K121284 · Chabner Xrt, LLC · IYE · Jun 14, 2013 · Radiology
Device Facts
Record ID
K121284
Device Name
CHABNER XRT GARMENTS
Applicant
Chabner Xrt, LLC
Product Code
IYE · Radiology
Decision Date
Jun 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures. The Chabner XRT™ Garments are not intended for pediatric use.
Device Story
Chabner XRT™ Garments are non-sterile, wearable compression garments (brassiere, shorts, or girdle) designed to aid patient support and positioning during radiological procedures (X-ray, CT, MRI, radiotherapy). Constructed from radiotranslucent materials (<1 mm thick) without metal, the garments include clear thermoplastic polyurethane (TPU) windows allowing clinicians to visualize index tattoos, midlines, and bony landmarks projected by machine light fields. The compressive nature and adjustment features assist in reproducing body geometry and patient-to-machine alignment. Used in clinical settings by healthcare providers, the garments provide the secondary benefit of protecting patient dignity and comfort during procedures. Radiation passes through the device without causing attenuation or bolus effects.
Indicated for adult patients requiring support and positioning during X-ray, CT, MRI, radiotherapy, and other diagnostic radiological procedures. Not indicated for pediatric use.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K121284
Page 1 of 4
# 510(k) Summary
"510(k) Summary" As required by section 807.92(c) For Chabner XRT™ Garments
June 3, 2013
1. 'Company Name and Address
a. Sponsor/Manufacturer
Bffl Co. Inc. 20 Kensington Road Scarsdale, NY 10583
b. Consultant/Contact
Evan P. Phelps OFW Law 600 New Hampshire Ave, Ste. 500 Washington D.C. 20037 ephelps@ofwlaw.com Tel: (202) 789-1212 Fax: (202) 234-3550
2. Establishment Registration Number:
Not yet assigned
- 3. Device Name:
- a. Trade Name:
Chabner XRT™ Garments
Therapy System
Patient Support and Positioning Garments
Accessory to Medical Charged-Particle Radiation
Common/Usual Name: b.
Classification Name: C.
## 4. Device Classification:
a. 21 C.F.R. § 892.5050 (Class 11)
b. Product Code: IYE
c. Classification Panel: Radiology JUN 1 4 2013
{1}------------------------------------------------
#### 5. Legally Marketed Predicate Devices:
- a. MR Patient Positioning Devices
- i. 510(k) Owner: Med-Tech, Inc. (d.b.a. CIVCO Medical Solutions)
- ii. 510(k) Number: K111340
- iii. 21 C.F.R. § 892.5050 (Class II)
- iv. Product Code: IYE
- b. Moldcare Head & Neck Cushion
- i. 510(k) Owner: Med-Tec., Inc.
- ii. 510(k) Number: K982624
- iii. 21 C.F.R. § 892.5050 (Class II)
- iy. Product Code: IYE
# 6. Device Description
The device is a non-sterile garment worn by patients as an aid in the support and positioning of patients during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures.
There are three versions of the garment: 1) brassiere; 2) athletic style shorts; and 3} girdle. All versions of the device are devoid of metal and can be worn for use with all treatment or diagnostic machines without causing a bolus (i.e., the materials do not mimic the properties of tissue during a radiological procedure) or attenuation effect. All materials used in the device's construction are < 1 mm in thickness.
Clear thermoplastic polyurethane (TPU) windows are incorporated into the device to enable clinicians to view index tattoos, the midline, bony landmarks, and other treatment parameters, as created by the light field that is projected by the treatment or diagnostic machine onto the patient's body. The compressive nature of the materials used in the device's construction, along with various adjustment features incorporated into the device, assist the clinician in positioning the body and to reproduce the desired body geometry, treatment position, and patient-to-machine alignment prior to each radiological procedure. Radiation passes through the TPU material without affecting therapy or diagnosis.
Because the device is worn as a garment, they offer the added benefit of protecting the patient's dignity during such procedures that, in turn, increases the patient's overall comfort during the process.
- 7. Intended Use
Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures.
{2}------------------------------------------------
The Chabner XRT™ Garments are not intended for pediatric ,use.
# 8. Technological Characteristics and Substantial Equivalence Evaluation
The Chabner XRT™ Garments are substantially equivalent in terms of intended use, operating principles, and operational specifications to CIVCO's MR Patient Positioning Devices (K111340) and Med-Tec. Inc.'s Moldcare Head and Neck Cushion (K982624) which are both legally marketable patient support and positioning aids classified by the agency as accessories to medical charged-particle radiation therapy systems. Like the CIVCO MR Patient Positioning Devices and Med-Tec. Inc. Moldcare Head and Neck Cushion, the Chabner XRT™ Garments are intended to be used as an aid in the support and positioning of patients during MR, radiological, and other procedures, are radiotranslucent, conform to the individual patient's body shape, and contact the body for less than a 24 hour period of time.
| Feature | Chabner XRT™<br>Garments | MR Patient<br>Positioning Devices | Moldcare Head &<br>Neck Cushion |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K111340 | K982624 |
| | | CIVCO | Med-Tec, Inc. |
| Indications for Use | Chabner XRT™<br>Garments are<br>intended to be used<br>as an aid in the<br>support of the<br>patient during X-ray,<br>computed<br>tomography,<br>magnetic resonance<br>imaging,<br>radiotherapy, and<br>other diagnostic<br>radiological<br>procedures.<br>The Chabner XRT™<br>Garments are not<br>intended for<br>pediatric use. | CIVCO patient<br>Positioning devices<br>are used to aid in the<br>support and<br>positioning of patients<br>during MR,<br>radiological, and other<br>procedures. | The intended use of<br>this device is to<br>provide an additional<br>aid to the fast and<br>accurate repeat<br>positioning of the<br>patient for radiation or<br>other treatment. The<br>device forms to the<br>back of the patient's<br>head and eliminates<br>the usual range of<br>movement of ordinary<br>head cups. |
| Radiotranslucent | Yes | Yes | Yes |
| Conforms to patient's<br>body shape | Yes | Yes | Yes |
{3}------------------------------------------------
K121284
Page 4 of 4
| Device Body Contact<br>Category (ISO 10993-1) | Surface devices,<br>intact skin; limited<br>contact duration<br>(<24 hours). | Surface devices, intact<br>skin; limited contact<br>duration (<24 hours). | Surface devices, intact<br>skin; limited contact<br>duration (<24 hours). |
|-----------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
|-----------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
There are no significant differences between the Chabner XRT™ Garments and the predicate devices. However, they do differ from the predicate devices in that they are produced as wearable compression garments that have the added benefit of protecting patient dignity; whereas, the predicate devices are not worn on the body. This difference does not affect the device's safety and efficacy.
- 9. Substantial Equivalence Conclusion
The Chabner XRT™ Garments are substantially equivalent to the CIVCO MR Patient Positioning Devices and Med-Tec. Inc. Moldcare Head and Neck Cushion with respect to intended use, safety, and effectiveness.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped with a pair of wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
June 14, 2013
BFFL Co., Inc. % Mr. Evan P. Phelps Counsel OFW Law 600 New Hampshire Avenue, NW, Suite 500 WASHINGTON DC 20037
Re: K121284
Trade/Device Name: Chabner XRT Garments Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: May 10, 2013 Received: May 15, 2013
Dear Mr. Phelps:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls.provisions.of.the.Act.include.requirements.for.annual.registration,.listing.of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Phelps
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
Michael D. O'Hara for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure-
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K121284
Device Name: Chabner XRT Garments
Indications for Use:
Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures.
The Chabner XRTTM Garments are not intended for pediatric use.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Porti
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K121284
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.