K060737 · Sinmed BV · IYE · Mar 31, 2006 · Radiology
Device Facts
Record ID
K060737
Device Name
SINMED POSITIONING DEVICES
Applicant
Sinmed BV
Product Code
IYE · Radiology
Decision Date
Mar 31, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Fixation and (re)positioning of the head- and neck area of the patient during radiotherapy and diagnostics. Fixation and (re)positioning of the head- and neck area of the patient during radiotherapy and diagnostics. Positioning of the lung- and thorax area of the patient during radiotherapy and diagnostics. Positioning of the lower abdomen of a patient during radiotherapy and diagnostics. Positioning of hip and lower extremities of a patient during radiotherapy and diagnostics. Fixation and (re)positioning of the patient during radiotherapy and diagnostics. Fixation and (re)positioning of the patient during radiotherapy and diagnostics.
Device Story
Sinmed immobilization systems consist of various baseplates, cushions, and thermoplastic masks used to secure patients in specific positions (prone, supine) during radiotherapy and diagnostic imaging. Baseplates (carbon fiber/acrylic) attach to treatment couches via rails; thermoplastic masks and foam supports (PE-foam) are used to immobilize specific body regions. Repovac vacuum cushions (polyurethane bags with polystyrene beads) create rigid molds of patient contours. Devices are used in clinical radiotherapy/diagnostic environments by healthcare professionals to ensure reproducible patient setup, improving treatment accuracy and patient comfort. No automated processing or software is involved.
Clinical Evidence
No clinical data or comparative investigations were performed. Substantial equivalence is based on design, material, and functional similarity to legally marketed predicate devices and established clinical experience.
Technological Characteristics
Materials: Carbon fiber sandwich, acrylic, PE-foam, polyurethane, polystyrene beads. Principle: Mechanical immobilization via baseplates, thermoplastic masks, and vacuum-formed cushions. Energy: None (passive). Form factor: Modular baseplates, cushions, and masks. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for radiotherapy and diagnostic patients requiring immobilization and positioning of the head, neck, thorax, lung, abdomen, pelvis, or lower extremities. Contraindications not specified.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K060737
# MAR 3 1 2006
510(k) summary, IYE- positioning devices
Image /page/0/Picture/3 description: The image shows the word "Sinmed" in bold, black letters. Below the word "Sinmed" are the words "Radiotherapy products" in a smaller font. The word "Sinmed" appears to be a company name or brand, and the words "Radiotherapy products" indicate the type of products the company offers.
Sinmed head and shoulder positioning and immobilization systems. (Baseplates)
| Date prepared | 23-09-05 | |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Sinmed BV | |
| Address: | Pasteurstraat 6<br>2811 DX Reeuwijk<br>Phone: 0031-182-394495<br>Fax: 0031-182-395014 | |
| Contact person: | Caroline de Keijzer (Quality Manager)<br>Leo de Mooy (Technical Director) | |
| Tradename: | -Posifix 1,2,4,5,7 head and neck systems<br>-Positilt head inclination system<br>-Posifix Pediatric<br>-Headsupports<br>-Blocks and wedges<br>-Shouldersupport cushion | |
| Common name: | Head and shoulders immobilization systems. | |
| Classification name: | Linear medical accelerator | |
| Classification: | Class 2 devices, 892.5050 IYE | |
| Intended use: | Fixation and (re)positioning of the head- and neck during radiotherapy and<br>diagnostics. | |
| Substantial Equivalence Device: | The Sinmed head and neck immobilization systems are defined as<br>Substantially Equivalent (SE) to the Bionix 3-Way Head immobilizator; 3Way<br>DeLuxe Reusable Frame, manufactured by Bionix Development corporation.<br>Registration number: 1526854) and cleared by FDA with K933613 | |
| Description of the devices: See Sinmed<br>product Catalogue 2006 | Page: 4<br>8<br>4<br>9<br>12<br>13 | Product:<br>-Posifix 1,2,4,5,7 head and neck systems<br>-Positilt head inclination system<br>-Posifix Pediatric<br>-Headsupports<br>-Blocks and wedges<br>-Shouldersupport cushion |
| Summary of the technological<br>characteristics of your device<br>compared to the predicate device | The Sinmed devices indicated above are all designed to position the head-<br>and neck of a patient for diagnostics and radiotherapy and exactly reposition<br>it several times. All these products can be used in combination with each<br>other to optimize the set-up of a patient. This is exactly the purpose of the<br>Bionix 3-Way Head Immobilizator. The same kind of construction and<br>materials are used to reach this. Products from both companies are used for<br>the same kind of radiotherapy treatments. These products from both | |
companies are really competitive systems
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "Sinmed" in bold, black letters. The first letter, "S", is enclosed in parentheses. The word is printed in a simple, sans-serif font. The background is white.
