K180021 · Medtec Inc Dba Civco Medical Solutions and Civco Radiotherap · IYE · Jun 29, 2018 · Radiology
Device Facts
Record ID
K180021
Device Name
Proton Positioing and Immobilization Devices
Applicant
Medtec Inc Dba Civco Medical Solutions and Civco Radiotherap
Product Code
IYE · Radiology
Decision Date
Jun 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
General Positioning Devices: The device is indicated to aid in supporting and positioning adult and pediatric patients undergoing radiation therapy including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. Devices include Couchtops, Lok-Bars, Extensions, Cushions, and Accessories. Head, Neck, and Shoulder Devices: The device is indicated to aid in supporting, positioning, and/or immobilization of adult and pediatric patients undergoing radiation therapy of the head, brain, neck, and spine including radiosurgery and electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. Devices include Overlays, Extensions, Thermoplastic Masks, and Cushions. Support Garments: The device is indicated to aid in supporting adolescent and adult patients undergoing radiation therapy including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age. Breast Positioning Devices: The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age.
Device Story
Device family includes positioning/immobilization aids: couch tops, overlays, extensions, thermoplastic masks, cushions, support garments, and breast boards. Used in radiation oncology (IMPT, IMRT, PBPT, IGRT, VMAT, SRS, SRT, SBRT) and diagnostic imaging (CT, MR). Operated by radiation oncology therapists, medical physicists, and oncologists. Devices provide mechanical support and immobilization to ensure reproducible patient positioning during simulation, planning, and treatment delivery. Some devices (AccuForm cushions) use chemical activation to mold to patient anatomy; others (thermoplastic masks) are softened in heated water. Output is physical patient stabilization. Benefits include improved treatment accuracy, reduced healthy tissue exposure, and patient comfort. Used in healthcare facilities.
Clinical Evidence
No clinical data. Bench testing only. Performance testing confirmed that devices used in the proton beam path do not perturb or impact the beam in a way that cannot be accounted for in treatment planning. Risk analysis confirms safety for pediatric/adolescent use.
Technological Characteristics
Mechanical, non-powered, static devices. Materials include carbon fiber, thermoplastic, nylon, and polystyrene beads. Limited skin contact (<24 hours). Non-sterile. MR-safe as labeled. No software or electronics.
Indications for Use
Indicated for adult and pediatric/adolescent patients undergoing radiation therapy (electron, photon, proton) and image acquisition for treatment planning. Specific anatomical regions include head, brain, neck, spine, breast, and chest. Support garments and breast positioning devices are contraindicated for patients under 12 years of age.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MEDTEC, Inc. d/b/a "CIVCO Medical Solutions" and "CIVCO Radi % Ms. Alena Newgren Regulatory Specialist 140 1 8th Street SE ORANGE CITY IOWA 51041
June 29, 2018
# Re: K180021
Trade/Device Name: Proton Positioning and Immobilization Devices Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: December 22, 2017 Received: January 3, 2018
Dear Ms. Alena:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Ms. Alena Newgren
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D.'Hara
For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180021
#### Device Name
General Positioning Devices; Head, Neck, and Shoulders Devices; Support Garments; Breast Positioning Devices
#### Indications for Use (Describe)
General Positioning Devices: The device is indicated to aid in supporting and positioning adult and pediatric patients undergoing radiation therapy including electron, photon treatments. The device is also used during image acquisition to support treatment planning. Devices include Couchtops, Cushions, Cushions, and Accessories. The following are other positioning devices:
Head. Neck. and Shoulders Device is indicated to aid in supporting, positioning, and/or immobilization of adult and pediatric patients undergoing radiation therapy of the head, brain, neck, and spine including radiosurgery and electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. Devices include Overlays, Extensions, Thermoplastic Masks, and Cushions.
Support Garments: The device is indicated to aid in supporting adolescents undergoing radiation therapy including electron, photon treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age.
Breast Positioning Devices: The device is indicated to aid in supporting adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font, also in blue. The logo is clean and modern, and the use of blue gives it a professional and trustworthy feel.
