K173032 · Paxman Coolers Limited · PMC · Jun 7, 2018 · General, Plastic Surgery
Device Facts
Record ID
K173032
Device Name
Paxman Scalp Cooler
Applicant
Paxman Coolers Limited
Product Code
PMC · General, Plastic Surgery
Decision Date
Jun 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4360
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K173032 · Jun 7, 2018
Paxman Scalp Cooler
Paxman Coolers Limited
Dutch Scalp Cooling Registry; Observational clinical studies; Published clinical literature; Non-published clinical study reports
RWE is used to support the expansion of the device's indications for use from 'women with breast cancer' to 'cancer patients with solid tumors' by demonstrating safety and efficacy across various cancer types and chemotherapy regimens in routine clinical practice.
Dutch Scalp Cooling Registry; Registry; Study Period: 2006-2017
Over 6000 cancer patients with multiple cancer types; Sample Size: 6000+
Not applicable for this study
Hair preservation (head cover use)
Sensor-controlled scalp cooling in the clinical routine; Retrospective analysis; Study Period: 2014-2016
99 female patients with breast or female genital tract cancer; Sample Size: 99
Not applicable for this study
Effectiveness and safety of sensor-controlled scalp cooling
No effect of scalp cooling on survival among women with breast cancer; Retrospective cohort; Follow-up/Duration: 6.3 to 8.0 years
1370 women with non-metastatic breast carcinoma; Sample Size: 1370
Non-scalp-cooled patients
Overall survival
Indications for Use
The Paxman Scalp Cooler is indicated to reduce the likelihood of chemotherapy-induced alopecia (CIA) in cancer patients with solid tumors.
Device Story
The Paxman Scalp Cooler is a mobile, electrically-powered refrigeration unit that circulates refrigerated liquid coolant through a cooling cap fitted to the patient's head. The system extracts heat from the scalp to maintain a low temperature, which reduces the likelihood of chemotherapy-induced alopecia. It is used in clinical settings by trained healthcare professionals. The device features a touchscreen controller with a graphical user interface (GUI) that allows the user to initiate, monitor, and complete the cooling process. The GUI provides status updates, timer countdowns, and alerts. The system uses pre-established programs for cooling administration. By maintaining scalp hypothermia during chemotherapy infusion, the device aims to reduce hair loss, potentially improving patient quality of life. The device is not intended for use with Taxanes plus anthracyclines in combination or sequence.
Clinical Evidence
Evidence includes registry data (Dutch Scalp Cooling Registry, n>6000), a randomized multi-center trial (NCT01986140, n=182), and various observational/retrospective studies. The SCALP trial demonstrated 50.5% successful hair preservation in the cooling group vs 0% in control (p=0.0061). Registry data and literature across various solid tumors (breast, lung, prostate, etc.) support efficacy, with variable success rates depending on chemotherapy regimen (e.g., higher success with taxanes, lower with anthracyclines). Adverse events were generally mild (grades 1-2), primarily cold sensation and headache.
Technological Characteristics
Mobile refrigeration unit; circulates liquid coolant (OrbisC organic salt-based) through silicone cooling caps (Primasil Sil 100, 20-80 Shore Durometer). Dimensions: 640mm x 320mm x 420mm; Weight: 29.5kg. Power: 100-120V, 50/60Hz. Coolant temperature range: -15°C to 5°C. Features touchscreen controller, recirculation pump, and quick-disconnect lines. Non-sterile, reusable system. Software-controlled with pre-established programs.
Indications for Use
Indicated for cancer patients with solid tumors to reduce the likelihood of chemotherapy-induced alopecia (CIA). Contraindicated in pediatric patients; patients with scalp metastases or suspected scalp metastases; head and neck cancers; CNS malignancies; cold sensitivity, cold agglutinin disease, cryoglobulinemia, cryofibrinogenemia, cold migraine, cold urticaria, and post-traumatic cold dystrophy; hematological malignancies; imminent bone marrow ablation chemotherapy; imminent or previous skull irradiation; severe liver or renal disease; skin cancers (melanoma, squamous cell, Merkel cell); small cell or squamous cell lung carcinoma; and solid tumors with high likelihood of transit metastasis.
Regulatory Classification
Identification
A scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia is a prescription device intended to reduce the frequency and severity of alopecia during chemotherapy in which alopecia-inducing chemotherapeutic agents are used.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use. This information must include testing to demonstrate accuracy of the temperature control mechanism.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical safety of the device.
