K142774 · Neurotronics, Inc. · OLV · Feb 18, 2015 · Neurology
Device Facts
Record ID
K142774
Device Name
Polysmith Sleep System
Applicant
Neurotronics, Inc.
Product Code
OLV · Neurology
Decision Date
Feb 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K142774 · Feb 18, 2015
Polysmith Sleep System
Neurotronics, Inc.
Existing sleep studies from accredited clinical laboratories
Retrospective clinical sleep studies were used to evaluate the performance and accuracy of the device's automated analysis algorithm (sleep staging, microarousal, apnea, hypopnea, desaturation, and limb movement) by comparing automated results against human expert scoring.
Retrospective data; Automated analysis validation; Sleep study records
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of existing clinical sleep studies
Patients undergoing sleep studies in accredited labs; Sample Size: 8,053 recorded epochs (7,606 scored); Number of Sites: Independent sleep lab
Human expert scoring (two registered sleep clinicians)
Agreement between automated analysis and human scoring for sleep staging and event detection
Retrospective analysis of 20 sleep studies
20 patients (14 male, 6 female; age range 16.0-74.3); Sample Size: 20 studies; Number of Sites: At least three different sources
Consensus of three independent human scorers
Accuracy of automated analysis algorithm compared to human consensus
Indications for Use
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in real-time or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of sleep disorders. The device is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional. The device is intended for use on both adults and children only under the direction of a physician or qualified sleep technician. The device, or any accessory, does not include alarms, and is not intended to be used as a critical component of an alarm system. The device, or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system. The device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
Device Story
Polysmith Sleep System (NTI6600) records/amplifies physiological signals (EEG, ECG, EMG, respiration, SpO2, CO2, ocular motility) for sleep studies. System includes amplifier, head box, PC, sensors, and optional audio/video inputs. Used in hospitals, clinics, or supervised home settings by clinicians/technicians. Software processes data for real-time or offline display/scoring. Automated analysis algorithms assist in identifying sleep stages, arousals, apneas, hypopneas, desaturations, and limb movements. Clinicians review waveforms and automated scores to diagnose sleep disorders. Benefits include standardized data acquisition, efficient scoring, and diagnostic support.
Clinical Evidence
Bench testing and performance validation. Automated analysis algorithm evaluated against human expert scoring using 8,053 epochs (initial study) and 20 sleep studies (secondary study). Performance metrics: Sleep Staging (71.53% agreement), Microarousal (80.15%), Apnea (97.28%), Hypopnea (95.44%), Desaturation (95.72%), Limb Movement (92.57%). Secondary study showed 70.35% total agreement with human consensus (kappa 0.58).
Technological Characteristics
System includes amplifier, head box, PC, and sensors. 16-bit A-D conversion; 1000 Hz sampling. Input impedance >= 10 M Ohm; CMRR >= 90 dB. Connectivity via Ethernet. Software supports Windows (32/64-bit). Standards: AAMI/ANSI/IEC 62304, 62366, 60601-1, 60601-1-2, 60601-2-26; AAMI/ANSI/ISO 14971.
Indications for Use
Indicated for adults and children requiring polysomnography (PSG) or sleep studies to aid in the diagnosis of sleep disorders. Used by healthcare professionals in clinical settings or under professional supervision. Contraindicated for use as a standalone apnea monitor, life support device, or critical alarm system component.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them, representing the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2015
Neurotronics, Inc. David Pezet Quality Manager 3600 NW 43td Street, Suite F1 Gainesville, FL 32606
Re: K142774
Trade/Device Name: Polysmith Sleep System NTI6600 Regulatory Class: II Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Product Code: OLZ, OLV, DQA Dated: January 12, 2015 Received: January 13, 2015
Dear Mr. Pezet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K142774
Device Name Polysmith Sleep System, Model NTI6600
#### Indications for Use (Describe)
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals and conditioned for analysis and display. The data may be analyzed in real-time or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
The device is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional.
The device is intended for use on both adults and children only under the direction of a physician or qualified sleep technician.
The device or any accessory, does not include alarms, and is not intended to be used as a critical component of an alarm system.
