The Serenity Piezo and Serenity Thermocouple Sensors are intended to measure and output snore and limb movement and thermal respiratory flow signals, respectively, from a patient for archival in a polysomnography study. The sensors are accessories to a polysomnography system which records and conditions the physiological signals for analysis and display, such that the data may be analyzed by a qualified sleep clinician to aid in the diagnosis of sleep disorders. The Serenity Piezo and Serenity Thermocouple Sensors are intended to be used by healthcare professionals for adults and children within a hospital, laboratory, clinic, or nursing home, or outside of a medical facility under the direction of a medical professional. The Serenity Piezo and Serenity Thermocouple Sensors are not intended for the life monitoring of high risk patients, do not include or trigger alarms, and are not intended to be used as a critical component of: an alarm or alarm system; an apnea monitor or apnea monitoring system; or life monitor or life monitoring system.
Device Story
Serenity Piezo and Thermocouple sensors are polysomnography (PSG) accessories; they collect physiological signals for processing by a host PSG amplifier and PC. The Piezo sensor uses an embedded piezo element to detect vibrations (snoring) or limb movement; it is placed on skin or worn in a heel strap. The Thermocouple sensor uses thermocouple junctions to detect thermal airflow under nostrils and in front of the mouth. Used in hospitals, clinics, or home settings by healthcare professionals. Output signals are recorded, conditioned, and displayed by the host system for review by sleep clinicians to score polysomnograms and diagnose sleep disorders. Devices do not include alarms or life-monitoring capabilities; they serve as passive signal acquisition components.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical safety (IEC 60601-1), dielectric strength, ingress protection (IPX2), biocompatibility (ISO 10993-5, ISO 10993-10), and mechanical integrity (tensile strength, mating cycles). Comparative performance testing demonstrated equivalent signal output to reference devices using identical host system configurations.
Technological Characteristics
Sensors consist of TPE enclosures and PVC cables. Piezo sensor uses piezo-electric sensing; Thermocouple sensor uses thermocouple junctions. Non-sterile. Operating temperature 10°C to 40°C. Connectivity is via direct connection to compatible PSG amplifier inputs. No internal software or active processing; signal conditioning occurs in the host amplifier.
Indications for Use
Indicated for adults and children undergoing polysomnography studies to measure limb movement, snore vibrations, and thermal respiratory flow for sleep disorder diagnosis.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
November 16, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, and the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
Neurotronics, Inc. David Pezet Quality Manager 4500 NW 27th Ave Ste C2 Gainesville, Florida 32606
Re: K181709
Trade/Device Name: Serenity Piezo Sensor, Serenity Thermocouple Sensor Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLV, GWL, MNR Dated: October 17, 2018 Received: October 18, 2018
Dear David Pezet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181709
Device Name
Serenity Piezo Sensor, Serenity Thermocouple Sensor
#### Indications for Use (Describe)
The Serenity Piezo and Serenity Thermocouple Sensors are intended to measure and limb movement and thermal respiratory flow signals, respectively, from a patient for archival in a polysomnography study.
The sensors are accessories to a polysomnography system which records and conditions the physiological signals for analysis and display, such that the data may be analyzed by a qualified sleep clinician to aid in the diagnosis of sleep disorders.
The Serenity Piezo and Serenity Thermocouple Sensors are intended for use on both adults and children by healtheare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under the direction of a medical professional.
The Serenity Piezo and Serenity Thermocouple Sensors are not intended for the life monitoring of high risk patients, do not include or trigger alarms, and are not intended to be used as a critical component of:
- an alarm or alarm system;
- · an apnea monitor or apnea monitoring system; or
- · life monitor or life monitoring system.