Radiotherapy products
| | Sinmed head and neck immobilization baseplates | Bionix 3-WAY HEAD IMMOBILIZER |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Fixation and (re)positioning of the head- and neck during radiotherapy and diagnostics | Fixation and (re)positioning of the head and neck during radiotherapy and diagnostics |
| Target population. | Radiotherapy patients with tumors in head- and neck area. | Radiotherapy patients with tumors in head- and neck area. |
| Position of the patient | Lying on a couch in prone or supine position.<br>Shoulders and head are positioned by in a certain position using accessories for an optimized treatment. | When you're treating tumors in the head and neck region, accurate, reproducible positioning is essential. Our 3-Way Immobilizer works on prone, supine and lateral positioned patients. |
| Material | Carbon fiber and acrylic, polyethylene foam.<br>The skin contact materials carbon fiber and acrylic are exactly the same as those used in the predicate device. The cytotoxicity test of the carbon fiber top layer we use is enclosed in Chapter 11.<br>The PE-Foam material is commonly used in radiotherapy, the toxicity report is enclosed in Chapter 11. | Carbon fiber and acrylic version baseplates |
| Dimensions | Square shaped baseplate which follows the patients contours of head (Posifix -4, -5, -7, Positilt) or head and shoulders (Posifix-1 and - 2 and - Pediatric). | Square shaped baseplate which follows the patients contours of the head. |
| Design / Shape | Flat baseplate on which the thermoplastic mask and various headsupports can be placed.<br>Using the Positilt or other accessories, the baseplate can be inclined to change the position of the head. | Flat baseplate on which the thermoplastic mask and various headsupports can be placed. The baseplate can be inclined to change the position of the head. |
| Possible adjustments | Head can be lifted or inclined by using the blocks and wedges or the Positilt system. | Head can be tilted or inclined by using the 3-way head immobilizator. |
| Couchfixation | The Aluminium fixationrails connects the baseplate to the couch.<br>The adjustable aluminium rails can be mounted on all of our baseplates. They allow the baseplate to be locked down in the desired position on the simulator or treatment couch.<br>Depending on the type of couch, one can choose from several mounting brackets. | The SecureFit bar connects the baseplate to the couch. Our SecureFit bar was designed to index our immobilization devices to the Exact™ Couch for increased accuracy. |
| Accessories | Foam headsupports, blocks and wedges, which can position the head for an optimized treatment. Foam Shouldersupport cushion for a better and more comfortable positioning of the patient.<br>Thermoplastic mask-material which can be placed on the baseplate with plastic profiles. | Headsupports designed for a dedicated treatment can be used on these baseplates.<br>Foam blocks and wedges can be used for a better and more comfortable positioning of the patient.<br>Thermoplastic mask-material which can be placed on the baseplate with plastic profiles. |
| Compatibility with the environment and other devices | Can be used on all brands of couches in diagnostic and radiotherapy environment. | Can be used on an Exact™ couch in diagnostic and radiotherapy environment. |
Brief discussion of the nonclinical tests submitted
The Sinmed described devices above are competitive devices to the device to which substantial equivalence is claimed. Both are designed and used for the same purpose. No comparative investigations are performed, but from clinical experience is known that both products function in the same way.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "Sinmed" in a bold, sans-serif font. Below the word "Sinmed" is the phrase "Radiotherapy products" in a smaller, lighter font. The word "Sinmed" is the main focus of the image, with the phrase below providing additional context.
