K182301 · DePuy Orthopaedics, Inc. · JWH · Nov 8, 2019 · Orthopedic
Device Facts
Record ID
K182301
Device Name
PFC SIGMA Knee System
Applicant
DePuy Orthopaedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Nov 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total knee arthroplasty is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee replacement may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely crippled patients with multiple joint involvement for whom a gain in knee mobility may lead to an expectation of significant improvement in the quality of their lives.
Device Story
PFC SIGMA Knee System is a total knee prosthesis consisting of femoral, tibial, and patellar components; replaces natural articular surfaces of knee joint. Femoral component is metal, with or without porous coating; tibial component is all-polyethylene or metal tray with polyethylene insert; patellar component is all-polyethylene or metal-backed polyethylene. Used by orthopedic surgeons in clinical settings for total or unicompartmental knee replacement. This submission updates labeling to include MR compatibility information. Device remains unchanged in design, materials, and manufacturing. MR safety evaluated via bench testing for magnetically induced displacement, torque, RF heating, and image artifacts per ASTM standards. Output is MR conditional labeling, allowing patients with these implants to undergo MRI under specified conditions, reducing risks of device dislodgement or heating during imaging.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence is supported by non-clinical bench testing regarding MR compatibility.
Technological Characteristics
Total knee prosthesis components (femoral, tibial, patellar). Materials include metal (e.g., Co-Cr) and polyethylene. Features include modular stems, sleeves, and wedges. Porous coating options available. MR compatibility established via ASTM F2052-14 (displacement), F2213-06 (torque), F2182-11a (RF heating), F2219-07 (artifacts), and F2503-13 (marking).
Indications for Use
Indicated for patients with severely painful and/or disabled joints due to osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or failed previous implants. For unicompartmental arthroplasty, only one side (medial or lateral) is affected. SIGMA C/R Porocoat Femoral Components are for cemented or cementless use; all other porous or non-porous components are for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
November 8, 2019
Image /page/0/Picture/1 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
DePuy Orthopaedics, Inc. Mr. Brian Kincaid Associate Director, Regulatory Affairs 700 Orthopaedic Drive Warsaw, Indiana 46582
Re: K182301
Trade/Device Name: PFC SIGMA Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: November 1, 2019 Received: November 4, 2019
Dear Mr. Kincaid:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ting Song, PhD Assistant Director (Acting) DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182301
Device Name PFC SIGMA Knee System
#### Indications for Use (Describe)
Candidates for total or unicompartmental knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant. In candidates for unicompartmental knee arthroplasty, only one side of the medial or lateral compartment) is affected.
The Sigma® C/R Porocoat® Femoral Components are intended for cementless use as the femoral component of a Total Knee Replacement system.
In the US, all other porous coated components have been cleared for CEMENTED USE ONL Y. Any Non-Porous coated components are intended for CEMENTED USE ONLY.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510 (k) Summary
(As required by 21 CFR 807.92)
| Submitter Information | |
|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Orthopaedics |
| Address | 700 Orthopedic Drive<br>Warsaw, IN 46582 |
| Phone number | 574-372-7337 |
| Fax number | 574- 371-4987 |
| Establishment Registration Number | 1818910 |
| Name of contact person | Brian Kincaid |
| Date prepared | May 23, 2019 |
| Name of device | |
| Trade or proprietary name | PFC SIGMA Knee System |
| Common or usual name | Total Knee Replacement Prothesis |
| Classification name | Knee joint patellofemorotibial polymer/metal/polymer semi-contrained<br>cemented prosthesis<br>Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3560, 21 CFR 888.3565 |
| Product Code(s) | JWH, MBH |
| Legally marketed device(s)<br>to which equivalence is<br>claimed. | K882234, K884796, K952830, K961685,<br>K971189<br>PFC SIGMA Knee System, Tibial Tray |
| | K952830<br>PFC SIGMA Knee System, Fluted Tibial Rod |
| Note: The primary predicate<br>device is K032115. All<br>others are additional<br>predicate devices. | K882234<br>PFC SIGMA Knee System, Taper Femoral Stem |
| | K063633<br>PFC SIGMA Knee System, Universal Stem |
| | K943462, K961685, K062654<br>PFC SIGMA Cruciate Knee System |
| | K950010, K952830, K971189<br>PFC SIGMA Knee System |
| | K952830, K060515<br>PFC SIGMA Knee System, adaptors and bolts |
| | K991106<br>PFC SIGMA Porous Modular Keel Tray |
| | K984158<br>PFC SIGMA Plus Offset Tibial Tray |
| | K032115 (Primary)<br>PFC SIGMA Co-Cr Tibial Tray |
{4}------------------------------------------------