Type of Use (Select one or both, as applicable)
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K181709 510(k) Summary
November 16, 2018
| SUBMITTER | | | |
|---------------------------------|-------------------------------------------------------|-----------------------------------------------------------------|---------------|
| Neurotronics, Inc. | | Phone | (352)372-9955 |
| 4500 NW 27th Ave Ste C2 | | Fax | (815)550-2871 |
| Gainesville, FL 32606 | | | |
| Contact Person | | | |
| David Pezet, Quality Manager | | | |
| E-mail quality@neurotronics.com | | Phone | (352)372-9955 |
| SUBJECT DEVICE | | | |
| Trade Name | Serenity Piezo Sensor<br>Serenity Thermocouple Sensor | | |
| Common Name | Polysomnography (PSG) Accessories: | Piezo Sensor<br>Thermocouple Sensor | |
| Classification | 21 CFR 882.1400 Electroencephalograph | | |
| Product Codes | OLV | (primary) Standard Polysomnograph with<br>Electroencephalograph | |
| | GWL | Physiological Signal Amplifier | |
| | MNR | Ventilatory Effort Recorder | |
#### Predicate Device K142774
The subject device includes two sensors intended for use with a parent device (Standard Polysomnograph with Electroencephalograph) and is intended to support, supplement, and/ or augment the performance of a parent device. Therefore, the primary predicate is appropriately classified pursuant to the same regulation as its corresponding parent device.
| 510(k)<br>Number | Submitter/Holder | Device<br>Name | Regulation | Classification<br>Product Codes |
|------------------|--------------------|------------------------------|------------------------------------------------------|---------------------------------|
| K142774 | Neurotronics, Inc. | Polysmith<br>Sleep<br>System | 21 CFR 882.1400<br>Electroencephalograph<br>Class II | OLV (primary),<br>OLZ, DQA |
{4}------------------------------------------------
#### Reference Devices
These references devices are identified to support the scientific methodology and standard reference values associated with the subject device PSG accessories.
| Submitter/Holder | Device Name | Regulation | Classification<br>Product Codes | 510(k)<br>Number |
|---------------------------------------------------|----------------------------------------------|------------|---------------------------------|------------------|
| Neurotronics, Inc. | Nomad Sleep System<br>Recorder, Model PMU800 | 882.1835 | GWL, DQA,<br>MNR | K092699 |
| S. L. P. Scientific<br>Laboratory<br>Products LTD | Sleepsense Sleep Sensors | 868.2375 | MNR | K042253 |
| Pro-Tech, Inc. | PLM Sensor | 868.2375 | BZQ | K940014 |
#### DEVICE DESCRIPTION
Serenity sleep sensors are intended to measure and output physiologic signals used for Polysomnography (PSG) or Sleep Studies. These devices are to be used as an accessory to compatible amplifiers.
Typical sleep amplifiers use sensors and electrodes to collect physiological signals to further digitize, and the amplifiers send these signals to a host PC.
Serenity sleep sensors are worn by the patient and connected directly to compatible inputs of an amplifier. The amplifier and related software then processes the signal for review by qualified practitioners to score polysomnograms and diagnose Sleep Disorders.
Image /page/4/Figure/7 description: The image is a block diagram titled "Figure 1 Basic Sleep Study Block Diagram". The diagram shows the flow of information and equipment used in a sleep study. The diagram includes blocks for Patient Area, Patient Sensors, Audio/Video (Optional), External Auxiliary Devices (Optional), Amplifier, PC, and Sleep Clinician. The diagram shows the connections between these blocks, indicating the flow of information and signals.
{5}------------------------------------------------
The Serenity Piezo Sensor uses an embedded piezo sensing element to detect the vibrations of snoring or to sense a patient's limb movement. The sensor outputs a signal which corresponds to movements of the limbs or snore vibrations. The Piezo sensor can be placed on the skin or worn in a heel strap.
Image /page/5/Picture/1 description: The image shows two separate objects. The first object is a white medical sensor with a circular pad and two blue connectors. The second object is a dark blue garment with black straps and a white wire running through it.
Image /page/5/Picture/2 description: The image shows a medical device, specifically a nasal cannula, against a dark background. The cannula is white and consists of a plastic connector with two prongs that are inserted into the nostrils. Attached to the connector are two clear tubes that connect to a blue connector. The device is used to deliver supplemental oxygen to a patient.
Image /page/5/Picture/3 description: The image shows a close-up of a nasal cannula, a medical device used to deliver supplemental oxygen to a patient. The cannula consists of two small prongs that are inserted into the nostrils and connected to a tube that carries oxygen from a source. The prongs are made of a soft, flexible material to ensure patient comfort. The nasal cannula is a common and effective method for providing oxygen therapy to individuals with respiratory conditions.
