K151872 · Molnlycke Health Care Us, LLC · OMP · Nov 2, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151872
Device Name
Avance Y-Connector S
Applicant
Molnlycke Health Care Us, LLC
Product Code
OMP · General, Plastic Surgery
Decision Date
Nov 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Avance® Y-connector S is intended for use in the Avance® Negative Pressure Wound Therapy (NPWT) system, to connect and allow therapy of multiple wounds simultaneously using one pump. The Avance® NPWT system, with associated products, are indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via the removal of fluids, including irrigation and body fluids, wound exudate and infectious materials. Examples of appropriate wound types include: chronic, acute, traumatic, sub-acute and dehisced wounds, ulcers (such as pressure or diabetic), partial-thickness burns, flaps and grafts.
Device Story
Avance® Y-Connector S is a single-lumen accessory for the Avance® NPWT system; enables connection of multiple dressing kits to a single suction pump. Input: wound exudate/fluids from multiple wound sites. Transformation: provides fluid pathway from two transfer pads to one canister tubing line; allows simultaneous negative pressure application to multiple wounds. Output: fluid transport to pump canister. Used in clinical/home settings; operated by healthcare providers or patients. Benefit: facilitates treatment of large or multiple wounds with one device; promotes healing via fluid removal.
Clinical Evidence
No clinical data was required. Bench testing only: biocompatibility (ISO 10993-5, ISO 10993-10) and internal fluid transport performance testing confirmed functionality with multiple wound models.
Technological Characteristics
Polyurethane/polypropylene single-lumen Y-connector; non-sterile; single-use. Biocompatibility per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Connects Avance® canister tubing to two single-lumen transfer pads.
Indications for Use
Indicated for patients requiring negative pressure wound therapy for chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, flaps, and grafts; allows simultaneous therapy of multiple wounds using one pump.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, with three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
Molnlycke Health Care, US LLC Ms. Angela Bunn Director, Regulatory Affairs for the Americas 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092
Re: K151872
Trade/Device Name: Avance® Connector S Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: September 18, 2015 Received: September 21, 2015
Dear Ms. Bunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| <b>Indications for Use</b> |
| Form Approved: OMB No. | 0910-0120 |
|------------------------|--------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number (if known) | K151872 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Avance Y-Connector S |
| Indications for Use (Describe) | Avance Y-connector S is intended for use in the Avance Negative Pressure Wound Therapy (NPWT) system, to connect and allow therapy of multiple wounds simultaneously by using one pump. The Avance NPWT system, with associated products are indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via the removal of fluids, including irrigation and body fluids, wound exudate and infectious materials. Examples of appropriate wound types include: chronic, acute, traumatic, sub-acute and dehisced wounds, ulcers (such as pressure or diabetic), partial-thickness burns, flaps and grafts. |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{3}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFICACY
This 510(k) summary of safety and efficacy information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
| Date Prepared: | October 26, 2015 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway, Suite 500<br>Norcross, GA 30092<br>Registration number:<br>3004763499<br>Owner/Operator Number:<br>8030877 |
| Official Correspondent: | Angela L. Bunn, RAC<br>Director, Regulatory Affairs of the Americas<br>Tel: 678-250-7930<br>Fax: 678-250-7981<br>e-mail: angela.bunn@Mölnlycke.com |
| Trade/Proprietary Name: | Avance® Y Connector S |
| Common Name: | NPWT Dressing Kits |
| Classification Name: | Powered Suction Pump |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4780 |
| Product Code: | OMP |
| Predicate Device Name(s): | K141847 - Avance® Foam Dressing Kit<br>K142626 - Invia Liberty System with Y-Connector |
### Description of Device:
The subject of this submission is the Avance® Y-Connector S, which is an accessory to the Avance® Negative Pressure Wound Therapy (NPWT) System. It allows connection of multiple dressing kits to one pump in order to accommodate large or multiple wounds. We have classified this device as Class II as it is directly indicated for use with an NPWT Pump which falls under that classification, and the Avance® Y-Connector S forms an integral part of the overall system when multiple wound sites require therapy.
The Avance® Y-connector S is a polyurethane/polypropylene based single lumen Yconnector. It is designed to function with existing Avance® NPWT pumps and dressing kits by connecting the Avance® Canister Tubing to two single lumen transfer pads which
{4}------------------------------------------------
are applied to the Avance® Foam dressings. Use of the Avance® Y-Connector S allows therapy and transportation of exudate from more than one wound simultaneously using one pump. A maximum of two Avance® Y-Connector S devices may be used to connect three wounds to one pump.
