RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB Abdominal Dressing Kit with Soft Port Device
K143133 · Smith & Nephew, Inc. · OMP · Jul 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143133
Device Name
RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB Abdominal Dressing Kit with Soft Port Device
Applicant
Smith & Nephew, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Jul 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types include: - Chronic - Acute - Traumatic - Sub-acute and Dehisced wounds - Ulcers (such as pressure or diabetic) - Partial-thickness burns - Flaps and grafts RENASYS EZ MAX professional healthcare facility model (REF 66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional. RENASYS AB Abdominal Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
Device Story
RENASYS EZ MAX is a lightweight, powered suction pump for NPWT; provides continuous or intermittent negative pressure to remove wound exudates, irrigation fluids, and infectious materials. System comprises pump, disposable canister (with/without solidifier), and dressing kit (foam, organ protection layer, film drapes, Soft Port). Used in acute care/hospital settings by healthcare professionals. Pump connects to wound dressing via tubing; exudate collected in canister. Soft Port acts as conduit for pressure transmission and fluid collection. Device operates via mains power or internal battery. Healthcare providers monitor system via alarms (e.g., canister full, leak detection). Therapy promotes wound healing by managing wound environment. Abdominal kit specifically designed for open abdominal wounds with exposed viscera.
Clinical Evidence
Bench testing only. Verification testing confirmed pumping capacity, pressure settings, operating modes, alarm functionality (canister full, high leak), and system performance under fault conditions and varying exudate viscosities. Soft Port functionality verified for 48-hour duration. Compliance with ISO 13485, ISO 14971, IEC 60601-1, and IEC 60601-1-8 demonstrated.
Technological Characteristics
Powered suction pump; mains or battery-operated. Components: polyurethane foam, polyurethane film drapes, acrylic adhesive, Soft Port assembly. Sterilization: Ethylene oxide (ISO 11135). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility). Connectivity: None (standalone).
Indications for Use
Indicated for patients requiring negative pressure wound therapy (NPWT) for chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers, burns, flaps, and grafts. Abdominal kit indicated for temporary bridging of abdominal wall openings with exposed viscera where primary closure is not possible or repeat entry is required. Intended for use in acute care/hospital settings by qualified professionals.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
RENASYS EZ MAX Negative Pressure Wound Therapy Device (K132446)
RENASYS EZ PLUS Negative Pressure Wound Therapy Device (K102001)
RENASYS AB Abdominal Dressing Kit with Soft Port (K112784)
RENASYS EZ MAX Negative Pressure Wound Therapy Device and RENASYS Foam and Gauze NPWT Wound Dressing Kits with Soft Port (K142979)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of an eagle with three human profiles incorporated into its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2016
Smith and Nephew Incorporated Ms. Laura Reynolds Director of Regulatory Affairs 970 Lake Carillon Drive, Suite 110 Saint Petersburg, Florida 33716
Re: K143133
Trade/Device Name: RENASYS™ EZ MAX Negative Pressure Wound Therapy Device, RENASYS™ AB Abdominal Dressing Kit with Soft Port Device Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP, FTL Dated: June 8, 2015 Received: June 9, 2015
Dear Ms. Reynolds:
This letter corrects our substantially equivalent letter of July 23, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K143133
#### Device Name
RENASYS™ EZ MAX Negative Pressure Wound Therapy Device
#### Indications for Use (Describe)
RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- Sub-acute and Dehisced wounds
- · Ulcers (such as pressure or diabetic)
- Partial-thickness burns
- Flaps and grafts
RENASYS EZ MAX professional healthcare facility model (REF 66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143133
Device Name
RENASYSTM AB Abdominal Dressing Kit with Soft Port Device
Indications for Use (Describe)
RENASYS AB Abdominal Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary.
It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome.
The use of RENASYS Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K143133" in a simple sans-serif font. Below this text, the phrase "Page 1 of 7" is displayed, indicating this is the first page of a document. The text is aligned to the left and presented against a plain white background.
Image /page/4/Picture/1 description: The image shows the logo for Smith & Nephew. The logo features a stylized orange flower-like symbol on the left. To the right of the symbol is the text "We are smith&nephew", with "We are" in gray and "smith&nephew" in orange.
