K991106 · DePuy Orthopaedics, Inc. · JWH · Jun 9, 1999 · Orthopedic
Device Facts
Record ID
K991106
Device Name
P.F.C. SIGMA POROUS MODULAR KEEL TIBIAL TRAY
Applicant
DePuy Orthopaedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jun 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The P.F.C. Sigma Porous Modular Keel Tibial Tray is indicated for use in total knee replacement in patients suffering from severe pain and disability due to permanent structural damage in the knee joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, or pseudogout. This damage may also be the result of trauma or failed prior surgical intervention. In accordance with Federal (USA) requirements, the P.F.C. Sigma Knee Modular Keel Tibial Tray is indicated for use only with PMMA bone cement.
Device Story
P.F.C. Sigma Porous Modular Keel Tibial Tray; titanium alloy (Ti-6Al-4V) tibial component for total knee arthroplasty. Distal surface features commercially pure titanium porous coating for cement fixation; includes four screw holes for optional supplemental fixation. Distal tip contains UHMWPE plug, removable to attach metallic extensions for long-stem applications. Designed for use with P.F.C. Modular and P.F.C. Sigma tibial inserts. Used in surgical setting by orthopedic surgeons. Device provides structural replacement of knee joint surfaces; requires PMMA bone cement for fixation. Benefits include pain relief and restoration of joint function for patients with severe structural knee damage.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through conformance with voluntary standards including ASTM F-1580 (porous coatings), ASTM F620 (titanium alloy), and ASTM F1044 (shear testing of porous coatings).
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V) per ASTM F620; porous coating: commercially pure titanium per ASTM F-1580. Features: Modular keel, four screw holes for fixation, removable UHMWPE distal plug for extension attachment. Fixation: Cemented only (PMMA). Sterilization: Not specified.
Indications for Use
Indicated for total knee replacement in patients with severe pain and disability due to permanent structural damage in the knee joint caused by rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma, or failed prior surgical intervention. Must be used with PMMA bone cement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
P.F.C. Modular Tibial Tray -Porous (K892394 and K961685)
{0}------------------------------------------------
# K991106
## 510(k) Summary P.F.C. Sigma Porous Modular Keel Tibial Tray
DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN 46581
### A. Contact Person:
Janet G. Johnson, RAC Senior Regulatory Associate (219) 372-7484
#### B. Device Information:
| Proprietary Name: | P.F.C. Sigma Porous Modular Keel Tibial Tray |
|----------------------|-------------------------------------------------------------------------------------------|
| Common Name: | Tibial Tray |
| Classification Name: | Knee Joint Patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis |
| Regulatory Class: | Class II, per 21 §CFR 888.3560 |
| Product Code: | 87 JWH |
#### C. Indications for Use:
The P.F.C. Sigma Porous Modular Keel Tibial Tray is indicated for use in total knee The I .. . . Sigma Forde moveing from severe pain and disability due to permanent structural damage in the knew ing from rheumatoid arthritis, osteoarthritis, post-traumatic surthritis, collagen disorders, or pseudogout. This damage may also be the result of trauma or failed prior surgical intervention.
In accordance with Federal (USA) requirements, the P.F.C. Sigma Knee Modular Kccl Tibial Tray is indicated for use only with PMMA bone cement.
{1}------------------------------------------------
#### D. Device Description:
The P.F.C. Sigma Porous Modular Keel Tibial Tray is manufactured from titanium alloy (Ti-6Al-4V). The distal surface, including approximately one-third of the stem and P.F.C. Sigma Porous Modular Keel Tibial Tray portions, is coated with commercially pure titanium porous coating to enhance cement fixation. If additional fixation is required, the tibial tray is designed with four screw holes that accept bone screws.
The distal tip of the P.F.C. Sigma Porous Modular Keel Tibial Tray is manufactured with a UHMWPE plug. When a long stem device is required the UHMWPE plug is removed to allow the tibial tray to accept attachment of a metallic extension.
The P.F.C. Sigma Porous Modular Keel Tibial Tray is designed for use with both P.F.C. Modular and P.F.C. Sigma tibial inserts, including curved posterior lipped and stabilized.