(3) Software verification, validation, and hazard analysis must be performed.
(4) The patient contacting components of the device must be demonstrated to be biocompatible. Material names must be provided.
(5) Labeling must include the following:
(i) A statement describing the potential risk of developing scalp metastasis.
(ii) Information on the patient population and chemotherapeutic agents/regimen for which the device has been demonstrated to be effective.
(iii) A summary of the non-clinical and/or clinical testing pertinent to use of the device.
(iv) A summary of the device technical parameters, including temperature cooling range and duration of cooling.
(v) A summary of the device- and procedure-related adverse events pertinent to use of the device.
(vi) Information on how the device operates and the typical course of treatment.
(6) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, and adverse effects/complications.
(ii) Information on how the device operates and the typical course of treatment.
(iii) Information on the patient population for which there is clinical evidence of effectiveness.
(iv) The potential risks and benefits associated with use of the device.
(v) Postoperative care instructions.
(vi) A statement describing the potential risk of developing scalp metastasis.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 7, 2018
Paxman Coolers Limited % Heather Crawford Senior Consultant, Quality and Regulatory Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746
Re: K173032
Trade/Device Name: Paxman Scalp Cooler Regulation Number: 21 CFR 878.4360 Regulation Name: Scalp Cooling System Regulatory Class: Class II Product Code: PMC Dated: May 2, 2018 Received: May 7, 2018
Dear Heather Crawford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173032
Device Name Paxman Scalp Cooler
Indications for Use (Describe)
The Paxman Scalp Cooler is indicated to reduce the likelihood of chemotherapy-induced alopecia (CIA) in cancer patients with solid tumors.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Paxman Scalp Cooler
# K173032
### 1. Submission Sponsor
Paxman Coolers Limited
International House
Penistone Road
Fenay Bridge
Huddersfield
HD8 OLE
United Kingdom
Contact: Richard PAXMAN
Title: Managing Director
### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327.9997
Contact: Heather Crawford, Senior Consultant, Quality and Regulatory
Email: project.management@emergogroup.com
#### 3. Date Prepared
June 03, 2018
{4}------------------------------------------------
#### 4. Device Identification
| Trade/Proprietary Name: | Paxman Scalp Cooler |
|-------------------------|------------------------------------------------------------------------------------|
| Common/Usual Name: | Scalp Cooling System |
| Classification Name: | Scalp Cooling System to Reduce the Likelihood of Chemotherapy-<br>Induced Alopecia |
| Regulation Number: | 878.4360 |
| Product Code: | PMC, Scalp Cooling System |
| Device Class: | Class II |
| Classification Panel: | General and Plastic Surgery |
#### 5. Legally Marketed Predicate/Reference Devices
Predicate Device: K163484, Paxman Scalp Cooler, Paxman Coolers Limited Reference Device: K170871, DigniCap Scalp Cooling System, Dignitana AB
#### 6. Device Description
The Paxman Scalp Cooler is a scalp cooling system used to reduce the likelihood of chemotherapyinduced alopecia previously cleared by FDA in women with breast cancer (K163484). The only change from the previously cleared device is the indication for use from "women with breast cancer" to use in "cancer patients with solid tumors." There are no changes to the hardware or software in the subject device when compared to the device cleared under K163484.
The Paxman Scalp Cooler is a self-contained, mobile, electrically-powered refrigeration unit that circulates a refrigerated liquid coolant, at a pre-set temperature and flow rate, through a cooling cap, which is fitted to the top of the patient's head and connected to the refrigeration unit by a pair of coolant lines. A touchscreen controller with a menu-driven, graphical user interface, integrated into the refrigeration unit, allows the healthcare professional to initiate, monitor, and complete the scalp cooling process.
The touchscreen displays a menu-driven Graphical User Interface (GUI) that provides information to the user concerning the operational status of the scalp cooling unit; it also prompts the user to initiate some actions regarding the scalp cooling procedure and provides a timer count-down function for scalp cooling sessions. The GUI does not, however, directly control the scalp cooling process as there are pre-established programs for the scalp cooling administration.
The touchscreen controller provides feedback to the user concerning the status of the Paxman Scalp Cooler as it relates to the achievement of the pre-set temperature of the coolant, operation of the recirculation pump and connection of a cooling cap to the system. The software also
{5}------------------------------------------------
provides a timer count-down function for the initiated pre- and post-infusion cooling procedure. At the end of the pre-set time, a message is displayed on the touchscreen, and a buzzer sounds to alert the user to the fact that the scalp cooling time is complete.