#### Wound Management 727 392-1261
Smith & Nephew, Inc. F 727 392-6914 or 727 392-0797 970 Lake Carillon Drive Customer Care Center: 1 800 876-1261 www.smith-nephew.com Suite 110 St. Petersburg, FL 33716
# 510(k) Summary
| General Information | |
|------------------------------------|------------------------------------------------------------------------------------------|
| Submitters Name/Address: | Smith & Nephew, Inc.<br>970 Lake Carillon Drive<br>Suite 110<br>St. Petersburg, FL 33716 |
| Establishment Registration Number: | 3006760724 |
| Contact Person: | Laura Reynolds<br>Director Regulatory Affairs |
| Phone Number: | (727) 329-7702 |
| Date Prepared: | July 1, 2015 |
| Device Description | |
| Trade Name: | RENASYST™ EZ MAX Negative Pressure Wound<br>Therapy (NPWT) Device |
| Generic/Common Name: | Powered Suction Pump |
| Classification Name: | Powered Suction Pump; 21 CFR 878.4780<br>Product Code: OMP |
| Trade Name: | RENASYS™ AB Abdominal Dressing Kit with Soft<br>Port™ Device |
| Generic/Common Name: | Mesh, Surgical Polymeric |
| Classification Name: | Mesh, Surgical Polymeric; 21 CFR 878.3300<br>Product Code: FTL |
#### Predicate Device Information
| Predicate Device | 510k# | Clearance<br>Date |
|--------------------------------------------------------|---------|-------------------|
| RENASYS EZ MAX Negative Pressure Wound Therapy Device | K132446 | 10/23/2013 |
| RENASYS EZ PLUS Negative Pressure Wound Therapy Device | K102001 | 08/06/2010 |
| RENASYS AB Abdominal Dressing Kit with Soft Port | K112784 | 11/22/2011 |
{5}------------------------------------------------
| Predicate Device | 510k# | Clearance<br>Date |
|--------------------------------------------------------------------------------------------------------------------------------|---------|-------------------|
| RENASYS EZ MAX Negative Pressure Wound Therapy Device and<br>RENASYS Foam and Gauze NPWT Wound Dressing Kits with Soft<br>Port | K142979 | 04/29/2015 |
#### Device Description
#### RENASYS EZ MAX Negative Pressure Wound Therapy Device
The RENASYS EZ MAX Negative Pressure Wound Therapy (NPWT) device is a lightweight, suction device intended for wound management via application of continuous or intermittent neqative pressure wound therapy to the wound for removal of fluids, including wound exudates, irrigation fluids, and infectious materials. The pump is connected to the wound dressing via a tube connected to a disposable canister. The device provides negative pressure wound therapy to the wound at a range of pressure settings and removes exudates from the wound site to the disposable canister. The device can operate either by a mains power supply or internal battery. The technological characteristics of the new device have not changed.
#### RENASYS Canisters
The 800ml and 250ml canisters are non-sterile, single use devices with a lid that is ultrasonically welded on. The canister kits contain a combination of solidifier gel pack, bacterial overflow guard that attaches to the pump and canister tubing that attaches to the Soft Port assembly.
The 800ml canister without solidifier is a non-sterile, single use device with a lid that is ultrasonically welded on. The canister kit contains a bacterial overflow quard that attaches to the pump and canister tubing that attaches to the Soft Port assembly.
#### RENASYS AB Abdominal Dressing Kit with Soft Port
RENASYS AB Abdominal Dressing Kit with Soft Port consists of the following components:
- Two large hydrophobic, reticulated polyurethane foam dressings that incorporate several ● cuts to facilitate custom sizing, if needed.
- One organ protection layer made of non-adherent clear polyurethane film perforated with slits. The slits in the organ protection layer facilitate fluid removal under negative pressure wound therapy.
- . Six 12" x 8" polyurethane transparent film drapes with acrylic adhesive backing. The drapes are placed over the foam and surrounding intact skin to create an airtight seal for the application of negative pressure wound therapy.
- . One Soft Port suction assembly with tubing that attaches to the exudate collection canister. The Soft Port assembly is placed over a hole cut in the drape to facilitate application of negative pressure wound therapy to the wound.
The abdominal kit is designed specifically for use with the RENASYS EZ MAX negative pressure wound therapy devices and canisters.
#### Indications for Use
#### RENASYS EZ MAX Negative Pressure Wound Therapy Device
{6}------------------------------------------------
The RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials
Appropriate wound types include:
- . Chronic
- Acute .
