K092699 · Neurotronics, Inc. · GWL · Mar 18, 2010 · Neurology
Device Facts
Record ID
K092699
Device Name
NOMAD SLEEP SYSTEM RECORDER, MODEL PMU800
Applicant
Neurotronics, Inc.
Product Code
GWL · Neurology
Decision Date
Mar 18, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1835
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a Sleep Study, such as Limb Movement, Body Position, Respiration fffort, and SpO2. The data may be analyzed in real-time or offline on dedicated polystemporably software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders. This device, or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system. This device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system. Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
Device Story
Nomad is a digital amplifier for polysomnography (PSG) or sleep studies; captures bio-potential signals including limb movement, body position, respiration effort, and SpO2. Device transfers raw physiological data to Polysmith software on a personal computer. Operated by qualified sleep clinicians or technicians in clinical settings; used under physician direction. Clinicians review recorded data to score polysomnograms and diagnose sleep disorders. Device does not directly contact patients; uses standard accessories (e.g., oximeter finger probes). Benefits include standardized signal acquisition for sleep disorder assessment.
Clinical Evidence
Bench testing only. Device underwent electromagnetic, environmental, mechanical stress, and medical device safety testing. Software functionality was verified and validated against product specifications. No clinical data required.
Technological Characteristics
Digital physiological signal amplifier. Connectivity: transfers data to PC via Polysmith software. Complies with IEC 60601-1 and 21 CFR 898 for electrode lead wires and patient cables. Non-sterile. Class II device.
Indications for Use
Indicated for adults and children requiring measurement, amplification, and recording of physiological signals (Limb Movement, Body Position, Respiration Effort, SpO2) for polysomnography/sleep studies to aid in sleep disorder diagnosis. Not for use as a standalone apnea monitor or life support system.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
Predicate Devices
Neurotronics Sphinx
Submission Summary (Full Text)
{0}------------------------------------------------
K092699
#### 510(K) SUMMARY
## Name and Address of Applicant
Nomad is designed and distributed by: Neurotronics®, Inc. 912 NE 2nd Street Gainesville, FL, 32601 Phone.: 352.372.9955
Contact Information:
James Schubert Vice President Tel: 352.372.9955 Ext. 309 Email: James.schubert@neurotronics.com
Fax: 815.550.2871
#### Indications for Use:
The Nomad device is a digital amplifier capable of measuring bio-potential signals that may be incorporated into a Polysomnogram.
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a Sleep Study, such as Limb Movement, Respiration Effort, and SpO2. The data may be analyzed on dedicated Polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Ding on s
This device, or any associated accessories, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
This device, or any associated accessories, is not to be used as a life support device or as a critical component of a life support system.
The device is not sterile.
The device does not directly contact patients. Accessories that contact patients, such as oximeter finger probe, are the same accessories used with other legally marketed products. Jan a oxime, good laboratory practice studies were not required per 21 CFR Part 58.
The device was subjected to electromagnetic, environmental, safety and performance testing. The results of validation and verification confirmed that the device performed within expected specifications. The risk evaluation also confirmed all the risks have been properly mitingted.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
***** The device is substantially equivalent to the predicate as stated.
{1}------------------------------------------------
#### Description
្រី
ﺮﺓ ﺍﻟ
The device is intended to amplify and record physiologic potentials used for Polysomnography (PSG) or Sleep Studies. The bio-potentials are transferred to Polysmith polysomnography software running on a personal computer. Qualified practitioners use the information to score Polysomnograms and diagnose Sleep Disorders. The device is intended for use on both adults and children under the direction of a physician or qualified sleep technician.
#### Intended Use
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a Sleep Study, such as Limb Movement, Body Position, Respiration fffort, and SpO2. The data may be analyzed in real-time or offline on dedicated polystemporably software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
This device, or any accessory, is not to be used alone as an apnea monitor or as a component in an apnea monitoring system.
This device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
#### Performance and Safety Testing
The accessories of the device comply with IEC 60601-1 Subclause 56.3(c) as determined by 21 CFR Part 898: Performance Standard for Electrode Lead Wires and Patient Cables.
The device was verified and validated according to the product specifications. The test criteria consists of standardized levels and internal product requirements. Tests performed on the device include environmental and mechanical stress testing, electromagnetic immunity and emissions. testing, and medical device safety testing. Software on the device was verified and validated according to the functionality of the operations of the device. The test results confirm that the device is in accordance with its specifications.
### STATEMENT OF SUBSTANTIAL EQUIVALENCE
Based on the statements made above, the device and the predicates have the same intended use, the same principles of operation, similar physical/functional descriptions, and equivalent voluntary standards.
Therefore, Neurotronics, Inc. believes that the Nomad is substantially equivalent to the predicate device, the Neurotronics Sphinx.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight. The symbol is composed of three curved lines that converge at the bottom.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
# MAR 1 8 2010
Neurotronics®, Inc. c/o Mr. James Schubert Vice President 912 NE 2nd Street Gainesville, FL 32601
Re: K092699
Trade/Device Name: NOMAD, Polysmith Sleep System Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological Signal Amplifier Regulatory Class: Class II Product Code: GWL, MNR, and DQA Dated: February 12, 2010 Received: February 12, 2010
Dear Mr. Schubert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Neurotronics Special 510(k)
#### G. Indications for Use Statement
K092699 510(k) Number (if known): _
Device Name: NOMAD, Polysmith Sleep System
Indications for Use:
The Nomad device is a digital amplifier capable of measuring bio-potential signals that may be incorporated into a Polysomnogram.
The device is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a Sleep Study, such as Limb Movement, Respiration Effort, and SpO2. The data may be analyzed on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
This device is not to be used alone as an apnea monitor or as a critical component in an apnea monitoring system.
This device is intended for use on both adults and children on the order of physician.
This device, or any accessory, is not to be used alone as a life support device or as a critical component of a life support system.
The device is not sterile.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
JOE HUTTER
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
NOMAD
Page 8 of 28
510(k) Number K092699
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.