#### 7. Indication for Use Statement
The Paxman Scalp Cooler is indicated to reduce the likelihood of chemotherapy-induced alopecia (CIA) in cancer patients with solid tumors.
### 8. Limitations
The sale, distribution, and use of Paxman Scalp Cooler are restricted to prescription use in accordance with 21 CFR §801.109.
Limitations on device use are also achieved through the following contraindications, warnings and precautions included in the instructions for use.
### 9. Contraindications
Scalp cooling is contraindicated in pediatric patients.
Scalp cooling is contraindicated in patients with:
- . An existing history of scalp metastases or the presence of scalp metastasis is suspected.
- Cancers of the head and neck.
- CNS malignancies (either primary or metastatic).
- Cold sensitivity, cold agglutinin disease, cryoglobulinemia, cryofibrinogenemia, cold migraine, cold urticaria, and post-traumatic cold dystrophy.
- . Hematological malignancies (leukemia, non-Hodgkin and other generalized lymphomas) or hematological malignancies that are being treated for cure.
- Imminent bone marrow ablation chemotherapy.
- . Imminent skull irradiation.
- . Previously received, or scheduled to undergo skull irradiation.
- . Scalp metastases have rarely been reported in the literature, but caution regarding their development has been a limitation for the broad-scale application of scalp cooling during chemotherapy. Theoretically, tumor cells that have seeded in the scalp might not receive adequate chemotherapy during hypothermia, thus allowing them to grow at a later date.
- . Severe liver or renal disease from any etiology who may not be able to metabolize or clear the metabolites of the chemotherapeutic agent.
- . Skin cancers including melanoma, squamous cell carcinoma, and Merkel cell carcinoma.
{6}------------------------------------------------
- Small cell carcinoma of the lung.
- . Solid tumors that have a high likelihood for metastasis in transit.
- Squamous cell carcinoma of the lung.
#### 10. Warnings and Precautions
Scalp and/or cutaneous metastases have been reported in patients with non-small cell lung cancer, colon cancer, renal cell carcinoma, ovarian cancer, and bladder cancer. Patients with advanced forms of these tumors may be more likely to experience scalp metastases with the scalp cooling system.
It cannot be guaranteed that scalp cooling will prevent all patients undergoing chemotherapy from losing any or all of their hair. The success rate of scalp cooling in reducing chemotherapyinduced hair loss varies from patient and according to the chemotherapy regimen administered.
Long-term effects of scalp-cooling and scalp metastasis have not been thoroughly studied.
Use of scalp cooling in the palliative setting in patients with metastatic cancer may also increase the risk for scalp metastases.
Use of scalp cooling with Taxanes plus anthracyclines when used together or in sequence has not been shown to be successful in preventing chemotherapeutic drug induced alopecia. The Paxman Scalp Cooler should not be used in these patients.
The effectiveness of this device in patients who have received previous chemotherapy has not been evaluated.
Clinical studies have demonstrated variable success rates in patient reduction of chemotherapyinduced alopecia with scalp cooling since the outcome is dependent on multiple factors including chemotherapy regimen, dose, duration of drug infusion, chemotherapy drug metabolism, and concomitant comorbidities. Data have shown that women who experience hair loss despite using scalp cooling might have worse quality of life than women who did not have scalp cooling.
The Paxman Scalp Cooler should only be used by appropriately qualified healthcare professionals who have been trained in the operation of the device.
Do not allow any liquids to be placed on the scalp cooler or near the touch screen controller, including drips from the cooling caps.
Avoid use at ambient temperatures of over 30°C/86°F.
Do not touch the side ventilation grills whilst the device is in use.
{7}------------------------------------------------
#### 11. Substantial Equivalence Discussion
Aside from the change in the Indications for Use, the Paxman Scalp Cooler is identical in technological characteristics, design, and performance to the predicate.