# 510(k) Summary
# A. Submitter Information
| | Submitter Name & Address: MEDTEC, Inc. d/b/a "CIVCO Medical Solutions" and<br>"CIVCO Radiotherapy"<br>1401 8th St. SE<br>Orange City, IA 51041 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Alena Newgren, Regulatory Specialist<br>Telephone: 712-737-8688, Fax: 877-613-6300<br>regulatory@civcort.com |
| Date Summary Prepared: | June 13, 2018 |
| Trade Name: | Proton Positioning and Immobilization Solutions |
| Common Name: | Head, Neck and Shoulders Devices<br>General Positioning Devices<br>Support Garments<br>Breast Positioning Devices |
| Classification Names & | Medical charged-particle radiation therapy system<br>(892.5050) |
| Numbers: | System, Nuclear Magnetic Resonance Imaging (892.1000) |
| Device Class: | Class II |
| Review Panels: | Radiology |
| Product Codes: | IYE, LNH |
# B. Predicate Devices
The subject devices are substantially equivalent to the predicate devices in the following 510(k)s:
| Predicate Device: | Manufacturer |
|-----------------------------------------------|--------------|
| K973842: Carbon Fiber Conformal Couch Top | MEDTEC, Inc. |
| Reference Devices: | |
| K933227: Uni-Frame Head Immobilization System | |
| K080072: MRI Patient Positioning Devices | |
| K060737: Sinmed Positioning Devices | |
| K982624: Moldcare Head & Neck Cushion | |
| K935300: Vac-Lok Immobilization System | |
| K121284: Chabner XRT Garments | |
| K974703: Carbon Fiber Breast Board | |
The purpose of this 510(k) is to modify the intended use/indications for use for the subject devices including specification of proton treatment and pediatric/adolescent use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CIVCO Radiotherapy. The logo is blue and features the word "CIVCO" in a stylized font. The "V" in CIVCO has a blue circle above it. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font.
### D. Device Descriptions
This submission does not include in-vitro diagnostic devices.
The CIVCO Positioning and Immobilization devices have been used for many years and were previously cleared under other 510(k) is to have these identical Class II products cleared for use in Proton Therapy environment.
The Head, Neck, and Shoulder Devices are used to treat cancer of the head, neck, brain and spine. Organs that are located in these anatomical regions are applicable.
The General Positioning Devices are used to treat cancer of the spine, thorax, breast, abdomen, pelvis and prostate. Organs that are located in these anatomical regions are applicable.
The Support Garments are used to treat cancer of the breast, thorax and abdominal area. Organs that are located in these anatomical regions are applicable.
The Breast Positioning Devices are used to treat cancer of the breast, thorax and abdominal area. Organs that are located in these anatomical regions are applicable.
The devices are intended for therapeutic treatment (including proton, photon, and electron beam therapy) and diagnostics (including CT and MR as indicated). The devices will be used in a variety of types of external beam radiation therapy including Intensity Modulated Proton Therapy (IMPT), Intensity Modulated Radiation Therapy (IMRT), Pencil Beam Proton Therapy (PBPT), Image Guided Radiation Therapy (IGRT), Volumetric Modulated Arc Therapy (VMAT), Stereotactic Radiosurgery (SRS), Stereotactic Radiotherapy (SRT), Stereotactic Body Radiation Therapy (SBRT), and Craniospinal Treatment.
The intended users of the devices are radiation oncology therapists, medical physicists, and oncologists. The provider selects the appropriate equipment based on a patient's individual clinical goals and unique constraints.
Thermoplastic mask, AccuForm Cushion, and Support Garment devices are single patient use. All devices are non-powered and static. Detailed device specific descriptions are located in Section 11.
The subject devices are reusable or disposable as provided in product labeling, provided non-sterile, and are used in a healthcare facility/hospital. The devices are labeled for MR safety, as applicable. The following models are included in this submission:
### www.civcort.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is in blue and features the word "CIVCO" in large, bold letters. The "V" in CIVCO has a light blue circle above it. Below the word "CIVCO" is the word "Radiotherapy" in a smaller, thinner font.