The Serenity Thermocouple Sensor uses thermocouple wire that is ioined together to form sensing elements. Thermocouple junctions under each nostril and in front of the mouth output a signal which corresponds to the patient's thermal airflow. The Serenity Thermocouple sensor is available with an optional cannula hanger to aid in patient usability when worn with an airflow pressure cannula.
#### INTENDED USE
The Serenity Piezo and Serenity Thermocouple Sensors are intended to measure and output snore and limb movement and thermal respiratory flow signals, respectively, from a patient for archival in a polysomnography study. The sensors are accessories to a polysomnography system which records and conditions the physiological signals for analysis and display, such that the data may be analyzed by a qualified sleep clinician to aid in the diagnosis of sleep disorders.
The Serenity Piezo and Serenity Thermocouple Sensors are intended to be used by healthcare professionals for adults and children within a hospital, laboratory, clinic, or nursing home, or outside of a medical facility under the direction of a medical professional.
The Serenity Piezo and Serenity Thermocouple Sensors are not intended for the life monitoring of high risk patients, do not include or trigger alarms, and are not intended to be used as a critical component of:
- an alarm or alarm system;
- an apnea monitor or apnea monitoring system; or
- life monitor or life monitoring system.
{6}------------------------------------------------
#### COMPARISON OF INTENDED USE TO THE PREDICATE DEVICE (PARENT DEVICE TYPE) AND REFERENCE DEVICES (PSG ACCESSORIES AND PHYSIOLOGICAL SIGNAL AMPLIFIER)
#### Indications for Use
## Predicate Device K142774 Neurotronics, Inc. Polysmith Sleep System (includes Sphinx PMU710- Sleep Amplifier)
Rx; The device is intended to measure, amplify, and record physiological signals acauired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in realtime or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of sleep disorders.
The device is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional.
The device is intended for use on both adults and children only under the direction of a physician or qualified sleep technician.
The device, or any accessory, does not include alarms, and is not intended to be used as a critical component of an alarm system.
The device, or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
The device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
#### Discussion: Comparison of Intended Use to the Predicate Device K142774
The intended use of the Serenity Piezo and Serenity Thermocouple Sensors is similar to the primary predicate K142774, the parent device type for which they are intended to be an accessory. Both the predicate and subject devices:
- Are intended to be used for archival of a sleep study, ●
- Are intended for use with adults and children ●
- Do not include alarms and are not intended to be used as a critical component of . an alarm system
- . Are not to be used as an apnea monitor or as a component in an apnea monitoring system
- . Not to be used as a life support device or a critical component of a life support system
The Sphinx Sleep Amplifier is intended to be used in a medical facility or outside a medical facility under direct supervision of a medical professional, whereas the Serenity sensors are intended to be used in a medical facility or outside at the direction of a medical professional. The limitation of direct supervision by the Polysmith Sleep System is due to the amplifier parent device and not because of the sensors, therefore the Nomad
{7}------------------------------------------------
PMU800 is identified as a reference device since it can be used with the sensors without the limitation of direct supervision.
#### Reference Device K092699 Neurotronics, Inc. Nomad Sleep System Recorder, Model PMU800
Rx; The Nomad device is a digital amplifier capable of measuring bio-potential signals that may be incorporated into a Polysomnogram.
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a Sleep Study, such as Limb Movement, Respiration Effort, and SpO2.
The data may be analyzed on dedicated Polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
This device, or any associated accessories, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
This device, or any associated accessories, is not to be used as a life support device or as a critical component of a life support system.
The device is not sterile.
#### Discussion: Comparison of Intended Use to Reference Device K092699, also a PSG parent device type cleared under 882.1835 Physiological signal amplifier
The intended use of the Serenity Piezo and Serenity Thermocouple Sensors is similar to the intended use of the reference device Nomad PMU800, also a device type with which the sensors are intended to be as an accessory. Both are intended to be used for archival of a sleep study, not to be used as an apnea monitor or as a component in an apnea monitoring system, not to be used as a life support device or a critical component of a life support system, and are not sterile.
The Nomad PMU800 is also intended to be used in or outside a medical facility at the direction of a medical professional, this is consistent with the intended use of the Serenity Piezo Sensor and Serenity Thermocouple Sensor. The Nomad 510k and Indications for Use statement includes thermal airflow as respiration effort. Use for children or adults, is also described in the 510k summary for K092699.
