K191999 · Sterilucent, Inc. · JOJ · Oct 21, 2019 · General Hospital
Device Facts
Record ID
K191999
Device Name
Sterilucent VH2O2 Chemical Indicators
Applicant
Sterilucent, Inc.
Product Code
JOJ · General Hospital
Decision Date
Oct 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
A chemical indicator for monitoring all cycles within the Sterilucent™ HC 80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-Lumen). VPRO® maX (Flexible. Lumen & Non-Lumen) and Sterilucent PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed units. Colors other than blue such as yellow/green should be treated as a process failure.
Device Story
Chemical process indicator (Class 1/Type 1) for vaporized hydrogen peroxide (VHP) sterilization; monitors exposure to sterilization cycles. Provided as strips or labels with pressure-sensitive adhesive. Operates via chemical reaction; indicator ink (tetraethylammonium bromide) changes color from red/pink to blue upon VHP exposure. Used by healthcare providers in clinical settings; placed with sterilization wraps, containers, or pouches. Visual color change confirms load exposure to sterilization process; blue indicates successful exposure, while yellow/green indicates process failure. Benefits patient by providing visual verification that medical devices have undergone sterilization processing.
Clinical Evidence
No clinical data. Bench testing only. Performance demonstrated via compliance with ANSI/AAMI/ISO 11140-1:2014, including shelf-life testing, indicator functionality in specified sterilizers, endpoint color stability, and biocompatibility.
Technological Characteristics
Class 1/Type 1 chemical indicator. Materials: polymeric substrate (Tyvek, polypropylene, polystyrene). Indicator agent: tetraethylammonium bromide. Ink components: water-based carrier, activator, indicator agent, modulator. Form factor: strip or label. Conforms to ANSI/AAMI/ISO 11140-1:2014. Shelf life: 2 years. Storage: 15-30°C, dry, light-protected.
Indications for Use
Indicated for health care providers to monitor vaporized hydrogen peroxide sterilization cycles in specified sterilizers (Sterilucent HC 80TT, STERRAD, STERIS V-PRO, Sterilucent PSD-85) by distinguishing between processed and unprocessed units using sterilization wraps, containers, cassettes, or pouches.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 21, 2019
Sterilucent, Inc. Peter Kalkbrenner Director of Engineering 1400 Marshall St. NE Minneapolis, Minnesota 55413
Re: K191999
Trade/Device Name: Sterilucent VH2O2 Chemical Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: July 25, 2019 Received: July 26, 2019
Dear Peter Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K191999
Device Name Sterilucent VH2O2 Chemical Indicator
#### Indications for Use (Describe)
A chemical indicator for monitoring all cycles within the Sterilucent™ HC 80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-Lumen). VPRO® maX (Flexible. Lumen & Non-Lumen) and Sterilucent PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed units. Colors other than blue such as yellow/green should be treated as a process failure.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
× | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for the Sterilucent VH2O2 Chemical Indicator
| Submitted by: | Sterilucent, Inc. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Peter R. Kalkbrenner |
| | Director of Engineering |
| | peter.kalkbrenner@sterilucent.com |
| | Phone: 612-767-3253 |
| | Fax: 612-767-3261 |
| Date of Summary: | 21 October 2019 |
| Device Trade Name: | Sterilucent VH2O2 Chemical Indicator |
| Common or Usual Name: | Chemical sterilization process indicator |
| Classification: | 21 CFR 880.2800(b) |
| Class: | II |
| Product Code: | JOJ |
| Predicate Device(s): | SPSmedical VH2O2 Indicators (K140566/ K110152) |
| Device Description: | The Sterilucent VH2O2 Chemical Indicator is a Class 1/Type 1 process<br>indicator that conforms to ANSI/AAMI/ISO 11140-1:2014, and is intended<br>to be used with currently cleared hydrogen peroxide sterilizers (STERRAD®<br>100S, 200, 100NX, NX, STERIS® VPRO™ 1, V-PRO™ 1 Plus, VPRO®<br>maX and the Sterilucent PSD-85) and the Sterilucent HC 80TT Hydrogen<br>Peroxide Sterilizer. The HC 80TT Hydrogen Peroxide Sterilizer has two pre-<br>programmed sterilization cycles ("Lumen" and "Flexible") which both<br>utilize vaporized hydrogen peroxide (VHP) to rapidly sterilize a variety of<br>reusable medical devices. |
| | The Sterilucent VH2O2 Chemical Indicator is provided in two different<br>formats: Strip and Label. Both formats consist of a polymeric material on<br>which indicator ink is deposited. A pressure-sensitive adhesive is provided<br>on the back of the Label, which allows the process indicator to be adhered to<br>various substrates including containers, pouches, tamper-proof arrows or<br>locks, etc. |
| | The Sterilucent VH2O2 Chemical Indicator provides a visual indication that a<br>sterilization load has been exposed to the VHP sterilization process. The<br>indicator works by means of a chemical reaction, which results in a<br>recognizable color change from red to blue. |
| Indication for Use: | A chemical indicator for monitoring all cycles within the Sterilucent™ HC<br>80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S<br>(Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard<br>& Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-<br>Lumen), VPRO® maX (Flexible, Lumen & Non-Lumen) and Sterilucent<br>PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be<br>used by health care providers with sterilization wraps, containers, cassettes,<br>or pouches to distinguish between processed and unprocessed units. Colors<br>other than blue such as yellow/green should be treated as a process failure. |
