Bashir™ Endovascular Catheter: The Bashir™ Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature. Bashir™ N-X Endovascular Catheter: The Bashir™ N-X Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids into the peripheral and pulmonary artery vasculature.
Device Story
Mechanical infusion catheters for peripheral and pulmonary artery vasculature; Bashir™ Endovascular Catheter features expandable infusion limbs forming a basket; Bashir™ N-X Endovascular Catheter features non-expandable, straight infusion limbs. Device delivered via sheath over guidewire under fluoroscopic guidance; physician-specified fluids delivered through laser-drilled holes in infusion limbs. Catheters allow direct blood pressure measurement via central lumen. Used in interventional settings by specialists; no electrical components or software. Benefits include controlled fluid distribution and potential for native blood flow through infusion channels. Device removed after procedure by retracting into sheath.
Clinical Evidence
Bench testing only; no human clinical trials. Bench evaluation included kink radius, trackability, advancement force, radial force, flow rates, pressure measurement, and mechanical integrity (tensile/torque/corrosion). GLP porcine animal study demonstrated no systemic or pathological adverse effects.
Technological Characteristics
Mechanical catheter; no electrical components or software. Materials are standard for vascular catheters. Features laser-drilled infusion holes. Bashir™ Model 7201 includes expandable infusion limbs; Bashir™ N-X Model 7200 has straight limbs. Sterilization via ISO 11135:2014 validated methods (SAL 10^-6).
Indications for Use
Indicated for controlled, selective infusion of physician-specified fluids (including thrombolytics) into peripheral vasculature (Bashir™ Endovascular Catheter) or peripheral and pulmonary artery vasculature (Bashir™ N-X Endovascular Catheter).
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Ekosonic Endovascular System with Rapid Pulse Modulation (K111705)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 3, 2022
Thrombolex, Inc. % Diane Horwitz Regulatory Consultant Eminence Clinical Research Inc. 2995 Steven Martin Dr. Fairfax. Virginia 22031
Re: K183290
Trade/Device Name: Bashir Endovascular Catheter Model 7201. Bashir N-X Endovascular Catheter. Mode 7200 Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA
Dear Diane Horwitz:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 25, 2019. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell@FDA.HHS.gov.
Sincerely.
Digitally signed by Gregory W. Gregory W. O'connell -S ○'connell -S Date: 2022.02.03
14:29:43 -05 14:29:43 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a bold, sans-serif font. The word "FDA" is in a blue box.
February 25, 2019
Thrombolex, Inc. % Diane Horwitz Regulatory Consultant Eminence Clinical Research Inc. 2995 Steven Martin Dr. Fairfax, Virginia 22031
Re: K183290
Trade/Device Name: Bashir Endovascular Catheter Model 7201, Bashir N-X Endovascular Catheter Model 7200 Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: January 31, 2019 Received: January 31, 2019
Dear Ms. Horwitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| For | Gregory O'Connell | 2019.02.25<br>16:29:55 -05'00' |
|-----|--------------------------------------------|--------------------------------|
| | Bram D. Zuckerman, M.D. | |
| | Director | |
| | Division of Cardiovascular Devices | |
| | Office of Device Evaluation | |
| | Center for Devices and Radiological Health | |
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183290
### Device Name
Bashir™ Endovascular Catheter Model 7201 Bashir™ N-X Endovascular Catheter Model 7200
Indications for Use (Describe) Bashir™ Endovascular Catheter:
The Bashir™ Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Bashir™ N-X Endovascular Catheter:
The Bashir™ N-X Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids into the peripheral and pulmonary artery vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "THROMBOLEX" in large, bold, red letters. The "X" in "THROMBOLEX" is stylized to look like a check mark. Below the word "THROMBOLEX" is the phrase "INNOVATIVE ENDOVASCULAR CATHETERS" in smaller, black letters.
