DEPUY KNEE PROSTHESIS SYSTEM UNIVERSAL STEM EXTENSIONS AND FEMORAL METAPHYSEAL SLEEVES
K063633 · DePuy Orthopaedics, Inc. · JWH · Mar 2, 2007 · Orthopedic
Device Facts
Record ID
K063633
Device Name
DEPUY KNEE PROSTHESIS SYSTEM UNIVERSAL STEM EXTENSIONS AND FEMORAL METAPHYSEAL SLEEVES
Applicant
DePuy Orthopaedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Mar 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are intended for use with the PFC, Sigma PFC, Sigma TC3 Revision Knee, or S-ROM knee prostheses in total knee replacement surgery for patients suffering from severe pain and disability due to permanent structural damage resulting from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma or failed prior surgical intervention. These devices are intended for cemented use only. The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are also intended for use with the DePuy LPS prosthesis for replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include: Malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) requiring extensive resection and replacement; patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis, requiring extensive resection and replacement; revision for failed previous prosthesis cases requiring extensive resection and replacement; severe trauma requiring extensive resection and replacement. The LPS prosthesis is also intended for use in bone loss postinfection, where the surgeon has elected to excise the bone and replacement is required. The Universal Stem and the Universal Metaphyseal Sleeve components are intended for cemented use only.
Device Story
Universal Femoral Metaphyseal Sleeves and Universal Stem Extensions function as modular orthopedic components for knee replacement systems (PFC, Sigma, S-ROM, LPS). Sleeves attach to femoral components via taper locking; stem extensions attach to sleeves or directly to femoral components via threaded connections. Stem extensions feature fluted, quadrifurcated designs for flexibility under physiologic loads. Components are implanted by orthopedic surgeons in clinical settings to restore joint function, address bone loss, or facilitate limb salvage. Benefits include modularity for complex revisions and extensive resections. Output is a stabilized, stemmed femoral knee construct. Cemented use only.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: Ti6-Al4-V alloy; porous coating of sintered commercially pure (CP) titanium microbeads. Design: Modular femoral metaphyseal sleeves (5 sizes) and stem extensions (3 lengths, 8 diameters). Attachment: Taper locking and threaded connections. Energy source: None (mechanical). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring total knee replacement or extensive bone resection/replacement due to severe pain, disability, structural damage from arthritis (rheumatoid, osteoarthritis, post-traumatic), collagen disorders, pseudogout, trauma, failed prior surgery, malignant tumors, or bone loss post-infection.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
14063633
MAR 0 2 2007
# SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF FIRM: | DePuy Orthopaedics Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46581-0988<br>EST REG No.: 1818910 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Steve Wentworth<br>Regulatory Affairs Project Manager<br>Tel: (574) 371-4913<br>Fax: (574) 371-4987 |
| TRADE NAME: | DePuy Knee Prosthesis System Universal Stem Extensions and<br>Universal Femoral Metaphyseal Sleeves |
| COMMON NAME: | Tricompartmental Knee Prosthesis |
| CLASSIFICATION: | Knee joint patellofemorotibial, polymer/metal/polymer semi-<br>constrained cemented prosthesis (21 CFR 888.3560), Class II Device |
| DEVICE PRODUCT CODE: | 87 JWH |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES: | Universal Femoral Metaphyseal Sleeve Components: DePuy LPS<br>Knee System (K040281, cleared July 9, 2004)<br>Universal Stem Extensions: S-ROM/Noiles™ PS Total Knee System<br>(K941769, cleared February 1, 1995) |
### DEVICE DESCRIPTIONS:
DePuy is offering the DePuy Universal Femoral Metaphyseal Sleeve and the DePuy Universal Stem Extenstion components as line extensions to its previously cleared PFC, PFC Sigma, S-ROM, Sigma TC3 Revision Knee and Limb Preservation System (LPS) knee replacement prostheses systems. These additional components are designed to be interchangeable within these four knee systems.
The DePuy Universal Femoral Metaphyseal Sleeve components are a modification to the design of the previously cleared DePuy LPS Metaphyseal Sleeve included in K040281. The Universal Femoral Metaphyseal Sleeve components are available in five sizes. They are also available in nonporous textured finish, fully porous coated and distally porous coated options. The femoral metaphyseal sleeve components are optional components that attach to the femoral knee prostheses components by a taper locking mechanism. Use of adapter components are required for the attachment of the metaphyseal sleeve component to the PFC Sigma, LPS and Sigma TC3 femoral components. The
# 0000004
{1}------------------------------------------------
metaphyseal sleeve attaches directly to the S-ROM femoral component. The threaded distal female portion of the femoral metaphyseal sleeve component receives the male threaded portion of the Universal Stem Extension to provide a stemmed femoral knee prosthesis construct. An optional polyethylene plug component is assembled to the metaphyseal sleeve.