#### Reference Device: K042253 S.L.P. LTD. Sleepsense Sleep Sensors
Rx; SleepSense sensors provide a qualitative measure of a patient's physiological parameters for recording onto an FDA-cleared data acquisition system. Their target population: Children and adult patients who are screened during sleep disorder studies. Their environment of use is usually at a sleep laboratory or sometimes at the patient's home.
{8}------------------------------------------------
#### Discussion: Comparison of Intended Use to Reference Device K042253 Sleep Sense Sensors, cleared under 868.2375 Breathing frequency monitor
The intended use of the Serenity Piezo and Thermocouple Sensor is identical to the S.L.P. LTD. Sleepsense Sleep Sensors. K042253 is also intended to provide a qualitative measure of a patient's physiological parameters for recording onto an FDA-cleared data acquisition system. The target population is also children and adults. The environment of use also includes sleep laboratories and the home.
#### Reference Device: K940014 Pro-Tech, Inc. PLM Sensor
Rx; The PLM Sensor is intended for use during sleep disorder studies to detect periodic limb movements for recording onto a physiological recorder.
#### Discussion: Comparison of Intended Use to Reference Device K940014 cleared under 868.2375 Breathing frequency monitor
The Serenity Piezo Sensor can be used as a limb movement sensor, which is the intended use of K940014 Pro-Tech PLM Sensor.
#### Comparison of Technological Characteristics to Reference Devices (PSG Sensors/ Accessories)
The Neurotronics Serenity Piezo Sensor and Serenity Thermocouple Sensor are similar in design and characteristics. Differences between these devices do not affect the intended use of the subject device for use as PSG accessories.
| Piezo Sensor Comparison | | K042253 | K940014 | K181709 |
|---------------------------------|--------------------------------|------------------------------------|---------------------------------|-------------------------------------------|
| | | S.L.P. Piezo<br>Sensor (Snore) | PRO-TECH PLM<br>SENSOR | Neurotronics<br>Serenity Piezo<br>Sensor |
| Image | | Image: S.L.P. Piezo Sensor (Snore) | Image: PRO-TECH PLM SENSOR | Image: Neurotronics Serenity Piezo Sensor |
| Physical<br>Properties | Size | ~0.80" (21mm)<br>Diameter | ~ 0.6" (16mm)<br>Diameter | ~ 0.6" (16mm)<br>Diameter |
| | Sensor Enclosure<br>Material | TPE | TPE | TPE |
| | Cable Material | PVC | PVC | PVC |
| Environmental<br>Specifications | Operating<br>Temperature | 40°F to 104°F<br>(5°C to 40°C) | 40°F to 104°F<br>(5°C to 40°C) | 50°F to 104°F<br>(10°C to 40°C) |
| | Operating<br>Relative Humidity | 5% to 95% RH<br>non-condensing | 15% to 95% RH<br>non-condensing | 0 to 93% RH non-<br>condensing |
## Piezo Sensor Comparison
{9}------------------------------------------------
| Piezo Sensor Comparison | | K042253 | K940014 | K181709 |
|--------------------------|---------------------------------------------------|----------------------------------|----------------------------------|------------------------------------------|
| | | S.L.P. Piezo<br>Sensor (Snore) | PRO-TECH PLM<br>SENSOR | Neurotronics<br>Serenity Piezo<br>Sensor |
| | Storage<br>Temperature | -4°F to 140°F<br>(-20°C to 60°C) | -4°F to 140°F<br>(-20°C to 60°C) | -4°F to 140°F<br>(-20°C to 60°C) |
| | Storage Relative<br>Humidity | 5% to 95% RH<br>non-condensing | 15% to 95% RH<br>non-condensing | 0 to 93% RH non-<br>condensing |
| Output<br>Specifications | Comparison<br>Output Signal<br>Testing<br>(Limb) | Equivalent | Not Applicable | Equivalent |
| | Comparison<br>Output Signal<br>Testing<br>(Snore) | Not Applicable | Equivalent | Equivalent |
# Thermocouple Sensor Comparison
| Thermocouple Sensor Comparison | | | |
|---------------------------------|-------------------------------------|-------------------------------------|--------------------------------------------------|
| | | | |
| | | K042253 | K181709 |
| | | S.L.P.<br>Thermocouple<br>Sensor | Neurotronics Serenity<br>Thermocouple Sensor |
| Image | | Image: S.L.P. Thermocouple Sensor | Image: Neurotronics Serenity Thermocouple Sensor |
| Physical<br>Properties | Material | TPE | TPE |
| Environmental<br>Specifications | Operating<br>Temperature | 40°F to 104°F<br>(5°C to 40°C) | 50°F to 104°F<br>(10°C to 40°C) |
| | Operating Relative<br>Humidity | 15% to 95%<br>RH non-<br>condensing | 0% to 93% RH non-<br>condensing |
| | Storage<br>Temperature | -4°F to 140°F<br>(-20°C to 60°C) | -4°F to 140°F<br>(-20°C to 60°C) |
| | Storage Relative<br>Humidity | 10% to 95% RH<br>non-condensing | 0% to 93% RH non-<br>condensing |
| Output<br>Specifications | Comparison Output<br>Signal Testing | Equivalent | Equivalent |
{10}------------------------------------------------
# Performance Data
## Non-Clinical Testing
Each test identified below includes methodology, criteria, and results for demonstrating performance and safety of the subject device. Neither the S.L.P. Sleepsense Sleep Sensors or Pro-tech sensors are found to publish testing to a standard.