### 510(k) SUMMARY K183290
#### GENERAL INFORMATION 1.
#### Submitter and 510(k) Owner 1.1
Thrombolex, Inc. 51 Britain Dr. New Britain PA 18901
#### 1.2 Official Correspondent
Diane Horwitz, Ph.D., RAC 2995 Steven Martin Dr. Fairfax, VA 22031 Telephone: 703.307.2921 Email: dhorwitz@ecr-inc.com
#### 1.3 Date of Preparation
February 21, 2019
### 2. NAME OF THE DEVICE
#### Trade/Proprietary Name 2.1.1
BashirTM Endovascular Catheter, Model 7201 Bashir™ N-X Endovascular Catheter, Model 7200
#### 2.1.2 Common/Usual Name
Continuous Flush Catheter
#### 2.1.3 Classification Information
| Classification Name: | Continuous Flush Catheter |
|----------------------------|---------------------------------|
| Classification Regulation: | 21 CFR 870.1210 |
| Class: | 2 |
| Product Code: | KRA – Continuous Flush Catheter |
| Panel: | Cardiovascular |
#### 3. PREDICATE DEVICE AND REFERENCE DEVICES
Ekosonic Endovascular System with Rapid Pulse Modulation (cleared under K111705).
{5}------------------------------------------------
Reference Devices are AngioJet Zelante (K151313), Cordis Maxi LD 14 mm PTA Balloon K963000), and 5 mm balloon catheters [Boston Scientific Sterling PTA Balloon ollatation Catheter (K132430), Abbott Fox SVA PTA Catheter (K092286), Cordis Savvy _ong PTA Balloon (K942094)].
# 4. DESCRIPTION OF THE DEVICE
"hrombolex developed the Bashir™ Endovascular Catheter (Model 7201) to enable elivery of physician-specified fluids, including thrombolytics, into the peripheral circulation. The Bashir™ N-X Endovascular Catheter (Model 7200) is a second model developed for infusion of physician-specified fluids in the pulmonary artery. The Bashir™ N-X differs from the Bashir™ Endovascular Catheter in that the infusion limbs do not expand into an infusion basket but remain collapsed in a straight configuration. The two levice models have a different appearance, namely in the handle design and color, which ensures appropriate use and identification.
The design of the Bashir™ Endovascular Catheter allows for delivery and positioning of he catheter at the target peripheral location via a sheath, over a guidewire using fluoroscopic guidance. Depending on the target location, the infusion can be initiated either n the non-expanded position (for vessels with smaller diameters), or by controlled expansion of the infusion basket to a desired dimension inside the vessel by the interventional specialist under fluoroscopic guidance. The physician-specified solution is then delivered via multiple laser-drilled holes through the infusion limbs of the Bashir™ Endovascular Catheter. After the completion of solution delivery, the infusion limbs are collapsed to a straight position in the case of peripheral vasculature, and then the Bashir™ Endovascular Catheter is retracted into the sheath and removed.
'he handle of the BashirTM N-X Endovascular Catheter has been modified to disable expansion of the basket. The Bashir™ N-X Endovascular Catheter, unexpanded, is thus similar to other straight infusion catheters including the predicate device.
Delivery of the Bashir N-X™ Endovascular Catheter into the pulmonary artery is accomplished using a long guide sheath using standard interventional techniques. The Bashir™ Endovascular Catheter is placed over a guidewire inside the sheath to the desired location under fluoroscopic guidance. The catheter does not come in direct contact with the right atrium, tricuspid valve, right ventricle or pulmonic valve at any time, as it is within a guide sheath.
The Bashir™ Endovascular Catheter and Bashir™ N-X Endovascular Catheter are comprised of common materials used in vascular catheters. The catheter is mechanical and contains no electrical components.
# 5. INDICATION FOR USE
he indications for use / intended use statement for the Bashir™ Endovascular Catheter and the Bashir™ N-X Endovascular Catheter are as follows:
{6}------------------------------------------------
Bashir™ Endovascular Catheter: The Bashir™ Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Bashir™ N-X Endovascular Catheter: The Bashir™ N-X Endovascular Catheter is intended for the controlled and selective infusion of physician-specified fluids into the peripheral and pulmonary artery vasculature.
# 6. COMPARISON OF THE INTENDED USE WITH THE PREDICATE DEVICE
The Bashir™ Endovascular Catheter and the Bashir™ N-X Endovascular Catheter and the predicate device, the Ekosonic Endovascular System, are intended to be used in the same interventional procedures in the same target population (including patients with thrombosis in the vasculature) and have the same primary function of providing a conduit for infusion of solutions. Thus, the Bashir™ Endovascular Catheter and Bashir™ N-X Endovascular Catheter have almost identical intended use / indications for use statements when compared to the predicate device, Ekosonic Endovascular System.
