The SPSmedical VH2O2 Indicators are intended for use as process indicators for all vaporized hydrogen peroxide cycles in the STERRAD® 100S, 200, 100NX, NX, STERIS® V-Pro 1, V-Pro 1 Plus, V-Pro maX and Sterilucent PSD-85 sterilization processes. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items. Indicators change to a signal color of Blue after exposure to vapor hydrogen peroxide.
Device Story
Single-use chemical process indicators; monitor vaporized hydrogen peroxide (H2O2) sterilization cycles. Input: exposure to H2O2 gas during sterilization. Transformation: chemical indicator ink changes color from pink/yellow to blue upon exposure. Output: visual color change on strips, labels, tape, or cards. Used by healthcare providers in clinical settings; placed with sterilization wraps, containers, cassettes, or pouches. Provider observes color change to distinguish processed from unprocessed items; blue indicates successful exposure, while other colors indicate process failure. Benefits: provides visual confirmation of sterilization cycle exposure, ensuring processed items are identified correctly.
Clinical Evidence
Bench testing only. Validation performed per ANSI/AAMI/ISO 11140-1:2005(R)2010 requirements for indicators in vaporized hydrogen peroxide processes. Testing included simulated use in Sterilucent PSD-85 sterilizers and multiple lots with varying shelf life. All lots met performance requirements.
Technological Characteristics
Chemical indicator ink on synthetic substrates (e.g., Tyvek, polypropylene, polystyrene). Meets ANSI/AAMI/ISO 11140-1 requirements for process indicators. Form factors include strips, labels, tape, and cards. Non-toxic, no MSDS required. Passive colorimetric sensing principle.
Indications for Use
Indicated for monitoring all vaporized hydrogen peroxide sterilization cycles in STERRAD 100S, 200, 100NX, NX; STERIS V-PRO 1, 1 Plus, maX; and Sterilucent PSD-85 sterilizers. Used by healthcare providers with sterilization wraps, containers, cassettes, or pouches to distinguish processed from unprocessed units. Colors other than blue (e.g., yellow/green) indicate process failure.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators (K110152)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure in a circular pattern.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24, 2014
Crosstex International Mr. Michael G. Nolan Research and Development Coordinator 6789 West Henrietta Road Rush, NY 14543
Re: K140566
Trade/Device Name: SPSmedical VH2O2 Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: October 21, 2014 Received: October 22, 2014
### Dear Mr. Nolan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Nolan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(K) NUMBER (IF KNOWN):
DEVICE NAME: SPSmedical VH2O2 Indicators
#### INDICATIONS FOR USE:
A chemical indicator for monitoring all cycles within the STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® V-PRO™ 1, V-PRO" 1 Plus (Lumen & Non-lumen), V-PRO® maX (Flexible, Lumen & Non-lumen) and Sterilucent™ PSD-85 (Lumen & Non-lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed units. Colors other than blue such as yellow/green should be treated as a process failure.
| CATALOG NUMBER | PRODUCT NAME |
|----------------|-----------------|
| GPS-250R | Indicator Strip |
| GPS-250Y | Indicator Strip |
| GPL-2000R | Indicator Label |
| GPL-2000Y | Indicator Label |
| HT-048 | Indicator Tape |
| HT-036 | Indicator Tape |
| 5093 | Indicator Card |
Prescription Use: (Part 21 CFR 801 Subpart D) and/or
Over-the-Counter Use: X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page
{3}------------------------------------------------
#### SUBMITTER INFORMATION
| SPSmedical Supply Corp.<br>a division of Crosstex International<br>6789 West Henrietta Road<br>Rush, NY 14543 U.S.A. | | Contact: | Michael G. Nolan<br>Research and Development Coordinator |
|----------------------------------------------------------------------------------------------------------------------|-------------------|----------|----------------------------------------------------------|
| | | Phone: | (800) 722-1529 |
| Date of Summary: | November 24, 2014 | Fax: | (585) 359-0167 |
#### DEVICE NAME AND CLASSIFICATION
| Device Trade Name: | SPSmedical VH2O2 Indicators |
|----------------------|------------------------------------------------------------------------|
| Common Name: | Vaporized Hydrogen Peroxide Chemical Indicators |
| Classification Name: | Physical/chemical sterilization process indicator 21 CFR § 880.2800(b) |
| Review Panel: | General Hospital |
| Product Code: | JOJ |
| Device Class: | II |
#### PREDICATE DEVICE:
The SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators cleared under K110152.