The device or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
The device or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(k) Summary
## SUMMARY DATE:
## February 18, 2015
## 510K SUBMITTER:
| Neurotronics, Inc. | | Phone: | 352.372.9955 |
|---------------------------------------------------------|-----------------------------------------------|--------|----------------------|
| 3600 NW 43rd Street, Suite FL<br>Gainesville, FL, 32606 | | Fax: | 815.550.2871 |
| Primary Contact: | | | |
| Name: | David Pezet | Title: | Quality Manager |
| Email: | quality@neurotronics.com | Phone: | 352.372.9955 Ext 355 |
| | Establishment Registration Number:<br>1063925 | | |
## DEVICE TYPE (COMMON NAME):
Polysomnography Monitoring System
PROPRIETARY NAME OF THE DEVICE:
Polysmith Sleep System, Model NTI6600
{4}------------------------------------------------
## CLASSIFICATION:
# Product Code:
| Product<br>Code | Device | Regulation<br>Description | Regulation<br>Number |
|-----------------|----------------------------------------------------------------------------------------|---------------------------|----------------------|
| OLZ | Automatic Event Detection Software<br>For Polysomnograph With<br>Electroencephalograph | Electroencephalograph. | 882.1400 |
# Subsequent Product Code:
| Product<br>Code | Device | Regulation<br>Description | Regulation<br>Number |
|-----------------|-------------------------------------------------------|---------------------------|----------------------|
| OLV | Standard Polysomnograph With<br>Electroencephalograph | Electroencephalograph. | 882.1400 |
| DQA | Oximeter | Oximeter. | 870.2700 |
## PREDICATE DEVICES
| Submitter/Holder | Device Name | Model | Product Codes | 510(K) |
|--------------------|-------------------------------------|----------|---------------|---------|
| Neurotronics, Inc. | Polysmith Sleep<br>System | NTI5498 | OLZ, DQA, OLV | K062943 |
| Nihon Kohden Corp. | PSG-1100 Sleep<br>Diagnostic System | PSG-1100 | GWQ, OLV | K120888 |
{5}------------------------------------------------
#### DESCRIPTION
The Polysmith Sleep System, Model NTI6600 is intended to amplify and record physiologic potentials used for Polysomnography (PSG) or Sleep Studies. The device consists of a compatible amplifier, head box, PC, patient sensors, and may include optional external devices, USB DC Box, and audio/video input devices.
Compatible amplifiers may use commercially available sensors and electrodes, an internal SpO2 module, and internal pressure transducers to collect, digitize, and send physiological signals to the host PC.
The Polysmith software may record from video, speaker and microphone equipment. The Polysmith software may also record auxiliary signals from compatible amplifiers or USB DC Box which allow for data inputs from compatible sources.
Polysmith records and displays the data for online or offline review. Qualified practitioners use the information to score Polysomnograms and diagnose Sleep Disorders.
Image /page/5/Figure/5 description: This image is a block diagram of a patient monitoring system. The diagram shows the different components of the system, including patient sensors, an amplifier, a PC, and optional audio/video equipment. The patient sensors are connected to the amplifier, which is connected to the PC. The PC is also connected to the audio/video equipment and a USB DC box.
1 Polysmith Sleep System Basic Block Diagram
{6}------------------------------------------------
#### INTENDED USE
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in real-time or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of sleep disorders.
The device is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional.
The device is intended for use on both adults and children only under the direction of a physician or qualified sleep technician.
The device, or any accessory, does not include alarms, and is not intended to be used as a critical component of an alarm system.
The device, or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
The device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
{7}------------------------------------------------
#### PREDICATE COMPARISON
#### Intended Use Comparison
The Polysmith Sleep System, Model NTI6600 improves upon the predicates' current capabilities access and review of on-line or previously recorded acquisitions. These improvements do not change the indications for use for the devices which it is an accessory. This device is intended to aid in diagnosis of sleep disorders, and other related disorders.
#### Technical Comparison
The Polysmith Sleep System. Model NTI6600 includes and improves upon the same feature set as found in the predicates. Therefore Neurotronics believes the Polysmith Sleep System, Model NTI6600 is substantially equivalent to the Polysmith Sleep System, Model NTI5498 (K062943), and PSG-1100 Sleep Diagnostic System (K120888).