## Technological Characteristics Table:
| | Proposed<br>Sterilucent VH2O2 Chemical<br>Indicators | Predicate<br>SPSmedical VH2O2 Indicators<br>(K140566/ K110152) | Comparison |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Crosstex-SPSmedical | Crosstex-SPSmedical | Same |
| Proposed<br>Sterilucent VH2O2 Chemical<br>Indicators | | Predicate<br>SPSmedical VH2O2 Indicators<br>(K140566/ K110152) | Comparison |
| Intended Use | Single use sterilization process indicator | Single use sterilization process indicator | Same |
| Indications for Use | A chemical indicator for monitoring all cycles within the Sterilucent™ HC<br>80TT Hydrogen Peroxide Sterilizer<br>(Lumen & Flexible), STERRAD®<br>100S (Standard & Long), 200, 100NX<br>(Standard, Flex & Express), NX<br>(Standard & Advanced), STERIS®<br>VPRO™ 1, V-PRO™ 1 Plus (Lumen<br>& Non-Lumen), VPRO® maX<br>(Flexible, Lumen & Non-Lumen) and<br>Sterilucent PSD-85 (Lumen & Non-<br>Lumen). The VH2O2 Indicators are<br>intended to be used by health care<br>providers with sterilization wraps,<br>containers, cassettes, or pouches to<br>distinguish between processed and<br>unprocessed units. Colors other than<br>blue such as yellow/green should be<br>treated as a process failure. | A chemical indicator for monitoring<br>all cycles within the STERRAD®<br>100S (Standard & Long), 200, 100NX<br>(Standard, Flex & Express), NX<br>(Standard & Advanced), STERIS®<br>VPRO™ 1, V-PRO™ 1 Plus (Lumen<br>& Non-Lumen), VPRO® maX<br>(Flexible, Lumen & Non-Lumen) and<br>Sterilucent™ PSD-85 (Lumen & Non-<br>Lumen). The VH2O2 Indicators are<br>intended to be used by health care<br>providers with sterilization wraps,<br>containers, cassettes, or pouches to<br>distinguish between processed and<br>unprocessed units. Colors other than<br>blue such as yellow/green should be<br>treated as a process failure. | Similar |
| Device Design | Strip, Label | Strip, Label, Tape | Similar |
| Endpoint Specifications -<br>Color Change Upon<br>Exposure to H2O2 | Red/Pink changes to Blue | Red/Pink changes to Blue | Same |
| Indicator Agent | Tetraethylammonium<br>Bromide(ammonium salt) | Tetraethylammonium<br>Bromide(ammonium salt) | Same |
| Indicator Ink<br>Components | Water-based Carrier; Activator,<br>Indicator Agent, & Modulator | Water-based Carrier; Activator,<br>Indicator Agent, & Modulator | Same |
| Substrate | Polymer (Tyvek, Polypropylene,<br>polystyrene) | Polymer (Tyvek, Polypropylene,<br>polystyrene) | Same |
| Conformance Testing<br>under ISO 11140-1 for<br>VH2O2 Indicators | Comply with Standard | Comply with Standard | Same |
| Full Cycle In-Use<br>Validation Testing of CI<br>with listed sterilizers | Yes | Yes | Same |
| Shelf Life | Up to 2 years from date of manufacture | Up to 2 years from date of<br>manufacture | Same |
| Distributor | Sterilucent & SPSmedical | Sterilucent & SPSmedical | Same |
| Storage Conditions | Dry area, ambient temperature of 15° to<br>30° C, away from alkaline chemicals,<br>acids, and sources of light | Dry area, ambient temperature of 15°<br>to 30° C, away from alkaline<br>chemicals, acids, and sources of light | Same |
{4}------------------------------------------------
The Sterilucent VH2O2 Chemical Indicator and predicate device are both single-use process indicators intended to monitor exposure to sterilization cycles utilizing vaporized hydrogen peroxide as a sterilant. Both devices have the same intended use, design, materials, specifications, and modeof-action, and are manufactured using the exact same production methods.
{5}------------------------------------------------
### Summary of Non-Clinical Testing:
Performance testing was conducted to demonstrate the functionality of the Sterilucent VH2O2 Chemical Indicator and general conformance with the requirements for Class 1/Type 1 vaporized hydrogen peroxide sterilization indicators as detailed in ANSI/AAMI/ISO 11140-1:2014. A summary of nonclinical tests performed follows:
| Test | Standard Used | Goal | Result |
|-------------------------------------------------------|---------------|----------------------------------------------------------------------------------------------------|--------|
| ISO 11140-1 compliance | ISO 11140-1 | Demonstrate compliance to ISO<br>11140-1 pass/fail criteria | Passed |
| Shelf-life | ISO 11140-1 | Demonstrate compliance to ISO<br>11140-1 pass/fail criteria on devices<br>at the stated shelf life | Passed |
| Sterilucent VH2O2 Chemical<br>Indicator Functionality | NA | Demonstrate appropriate color<br>change in applicable sterilizer | Passed |
| Endpoint Color Stability | NA | Demonstrate processed indicator<br>endpoint color stability | Passed |
| Biocompatibility | NA | Demonstrate CI does not release<br>any know toxic substances | Passed |
Summary of Clinical Clinical evaluations were not required and therefore are not submitted with this Testing: 510(k).
Overall Performance Based on the nonclinical tests performed, the subject device, the Sterilucent Conclusion: VH2O2 Chemical Indicator, is as safe, as effective and performs as well as or better than the legally marketed predicate device, the SPSmedical VH2O2 Indicators.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.