The following table compares the Paxman Scalp Cooler to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| | Subject Device | Predicate Device | |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Paxman Coolers Limited | Paxman Coolers Limited | Significant Differences |
| Trade Name | Paxman Scalp Cooler | Paxman Scalp Cooler | |
| 510(k) Number | K173032 | K163484 | Not applicable |
| Product Code | PMC | PMC | Same |
| Regulation Number | 878.4360 | 878.4360 | Same |
| Regulation Name | Scalp Cooling System | Scalp Cooling System | Same |
| Indications for Use | The Paxman Scalp Cooler is indicated to reduce the likelihood of chemotherapy-induced alopecia (CIA) in cancer patients with solid tumors. | The Paxman Scalp Cooler is indicated to reduce the likelihood of chemotherapy-induced alopecia (CIA) in women with breast cancer. | The subject device is identical in technological characteristics to the predicate device K163484 apart from the indications for use, which are being broadened from “in women with breast cancer” to “in cancer patients with solid tumors.” The subject device indications for use are identical to DigniCap, |
#### Table 5A – Comparison of Characteristics
{8}------------------------------------------------
| | Subject Device | Predicate Device | Significant Differences |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Paxman Coolers Limited | Paxman Coolers Limited | |
| Trade Name | Paxman Scalp Cooler | Paxman Scalp Cooler | K170871. |
| Mechanism of Action | The unit is a compact, mobile refrigeration unit which circulates liquid coolant at low pressure through a special cooling cap on the patient's head. The circulation of the refrigerated coolant through the cap extracts heat from the patient's scalp maintaining temperature. | The unit is a compact, mobile refrigeration unit which circulates liquid coolant at low pressure through a special cooling cap on the patient's head. The circulation of the refrigerated coolant through the cap extracts heat from the patient's scalp maintaining temperature. | Same |
| Technology Overview | The unit is composed of the main unit that contains the refrigeration components, touch screen controller, and coolant tank. There are detachable coolant lines with covers, detachable cooling cap with covers, and proprietary coolant. | The unit is composed of the main unit that contains the refrigeration components, touch screen controller, and coolant tank. There are detachable coolant lines with covers, detachable cooling cap with covers, and proprietary coolant. | Same |
| Patient Population | Cancer patients with solid tumors | Women with Stage I - II breast cancer undergoing neoadjuvant or adjuvant chemotherapy. | Differs - The expanded patient population includes all cancer patients with solid tumors, which includes women with breast cancer. |
| Set Cooling Time | Yes | Yes | Same |
| Pre/Post Cooling Time | Yes | Yes | Same |
{9}------------------------------------------------
| | Subject Device | Predicate Device | Significant Differences |
|------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|-------------------------|
| Manufacturer | Paxman Coolers Limited | Paxman Coolers Limited | |
| Trade Name | Paxman Scalp Cooler | Paxman Scalp Cooler | |
| Material of Cooling Cap | Silicone; Primasil Sil 100 silicone from 20 – 80 Shore Duromater | Silicone; Primasil Sil 100 silicone from 20 - 80 Shore Duromater | Same |
| Size of Cooling Cap | Small, Medium, Large | Small, Medium, Large | Same |
| Number of Cooling Caps/Lines | 2 | 2 | Same |
| Quick Disconnect | Yes | Yes | Same |
| Coolant Temperature Range | -15°C to 5°C | -15°C to 5°C | Same |
| Refrigerant Type | OrbisC (organic salt based) | OrbisC (organic salt based) | Same |
| Coolant Refilling | Yes | Yes | Same |
| Sterile | No | No | Same |
| Single-Use | No | No | Same |
| Main Unit Dimensions | Height: 640 mm<br>Width: 320 mm<br>Depth: 420 mm | Height: 640 mm<br>Width: 320 mm<br>Depth: 420 mm | Same |
| Weight | 29.5 kg | 29.5 kg | Same |
| AC Powered | 100 – 120 V, 50/60 Hz | 100 – 120 V, 50/60 Hz | Same |
| Touch Screen Interface | Yes | Yes | Same |
| Software Controlled | Yes | Yes | Same |
| Complies with | Yes | Yes | Same |
{10}------------------------------------------------
| | Subject Device | Predicate Device | Significant Differences |
|----------------------------------------|------------------------|------------------------|-------------------------|
| Manufacturer | Paxman Coolers Limited | Paxman Coolers Limited | |
| Trade Name | Paxman Scalp Cooler | Paxman Scalp Cooler | |
| ISO 10993-1 | | | |
| Electrical<br>Safety Testing<br>Passed | Yes | Yes | Same |
#### 12. Non-Clinical Performance Data
No additional testing was conducted for this 510(k) to support substantial equivalence. The device in this submission is identical to the previously cleared device, as the purpose of this application was for the expansion of the treatment population only. Testing of the predicate device included biocompatibility, shelf-life, shipping and packaging, electromagnetic compatibility and electrical safety, software, and bench testing. All tests met the pre-determined specifications and acceptance criteria and demonstrated the Paxman Scalp Cooler to be safe and effective as labeled.