## Table A. Head, Neck, and Shoulders Devices.
| Device<br>Family | Part No. | Device Name | Reference<br>510(k) |
|-----------------------------|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Overlay | MTPT210002 | Type-S™ Overlay (Reusable non-sterile overlay with thermoplastic frame for use with Hitachi proton system) | K933227 |
| Overlay | MTMRHNTYPES | Type-S™ Overlay (Reusable non-sterile Type-S™ overlay with strap) | K080072 |
| Thermoplastic<br>Mask | MTAPUDMP | Disposable non-sterile head only FirmFit™ Thermoplastic (2.3mm) | K060737 |
| Thermoplastic<br>Mask | MTAPSDMP | Disposable non-sterile head neck, and shoulder FirmFit™ Thermoplastic (2.3mm) | |
| Thermoplastic<br>Mask | MTAPUDMPSG | Disposable non-sterile head only open face FirmFit™ Thermoplastic (2.3mm) | |
| Extension | MTPT210001 | Type-S™ Extension (Reusable non-sterile head extension for use with Hitachi proton system) | K933227 |
| Extension | MTIL6665 | Type-S Extension (Reusable non-sterile extension for use with Universal Couchtop Long Extension) | K973842 |
| Extension | MTIL6668 | SRS Posifix Extension (Reusable non-sterile SRS compatible extension for use with Universal Couchtop Long Extension) | K973842 |
| Cushion | MTACL1520 | AccuForm™ Cushion (15 x 20 cm) | |
| Cushion | MTACL2035 | AccuForm™ Cushion (20 x 35 cm) chamfered | |
| Cushion | MTACL2035HF | AccuForm™ Cushion (20 x 35 cm) chamfered, half fill | |
| Cushion | MTACL2045 | AccuForm™ Cushion (20 x 45 cm) | K982624 and |
| Cushion | MTACL4060 | AccuForm™ Cushion (40 x 60 cm) | K080072 |
| Cushion | MTACL6060 | AccuForm™ Cushion (60 x 60 cm) | |
| Cushion | MTACL14242 | AccuForm™ Cushion (20 x 25 cm) | |
| Cushion | MTACLTYPES1 | AccuForm™ Cushion (Types-S) | |
| Couchtop &<br>Lok-Bar | MTIL6720 | IPPS™ Couchtop (Reusable non-sterile one piece couchtop<br>with Lok-Bar) | |
| | MTIL6740 | Universal Couchtop (Reusable non-sterile couchtop with Lok-<br>Bar, rectangular extension, and storage rack) | K973842 |
| Extension | MTIL6660 | Rectangular Extension (Reusable non-sterile rectangular<br>extension for use with Universal Couchtop Long Extension) | K973842 |
| Accessories<br>(Hand Grips) | MTIL662501 | Patient Hand Grip (Reusable non-sterile SBRT Body Pro-<br>Lok™ hand grip (right) for use with Universal Couchtop) | K973842 |
| | MTIL662502 | Patient Hand Grip (Reusable non-sterile SBRT Body Pro-<br>Lok™ hand grip (left) for use with Universal Couchtop) | |
| Cushion<br>(Vac-Lok) | MTVLG35FC | Vac-Lok ™ Cushion (Reusable, non-sterile, 70 x 100 cm, 35L,<br>nylon bag with indexing safe for use in 1.5T and 3.0T MR<br>environments) | K080072<br>K935300 |
| Cushion<br>(Combifix) | 151004 | Combifix™ (Reusable non-sterile Combifix™ baseplate with<br>(2) elevation blocks, Feetfix™ bracket, Feetfix cushion, and<br>Kneefix™ cushion) | K060737 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in large, bold letters. A blue circle is above the V in CIVCO. Below the word "CIVCO" is the word "Radiotherapy" in a smaller, thinner font.
# Table B. General Positioning Devices.
### www.civcort.com
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font. The logo is clean and modern, and the blue color gives it a sense of trust and reliability.
## Table C. Support Garments
| Support<br>Garments<br>(Brassiere) | CRTBRA01 | Chabner XRT® Radiation Bra, Size 1 | |
|------------------------------------|----------|------------------------------------|---------|
| | CRTBRA02 | Chabner XRT® Radiation Bra, Size 2 | |
| | CRTBRA03 | Chabner XRT® Radiation Bra, Size 3 | |
| | CRTBRA04 | Chabner XRT® Radiation Bra, Size 4 | |
| | CRTBRA05 | Chabner XRT® Radiation Bra, Size 5 | |
| | CRTBRA06 | Chabner XRT® Radiation Bra, Size 6 | K121284 |
| | CRTBRA07 | Chabner XRT® Radiation Bra, Size 7 | |
| | CRTBRA08 | Chabner XRT® Radiation Bra, Size 8 | |
| | CRTBRA09 | Chabner XRT® Radiation Bra, Size 9 | |
| | CRTEXT01 | Bra Extender | |
## Table D. Breast Positioning Devices
| Breastboard | MTM400 | C-Qual™ (Reusable non-sterile breastboard with head support and Monarch™ Overhead Arm Positioner) | K080072 |
|-------------|--------|---------------------------------------------------------------------------------------------------|---------|
|-------------|--------|---------------------------------------------------------------------------------------------------|---------|
# E. Indications for Use/Intended Use Statements
### General Positioning Devices:
The device is indicated to aid in supporting and positioning adult and pediatric patients undergoing radiation therapy including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. Devices include Couchtops, Lok-Bars, Extensions, Cushions, and Accessories. The following are other positioning devices:
### Head, Neck, and Shoulder Devices:
The device is indicated to aid in supporting, positioning, and/or immobilization of adult and pediatric patients undergoing radiation therapy of the head, brain, neck, and spine including radiosurgery and electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. Devices include Overlays, Extensions, Thermoplastic Masks, and Cushions.