| Test Type | Test Method Summary | Results |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | 60601-1:2003 Ed.1 + R:26 Apr 2006 Dielectric Strength Ingress of Liquids - IPX2 Patient Leads (21CFR898 Performance<br>Standard for Electrode Lead Wires and<br>Patient Cables) | All samples passed the<br>acceptance criteria.<br>Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish testing to an<br>electrical safety standard. |
| Piezo Sensor<br>Verification | Dielectric strength 1.5 kVAC, 10s ramp, 1 min Signal Level Movement and vibration clearly visible<br>with recommended configuration. Output signal within listed specifications. Wire Test Connector Retention >= 4.5N Tensile Strength >= 50 N Leadwire Resistance <= 50 Ohms Mating Cycles >= 3,650 | All samples passed the<br>acceptance criteria.<br>Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish testing details. |
| Thermocouple<br>Sensor<br>Verification | Dielectric strength 1.5 kVAC, 10s ramp, 1 min Signal Level Oral and Nasal breathing clearly visible<br>at sensitivity of 20. Output signal within listed specifications. Wire Test Connector Retention >= 4.5N Tensile Strength >= 50 N Leadwire Resistance <= 50 Ohms Mating Cycles >= 3,650 | All samples passed the<br>acceptance criteria.<br>Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish testing details. |
| Reference<br>Device<br>Comparison | Piezo Sensor Reference Device<br>Comparison Sensor snore response to vibration<br>relative to noise floor (SNR) at varied<br>frequencies and complex waveforms<br>using recommended polysomnography<br>montage configuration. Sensor limb movement response to<br>movement. Output signal within listed specifications. Thermocouple Sensor Reference Device<br>Comparison Sensor response as warm air passes | Comparison testing shows<br>equivalent performance of<br>the Serenity sensors and<br>the reference devices using<br>the same host system<br>configurations.<br>Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish comparison<br>information. |
| Test Type | Test Method Summary | Results |
| | recommended polysomnography<br>montage configuration.<br>• Signal cessation attenuated by >=90%<br>of pre-event baseline.<br>• Output signal within listed specifications. | |
| Biocompatibility | • ISO 10993-1 Fourth edition 2009-10-15<br>Biological evaluation of medical devices<br>- Part 1: Evaluation and testing within a<br>risk management process<br>• ISO 10993-5:2009 Biological<br>Evaluation of Medical Devices - Part 5,<br>Tests for in vitro Cytotoxicity<br>• ISO 10993-10:2010 Biological<br>Evaluation of Medical Devices - Part 10,<br>Tests for Irritation and Skin Sensitization | All samples passed the<br>acceptance criteria for the<br>performed biocompatibility<br>testing.<br>Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish biocompatibility. |
| Sterility | Not applicable | Neither the S.L.P. or Pro-<br>tech sensors were found to<br>publish sterility information. |
{11}------------------------------------------------
#### CONCLUSION
Based on the comparisons of the intended use and technological characteristics, as well as the results of the comparative testing, the Serenity Piezo Sensor and Serenity Thermocouple Sensors are substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.