#### 13. Clinical Performance Data
Published and non-published clinical data support use of the Paxman Scalp Cooling System in cancer patients with solid tumors. Clinical data in this application comprises registry data, a manufacturer-sponsored clinical study, clinical literature and non-published studies as described below.
#### Dutch Scalp Cooling Registry
The Dutch Scalp Cooling Registry has enrolled over 6000 cancer patients, including multiple cancer types and varies chemotherapy regimens. Seventy-one percent of all solid tumor patients wore no head cover at their last cooling session (p < 0.05) compared to 50% of all breast cancer patients at their last cooling session. Registry data demonstrate some subgroups, e.g. patients treated with certain chemotherapeutic agents, including AC, DAC and Irino mono, as well as patients with Asian or chemically-colored hair may be least likely to benefit from scalp cooling.
#### Paxman Coolers Limited SCALP study
The Paxman Coolers Limited SCALP (Scalp Cooling to Prevent Chemo-induced Hair Loss) clinical study (ClinicalTrials.gov identifier: NCT01986140) was initiated to evaluate the safety and efficacy of the Paxman Scalp Cooler, in support of its initial indication to reduce the likelihood of CIA in women with breast cancer (K163484).
Successful hair preservation was demonstrated in 69 of 130 women with cooling (53.1%; 95%CJ, 44.5%-61.4%) compared with 0 of 54 women in the no cooling (control) group (0%; 95%C), 0%-
{11}------------------------------------------------
6.6%) (success rate difference, 53.1%; 95%Cl, 44.5%-61.7%). Seventy-one device-related adverse events were reported in the cooling group, all grades 1 and 2. No serious device-related adverse events occurred.
Primary and secondary study endpoints are complete. Enrolled subjects are currently in long-term follow-up for disease status and survival.
#### Clinical literature and non-published studies
The performance and/or safety of scalp cooling with the Paxman Scalp Cooler is reported in the clinical literature and non-published studies. These documents report the effects of scalp cooling on a range of cancer types, chemotherapeutic agents and regimens, and patient characteristics. Scalp cooling treatments appear well-tolerated and were infrequently discontinued by patients. The sensation of cold was sometimes reported, no serious adverse events occurred, and patient ratings of cooling therapy were generally favorable.
Findings from the clinical literature include:
- . No unfavorable development of disease due to scalp cooling has been documented in patients with solid tumors, scalp cooling can safely be offered to patients treated with alopeciainducing chemotherapy;
- . Among women with stage I to II breast cancer receiving chemotherapy with a taxane, anthracycline, or both, those who underwent scalp cooling were significantly more likely to have < 50% hair loss after the 4th chemotherapy cycle compared with those who received no scalp cooling; and
- Solid tumor cancer patients undergoing docetaxel chemotherapy with Paxman scalp cooling had a lower occurrence of alopecia compared to patients without cooling. Alopecia occurrence among Paxman Scalp Cooler patients in the 3-weekly docetaxel regimen group was 23% compared with 74% in no cooling (control) patients. Similarly, alopecia occurrence was 7% among Paxman Scalp Cooler patients undergoing weekly docetaxel treatments compared with 17% in the no cooling patient group.
Findings from non-published studies include:
- . An observational multi-center study of patients with several types of cancers reported a head cover was used by 51% of scalp-cooled patients; 38% of scalp-cooled patients were thought to have needlessly purchased a wig;
- A multi-center trial of breast, prostate and lung cancer patients demonstrated no head cover or wig was required in 88% of patients following 45 minutes of post-infusion cooling after 3weekly docetaxel, compared with 74% of patients after 90 minutes post-infusion cooling; and
- A multi-center observational study of Lebanese patients reported a 91% success rate including 6 of 6 (100%) patients undergoing treatment with TAC who showed no signs of hair loss.
Table 5B and Table 5C, respectively, summarize published and non-published studies using the Paxman Scalp Cooling System on patients receiving chemotherapy.
{12}------------------------------------------------
#### 14. Statement of Substantial Equivalence
The Paxman Scalp Cooler described in this application is unchanged from the device previously cleared in K163484. This application is limited to a revision of the indications for use. The clinical data from the Dutch Scalp Cooling Registry, Paxman Coolers Limited SCALP study, clinical literature and non-published studies provided in this application demonstrate the Paxman Scalp Cooler can be used to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.
