K181204 · Smith & Nephew Medical Limited · OMP · Jan 10, 2019 · General, Plastic Surgery
Device Facts
Record ID
K181204
Device Name
RENASYS Y-Connector; RENASYS TOUCH
Applicant
Smith & Nephew Medical Limited
Product Code
OMP · General, Plastic Surgery
Decision Date
Jan 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RENASYS Y-Connector can only be used with RENASYS TOUCH and RENASYS GO systems. RENASYS TOUCH and RENASYS GO are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. The following wound types may be used with a RENASYS system that is utilizing a RENASYS Y-connector: Flaps and grafts (only in one wound configuration) Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit) Chronic Acute Traumatic Sub-Acute and dehisced wounds Ulcers (such as pressure or diabetic) Partial-thickness burns The RENASYS TOUCH is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional. When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary bridging of abdominal wall openings where is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Device Story
RENASYS Y-Connector is an accessory for RENASYS TOUCH and RENASYS GO negative pressure wound therapy (NPWT) pumps. Device allows connection of two wounds to a single pump via PVC tubing and soft ports. Used in acute care settings by healthcare professionals. System applies negative pressure to wound beds to remove fluids, exudates, and infectious materials, promoting healing. RENASYS TOUCH GUI updated to include 'Y-Connector' symbol when in use. Device performance verified via bench testing for pressure maintenance, alarm functionality (leak/blockage), and tensile strength. Benefits include efficient management of multiple wounds or complex abdominal openings using existing NPWT infrastructure.
Clinical Evidence
Bench testing only. Simulated cavity wound models confirmed the NPWT system maintains required negative pressure (min 95% duration) across all configurations, including intermittent therapy and viscous fluids. Alarm characterization confirmed leak and blockage alarms function as intended. Tensile strength testing met ISO 8536-4 standards.
Technological Characteristics
Constructed of PVC tubing. Interfaces with RENASYS TOUCH/GO pumps and RENASYS Foam/Gauze/Abdominal dressing kits. Uses disposable canisters for exudate collection. Tensile strength meets ISO 8536-4. Biocompatibility per ISO 10993-1. Software update to RENASYS TOUCH GUI (model 66801281) to include Y-Connector symbol.
Indications for Use
Indicated for patients requiring negative pressure wound therapy (NPWT) to promote healing via removal of fluids, exudates, and infectious materials. Applicable to chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, and flaps/grafts. Also indicated for temporary bridging of open abdominal wounds with exposed viscera (e.g., abdominal compartment syndrome) when used with RENASYS AB Abdominal Kit. Contraindicated for flaps/grafts or open abdomen in two-wound configurations, and for connecting infected and non-infected wounds together.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
RENASYS AB Abdominal Dressing Kit with Soft Port (K143133)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 10, 2019
Smith & Nephew Medical Limited Sam Greenhalgh Senior Regulatory Affairs Specialist 101 Hessle Road Hull, HU3 2BN Gb
Re: K181204
Trade/Device Name: RENASYS Y-Connector; RENASYS TOUCH Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: November 30. 2018 Received: December 11, 2018
Dear Sam Greenhalgh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kimberly Ferlin -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
#### K181204
Device Name RENASYS Y-Connector
Indications for Use (Describe)
The RENASYS Y-Connector can only be used with RENASYS TOUCH and RENASYS GO systems.
RENASYS TOUCH and RENASYS GO are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids. wound exudates and infectious materials.
The following wound types may be used with a RENASYS system that is utilizing a RENAS YS Y-connector:
- Flaps and grafts (only in one wound configuration)
- Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit)
- Chronic .
- . Acute
- . Traumatic
- . Sub-Acute and dehisced wounds
- . Ulcers (such as pressure or diabetic)
- Partial-thickness burns
The RENASYS TOUCH is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary bridging of abdominal wall openings where is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181204
#### Device Name RENASYS TOUCH
Indications for Use (Describe)
RENASYS TOUCH is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
Appropriate wound types include:
- · Chronic
- · Acute
- · Traumatic
- · Sub-Acute and dehisced wounds
- · Ulcers (such as pressure or diabetic)
- · Partial-thickness burns
- · Flaps and grafts
The RENASYS TOUCH is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When used with the RENASYS AB Abdominal Kit with Soft Port, RENASYS TOUCH is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### K181204
## General Information
| Submitter Name/ Address: | Smith & Nephew Medical Limited<br><br>101 Hessle Road,<br>Hull,<br>HU3 2BN<br>United Kingdom |
|------------------------------------|----------------------------------------------------------------------------------------------|
| Establishment registration Number: | 8043484 |
| Contact Person: | Sam Greenhalgh, Senior Regulatory Affairs<br>Specialist |
| Phone Number: | +44 1482 673436 |
| Date Prepared: | May 4th, 2018 |
| Device Description | |
| Trade Name: | RENASYS Y-Connector; RENASYS<br>TOUCH |
| Common or Usual Name: | Negative Pressure Wound Therapy Suction<br>Pump |
| Device Classification: | Powered suction pump (21 CFR 878.4780) |
| Regulatory Class: | II |
| Product Code: | OMP |
{5}------------------------------------------------
## Predicate Device Information
| 510(k)# | Device | Clearance Date |
|---------|----------------------------------------------------------------------------------------|-------------------|
| K153209 | RENASYS Y-Connector<br>(Primary Predicate) | August 04, 2016 |
| K151872 | Avance Y-Connector S<br>(Secondary Predicate) | November 02, 2015 |
| K143133 | RENASYS AB Abdominal<br>Dressing Kit with Soft Port<br>Device (Secondary<br>Predicate) | July 23, 2015 |
## Device Description
The RENASYS Y-connector is an accessory device to the previously cleared RENASYS TOUCH (K153209) pump for use on one wound only. It is only intended to be used in conjunction with RENASYS Negative Pressure Wound Therapy (NPWT) system which is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
{6}------------------------------------------------
The RENASYS Y-Connector can be used to connect one or two wounds of the same etiologies through two RENASYS Soft Ports to a single pump. The Y-Connector is constructed of PVC tubing and is compatible with Smith & Nephew RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits with Soft Port (it can only be used on one wound with the RENASYS Abdominal dressing kits and only with RENASYS TOUCH).