### www.civcort.com
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is in blue and features the word "CIVCO" in large, rounded letters. Above the "V" is a small, light blue circle. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font.
### Support Garments:
The device is indicated to aid in supporting adolescent and adult patients undergoing radiation therapy including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age.
#### Breast Positioning Devices:
The device is indicated to aid in supporting and positioning adult and adolescent patients undergoing radiation therapy of the breast and chest region including electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. The device is not intended for use with patients under 12 years of age.
### F. Comparison of Technological Characteristics
This 510(k) is being submitted only to add the Proton Intended Use and pediatric/adolescent use for these previously cleared devices.
| MTPT210002 | Type-S TM Overlay | | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component/<br>Parameter | Proposed Device | Predicate Device | Reference Device |
| 510k | K180021 | K973842 | K933227 |
| Intended Use/<br>Indications for<br>Use | The device is indicated to aid<br>in supporting, positioning,<br>and/or immobilization of adult<br>and pediatric patients<br>undergoing radiation therapy<br>of the head, brain, neck, and<br>spine including radiosurgery<br>and electron, photon, and<br>proton treatments. The device<br>is also used during image<br>acquisition to support<br>treatment planning. | The intended use of<br>the MED-TEC, INC.<br>Carbon Fiber<br>Conformal Couch Top<br>is to support and aid in<br>positioning a patient<br>during radiologic and<br>other medical<br>procedures. | The intended use of this device is<br>to immobilize the head during<br>radiation therapy. |
| Design<br>Features | The Type-S Overlay is placed<br>on top of the treatment table<br>via Lok-Bar and allows for<br>attachments of head or head,<br>neck, and shoulders<br>thermoplastic masks. The<br>Overlay also allows a<br>headrest to be used. The | The Couch Top is<br>placed onto the<br>treatment pedestal and<br>includes indexing for<br>accessories to be<br>attached onto to allow<br>for patient positioning<br>and immobilization. | The baseplate sets on the<br>treatment table and allows for<br>attachments of head or head,<br>neck, and shoulders thermoplastic<br>masks. The baseplate also allows<br>a headrest to be used. The<br>baseplate is fitted with two<br>registration pins that index to |
## Head, Neck, and Shoulders Devices
### www.civcort.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The word "CIVCO" is in large, bold, blue letters, with a light blue circle above the V. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font.
| | Overlay has holes that<br>correspond to thermoplastic<br>mask attachment pins. | | corresponding holes in the<br>thermoplastic frame to give<br>precision repositioning. Four<br>swivel clamps allow quick lock<br>down and removal of the<br>thermoplastic frame. |
|------------------------------------|--------------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>Operation | Mechanical devices without the use of software or electronics. | | |
| Device Body<br>Contact<br>Category | Limited contact duration (<24 hours) for surface devices (skin). | | |
| Sterility | N/A Devices are non-sterile | | |
| MTPT210001 | Type-S™ Extension | | |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component/<br>Parameter | Proposed Device | Predicate Device | Reference Device |
| 510k | K180021 | K973842 | K933227 |
| Intended Use/<br>Indications for<br>Use | The device is indicated to aid in supporting, positioning, and/or immobilization of adult and pediatric patients undergoing radiation therapy of the head, brain, neck, and spine including radiosurgery and electron, photon, and proton treatments. The device is also used during image acquisition to support treatment planning. | The intended use of the MED-TEC, INC. Carbon Fiber Conformal Couch Top is to support and aid in positioning a patient during radiologic and other medical procedures. | The intended use of this device is to immobilize the head during radiation therapy. |
| Design<br>Features | The device attaches to the end of the treatment table via a cantilever lock and pin mechanism. The Extensions allows for the placement of a headrest and has corresponding holes for thermoplastic mask attachment of head only or head, neck, and shoulders. | The Couch Top is placed onto the treatment pedestal and includes indexing for accessories to be attached onto to allow for patient positioning and immobilization. | The baseplate sets on the treatment table and allows for attachments of head or head, neck, and shoulders thermoplastic masks. The baseplate also allows a headrest to be used. The baseplate is fitted with two registration pins that index to corresponding holes in the thermoplastic frame to give precision repositioning. Four swivel clamps allow quick lock down and removal of the thermoplastic frame. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for CIVCO Radiotherapy. The logo is in blue and features the word "CIVCO" in a bold, sans-serif font. A light blue circle is above the V. Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font.