| Reason for 510(k) submission | The subject of this submission is to provide magnetic resonance imaging<br>(MRI) compatibility labeling for the PFC SIGMA Knee System. There are<br>no proposed changes to the indications for use, the design of the product,<br>the materials that comprise the product, or the use of the product. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description | A total knee prosthesis is composed of individually packaged femoral, tibial<br>and patellar components designed to replace the natural articular surface of<br>the knee joint. The femoral component is a metal implant, with or without<br>a porous coating. The tibial component may be an all polyethylene<br>component or comprised of a metal tibial tray with or without porous<br>coating, and a polyethylene insert and locking components. Some metal<br>components have modular stems, sleeves and/or modular wedges. The<br>patella component may be of an all polyethylene design or may be a metal<br>backed polyethylene component. |
| Intended use of the device | Total knee arthroplasty is intended to provide increased patient mobility and<br>reduced pain by replacing the damaged knee joint articulation in patients<br>where there is evidence of sufficient sound bone to seat and support the<br>components. Total knee replacement may be considered for younger<br>patients if, in the opinion of the surgeon, an unequivocal indication for total<br>knee replacement outweighs the risks associated with the age of the patient,<br>and if limited demands regarding activity and knee joint loading can be<br>assured. This includes severely crippled patients with multiple joint<br>involvement for whom a gain in knee mobility may lead to an expectation<br>of significant improvement in the quality of their lives. |
| Indications for use | Candidates for total or unicompartmental knee replacement include patients<br>with a severely painful and/or severely disabled joint resulting from<br>osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed<br>previous implant. In candidates for unicompartmental knee arthroplasty,<br>only one side of the joint (the medial or lateral compartment) is affected.<br>The SIGMA® C/R Porocoat® Femoral Components are intended for<br>cemented or cementless use as the femoral component of a Total Knee<br>Replacement system.<br>In the US, all other porous coated components have been cleared for<br>CEMENTED USE ONLY. Any Non-Porous coated components are<br>intended for CEMENTED USE ONLY. |
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED<br>TO THE PREDICATE DEVICE | |
The purpose of this submission is to change the device labeling to include information for safety and compatibility of the subject devices in the MR environment. The design features and materials of the subject devices are identical to those of the predicate devices. The indications for use are identical to the predicate devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this submission.
{5}------------------------------------------------
## PERFORMANCE DATA
## SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The purpose of the submission is to add MR compatibility to the device labeling for the PFC SIGMA Knee System. There are no changes to the intended use, design, materials, sterilization, packaging, shelf life, or manufacturing methods for the PFC SIGMA Knee System.
The MR environment could potentially interact with the subject devices that may impact patient safety and imaging, including movement or dislodgement of the device due to magnetically induced displacement force and/or torque, radiofrequency (RF) induced heating, and image artifacts. Methods used to evaluate MR compatibility included the Food and Drug Administration (FDA) guidance document entitled "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment', as well as the following standards from the American Society for testing and Materials (ASTM):
- Magnetically induced displacement force (ASTM F2052-14*) ●
- Torque (ASTM F2213-06)
- Radio frequency (RF) heating (ASTM F2182-11a)
- Image Artifacts (ASTM F2219-07)
- Standard Practice for marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment ● (F2503-13)
The tests determined the effects of the implants, and the effects of the implants on the image quality. The tests evaluated the worst-case components and constructs for RF Heating, and image artifacts. The testing concluded that there are no safety issues related to magnetic field interactions under specific conditions identified in the labeling.
As a result, the testing supports a MR conditional label as described within the FDA's Guidance Document and the ASTM standard, F2503-13.
*Note: External test reports have been written in compliance to ASTM F2052-14. However, a single change has been made to ASTM F2052-15 to voluntarily include the determination of the maximum allowable spatial gradient of the magnetic field. This has been incorporated in the respective reports. Henceforth, all references to the standard will be ASTM F2052-14.
## SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
## CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.