#### DEVICE DESCRIPTION:
The SPSmedical VH2O2 Indicators are single use process indicators which are cleared for use in verifying exposure to all vaporized hydrogen peroxide cycles in the STERRAD® 100S, 200, 100NX, NX, STERIS® V-Pro 1, V-Pro 1 Plus, and V-Pro maX Testing has been performed which validated the SPSmedical VH2O2 Indicators for use in the Sterilucent PSD-85 Sterilizer.
Indicators will identify if an item has seen H2O2 during the Sterilucent PSD-85 sterilization processes by changing to a Blue signal color. They provide a visual indication to help distinguish between processed and unprocessed items.
### INTENDED USE:
The SPSmedical VH2O2 Indicators are intended for use as process indicators for all vaporized hydrogen peroxide cycles in the STERRAD® 100S, 200, 100NX, NX, STERIS® V-Pro 1, V-Pro 1 Plus, V-Pro maX and Sterilucent PSD-85 sterilization processes. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items. Indicators change to a signal color of Blue after exposure to vapor hydrogen peroxide.
#### TECHNICAL CHARACTERISTICS:
The chemical indicator ink utilized on the SPSmedical VH2O2 Indicators has been formulated to meet the performance requirements ANSV/AAMI/ISO 11140-1 for Process Indicators. When applied to different synthetic substrates the Indicators are indicated for use as internal strips, external cards, labels or tape. Which all function as process indicators.
#### BIOCOMPATIBILITY
The information provided here is identical between to K110152, and no modifications have been made to the proposed device when compared with the K110152 device. SPSmedical VH2O2 Chemical Indicators are manufactured using nontoxic inks and substrates that will not alter the chemical composition of the products being sterilized and are safe for human contact. Indicators can be disposed in general waste receptacles as no Material Safety Data Sheet (MSDS) is required.
{4}------------------------------------------------
#### STORAGE CONDITIONS:
Store in a cool, dry place (15-30°C).
### NON-CLINICAL TESTING:
Verification and validation tests were performed as a result of a Failure Mode and Effects Analysis (FMEA).
Various testing, including testing to ANSVAAMI/ISO 11140-1:2005(R)2010 requirements for indicators being run in vaporized hydrogen peroxide sterilization processes was performed. Testing also included simulated use in the Sterilization processes. Multiple lots of indicators with various levels of shelf life were included in testing.
All lots of SPSmedical VH2O2 Indicators gave acceptable results for all tests performed.
#### COMPONENTS:
The components of the SPSmedical VH2O2 Indicators are all previously approved materials for the Vaporized Hydrogen Peroxide Sterilization Process. Substrates and manufacturing processes used for the Subject device are identical to the substrates and manufacturing processes used in K110152. Indicators are composed of commercially existing materials of synthetic paper or label stock (e.g. Tyvek®, polypropylene, polystyrene etc ... ). The indicator ink is a nontoxic sterilization indicator ink that changes from an initial color of pink or yellow to a blue signal color when exposed to the Hydrogen Peroxide Sterilization Process.
## SUBSTANTIAL EQUIVALENCE DISCUSSION:
SPSmedical Supply Corp. has identified the SPSmedical VH2O2 Indicators cleared under K110152 to show equivalence to the proposed SPSmedical VH2O2 Indicators. Both the predicate and proposed chemical indicators are the same indicators made with the same indicator ink, substrates and manufacturing process. There is no difference between the predicate and proposed VH-O2 indicators. Therefore, the proposed SPSmedical VH2O2 Indicators are substantially equivalent to the predicate in terms of their use and functional characteristics in VH2O2 sterilization processes.