## Sinmed head and shoulders immobilization systems. (Extensionplates)
| Date prepared | 23-09-05 | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Submitter | Sinmed BV | |
| Address: | Pasteurstraat 6<br>2811 DX Reeuwijk<br>Phone: 0031-182-394495<br>Fax: 0031-182-395014 | |
| Contact person: | Caroline de Keijzer (Quality Manager)<br>Leo de Mooy (Technical Director) | |
| Tradename: | -Posifix-EXT systems<br>-Shoulderretractor<br>-Couchtop Cover | |
| Common name: | Head and shoulders immobilization systems | |
| Classification name: | Linear medical accelerator | |
| Classification: | Class 2 devices, 892.5050 IYE | |
| Intended use: | Fixation and (re)positioning of the head- and neck during radiotherapy and<br>diagnostics. | |
| Substantial Equivalence Device: | The Sinmed head and neck immobilization systems are defined as<br>Substantially Equivalent (SE) to the Bionix, Versaboard, Model 7040,<br>manufactured by Bionix Development Corporation, Registration number:<br>1526854) and cleared by FDA with K030051 | |
| Description of the devices: See Sinmed<br>product Catalogue 2006 | Page:<br>8<br>35<br>5 | Product:<br>Posifix-EXT systems<br>Shoulderretractor<br>Couchtop Cover |
| Summary of the technological<br>characteristics of your device compared<br>to the predicate device | The Sinmed Posifix-EXT system, in combination with the Shoulderretractor<br>and the Couchtopcover, can exactly be compared with the Bionix<br>Versaboard. These products from both companies are intended to position<br>the head- and neck of the patient outside the couch, to optimize the<br>treatment possibilities. The Bionix baseplate is longer and also supports<br>the back and buttocks of the patient. Sinmed has chosen to design a<br>shorter version, and creates this total support by using the couchtop cover.<br>On both systems the mask-material can be placed as well as a choice of<br>headsupports for the positioning of the head.<br>In order to further increase the space around the head of the patient, the<br>shoulders should be suppressed. Sinmed uses a separate Shoulder<br>retractor for this, and Bionix uses the relocatable shoulder suppression<br>system which can be integrated in the baseplate.<br>Both systems are placed and locked on the cranial side of the radiotherapy<br>treatment couch. Both systems are specially developed for use during<br>radiotherapy treatment, and not during diagnostics. | |
{3}------------------------------------------------
-
Image /page/3/Picture/1 description: The image shows the word "Sinmed" in bold, black font. The letters are closely spaced together, creating a compact and easily readable word. The font appears to be sans-serif, giving it a modern and clean look. The word is likely a brand name or a company name.
Radiotherapy products
| | Sinmed head and neck immobilization | Bionix |
|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | baseplates (Extensionplate) | VERSABOARD, MODEL 7040 |
| | | |
| Intended use | Fixation and (re)positioning of the head- and<br>neck during radiotherapy, especially IMRT-<br>treatment. | Fixation and (re)positioning of the head- and neck<br>during diagnostics and radiotherapy.<br>This new board is ideal for IMRT and will solve many of<br>your hard to reach angles and maximize your treatment<br>angles. |
| Target population. | Patient with tumors in head- and neck-area,<br>treated with IMRT-techniques | Patient with tumors in head- and neck-area, treated with<br>IMRT-techniques |
| Special feature in<br>relation to head and<br>shoulder immob.<br>systems.(baseplates) | These extension plates position the head- and<br>shoulders of the patient outside the couch,<br>which enlarges the radiotherapy treatment<br>possibilities. | It extends off the couch for a broad range of treatment<br>options while improving therapy accuracy. |
| Position of the<br>patient | Shoulderretractor suppresses the shoulders for<br>an optimized treatment. | The optional relocatable shoulder suppression system<br>allows you to comfortably push down the shoulders for<br>better reproducibility. |
| Material | Carbon fiber sandwich construction, PE-foam.<br>The skin contact materials carbon fiber is<br>exactly the same as that used in the predicate<br>device. The cytotoxicity test of the carbon fiber<br>top layer we use is enclosed in Chapter 11.<br>The PE-Foam material is commonly used in<br>radiotherapy, the toxicity report is enclosed in<br>Chapter 11. | Carbon fiber sandwich construction.. |
| Dimensions | Baseplate supports the head and shoulders of<br>the patient. When also using the couchtop<br>cover, the patient support is completely flat<br>from head to buttocks. | Baseplate supports the patient from head to buttocks. |
