonary Embolism" (2016)
{10}------------------------------------------------
- . Piazza, et al. – "Prospective, Single-Arm, Multi-Center Trial of EkoSonic Endovascular System and Activase for Treatment of Acute Pulmonary Embolism – SEATTLE II Study" (2015)
- . Kuo, et al. – "Pulmonary Embolism Response to Fragmentation, Embolectomy, and Catheter Thrombolysis - PERFECT Study" (2015)
- . Kucher, et al. – "Ultrasound-Assisted Catheter Directed Thrombolysis for Acute Intermediate-Risk Pulmonary Embolism – ULTIMA Study" (2013)
- . Meyer, et al. – "Pulmonary Embolism Thrombolysis – PEITHO Study" (2014)
The analysis outlines the number of patients involved in each study type, the endpoints of each study, the arms of each study, and the findings/complications/etc. it also analyzes various types of catheter-directed therapies in the pulmonary artery vasculature, most notably "Catheter-Directed Thrombolysis via Infusion Catheters." It describes the benefits as well as the adverse aspects of each treatment option, and identifies the complications associated with each. Of the various options identified within the literature, none of the complications identified were related to vessel damage, endothelial cell destruction, or other types of complications synonymous with pulmonary artery vasculature injury. It concludes "there has been a reemergence of interest in catheter-directed techniques. Newer guidelines employ CDI [catheter-directed intervention] in treatment protocols. CDI has become a vital to at many institutions."
# 510(k) Summary Conclusions
In conclusion, assessment of the similarities and differences of the proposed UNI*FUSE™ Infusion System with Cooper Wire and the predicate Bashir™ N-X Endovascular Catheter Model 7200 led Angiodynamics, Inc. to determine that the two are substantially equivalent to one another; specifically:
- The proposed and predicate device have the identical ProCode, Regulation Number, Regulation ● Name, and Regulatory Class as one another;
- . The proposed and predicate device have substantially equivalent Indications for Use and/or Intended Uses;
- . The proposed and predicate devices each employ the same operating principle, mechanism of action, and are intended for the same patient populations; and,
- . The proposed and predicate exhibit equivalent overall design, materials of manufacture, performance testing, sizes, and configurations.
Additionally, results of human factors testing on the proposed UNI*FUSE™ Infusion System with Cooper Wire provide further evidence that the devices are able to be used in the pulmonary artery vasculature. Lastly, evaluation of publicly available clinical literature leads AngioDynamics, Inc. to conclude that:
- There is a lack of evidence indicating the pulmonary artery vasculature to be more markedly prone to damage than other vessels for this application; and,
- . The literature suggests that catheters that employ some type of mechanical component (or component that is in-addition-to a "typical" catheter design) may have a greater risk for the potential of vessel damage compared to catheters that do not.
The sum of these evaluations and determinations lead AngioDynamics, Inc. to conclude that substantial equivalence has been demonstrated, and that the existing data, additional testing, and clinical evaluation determinations have confirmed that there are no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.