- Traumatic ●
- Sub-acute and dehisced wounds ●
- Ulcers (such as pressure or diabetic) .
- Partial-thickness burns .
- . Flaps and grafts
RENASYS EZ MAX professional healthcare facility model (REF 66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
#### RENASYS AB Abdominal Dressing Kit with Soft Port
The RENASYS AB Abdominal Dressing Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome.
The RENASYS AB Abdominal Dressing Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.
#### Summary Comparison between New and Predicate Devices
#### RENASYS EZ MAX NPWT
| | New Device: 510(k) K143133 | Predicate Device: 510(k)# K142979 | Predicate Device: 510(k)# K132446 |
|--------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Trade Name: | RENASYS EZ MAX NPWT | RENASYS EZ MAX Negative Pressure Wound Therapy Device and RENASYS Foam and Gauze NPWT Wound Dressing Kits with Soft Port | RENASYS EZ MAX NPWT |
| Indications for Use: | Substantially equivalent | Substantially equivalent | Substantially equivalent |
| Components | Substantially equivalent | Substantially equivalent | Substantially equivalent |
| Principle of Operation | Identical | Identical | Identical |
| Operating Time (Battery) | Identical | Identical | Identical |
{7}------------------------------------------------
| | New Device: | Predicate Device: | Predicate Device: |
|-------------------------------------------------|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| | 510(k) K143133 | 510(k)# K142979 | 510(k)# K132446 |
| Trade Name: | RENASYS EZ MAX<br>NPWT | RENASYS EZ MAX<br>Negative Pressure<br>Wound Therapy<br>Device and RENASYS<br>Foam and Gauze<br>NPWT Wound<br>Dressing Kits with<br>Soft Port | RENASYS EZ MAX<br>NPWT |
| Negative Pressure<br>Range | Identical | Identical | Identical |
| High Flow/Leak Rate<br>Alarm Threshold<br>Limit | Identical | Identical | Identical |
| Sterilization | N/A | N/A | N/A |
| Biocompability | No patient contact<br>materials | No patient contact<br>materials | No patient contact<br>materials |
| Alarms and<br>Indicators | Identical | Identical | Identical |
| Software | N/A | N/A | N/A |
### RENASYS EZ MAX NPWT Canisters Only
| | New Device:<br>510(k) K143133 | Predicate Device:<br>510(k)# K142979 | Predicate Device:<br>510(k)# K102001 |
|--------------------------------|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Trade Name: | RENASYS EZ MAX<br>NPWT | RENASYS EZ MAX<br>Negative Pressure<br>Wound Therapy Device<br>and RENASYS Foam<br>and Gauze NPWT<br>Wound Dressing Kits<br>with Soft Port | RENASYS EZ PLUS<br>NPWT |
| Indications for<br>Use: | Substantially equivalent | Substantially equivalent | Substantially equivalent |
| Bacterial Overflow<br>Guard | Material change to softer<br>plastic outer casing of<br>filter | Material change to softer<br>plastic outer casing of<br>filter | Plastic outer casing of<br>filter |
| Material | Substantially equivalent | Substantially equivalent | Substantially equivalent |
| Canister without<br>solidifier | New | New | N/A |
| Canisters with<br>Solidifier | Substantially equivalent | Substantially equivalent | Substantially equivalent |
{8}------------------------------------------------
## RENASYS AB Abdominal Dressing Kit with Soft Port
| | New Device: | Predicate Device: |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Trade Name: | 510(k) K143133<br>RENASYS AB Abdominal<br>Dressing Kit with Soft Port | 510(k)# K112784<br>RENASYS AB<br>Abdominal Dressing Kit<br>with Soft Port |
| Indications for Use: | Indicated for temporary bridging<br>of abdominal wall openings<br>where primary closure is not<br>possible and/or repeat<br>abdominal entries are<br>necessary. It is intended to be<br>used in open abdominal wounds<br>with exposed viscera, including<br>but not limited to abdominal<br>compartment syndrome.<br><br>This dressing kit is intended for<br>use in acute hospital settings<br>(trauma, general and plastic<br>surgery wards) and should<br>ideally be applied in the<br>operating theatre. | Identical |
| Materials: | Substantially equivalent | Substantially equivalent |
| Soft Port Assembly: | Increased aperture size<br>Two slits in paper release liner | Smaller aperture size<br>One slit in paper release<br>liner |
| Single-use or<br>Reusable: | Single-use | Single-use |
| Method of<br>Sterilization: | Identical | Identical |
| Biocompatibility: | All components comply with ISO<br>10993 | All components comply<br>with ISO 10993 |
| Packaging: | Substantially equivalent | Substantially equivalent |
| Kit Shelf-life | 12 months | 12 months |
# Table of Modifications
| Modification | Reason For Change | Verification Testing<br>Performed |
|------------------------------------------------------------------------------------------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------|
| Modification to the<br>release paper liner of the<br>Soft Port assembly to add<br>an additional slit | Improved usability | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance and<br>improved usability |
{9}------------------------------------------------
| Modification | Reason For Change | Verification Testing<br>Performed |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Modification to increase<br>the Soft Port aperture size | Improved usability | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance |
| Modification of the hole<br>size and geometry in the<br>mid and bottom poly<br>sheet layers of the Soft<br>Port Assembly | Improve manufacturability | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance and<br>improved usability |
#### Non-Clinical Tests (Bench)
#### RENASYS EZ MAX Negative Pressure Wound Therapy Device
Testing has been conducted to verify the modifications to the RENASYS EZ MAX NPWT meet design specifications and demonstrate substantial equivalence to the predicate device.