| Software Technical<br>Comparison | Polysmith Sleep<br>System,<br>NTI6600<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System,<br>NTI5498<br>Polysmith<br>(K062943) | PSG-1100<br>(K120888) | |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Display electrical activity of<br>the brain and other<br>physiological signals on a<br>monitor | Yes | Yes | Yes | |
| Device uses Windows on a<br>32-bit or 64-bit computer | 32-bit & 64-bit | 32-bit | 32-bit & 64-bit | |
| Display waveforms in<br>acquisition and review | Yes | Yes | Yes | |
| Maximum number of<br>waveforms on screen | 32 | 32 | 32 | |
| Simultaneously view<br>waveforms from multiple<br>EEG/PSG instruments and<br>display values from<br>external instruments | On-line & Off-line | On-line & Off-line | On-line & Off-line | |
| Variable waveform display<br>duration | On-line & Off-line | On-line & Off-line | On-line & Off-line | |
| Change waveform<br>parameters (pattern,<br>montage, amplifier<br>conditions, AV induction) | On-line & Off-line | On-line & Off-line | On-line & Off-line | |
| The device can record and<br>measure vital signs<br>(including ECG, EMG, | On-line & Off-line | On-line & Off-line | On-line & Off-line | |
| Software Technical<br>Comparison | Polysmith Sleep<br>System,<br>NTI6600<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System,<br>NTI5498<br>Polysmith<br>(K062943) | PSG-1100<br>(K120888) | |
| Respiration, Ocular<br>Motility, SpO2, and CO2) | | | | |
| Nihon Kohden Amplifiers<br>Compatibility | Yes | Yes | Yes | |
| Lifelines Trackit Series<br>Amplifiers Compatibility | Yes | Yes | Yes | |
| USB DC Box Compatibility | Yes | Yes | Yes | |
| Sleep Stage Analysis | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Arrhythmia Analysis | Off-line<br>(integrated<br>K012686) | Off-line<br>(integrated<br>K012686) | Off-line<br>(integrated<br>K012686) | |
| Respiratory Event Analysis | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| SpO2 Desaturation Analysis | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Limb Movement Event<br>Analysis | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Snore Event Analysis | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| FFT Analysis | Off-line | Off-line | Off-line | |
| Manual Event Scoring | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Manual Custom Event<br>Scoring | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Trend Plots | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Display patient video along<br>with waveforms | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Record Patient Audio and<br>Video | Yes | Yes | Yes | |
| Integrated Intercom<br>Interface | Yes<br>(with PMU710) | No | No | |
| For use by medical<br>personnel in a medical<br>facility, physician's office,<br>laboratory, clinic, or<br>nursing home | Yes | Yes | Yes | |
| Software Technical<br>Comparison | Polysmith Sleep<br>System,<br>NTI6600<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System,<br>NTI5498<br>Polysmith<br>(K062943) | PSG-1100<br>(K120888) | |
| Any patient population<br>including adults and<br>children as determined by<br>a trained professional | Yes<br>Yes<br>Yes | | | |
| Patient Database and<br>Scheduling | Yes | Yes | Yes | |
| Reporting and Custom<br>Report Capabilities | Yes | Yes | Yes | |
| HL7 Interface Capabilities | Yes | Yes | Yes | |
| Remote Access Capabilities | On-line & Off-line | On-line & Off-<br>line | On-line & Off-<br>line | |
| Display Resolution | >=1600 x 1200 | >=1600 x 1200 | >=1600 x 1200 | |
| File Format | XDF (XML), EDF | XDF (XML),<br>EDF | XDF (XML),<br>EDF | |
| File Management | Yes | Yes | Yes | |