## Indications for Use
The RENASYS Y-Connector can only be used with RENASYS TOUCH and RENASYS GO systems.
RENASYS TOUCH and RENASYS GO are indicated for patients who would benefit from a suction device (neqative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
The following wound types may be used with a RENASYS system that is utilizing a RENASYS Y-connector:
- . Flaps and grafts (only in one wound configuration)
- . Open abdomen (only in one wound configuration & only with RENASYS TOUCH and RENASYS AB Abdominal Kit)
- . Chronic
- Acute
- . Traumatic
- Sub-Acute and dehisced wounds
- Ulcers (such as pressure or diabetic)
- . Partial-thickness burns
{7}------------------------------------------------
The RENASYS TOUCH is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When the RENASYS Y-Connector is used with the RENASYS AB Abdominal Kit with Soft Port (only in one wound configuration), it is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
#### Comparison of Technological Characteristics
The subject device, RENASYS Y-Connector is substantially equivalent to the primary and secondary predicate devices: RENASYS Y-Connector (K153209) and Avance® Y-Connector S (K151872). The subject and predicate devices have similar indications for use, contraindications and precautions and are intended to be used in conjunction with the same negative pressure wound therapy technology.
The subject device interfaces with RENASYS TOUCH and RENASYS GO NPWT pumps in conjunction with RENASYS Foam, RENASYS Gauze and RENASYS Abdominal Dressing Kits with Soft-Port which use disposable canisters for exudate collection. The technology utilised is the same as for the predicate devices.
Main differences between the subject device and the primary predicate device RENASYS Y-Connector (K153209) is an increase of number of wounds to which the cleared Y-
{8}------------------------------------------------
Connector can be connected i.e. being able to use the Y-Connector on 1 wound to 2 wounds and the inclusion of specific device instructions for use.
To facilitate the use of RENASYS Y-Connector on two wounds with the RENASYS TOUCH device the GUI screen wording and includes a new 'Y-Connector' symbol to the screen when the Y-Connector is being used. This has resulted in a new model number for RENASYS TOUCH; the new code is 66801281.
The Y-Connector cleared as part of K153209 was only cleared for use with RENASYS TOUCH systems. As part of this submission, RENASYS GO will be added as a compatible device.
The Y-Connector cleared as part of K153209 was not cleared for use with the RENASYS AB kit. As part of this submission, RENASYS AB Abdominal kit with Soft Port has been included as a compatible device in a one wound configuration with RENASYS TOUCH.
#### Comparison between New and Predicate Devices
The Indications for Use statement for the subject device, RENASYS Y-Connector is the same as for the primary predicate device (RENASYS Y-Connector K153209). The Indications for Use Statement for the primary predicate device will be transferred to the subject device with the addition of a new indication for use on two wounds.
The primary predicate device and the subject of this submission are identical in terms of materials used and technology used.
The main differences between the subject and predicate devices are:
- 1. Increase in number of wounds, which can be used with RENASYS Y-Connector from one to two wounds.
{9}------------------------------------------------
- 2. Addition of device specific Instructions for Use Leaflet.
As a consequence of increasing the number of wounds which can be used with the RENASYS Y-Connector from one to two wounds, the following differences can also be observed between the subject and the primary predicate device Indications for Use and Contraindications:
A. Clarification of wound configuration for the following indication for use:
- Flaps and grafts (only in one wound configuration). .
- B. Addition of the following indication for use:
- · Open abdomen (only in one wound configuration & only with RENASYS TOUCH & only in a hospital environment)
- C. Addition of the following contraindications:
- . Flaps/Grafts in a two wound configuration
- Open abdomen in a two wound configuration .
- An infected and non-infected wound connected together .
Additionally, the RENASYS TOUCH Negative Pressure Wound Therapy device manual has been updated to include use with the RENASYS AB Abdominal Dressing Kit and read as follows:
RENASYS TOUCH is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include:
- . Chronic
- Acute ●
- Traumatic .