The list below summarizes the bench testing undertaken and successfully completed for the RENASYS EZ MAX NPWT device:
- Pumping capacity is equivalent to the predicate device. ●
- . Device provides negative pressure wound therapy at individual pressure settings, identical to the predicate device.
- . Verification that the device delivers negative pressure wound therapy in a continuous and intermittent operating mode identical to the predicate device.
- Verification of Canister Full alarm functionality using wound fluid designed to simulate . chemistry and protein content of real exudate.
- Verification of system performance in foreseeable fault conditions. ●
- Verification of system performance when running with high air leaks at the dressing site ●
- Verification of system performance in worst case scenarios with ranges of exudate viscosity and protein content.
#### RENASYS AB Abdominal Dressing Kit with Soft Port
Testing has been conducted to verify the modifications to the RENASYS AB Abdominal Dressing Kit with Soft Port meet design specifications and demonstrate substantial equivalence to the predicate device.
- . Verification that the Soft Port serves as a conduit between RENASYS EZ MAX NPWT device and RENASYS AB Abdominal Dressing Kit with Soft Port by transmitting negative pressure wound therapy and collecting exudate flows.
- . Verification that the Soft Port functions throughout the recommended maximum elapsed time of 48 hours between dressing changes.
There have been no other changes to patient contacting materials from the predicate devices.
Device complies with the following standards:
- ISO13485:2003, Medical Devices Quality Management Systems .
- ISO 14971:2012 Medical Devices Application of Risk Management to Medical Devices .
{10}------------------------------------------------
- ISO 15223-1:2012 Medical devices Symbols to be used with medical device labels, . labelling, and information to be supplied Part 1: General requirements.(General)
- . ISO 15223-2:2010 Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied-Part 2: Symbol development, selection and validation. (General)
- BS EN 980:2008 Graphical Symbols for use in the labeling of Medical Devices ●
- . BS EN 1041:2008 +A1:2013 Information Supplied by the Manufacturer with Medical Devices
- IEC 60601-1-2:2007(3rd edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, Interpretation Sheet
- . IEC 60601-1:2005 (3rd edition). Medical Electrical Equipment - Part 1: General Requirements for Safety
- ANSI/AAMI ES60601-1:2005 Version (R2012) Medical Electrical Equipment Part 1: ● General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-8:2006 (2nd edition) Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical
- . IEC 60601-1-6:2010 (3rd Edition) Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62366:2007 (1st edition) Medical devices Application of usability engineering to . medical devices. (General)
- . ISO 11135: 2014 Medical Devices - Validation and routine control of ethylene oxide sterilization
- . ISO 11607-1:2006 Packaging for terminally sterilized medical devices Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing with a risk management process
- . AAMI/ANSI HE75:2009 Human Factor Engineering-Design of Medical Devices
#### Conclusions
In establishing substantial equivalence to the currently marketed predicate devices, Smith & Nephew, Inc. evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing, and electrical safety testing has been successfully completed to demonstrate that the RENASYS EZ MAX NPWT device and canisters, and the RENASYS AB Abdominal Dressing Kit with Soft Port are substantially equivalent to the predicate devices for the intended uses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.