| Compatible Amplifiers for<br>Recording | · Neurotronics<br>Sphinx<br>PMU710<br>(NEW)<br>Neurotronics<br>●<br>Sphinx<br>PMU700<br>(K062943)<br>Neurotronics<br>Nomad<br>PMU800<br>(K092699)<br>Nihon Kohden<br>●<br>PSG-1100<br>(K120888)<br>Nihon Kohden<br>●<br>EEG-1200a<br>Amplifiers<br>(K113117,<br>K080546)<br>Nihon Kohden<br>Wireless Input<br>Unit, Model<br>WEE-1000a | Neurotronics<br>●<br>Sphinx<br>PMU700<br>(K062943)<br>Nihon Kohden<br>●<br>Wireless Input<br>Unit, Model<br>WEE-1000a<br>Series<br>(K033475)<br>Nihon Kohden<br>Electrode<br>Junction Box,<br>Model JE-921a<br>(K050833)<br>Nihon Kohden<br>●<br>PSG Input<br>Box, Model<br>JE-912ak<br>(K022121)<br>Nihon Kohden<br>●<br>Neurofax,<br>Models EEG-<br>1100a, EEG-<br>9100a | Neurotronics<br>Sphinx<br>PMU700<br>(K062943)<br>Neurotronics<br>●<br>Nomad<br>PMU800<br>(K092699)<br>Nihon Kohden<br>PSG-1100<br>(K120888)<br>Nihon Kohden<br>EEG-1200a<br>Amplifiers<br>(K113117,<br>K080546)<br>Nihon Kohden<br>●<br>Wireless Input<br>Unit. Model<br>WEE-1000a<br>Series<br>(K033475)<br>Nihon Kohden<br>●<br>Electrode<br>Junction Box, | |
| Software Technical<br>Comparison | Polysmith Sleep<br>System,<br>NTI6600<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System,<br>NTI5498<br>Polysmith<br>(K062943) | PSG-1100<br>(K120888) | |
| | Series<br>(K033475)<br>• Nihon Kohden<br>Electrode<br>Junction Box,<br>Model JE-921a<br>(K050833)<br>• Nihon Kohden<br>PSG Input Box,<br>Model JE-<br>912ak<br>(K022121)<br>• Nihon Kohden<br>Neurofax,<br>Models EEG-<br>1100a, EEG-<br>9100a<br>(K011204,<br>K992742)<br>• Lifelines<br>Trackit Series<br>amplifiers<br>(K010460) | (K011204,<br>K992742)<br>• Lifelines<br>Trackit Series<br>amplifiers<br>(K010460) | Model JE-921a<br>(K050833)<br>• Nihon Kohden<br>PSG Input<br>Box, Model<br>JE-912ak<br>(K022121)<br>• Nihon Kohden<br>Neurofax,<br>Models EEG-<br>1100a, EEG-<br>9100a<br>(K011204,<br>K992742)<br>• Lifelines<br>Trackit Series<br>amplifiers<br>(K010460) | |
| Amplifier Technical<br>Comparison | Polysmith Sleep<br>System, NTI6600<br>Sphinx PMU710<br>(NEW) | Polysmith Sleep<br>System, NTI5498<br>Sphinx PMU700<br>(K062943) | PSG-1100<br>(K120888) | |
| Number of Channels | 23 | 33 | 42 | |
| Pulse Oximeter | Integrated Nihon<br>Kohden Oximeter<br>Module | Integrated Nonin<br>Oximeter Module | Integrated Nihon Kohden<br>Oximeter Module | |
| Input Impedance | >= 10 M Ohm | >= 10 M Ohm | 100 M ohm | |
| Calibration Check | Step square 50uV | Step square 50uV | Step square 50uV | |
| Impedance Check | Yes | Yes | Yes | |
| Common-Mode Rejection Ratio<br>(CMRR) | >= 90 dB | >= 90 dB | >= 105 dB | |
| Noise Level | < 2 µV p-p<br>(0.1 to 60 Hz) | < 2 uV p-p<br>(0.53 to 60 Hz) | < 1.5 uV p-p<br>(0.53 to 60Hz) | |
| Hardware Frequency Response | .072 Hz to 300 Hz | 0.1 Hz to 100 Hz | 0.08 to 300 Hz | |
| Software High-pass Filter | .072 Hz<br>DC Standard | .1 Hz<br>DC Standard | 0.08 to 53 Hz<br>DC Standard | |
| Software Low-pass Filter | 15 Hz to 300 Hz | 15 Hz to 100 Hz | 15 to 300 Hz | |
| AC Filter | 50 or 60 Hz | 50 or 60 Hz | 50 or 60 Hz | |
| Sensitivity | Off, 0.1 to 200<br>uV/mm (20 steps)<br>DC: Off, 10 to 200<br>mV/mm (10 steps) | Off, 0.1 to 200<br>uV/mm (20 steps)<br>DC: Off, 10 to 200<br>mV/mm (10 steps) | Off, 0.1 to 200 uV/mm (20<br>steps)<br>DC: Off, 10 to 200 mV/mm<br>(10 steps) | |
| A-D Conversion | 16 bits | 16 bits | 16 bits | |
| Sampling | 1000 Hz | 1000 Hz | All Channels<br>200, 250, 500,<br>1000, 2000 Hz | |
| Power | Mains | USB | 100-240 V +/- 10%<br>50 / 60 Hz<br>42 VA (Main Unit only) | |
| PC Connection | Ethernet | USB…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.