- . Sub-Acute and dehisced wounds
- . Ulcers (such as pressure or diabetic)
{10}------------------------------------------------
- . Partial-thickness burns
- . Flaps and grafts
The RENASYS TOUCH is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When used with the RENASYS AB Abdominal Kit with Soft Port, RENASYS TOUCH is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
#### Non-Clinical Tests (Bench)
The following non-clinical (bench) testing has been carried out: Simulated Cavity Wound Model Compatibility Testing
- Test results determined that the RENASYS Negative Pressure Wound Therapy . (NPWT) system was able to maintain negative pressure within the test requirements (as defined within the acceptance criteria) at the simulated wound bed for a minimum of 95% of the test duration for all test configurations. Testing showed that the use of the subject device, RENASYS Y-Connector does not impact the RENASYS NPWT system performance when tested in conjunction with the cleared RENASYS TOUCH and RENASYS GO NPWT pump devices, RENASYS Foam, RENASYS Gauze and RENASYS Abdominal dressing kits with soft port (only with RENASYS TOUCH) therefore demonstrating product compatibility.
{11}------------------------------------------------
Based on the results, it can be concluded that RENASYS Y-Connector can be used for its intended purposes.
#### Simulated Cavity Wound Model Testing using Intermittent Therapy Mode
- Test results determined that the RENASYS NPWT system was able to maintain . negative pressure within the test requirements (as defined within the acceptance criteria) at the simulated wound bed for a minimum of 95% of the test duration for all test configurations when in intermittent therapy mode. Testing showed that the use of the subject device, RENASYS Y-Connector does not impact the RENASYS NPWT system performance when tested in conjunction with the cleared RENASYS TOUCH and RENASYS GO NPWT pump devices RENASYS Foam and RENASYS Gauze/Foam dressing kits with soft port.
Based on the results, it can be concluded that RENASYS Y-Connector can be used for its intended purposes.
#### Simulated Cavity Wound Model Testing using Viscous Simulated Wound Fluid
- Test results determined that the RENASYS NPWT system was able to maintain . negative pressure within the test requirements (as defined within the acceptance criteria) at the simulated wound bed for a minimum of 95% of the test duration for all test configurations when using viscous simulated wound fluid and a worst case test device. Testing showed that the use of the subject device, RENASYS Y-Connector does not impact the RENASYS NPWT system performance when using viscous simulated wound fluid and tested in conjunction with the cleared RENASYS TOUCH and RENASYS GO NPWT pump devices and RENASYS Gauze/Foam dressing kits with soft port.
Based on the results, it can be concluded that RENASYS Y-Connector can be used for its intended purposes.
{12}------------------------------------------------
## RENASYS NPWT Pump Battery Characterisation Testing using a Simulated Cavity Wound Model
- . Test results determined that the RENASYS NPWT system was able to maintain negative pressure within the test requirements (as defined within the acceptance criteria) at the simulated wound bed for a minimum of 95% of the test duration for all test configurations when in continuous therapy mode. Testing showed that the use of the subject device, RENASYS Y-Connector does not impact the RENASYS TOUCH or RENASYS GO pump device batteries or NPWT system performance when tested in conjunction with the cleared RENASYS TOUCH and RENASYS GO NPWT pump devices and RENASYS/Gauze Foam dressing kits with soft port. Based on the results, it can be concluded that RENASYS Y-Connector can be used for its intended purposes.
## Alarm Characterisation Testing using a Simulated Cavity Wound Model
- Testing with the Y-Connector, in conjunction with the cleared RENASYS TOUCH . and RENASYS GO NPWT pump devices, using RENASYS Foam and RENASYS Gauze dressing kits with soft port demonstrates that both the leak and blockage alarms functioned as intended as defined within the acceptance criteria. Based on the results, it can be concluded that RENASYS Y-Connector can be used for its intended purposes.
## RENASYS Y-Connector Robustness Testing
- Tensile strength testing showed that none of the RENASYS Y-Connector units . came apart when pull forces were applied. All units tested met the acceptance criteria in accordance with ISO 8536-4, Section 6.3 Tensile Strength. Based on the results, it can be concluded that the RENASYS Y-Connector can be used for its intended purposes.
{13}------------------------------------------------
## Biocompatibility Testing
The RENASYS Y-Connector Assembly has been evaluated according to the Biological Evaluation of Medical Devices Standard BS EN ISO 10993, with particular reference to Part 1 (2009): Evaluation and testing within a risk management process and FDA guidance 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
The composition of the subject device, RENASYS Y-Connector is identical to that of the primary predicate device, RENASYS Y-Connector (K153209).
The biocompatibility of the RENASYS Y-Connector has been addressed and cleared in 510(k) K153209, August 04, 2016.
## Conclusion
In establishing substantial equivalence to the currently marketed predicate device, Smith & Nephew Medical Limited evaluated the indications for use, materials, technology and product specifications of the device. Performance testing and biocompatibility testing assessment testing has been successfully completed to demonstrate that the RENASYS Y-Connector is substantially equivalent to the predicate devices for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.