| Principle of<br>Operation | Mechanical devices without the use of software or electronics. |
|------------------------------------|------------------------------------------------------------------|
| Device Body<br>Contact<br>Category | Limited contact duration (<24 hours) for surface devices (skin). |
| Sterility | N/A Devices are non-sterile |
| MTMRHNTYPES | Type-ST™ Overlay | | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component/<br>Parameter | Proposed Device | Predicate Device | Reference Device |
| 510k | K180021 | K973842 | K080072 |
| Intended Use/<br>Indications for Use | The device is indicated to aid in<br>supporting, positioning, and/or<br>immobilization of adult and<br>pediatric patients undergoing<br>radiation therapy of the head,<br>brain, neck, and spine including<br>radiosurgery and electron, photon,<br>and proton treatments. The<br>device is also used during image<br>acquisition to support treatment<br>planning. | The intended use of<br>the MED-TEC, INC.<br>Carbon Fiber<br>Conformal Couch Top<br>is to support and aid in<br>positioning a patient<br>during radiologic and<br>other medical<br>procedures. | Patient Positioning<br>Devices are used to aid in<br>the support and<br>positioning of patients<br>during an MRI. |
| Design Features | The Overlay is attached to the<br>treatment couch top via Lok-Bar<br>and secured by a Velcro strap.<br>The overlay is contoured to follow<br>the anatomy of the head, neck,<br>and shoulder region by a 'S'<br>shape. The Overlay contains<br>holes for thermoplastic mask to<br>attach to in either head only or<br>head, neck, and shoulder style. A<br>headrest can be placed on the<br>Overlay for patient positioning and<br>comfort. | The Couch Top is<br>placed onto the<br>treatment pedestal and<br>includes indexing for<br>accessories to be<br>attached onto to allow<br>for patient positioning<br>and immobilization. | The device is attached to<br>the treatment pedestal via<br>Lok-Bar and allows<br>attachment of other<br>positioning and<br>immobilization devices or<br>to be used stand-alone.<br>The device allows a<br>headrest to be added and<br>thermoplastic masks to be<br>attached. |
| Principle of<br>Operation | Mechanical devices without the use of software or electronics. | | |
| Device Body<br>Contact Category | Limited contact duration (<24 hours) for surface devices (skin). | | |
| Sterility | N/A Devices are non-sterile | | |
Corporate Office | 2301 Jones Boulevard | Coralville, IA 52241 | USA | office +1 (319) 248.6757 | info@civcort.com Global Sales Office | 1401 8º Street SE | Orange City, IA 51041 | USA | office +1 (712) 737.8688 | fax +1 (877) 613.6300
510(k) Summary – Page 8
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for CIVCO Radiotherapy. The logo is in blue and features the word "CIVCO" in a stylized font. Above the V is a light blue circle. Below the logo is the word "Radiotherapy" in a smaller, sans-serif font.