| Design / Shape | Head- and shoulders are supported by the<br>baseplate and positioned outside the couch. | Head and shoulders are supported by the baseplate. It<br>extends off the couch for a broad range of treatment<br>options while improving therapy accuracy. |
| Possible<br>adjustments | No adjustments possible with the baseplate.<br>Adjustments in positioning of the head- and<br>shoulders can be done by using accessories<br>like different headsupports or the<br>shoulderretractor. | No adjustments possible with the baseplate.<br>Adjustments in positioning of the head- and shoulders<br>can be done by using accessories like different<br>headsupports or the shoulder suppression system. |
| Couchfixation | Flat baseplate which hooks onto the cranial<br>side of the treatment couch and can be locked.<br>After mounting onto the couch, the construction<br>is fixed with a locking mechanism. | Baseplate can be placed on the couch and attached<br>onto the cranial side. |
| Accessories | -Thermoplastic mask-material which can be<br>placed on the baseplate with fixation profiles.<br>-Various headsupports designed for a<br>dedicated treatment can be used on this<br>baseplate.<br>-Couchtop Cover to overcome the height<br>difference when also using other baseplates. | -Thermoplastic mask-material which can be placed on<br>the baseplate with plastic profiles.<br>- Headsupports designed for a dedicated treatment can<br>be used on this baseplate. |
| Compatibility with<br>the environment and<br>other devices | Three types of which each can be placed on a<br>Varian, Siemens or Elekta couch. | Can be placed on a Varian couch |
Brief discussion of the nonclinical tests submitted
The Sinmed described devices above are competitive devices to the device to which substantial equivalence is claimed. Both are designed and used for the same purpose. No comparative investigations are performed, but from clinical experience is known that both products function in the same way.
2-16
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Sinmed" in a bold, sans-serif font. Below the word "Sinmed" is the phrase "Radiotherapy products" in a smaller, regular font. The word "Sinmed" appears to be the name of a company or brand, and the phrase "Radiotherapy products" indicates the type of products that the company offers.
## Sinmed lung and thorax immobilization systems
| Date prepared | 23-09-05 | | | | | | | | | | |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|
| Submitter | Sinmed BV | | | | | | | | | | |
| Address: | Pasteurstraat 6<br>2811 DX Reeuwijk<br>Phone: 0031-182-394495<br>Fax: 0031-182-395014 | | | | | | | | | | |
| Contact person: | Caroline de Keijzer (Quality Manager)<br>Leo de Mooy (Technical Director) | | | | | | | | | | |
| Tradename: | -Thorawedge<br>-Posirest<br>-Posiboard<br>-PET-armsupport | | | | | | | | | | |
| Common name: | Lung and thorax immobilization systems | | | | | | | | | | |
| Classification name: | Linear medical accelerator | | | | | | | | | | |
| Classification: | Class 2 devices, 892.5050 IYE | | | | | | | | | | |
| Intended use: | Positioning the patient for irradiation in the lung- and thorax area. | | | | | | | | | | |
| Substantial Equivalence Device: | The Sinmed Lung and thorax immobilization systems are defined as<br>Substantially Equivalent (SE) to the MedTec CARBON FIBER BREAST<br>BOARD, manufactured by MED-TEC,INC. (Registration number:<br>1932738) and cleared by FDA with K974703. | | | | | | | | | | |
| Description of the devices: See Sinmed<br>product Catalogue 2006 | Page: Product: 27 Thorawedge 26 Posirest 23 Posiboard 29 PET-armsupport | | | | | | | | | | |
| Summary of the technological<br>characteristics of your device compared<br>to the predicate device | The Medtec Breastboard is a device intended to position patients<br>undergoing irradiation treatment in the breast- and lung area. This is<br>exactly the same purpose of the Sinmed devices in this group.<br>The patient support of the Medtec Breastboard can be inclined for better<br>treatment. This same inclination of the upper body is also possible with the<br>Sinmed Posiboard and the Thorawedge.<br>To keep the arms out of the treatment area, arms above the head, various<br>armsupports can fixed on the Medtec Breastboard. This fixation of<br>armsupports is also possible on the Sinmed Posiboard. This positioning of<br>the arms above the head to achieve more space around the breast and<br>lung-area is also the exact intended use of the Sinmed Posirest and PET-<br>armsupport and the Thorawedge system.<br>The features of the Medtec Breastboard can all be found in the Sinmed<br>Posiboard. The other three Sinmed lung and thorax immobilization<br>systems are less complicated devices with the same purpose, each with | | | | | | | | | | |
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Sinmed" in a bold, sans-serif font. The letters are black against a white background. The word appears to be a logo or brand name.