#### PREDICATE I.D.
| Trade Name: | SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators | |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Model No.: | GPS-250R<br>GPS-250Y<br>GPL-2000R<br>GPL-2000Y<br>HT-048<br>HT-036<br>5093 | Indicator Strip<br>Indicator Strip<br>Indicator Label<br>Indicator Label<br>Indicator Tape<br>Indicator Tape<br>Indicator Card |
| Submitter/holder: | SPSmedical Supply Corp.<br>a division of Crosstex International<br>6789 West Henrietta Road<br>Rush, NY 14543 U.S.A.<br>Phone: (585) 359-0130<br>Fax: (585) 359-0167 | |
| 510(k) No.: | K110152 | |
{5}------------------------------------------------
#### COMPARISON OF INDICATIONS FOR USE (IFU):
Predicate IFU-The SPSmedical H2O2 Chemical Indicator are indicated for use as process indicators for all vaporized hydrogen peroxide cycles in the STERRAD® 100S, 200, 100NX, NX and STERIS® V-Pro 1 Plus and V-Pro maX sterilizers. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items. Indicators change to a signal color of Blue after exposure to vapor hydrogen peroxide.
Subject IFU-A chemical indicator for monitoring all cycles within the STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® V-PRO™ 1, V-PRO" 1 Plus (Lumen & Non-lumen), V-PRO® maX (Flexible, Lumen & Non-lumen) and Sterilucent" PSD-85 (Lumen & Non-lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed units. Colors other than blue such as yellow/green should be treated as a process failure.
Discussion-The Predicate and Subject Device are the same device in materials and manufacturing, but the indications for use are slightly different. With the Subject IFU we have added the sterilization cycle names which they were always cleared for, but not clearly identified. The addition of a label claim and the grounds of this 510(k) submission the PSD-85 (Lumen & Nonlumen) cycles have been added.
#### DESIGN DIFFERENCES PREDICATE VS. SUBJECT DEVICE:
There are no changes in design from the Predicate to the Subject Device. Only an additional Indication for Use as been added which is the Sterilucent PSD-85 (Lumen and Non-lumen) cycle.
#### FUNCTIONAL CHARACTERISTICS
The SPSmedical VH2O2 Indicators are designed to monitor sterilization cycles in low temperature VH2O2 type sterilizers. They are a reliable tool used for the monitoring of VH2O2 sterilization processes and provide a visual indication that hydrogen peroxide (H2O2), an essential ingredient in the vaporized hydrogen peroxide sterilization process has been introduced into the sterilizer's chamber. Indicators will identify if an item has seen H2O2 during the sterilization process. Indicators change from an initial color of Yellow or Pink to a final signal color of Blue.
#### DISCUSSION
SPSmedical is claiming substantial equivalence for its proposed SPSmedical VH2O2 Indicators to the current SPSmedical H2O2 Indicators based on test data taken during comparison studies. We have demonstrated with testing that the SPSmedical VH2O2 Indicators perform consistently with results which indicate exposure to the VH2O2 sterilization process.
#### SUBSTANTIAL EQUIVALENCE CONCLUSIONS
The proposed SPSmedical VH2O2 Indicators have the same intended use and characteristics as the current SPSmedical H2O2 Indicators. Both provide a visual indication that the indicator has been exposed to the H2O2 sterilization process. The SPSmedical products are comprised of synthetic materials and are printed with an indicator ink which provides a signal color after exposure to the H2O2 sterilization process. The performance of the SPSmedical VH2O2 Indicators have been validated through testing directed by SPSmedical.