| MTAPUDMP<br>Component/<br>Parameter | Disposable non-sterile head only FirmFit™ Thermoplastic | Proposed Device | Predicate<br>Device | Reference Device | Reference Device |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| 510k | K180021 | K973842 | K080072 | K060737 | |
| Intended Use/<br>Indications for<br>Use | The device is indicated<br>to aid in supporting,<br>positioning, and/or<br>immobilization of adult<br>and pediatric patients<br>undergoing radiation<br>therapy of the head,<br>brain, neck, and spine<br>including radiosurgery<br>and electron, photon,<br>and proton treatments.<br>The device is also used<br>during image<br>acquisition to support<br>treatment planning. | The intended use<br>of the MED-TEC,<br>INC. Carbon Fiber<br>Conformal Couch<br>Top is to support<br>and aid in<br>positioning a<br>patient during<br>radiologic and<br>other medical<br>procedures. | Patient Positioning<br>Devices are used to<br>aid in the support and<br>positioning of patients<br>during an MRI. | Fixation and<br>(re)positioning of the<br>head- and neck area<br>of the patient during<br>radiotherapy and<br>diagnostics. | |
| Design<br>Features | Thermoplastic Masks<br>are composed of a<br>frame, mask, and push<br>pins. The thermoplastic<br>mask is submerged in<br>heated water to soften.<br>Once removed from the<br>water, the user quickly<br>dries off the water and<br>then forms the mask<br>around the patient's<br>head and locks the pins<br>into pinholes of the<br>Couchtop, Extension,<br>or Overlay. The<br>SnapFrame style allows<br>the user to pull the<br>sides down first and<br>then connect the top<br>portion with the sides to<br>make a continuous U-<br>shaped frame. | The Couch Top is<br>placed onto the<br>treatment pedestal<br>and includes<br>indexing for<br>accessories to be<br>attached onto to<br>allow for patient<br>positioning and<br>immobilization. | Thermoplastic Masks<br>are composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask is<br>submerged in heated<br>water to soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then forms<br>the mask around the<br>patient's head and<br>locks the pins into<br>pinholes of the<br>Couchtop, Extension,<br>or Overlay. | Thermoplastic Masks<br>are composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask is<br>submerged in heated<br>water to soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then forms<br>the mask around the<br>patient's head and<br>locks the pins into<br>pinholes of the<br>Couchtop, Extension,<br>or Overlay. | |
| Principle of<br>Operation | Mechanical devices<br>without the use of<br>software or electronics. | Mechanical<br>devices without<br>the use of | Mechanical devices<br>without the use of<br>software or<br>electronics | Mechanical devices<br>without the use of<br>software or<br>electronics | |
| | | software or<br>electronics. | | | |
| Device Body<br>Contact<br>Category | Limited contact duration<br>(<24 hours) for surface<br>devices (skin). | Limited contact<br>duration (<24<br>hours) for surface<br>devices (skin). | Limited contact<br>duration (<24 hours)<br>for surface devices<br>(skin). | Limited contact<br>duration (<24 hours)<br>for surface devices<br>(skin). | |
| Sterility | N/A Devices are non-<br>sterile | N/A Devices are<br>non-sterile | N/A Devices are non-<br>sterile | N/A Devices are non-<br>sterile | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is in blue and consists of the word "CIVCO" in a sans-serif font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller sans-serif font, also in blue.
| MTAPSDMP | Disposable non-sterile head neck, and shoulder FirmFit™ Thermoplastic | | | |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component/<br>Parameter | Proposed Device | Predicate<br>Device | Reference Device | Reference Device |
| 510k | K180021 | K973842 | K080072 | K060737 |
| Intended Use/<br>Indications for<br>Use | The device is indicated<br>to aid in supporting,<br>positioning, and/or<br>immobilization of adult<br>and pediatric patients<br>undergoing radiation<br>therapy of the head,<br>brain, neck, and spine<br>including radiosurgery<br>and electron, photon,<br>and proton treatments.<br>The device is also used<br>during image<br>acquisition to support<br>treatment planning. | The intended use<br>of the MED-TEC,<br>INC. Carbon Fiber<br>Conformal Couch<br>Top is to support<br>and aid in<br>positioning a<br>patient during<br>radiologic and<br>other medical<br>procedures. | The intended use of<br>this device is to<br>provide an additional<br>aid to the fast and<br>accurate repeat<br>positioning of the<br>patient for radiation or<br>other treatment. | Fixation and<br>(re)positioning of the<br>head- and neck area<br>of the patient during<br>radiotherapy and<br>diagnostics. |
| Design<br>Features | Thermoplastic Masks<br>are composed of a<br>frame, mask, and push<br>pins. The thermoplastic<br>mask is submerged in<br>heated water to soften.<br>Once removed from the<br>water, the user quickly<br>dries off the water and<br>then forms the mask<br>around the patient's<br>head and locks the pins<br>into pinholes of the<br>Couchtop, Extension, or<br>Overlay. The<br>SnapFrame style allows<br>the user to pull the | The Couch Top is<br>placed onto the<br>treatment pedestal<br>and includes<br>indexing for<br>accessories to be<br>attached onto to<br>allow for patient<br>positioning and<br>immobilization. | Thermoplastic Masks<br>are composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask is<br>submerged in heated<br>water to soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then forms<br>the mask around the<br>patient's head and<br>locks the pins into<br>pinholes of the<br>Couchtop, Extension,<br>or Overlay. | Thermoplastic Masks<br>are composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask is<br>submerged in heated<br>water to soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then forms<br>the mask around the<br>patient's head and<br>locks the pins into<br>pinholes of the<br>Couchtop, Extension,<br>or Overlay. |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for CIVCO Radiotherapy. The logo is blue and features the word "CIVCO" in a stylized font, with a blue circle above the "V". Below the word "CIVCO" is the word "Radiotherapy" in a smaller, sans-serif font, also in blue. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.