The Universal Stem Extension component is an optional component available in three lengths and eight diameters. The distal portion of the stem extension below the male threaded portion is fluted and terminates in a bullet shaped tip. The stem extension component is perpendicularly quadrifurcated to allow for flexibility of the stem component under physiologic loads. The two smallest component diameters are solid. The DePuy Universal Stem Extension components are a modification to the design of the stem extension components previously cleared in K941769. These components can be attached to the female threaded portion of the Universal Femoral Metaphyseal Sleeve or may be attached without the sleeve to the PFC Sigma and Sigma TC3 femoral components using the adapters. The sleeve component is required for the LPS and S-ROM femoral components. The same male threaded attachment is used to connect the stem extensions to the tibial tray components from the PFC, PFC Sigma and Sigma TC3 Revision knee prosthesis systems. The stem extension in not used for the S-ROM tibial trays.
The Universal Stem and Universal Femoral Metaphyseal Sleeve are both manufactured from Ti6-A14-V alloy. The porous coating on the Universal Femoral Metaphyseal Sleeve is comprised of sintered microbeads of commercially pure (CP) titanium. The Universal Stem Extension component is substantially equivalent to the S-ROM Knee System slotted stem extension component cleared previously under K941769 and the Universal Femoral Metaphyseal Sleeve component is substantially equivalent to the LPS Metaphyseal Sleeve cleared under K040281.
## INDICATIONS FOR USE:
The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are intended for use with the PFC, PFC Sigma, Sigma TC3 Revision Knee, or S-ROM knee prostheses in total knee replacement surgery for patients suffering from severe pain and disability due to permanent structural damage resulting from rheumatoid arthritis. osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma or failed prior surgical intervention. These devices are intended for cemented use only.
The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are also intended for use with the DePuy LPS prosthesis for replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include: Malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) requiring extensive resection and replacement; patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis, requiring
0000005 Regulatory_affairs/510(k)/2006/stmslvprmkt.doc
{2}------------------------------------------------
extensive resection and replacement; revision for failed previous prosthesis cases requiring extensive resection and replacement; severe trauma requiring extensive resection and replacement. The LPS prosthesis is also intended for use in bone loss postinfection, where the surgeon has elected to excise the bone and replacement is required.
The Universal Stem and the Universal Metaphyseal Sleeve components are intended for cemented use only.
## BASIS FOR SUBSTANTIAL EQUIVALENCE:
The design of the DePuy Universal Femoral Metaphyseal Sleeve is substantially equivalent to the LPS Metaphyseal Sleeve (K040281). Since the intended use and materials used for the subject and predicate device components are the same, the DePuy Universal Femoral Metaphyseal Sleeve is considered to be substantially equivalent to the predicate device.
The design of the DePuy Universal Stem Extension is substantially equivalent to the S-ROM/Noiles Knee slotted stem component. Since the materials and intended use for the subject device components are identical to those for the predicate device, the DePuy Universal Stem Extension is considered substantially equivalent to the predicate device.
00000006
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three intertwined snakes and a staff. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DePuy Orthopaedics, Inc. % Mr. Steven J. Wentworth Regulatory Affairs Project Manager 700 Orthopaedic Drive P.O. Box 988 Warsaw, Indiana 46581-0988
Re: K063633
Trade/Device Name: DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: December 5, 2006 Received: December 6, 2006
MAR 0 2 2007
Dear Mr. Wentworth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
## Page 2 - Mr. Steven J. Wentworth
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A Milkersen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): _ < 063633
# Device Name: DePuy Universal Metaphyseal Sleeve and Universal Stem Extension
Intended Use and Indications:
The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are intended for use with the PFC, Sigma PFC, Sigma TC3 Revision Knee, or S-ROM knee prostheses in total knee replacement surgery for patients suffering from severe pain and disability due to permanent structural damage resulting from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma or failed prior surgical intervention. These devices are intended for cemented use only.
The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are also intended for use with the DePuy LPS prosthesis for replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include: Malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) requiring extensive resection and replacement; patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis, requiring extensive resection and replacement; revision for failed previous prosthesis cases requiring extensive resection and replacement; severe trauma requiring extensive resection and replacement. The LPS prosthesis is also intended for use in bone loss postinfection, where the surgeon has elected to excise the bone and replacement is required.
The Universal Stem and the Universal Metaphyseal Sleeve components are intended for cemented use only.
Mark A. Millican
Division of General, Restorative, and Neurological Devices
Prescription Use × (Part 21 CFR 801 Subpart D) AND/
0000007
K063633
Concurrence of CDRH, Office of Device Evaluation (ODE)
Regulatory affairs/510(k)/2006/stmslvprmkt.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.