### Intended Use/Indication for Use:
The Avance® Y-connector S is intended for use in the Avance® Negative Pressure Wound Therapy (NPWT) system, to connect and allow therapy of multiple wounds simultaneously using one pump. The Avance® NPWT system, with associated products, are indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via the removal of fluids, including irrigation and body fluids, wound exudate and infectious materials. Examples of appropriate wound types include: chronic, acute, traumatic, sub-acute and dehisced wounds, ulcers (such as pressure or diabetic), partial-thickness burns, flaps and grafts.
{5}------------------------------------------------
# Predicate Device Comparison Table
| Predicate Device Comparison | | | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Avance® Y-Connector S | Invia Liberty NPWT System with Y connector | Avance® Foam Dressing Kits with ViewPad |
| 510(k) Clearance | Subject of Submission | K142626 | K141847 |
| Manufacturer | Mölnlycke Health Care | Medela | Mölnlycke Health Care |
| Common Name | NPWT Dressing Kits | NPWT Dressing Kits | NPWT Dressing Kits |
| Classification # | 21 CFR 878.4780 | 21 CFR 878.4780 | 21 CFR 878.4780 |
| Class Name | Powered Suction Pump | Powered Suction Pump | Powered Suction Pump |
| Class | II | II | II |
| Product Code | OMP | OMP | OMP |
| Indication For Use/Intended Use | Avance® Y-connector S is intended for use<br>in the Avance® Negative Pressure Wound<br>Therapy (NPWT) system, to connect and<br>allow therapy of multiple wounds<br>simultaneously by using one pump.<br><br>Avance® Y-connector S does not change<br>the intended use of indications for use of the<br>Avance® NPWT System, as stated in<br>K141847. | The Invia Liberty® NPWT system, with<br>associated products are indicated for patients<br>who would benefit from a suction device<br>(negative pressure wound therapy), as it may<br>promote wound healing via the removal of<br>fluids, including irrigation and body fluids,<br>wound exudates and infectious materials.<br>Examples of appropriate wound types<br>include: chronic, acute, traumatic, sub-acute<br>and dehisced wounds, ulcers (such as<br>pressure or diabetic), partial-thickness<br>burns, flaps and grafts. | The Avance® NPWT system, with<br>associated products are indicated for patients<br>who would benefit from a suction device<br>(negative pressure wound therapy), as it may<br>promote wound healing via the removal of<br>fluids, including irrigation and body fluids,<br>wound exudates and infectious materials.<br>Examples of appropriate wound types<br>include: chronic, acute, traumatic, sub-acute<br>and dehisced wounds, ulcers (such as<br>pressure or diabetic), partial-thickness<br>burns, flaps and grafts. |
| Product Feature | | | |
| Y Connector | Y connector used to connect more than one<br>dressing kit to a pump | Y connector used to connect more than one<br>dressing kit to a pump | N/A |
| Tubing | Same tubing as ViewPad; used to connect to<br>the tubing of the Avance® Canister | N/A | ViewPad tubing used to connect to the<br>tubing of the Avance Canister |
| Single Use | Yes | Yes | Yes |
| Non-sterile | Yes | Not known | N/A (no Y Connector component) |
| Biocompatibility | Components meet applicable requirements<br>of ISO 10993-5 and ISO 10993-10 | Components meet applicable requirements<br>of ISO 10993-5 and ISO 10993-10 | Components meet applicable requirements<br>of ISO 10993 |
{6}------------------------------------------------
#### Summary of Performance Testing
The Avance® Y-Connector S has been evaluated for biocompatibility and for performance within the Avance® NPWT System. The results are as follows.
| Test Data Summary: Avance® Y-Connector S | | |
|------------------------------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Standard/Test/FDA<br>Guidance | Results Summary |
| Biocompatibility : | | |
| Cytotoxicity | ISO 10993-5 | Under the conditions of the study, the<br>components are not cytotoxic. |
| Primary Skin Irritation | ISO 10993-10 | Under the conditions of the study, the<br>components are not irritants. |
| ISO Closed Patch<br>Sensitization | ISO 10993-10 | Under the conditions of the study, the<br>components are not sensitizers. |
| Performance : | | |
| NPWT Functionality | Internal Mölnlycke test<br>method for fluid transport in<br>NPWT | The fluid handling properties of the Avance®<br>NPWT System are sufficient to accommodate<br>the exudate flow rate from large exuding<br>wounds. The test setup included the use of<br>the Avance® Y-Connector S to connect<br>multiple wound models. |
### Clinical Testing:
No clinical data was required.
#### Conclusion:
Based on the information presented in this submission, it can be concluded that the Avance® Y-Connector S is substantially equivalent to the predicate device identified with respect to intended use, design and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.