Pulse* Spray Infusion System, Uni*Fuse Infusion System
K163356 · AngioDynamics, Inc. · QEY · May 30, 2017 · Cardiovascular
Device Facts
Record ID
K163356
Device Name
Pulse* Spray Infusion System, Uni*Fuse Infusion System
Applicant
AngioDynamics, Inc.
Product Code
QEY · Cardiovascular
Decision Date
May 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pulse*Spray System, PRO Infusion Catheter, and its related components, are intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature. The Uni*Fuse Infusion System is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature.
Device Story
Pulse*Spray and Uni*Fuse Infusion Systems are catheter-based devices for peripheral vascular fluid delivery. The Pulse*Spray system includes a PRO (Pressure Responsive Outlet) infusion catheter with longitudinal slits, an occluding guidewire/ball wire, syringes, Y-connector, and check valve. The Uni*Fuse system includes a catheter and occluding guidewire. The devices are used by clinicians to deliver thrombolytic agents or contrast media. The occluding wire blocks the distal end, forcing fluid through the catheter slits into the target vessel. The system is a manual, non-powered infusion tool. The modification in this submission involves changes to the catheter shaft and hub materials. The device provides a controlled delivery mechanism for therapeutic or diagnostic fluids, aiding in procedures such as thrombolysis.
Clinical Evidence
Bench testing only. Testing included catheter tensile strength, static burst pressure, flow rate, occlusion tip durability, temperature conditioning, dispersion and occlusion seal integrity, and biocompatibility per ISO 10993-1.
Technological Characteristics
Infusion catheter system; includes PRO infusion catheter with longitudinal slits, occluding guidewire/ball wire, syringes, Y-connector, and dual check valve. Materials: modified catheter shaft and hub materials. Sterilization: not specified. Connectivity: none (manual device).
Indications for Use
Indicated for administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
AngioDynamics Pulse*Spray and Uni*Fuse Infusion Systems (K961763)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 30, 2017
AngioDynamics, Inc. Robin Fuller Sr. Manager, Regulatory Affairs 26 Forest Street Marlborough, MA 01752
Re: K163356
Trade/Device Name: Pulse* Spray and Uni*Fuse Infusion Systems Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: April 4, 2017 Received: May 5, 2017
Dear Robin Fuller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Agnel
Fernando Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163356
Device Name Pulse* Spray Infusion System, Uni*Fuse Infusion System
#### Indications for Use (Describe)
The Pulse*Spray System, PRO Infusion Catheter, and its related components, are intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature.
The Uni*Fuse Infusion System is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature.
Type of Use (*Select one or both, as applicable*)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
AngioDynamics Pulse*Spray and Uni*Fuse Infusion Systems
Traditional 510(k)
K163356
Page 1 of 4
### 510(K) SUMMARY FOR THE
### ANGIODYNAMICS, INC. PULSE*SPRAY AND UNI*FUSE INFUSION SYSTEMS
Date Prepared: 10 January 2017
A. Sponsor:
AngioDynamics, Inc.
26 Forest Street
Marlborough, MA 01752
| B. Contact: | Robin Fuller<br>Sr. Manager Regulatory Affairs |
|-------------|---------------------------------------------------------------------------------|
| | Tel: 508-658-7986 |
| | Fax: 508-658-7976 |
| | Email: <a href="mailto:RFuller@angiodynamics.com">RFuller@angiodynamics.com</a> |
Subject Device:
| Trade Name: | AngioDynamics Pulse*Spray and Uni*Fuse<br>Infusion Systems |
|--------------------|------------------------------------------------------------|
| Common Name: | Infusion Catheter |
| Regulation Number; | 21 CFR 870.1210 |
| Regulation Name: | Continuous Flush Catheter |
Class II
Product Code:
Classification Panel:
Regulatory Class:
Cardiovascular
KRA
{4}------------------------------------------------
AngioDynamics Pulse*Spray and Uni*Fuse Infusion Systems
Traditional 510(k)
KJ63356
Page 2 of 4
# Predicate Device:
| Trade Name: | AngioDynamics Pulse*Spray and Uni*Fuse<br>Infusion Systems |
|-----------------------|------------------------------------------------------------|
| 510(k) Reference: | K961763 |
| Common Name: | Infusion Catheter |
| Regulation Number: | 21 CFR 870.1210 |
| Regulation Name: | Continuous Flush Catheter |
| Regulatory Class: | Class II |
| Product Code: | KRA |
| Classification Panel: | Cardiovascular |
{5}------------------------------------------------
Traditional 510(k)
K163356
Page 3 of 4
### C. Device Description:
The Pulse*Spray Catheter is packaged in a kit configuration. The following components are sold in the Pulse*Spray Infusion Catheter Kits:
·PRO (Pressure Responsive Outlet) Infusion Catheter with longitudinal slits at the distal end. Slits are located at 90° intervals around the catheter. Radiopaque markers on the catheter shaft indicate the active infusion pattern.
· Occluding Guidewire or Occluding Ball Wire which occludes the distal end of the PRO Infusion Catheter. The Occluding Guidewires or Occluding Ball Wires have proximal markers for occlusion position verification.
·1 mL Syringe and finger flange
- •10 mL Syringe
- ·20 mL Syringe
- · Y-Connector with Tuohy-Borst adapter
- ·Dual Check Valve
The Uni*Fuse Catheters are sold in only one configuration, with an occluding guidewire.
### D. Intended Use/Indications for Use:
The Pulse*Spray System, PRO Infusion Catheter, and its related components, are intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature.
The Uni*Fuse Infusion System is intended for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral vasculature
### E. Summary of Similarities and Differences in Technology Characteristics and Performance:
The proposed device has the same design and technical characteristics as the predicate device. The purpose of this 510(k) submission is to introduce into commercial distribution a modified AngioDynamics, Inc. Pulse*Spray and Uni*Fuse Infusion catheter. The modification is a change in the catheters shaft material and hub material.
Beyond the two material changes, the proposed and predicate devices are identical in regards to design, materials, packaging and sterilization to the predicate device. There are no changes being made to any of the accessories provided within the Pulse*Spray and Uni*Fuse Infusion catheters.
### F. Performance Data:
The performance testing included non-clinical bench testing. The following tests were performed.
{6}------------------------------------------------
Traditional 510(k)
K163356
Page 4 of 4
- CatheterTensile
- Static Burst
- Catheter Flow Rate
- Catheter Occlusion Tip Durability
- Catheter Temperature Condition
- Catheter Dispersion and Occlusion Seal
- Biocompatibility per ISO 10993-1
### G. Conclusion:
The results of the non-clinical testing and a comparison of similarities and differences demonstrate that the proposed and predicate devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.