SPSMEDICAL VAPORIZED HYDROGEN PEROXIDE CHEMICAL INDICATORS
K110152 · Spsmedical Supply Corp. · JOJ · Apr 15, 2011 · General Hospital
Device Facts
Record ID
K110152
Device Name
SPSMEDICAL VAPORIZED HYDROGEN PEROXIDE CHEMICAL INDICATORS
Applicant
Spsmedical Supply Corp.
Product Code
JOJ · General Hospital
Decision Date
Apr 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The SPSmedical Vaporized Hydrogen Peroxide (H2O2) Chemical Indicators are process indicators used to verify exposure to the STERIS® V-Pro maX flexible sterilization process. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items.
Device Story
Chemical indicator strips, labels, tapes, and cards; used to verify exposure to vaporized hydrogen peroxide in specific sterilization cycles (STERRAD and STERIS V-Pro). Device changes color from original to blue upon exposure to H2O2. Used by healthcare providers in clinical settings; placed with items inside sterilization wraps, containers, or pouches. Provides visual confirmation of processing; helps distinguish processed from unprocessed items; ensures quality control in sterilization workflows.
Clinical Evidence
Bench testing only. Testing performed per AAMI/ANSI/ISO 11140:2005 requirements for indicators in vaporized hydrogen peroxide sterilization processes. Simulated use testing conducted in the STERIS V-Pro maX flexible sterilization cycle using multiple lots with various shelf-life levels. All lots yielded acceptable results.
Technological Characteristics
Chemical process indicator (strip, label, tape, card). Sensing principle: chemical color change upon exposure to vaporized hydrogen peroxide. Standard: AAMI/ANSI/ISO 11140:2005. No electronic components, software, or energy source.
Indications for Use
Indicated for use as process indicators for all vaporized hydrogen peroxide cycles in STERRAD 100S, 200, 100NX, NX and STERIS V-Pro 1, V-Pro 1 Plus, and V-Pro maX sterilizers. Used by healthcare providers with sterilization wraps, containers, cassettes, or pouches to distinguish processed from unprocessed items.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators (K030680)
SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators (K090650)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
APR 1 5 2011 K110152
510k No.:
## Submitter:
- . SPSmedical Supply Corp. 6789 West Henrietta Road Rush, NY 14543 U.S.A. Phone: (585)-359-0130 Fax: (585)-359-0167
- Establishment FDA Registration No.: 1319130 .
- January 14th, 2011 Date Summary was Prepared .
- Gary J. Socola . Printed name of person submitting for 510(k)
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------>OCD Sare ડ Signature of person submitting for 510(k)
- Vice President, Scientific Affairs ● Title of person submitting for 510(k)
#### Device Name and Classification
| Trade Name: | SPSmedical Vaporized Hydrogen Peroxide<br>Chemical Indicators |
|------------------------|-------------------------------------------------------------------------------------|
| Classification Name: | Sterilization Process Chemical Indicator |
| Common Name: | Vaporized Hydrogen Peroxide Chemical Indicators |
| Device Classification: | General Hospital - Class II, Regulation Number 880.2800 |
| Product Code: | 80JOJ |
| Predicate Device: | SPSmedical Vaporized Hydrogen Peroxide Chemical<br>Indicators (K030680 and K090650) |
{1}------------------------------------------------
#### Device Description:
SPSmedical Vaporized Hydrogen Peroxide (HzOn) Chemical Indicators are process indicators which are cleared for use in verifying exposure to vapor hydrogen peroxide in the STERRAD® 100S, 200, 100NX, NX, STERIS® V-Pro lumen and non lumen sterilization processes. Testing has been performed which validates the SPSmedical Vaporized Hydrogen Peroxide (H2O2) Chemical Indicators for use in the STERIS® V-Pro flexible sterilization cycle and the STERIS® V-Pro maX sterilizer.
Indicators will identify if an item has seen H2O2 during the STERIS® V-Pro flexible sterilization process by changing to a Blue signal color. They provide a visual indication to help distinguish between processed and unprocessed items.
#### Intended Use:
The SPSmedical Vaporized Hydrogen Peroxide (H2O2) Chemical Indicators are process indicators used to verify exposure to the STERIS® V-Pro maX flexible sterilization process. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items.
Statement of Similarity to the Legally Marketed Predicate Device:
- Have the same intended use ●
- Have the same device design ●
- Incorporate the same technical characteristics ●
- Incorporate the same materials .
- . Have the same endpoint color change
- Have the same shelf life �
- Have the same storage conditions .
- Packaged using the same materials and processes .
- --------------------------------------------------------------------------------------
#### Non-Clinical Testing:
Verification and validation tests were performed as a result of a Failure Mode and Effects Analysis (FMEA) assessment.
Various testing, including testing to AAMI/ANSVISO 11140:2005 requirements for indicators being run in vaporized hydrogen peroxide sterilization processes was performed. Testing also included simulated use in the STERIS® V-Pro maX vaporized hydrogen peroxide flexible sterilization cycle. Multiple lots of indicators with various levels of shelf life were included in testing.
All lots of SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators gave acceptable results for all test performed.
#### Conclusion:
For all the foregoing reasons, SPSmedical believes that the SPSmedical Vaporized Hydrogen Peroxide Chemical Indicators are safe and effective when used for routine monitoring of STERIS® V-Pro maX sterilizer running a vaporized hydrogen peroxide flexible sterilization cycle and can be safely marketed in the United States.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a staff entwined by two snakes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WQ66-G609 Silver Spring, MD 20993-0002
Mr. Gary J. Socola Vice President, Scientific Affairs SPSmedical Supply, Corporation 6789 West Henrietta Road Rush, New York 14543
APR 1 5 2011
Re: K110152
Trade/Device Name: SPSmedical H2O2 Chemical Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: January 14, 2011 Received: January 19, 2011
Dear Mr. Socola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Socola
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucmm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRF1's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS for USE STATEMENT
SPSmedical Supply Corp. Applicant:
510(k) Number (if known):
Device Name: SPSmedical H2O2 Chemical Indicator
Indications For Use:
The SPSmedical H2O2 Chemical Indicators are indicated for use as process indicators for all vaporized hydrogen peroxide cycles in the STERRAD® 100S, 200, 100NX, NX and STERIS® V-Pro 1, V-Pro 1 Plus and V-Pro maX sterilizers. They are intended to be used by health care providers with articles such as sterilization wraps, containers, cassettes, or pouches to distinguish between processed and unprocessed items. Indicators change to a signal color of Blue after exposure to vapor hydrogen peroxide.
| Catalog Number | Product Name |
|----------------|-----------------|
| GPS-250R | Indicator Strip |
| GPS-250Y | Indicator Strip |
| GPL-2000R | Indicator Label |
| GPL-2000Y | Indicator Label |
| HT-048 | Indicator Tape |
| HT-036 | Indicator Tape |
| 5093 | Indicator Card |
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
urrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page __ of ___________________________________________________________________________________________________________________________________________________________________
510(k) Number: K110152
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.