#### E. Substantial Equivalence:
The P.F.C. Sigma Porous Modular Keel Tibial Tray is substantially equivalent in terms of intended use, materials, design, sterilization method, and packaging to the P.F.C. Modular Tibial Tray -Porous (K892394 and K961685), P.F.C. Modular Plus Tibial Tray (K923807 and K961685), AMK Total Knee (K864671) and CRX Tibial Tray (K961379).
The determination of substantial equivalence for this device was based on a detailed device description, performance testing and conformance with voluntary performance standards. e.g. ASTM F-1580, ASTM F620 and ASTM F1044.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird with stylized wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
d 1899 JUN
Janet G. Johnson, RAC Janet G. Johnson, Associate Senior Regardic, Inc. P.O. Box 988 700 Orthopaedic Drive 700 Orthopaedio Bossona 46581-0988
Warsaw, Indiana 46581-0988
K991106
P.F.C. Sigma Porous Modular Keel Tibial Tray
Classes Classes Classe, II Re : Regulatory Class : II Product Code: JHW Dated: March 30, 1999 April 1, 1999 Received:
Dear Ms. Johnson:
Deal nove reviewed your Section 510 (k) notification of intent to we have reviewed your Section 510(K) noce have determined the market the for no market the device referenced above and we indications for
device is substantially equivalent (for the indications for device is substantially equivalent (ib) the enables marketed in interestate
use stated in the May 28, 1976 the enactment date of the use stated in the enclosure) to devices manned date of the commerce prior to May 28, 1978, che chaobment Medical Device Amendents of to devices that work.
reclassified in accordance with the provisions of the based of reclassified in accordance with the provisions of the rood, Drug, and Cosmetic Act (Act). Thirs decises labeled
this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone and intended to be fixed within bone were your device subject to cement." You may, therefore, market your act and the following the general control limitations:
- This device may not be labeled or promoted for non-1. cemented use.
- All labeling for this device, including package label
s and for the led within the package, must 2. All labeling for this action the package, must and labeling included within the promises.
prominently state that the device is intended for cemented use only.
{3}------------------------------------------------
Page 2 Janet G. Johnson, RAC
- 3 . Any non-cemented fixation of this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for noncemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class TTT (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in viter diagnostic devices) , please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be
{4}------------------------------------------------
Page 3 - Janet G. Johnson, RAC
obtained from the Division of Small Manufacturers Assistance obtained from the Division of Small Manufacearch 100 mg 1501 https://www.bac
at its toll-free number (800) 638-2041 or (301) 443-6597 obj. at its toll-free number (800) 638-2041 of (301) 449-697 08-11-2
its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices - Reseord Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510(k) Number (if known) | |
|--------------------------|--|
| Device Name | |
1991106 P.F.C. Sigma Porous Modular Keel Tibial Tray
## Indications for Use
The P.F.C. Sigma Porous Modular Keel Tibial Tray is indicated for use in total knee
r in the many and commend simp from severe pain and disability due to permanent The P.F.C. Sigma Porous Modular Keel Tibal Tray Is Indiano Ior users and replacements of patients suffering from severe pain and only and on on or experientiality
structural danage in the knee as neardorout . This damage may also be the result of structural damage in the knee joint from rheumatoid artist. Shoolarines p
arthritis, collagen disorders, or pseudogout. This damage may also be the result of trauma
nt the r artifits, conagers surgical intervention.
In accordance with Federal (USA) requirements, the P.F.C. Sigma Knee Modular Keel
, In accordance with a more with RMMA bone cement. In accordance with Federal (USA) requirements,
Tibial Tray is indicated for use only with PMMA bone cement.
(Please do not write below this line - Continue on another page if necessary)
(Please of the would on and on and Device Eveluation (ODE) do not write below uns inis mis ims into into evaluation (ODE)
Prescription Use_ × (Per 21 CFR §801.109) OR
Over-the-Counter Use_
(Optional Format 1-2-96)
collefo
Division Sign-Off)
Division of General Restorative Devices K94104
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.