SPSmedical believes that the SPSmedical VH2O2 Indicators are substantially equivalent to the predicate device because they are the same device without any modifications. Both the current and proposed indicators have the same intended use, technical characteristics and performance. Because the ability to perform its intended function has been shown through validated testing and the SPSmedical line of VH2O2 Indicators raises no issues related to safety or effectiveness. Based on
{6}------------------------------------------------
the nonclinical tests performed the subject device is as effective and performs at least as safely and effectively as the legally marketed predicate device, Class II (21 CFR 880.2800(b)), product code JOJ. See Table 1 for a substantial equivalence comparison.
| ELEMENT | PROPOSED DEVICE | PREDICATE<br>(K110152) |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Process Indicator | Process Indicator |
| Indications for Use | A chemical indicator for<br>monitoring all cycles within<br>the STERRAD® 100S<br>(Standard & Long), 200,<br>100NX (Standard, Flex &<br>Express), NX (Standard &<br>Advanced), STERIS® V-<br>PRO™ 1, V-PRO 1 Plus<br>(Lumen & Non-lumen), V-<br>PRO® maX (Flexible, Lumen<br>& Non-lumen) and<br>Sterilucent™ PSD-85 (Lumen<br>& Non-lumen). The VH2O2<br>Indicators are intended to be<br>used by health care providers<br>with sterilization wraps,<br>containers, cassettes, or<br>pouches to distinguish<br>between processed and<br>unprocessed units. Colors<br>other than blue such as<br>yellow/green should be<br>treated as a process failure. | The SPSmedical H2O2<br>Chemical Indicator are<br>indicated for use as process<br>indicators for all vaporized<br>hydrogen peroxide cycles in<br>the STERRAD® 100S, 200,<br>100NX, NX and STERIS® V-<br>Pro1, V-Pro 1 Plus and V-Pro<br>maX sterilizers. They are<br>intended to be used by health<br>care providers with articles<br>such as sterilization wraps,<br>containers, cassettes, or<br>pouches to distinguish<br>between processed and<br>unprocessed items.<br>Indicators change to a signal<br>color of Blue after exposure<br>to vapor hydrogen peroxide. |
| Device Design | Strip, Label, Tape, Card etc. | Strip, Label, Tape, Card etc. |
| Endpoint Color | Signal Color of Blue | Signal Color of Blue |
| Indicator Agent | H2O2 Indicator Ink | H2O2 Indicator Ink |
| Sterilization Method | Vaporized Hydrogen<br>Peroxide | Vaporized Hydrogen<br>Peroxide |
| Device Materials | Same (Synthetic Substrate) | Same (Synthetic Substrate) |
| Performance under ISO<br>11140-1 Complete<br>reaction cycle | Equivalent | Equivalent |
| Performance under ISO<br>11140-1 Incomplete<br>reaction cycle | Equivalent | Equivalent |
| Biocompatibility | Equivalent | Equivalent |
| Non-clinical Performance-<br>Sterilucent PSD-85<br>(Lumen & Non-Lumen<br>cycles) | Complete Cycle Turns Blue<br>with yellow/green<br>intermediate<br>(Same) | Complete Cycle Turns Blue<br>with yellow/green<br>intermediate<br>(Same) |
| Same Shelf-life | Up to 2 years | |
## SUBSTANTIAL EQUIVALENCE COMPARISON TABLE
{7}------------------------------------------------
### CONCLUSION:
The subject device SPSmedical VH2O2 chemical indicators have the same intended use and characteristics as the predicate SPSmedical H2O2 chemical indicators. Both provide a visual indication that the indicator has been exposed to the PSD-85 sterilization process.
SPSmedical believes that the SPSmedical VH₂O₂ chemical indicators are substantially equivalent to the predicate device because they are the same device without any modifications. Both the subject and predicate indicators have the same intended use, technical characteristics and performance. Based on the non-clinical performance testing data the subject device has proven to be as safe and effective as the predicate device when monitoring the PSD-85 sterilization processes and can be safely marketed in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.