| | sides down first and<br>then connect the top<br>portion with the sides to<br>make a continuous U-<br>shaped frame. | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------|--|--|
| Principle of<br>Operation | Mechanical devices without the use of software or electronics. | | |
| Device Body<br>Contact<br>Category | Limited contact duration (<24 hours) for surface devices (skin). | | |
| Sterility | N/A Devices are non-sterile | | |
| MTAPUDMPSG | Disposable non-sterile head only open face FirmFit™ Thermoplastic | | | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component/<br>Parameter | Proposed Device | Predicate<br>Device | Reference Device | Reference Device |
| 510k | K180021 | K973842 | K982624 | K060737 |
| Intended Use/<br>Indications for<br>Use | The device is<br>indicated to aid in<br>supporting,<br>positioning, and/or<br>immobilization of adult<br>and pediatric patients<br>undergoing radiation<br>therapy of the head,<br>brain, neck, and spine<br>including radiosurgery<br>and electron, photon,<br>and proton treatments.<br>The device is also<br>used during image<br>acquisition to support<br>treatment planning. | The intended use<br>of the MED-TEC,<br>INC. Carbon Fiber<br>Conformal Couch<br>Top is to support<br>and aid in<br>positioning a<br>patient during<br>radiologic and<br>other medical<br>procedures. | The intended use of<br>this device is to<br>provide an additional<br>aid to the fast and<br>accurate repeat<br>positioning of the<br>patient for radiation<br>or other treatment. | Fixation and<br>(re)positioning of the<br>head- and neck area<br>of the patient during<br>radiotherapy and<br>diagnostics. |
| Design Features | Thermoplastic Masks<br>are composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask is<br>submerged in heated<br>water to soften. Once<br>removed from the<br>water, the user quickly<br>dries off the water and<br>then forms the mask<br>around the patient's<br>head and locks the | The Couch Top is<br>placed onto the<br>treatment<br>pedestal and<br>includes indexing<br>for accessories to<br>be attached onto<br>to allow for patient<br>positioning and<br>immobilization. | Thermoplastic<br>Masks are<br>composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask<br>is submerged in<br>heated water to<br>soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then | Thermoplastic<br>Masks are<br>composed of a<br>frame, mask, and<br>push pins. The<br>thermoplastic mask<br>is submerged in<br>heated water to<br>soften. Once<br>removed from the<br>water, the user<br>quickly dries off the<br>water and then forms |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the CIVCO Radiotherapy logo. The logo is in blue and features the word "CIVCO" in a stylized font. The "I" in CIVCO has a blue circle above it. Below the word "CIVCO" is the word "Radiotherapy" in a smaller font.
| | pins into pinholes of<br>the Couchtop,<br>Extension, or Overlay.<br>The SnapFrame style<br>allows the user to pull<br>the sides down first<br>and then connect the<br>top portion with the<br>sides to make a<br>continuous U-shaped<br>frame. | forms the mask<br>around the patient's<br>head and locks the<br>pins into pinholes of<br>the Couchtop,<br>Extension, or<br>Overlay. | the mask around the<br>patient's head and<br>locks the pins into<br>pinholes of the<br>Couchtop,<br>Extension, or<br>Overlay. |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>Operation | Mechanical devices without the use of software or electronics. | | |
| Device Body<br>Contact Category | Limited contact duration (<24 hours) for surface devices (skin). | | |
| Sterility | N/A Devices are non-sterile | | |
| MTIL6665 | Type-S Extension | Predicate Device |
|-…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.