## 7. TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE
The Bashir™ Endovascular Catheter is substantially equivalent to the predicate device regarding its similar intended use, fluid dispersion function, certain aspects of design, materials and sterility.
There are also some differences between the technological characteristics of the Bashir™ Endovascular Catheter, the Bashir™ N-X Endovascular Catheter and the predicate device. These differences do not raise new questions of safety or effectiveness, and there are standard methods and tests that can evaluate the effects of new or different characteristics on safety and performance.
- Infusion limbs at the distal end of the Bashir™ Endovascular Catheter can expand in the peripheral circulation, to facilitate the distribution of infusate in the peripheral vasculature. This is compared to the straight distal end of the predicate catheter. This difference does not raise different questions of safety or effectiveness, since appropriate design verification and validation testing was performed by comparison of the Bashir™ Endovascular Catheter to legally marketed reference devices with expandable features to demonstrate that the radial force exerted is similar for the Bashir™ Endovascular Catheter compared to the balloon reference devices.
- The ultrasound transducers and the electronic and software-controlled components ● of the predicate device are not present for the subject device.
- . The Bashir Catheter and Bashir N-X use the infusion limbs to aid in distribution of fluids. Both models of the Thrombolex device utilize the multiple lumens of the distal infusion region to provide a large surface for infusion of physician-specified fluids and, also to permit flow of native blood containing endogenous thrombolytics through the channel(s) into peripheral the blood vessel or pulmonary artery (Bashir™ N-X Endovascular Catheter).
{7}------------------------------------------------
- The Bashir™ Catheters have the capability for direct blood pressure measurement via the central lumen of the subject devices.
Through product testing, Thrombolex has demonstrated that the technological differences between the subject device and the predicate device do not raise different issues of safety or effectiveness.
# 8. PERFORMANCE TESTING
This 510(k) notification provides performance data to establish the substantial equivalence of the Bashir™ Endovascular Catheter and the Bashir™™™ N-X Endovascular Catheter to the predicate device. The following is a summary of the performance data.
Sterilization and Shelf Life: The devices are sterilized using standard methods and the sterilization cycles have been validated following international standards. The shelf life of the devices has been established through stability studies. The sterilization validation complied with ISO 11135:2014. The sterilization cycle was validated to a sterility assurance level of 106. The shelf life of the Bashir Catheters was established through accelerated aging studies, and support a shelf life of 1 year.
Biocompatibility: Biocompatibility evaluation has been performed to show the finished, sterilized device is biocompatible and suitable for its intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" have been taken into account to evaluate the biocompatibility of the subject device.
Human Factors Testing: Usability testing was performed using the to-be-marketed device and was found to be substantially equivalent for the intended users, uses and use environments. Participants were able to perform procedures with the Bashir™ Endovascular Catheter and the Bashir™ N-X Endovascular Catheter and meet the predefined acceptance criteria.
Performance Testing- Bench: Performance testing was performed to characterize the Bashir™ Endovascular Catheter and the Bashir™ N-X Endovascular Catheter in comparison to the predicate device.
The testing included an evaluation of: Kink radius, trackability, advancement force, slider actuator force, catheter retraction, radial force, delivery flow rate, infusion pressure at various flow rates, infusion pressure with pulse spray, pressure measurement through central lumen, guidewire compatibility, dimensional verification, compliance of injection hubs, air leakage, fluid leakage, stress cracking, resistance to separation, torque strength, corrosion resistance, joint tensile strengths and particulate generation. The performance data demonstrate that the new devices are substantially equivalent to the predicate device.
Performance Testing - Animal: The safety of the Bashir™ Endovascular Catheter was demonstrated in a GLP porcine study. Use of the catheter had no adverse effects systemically or pathologically, supporting the safe use of the device.
ব
{8}------------------------------------------------
## 9. CONCLUSIONS
The information presented in this 510(k) submission demonstrates that the Bashir™ Endovascular Catheter and the Bashir™ N-X Endovascular Catheter are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.