Radiotherapy products
| | Sinmed Lung and thorax | MedTec<br>Carbon Fiber Breast Board |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | immobilization systems | |
| Intended use | Positioning the patient for irradiation in the<br>lung- and thorax area. | Positioning the patient for irradiation in the lung-<br>and thorax area. |
| Target<br>population. | Patients with tumors in breast and thorax. | Patients with tumors in breast and thorax |
| Position of the<br>patient | Supine position, can be inclined, with arms<br>positioned above the head | Supine position, can be inclined, with arms<br>positioned above the head |
| Material | Carbon fiber sandwich construction, PE-<br>Foam, and acrylic. Only non-metal<br>materials are used, like foam, acrylic and<br>carbon fiber sandwich in the treatment<br>area.<br>The skin contact material carbon fiber is<br>exactly the same as that used in the<br>predicate device. The cytotoxicity test of<br>the carbon fiber top layer we use is<br>enclosed in Chapter 11.<br>The PE-Foam material is commonly used<br>in radiotherapy, the toxicity report is<br>enclosed in Chapter 11.<br>The acrylic material is the same used in the<br>Sinmed head and shoulder positioning and<br>immobilization systems. (baseplates) and<br>is identical to the material used in the<br>predicate device Bionix 3-WAY HEAD<br>IMMOBILIZER. | Carbon-fiber sandwich, there are no metal<br>components in the vital scanning and treatment<br>areas making it even more "CT-friendly"; |
| Dimensions | Optimized dimensions (width) to be CT-<br>scanner compatible | Available in both standard and CT-compatible<br>widths. |
| Design / Shape | Baseplate, with an aperture for placing<br>various headsupports, Above the head the<br>arms can be placed by using various<br>armsupports. | Baseplate, with an aperture for placing various<br>headsupports, Above the head the arms can be<br>placed by using various armsupports. |
| Possible<br>adjustments | 0-25 degrees inclination possible with<br>Posiboard and Thorawedge.<br>0 degrees inclination with Posirest and<br>PET-armsupport. | 0-25 degrees inclination possible. |
| Couchfixation | Posirest, Posiboard, PET-armsupport is<br>fixed onto the couch with fixationstrips. | The Medtec Carbon Fiber Breastboard is fixed onto<br>the couch using the MedTec fixation rails. |
| Accessories | Various headsupports for a comfortable<br>patient positioning. Various armsupports<br>for patient positioning. | Various headsupports for a comfortable patient<br>positioning. Various armsupports for patient<br>positioning. |
| Compatibility<br>with the<br>environment<br>and other<br>devices | Due to the chosen construction and<br>materials, these products can be used<br>during both diagnostics and radiotherapy. | Due to the chosen construction and materials,<br>these products can be used during both<br>diagnostics and radiotherapy. |
Brief discussion of the nonclinical tests submitted
The Sinmed described devices above are competitive devices to the device to which substantial equivalence is claimed. Both are designed and used for the same purpose. No comparative investigations are performed, but from clinical experience is known that both products function in the same way.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Sinmed" in a bold, sans-serif font. Below the word "Sinmed" is the phrase "Radiotherapy products" in a smaller, less bold font. The text is black against a white background.
### Sinmed pelvis and lower extremities immobilization systems (Bellyboard system)
| Date prepared | 23-09-05 |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Sinmed BV |
| Address: | Pasteurstraat 6<br>2811 DX Reeuwijk<br>Phone: 0031-182-394495<br>Fax: 0031-182-395014…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.