{13}------------------------------------------------
| Author | Type of Study/<br>Method | Patients | Chemotherapy Agents | Objective | Results<br>(Performance and Safety) | Conclusion | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| van den Hurk, CJ,<br>et al. (2012)<br>Note: Study data<br>is included 2006-<br>2010 Dutch<br>registry data and<br>Paxman<br>Netherlands<br>Clinical Study of<br>Efficacy/3 | Registry,<br>multi-center<br>(28)<br>Nurses and<br>patients<br>completed<br>questionnaires<br>on patients,<br>chemotherapy<br>and scalp<br>cooling;<br>logistic<br>regression<br>analysis was<br>used to<br>examine<br>associated<br>characteristics<br>of the scalp<br>cooling result | n=1411<br>Male n=50 (4%),<br>female n=1357<br>(96%), missing n=4<br>Types of cancer:<br>breast n=1216<br>(86%), female<br>genital n=65 (5%),<br>gastrointestinal/<br>colorectal n=63<br>(4%), lung n=19<br>(1%), prostate n=27<br>(2%), other n=16<br>(1%), missing n=5 | A60C600 (AC) (n=74),<br>A60C600/D100 (ACD)<br>(n=16), ACT overall<br>(n=50), D75A50C500<br>(TAC) (n=66), D overall<br>(n=120), F500A50C500<br>(FAC) (n=39), FEC<br>overall (n=752),<br>F500E100C500/D100<br>(FE100CD) (n=46),<br>TCarbo overall (n=68),<br>T70-90 (42), Irino 250<br>(n=42), other (n=64) | To estimate the results of<br>scalp cooling for currently<br>used chemotherapies to<br>provide patient information<br>and identify characteristics<br>associated with the results. | Overall, 50% of the 1411<br>scalp-cooled patients did not<br>wear a head cover during<br>their last chemotherapy<br>session.<br>Patients were satisfied with<br>the results in 8% of cases after<br>TAC chemotherapy and up to<br>95% after paclitaxel<br>treatment.<br>Besides the type of<br>chemotherapy, higher dose<br>and shorter infusion time,<br>older age, female gender and<br>non-West-European type of<br>hair significantly increased the<br>proportion head cover use.<br>Hair length, quantity,<br>chemical manipulation<br>(dyeing, waving, coloring),<br>wetting hair before scalp<br>cooling, and treatment with<br>chemotherapy ever before did<br>not influence the degree of<br>head covering among | Scalp cooling results as<br>recorded in this open<br>patient registry were<br>positive for most<br>regimens, justifying its use<br>by all eligible patients,<br>except for those needing<br>TAC.<br>Lengthening infusion time<br>may improve the results. | |
| Author | Type of Study/<br>Method | Patients | Chemotherapy Agents | Objective | Results<br>(Performance and Safety) | Conclusion | |
| | | | | | patients. | | |
| van den Hurk, CJ,<br>et al. (2013) | Review of<br>observational<br>studies,<br>autopsy<br>studies and<br>Munich cancer<br>registry | Studies of skin and<br>scalp skin<br>metastases in<br>patients with breast<br>cancer without<br>scalp cooling:<br>studies of scalp skin<br>metastases in scalp-<br>cooled patients with<br>(mainly) breast<br>cancer | Diverse | Using frequency data on<br>metastases in breast cancer to<br>discuss the risk of whether<br>scalp cooling might facilitate<br>hiding and disseminating scalp<br>skin metastases and thus<br>decrease survival. | The incidence of scalp skin<br>metastases in breast cancer<br>patients seems to be<br>comparable for scalp-cooled<br>(0.04-1%) and for non-scalp-<br>cooled (0.03-3%) patients. | In patients with solid<br>tumors, an unfavorable<br>development of the<br>disease due to scalp<br>cooling has never been<br>documented.<br>Scalp cooling can safely be<br>offered to patients treated<br>with alopecia-inducing<br>chemotherapy. | |
| Lemieux J, et al.<br>(2015) | Retrospective<br>cohort | n=1370 women<br>with non-metastatic<br>breast carcinoma<br>who received<br>chemotherapy in<br>neoadjuvant<br>(n=140) or adjuvant<br>setting (n=1230)<br>n=553 used scalp<br>cooling; n=817 were<br>treated in facilities<br>where scalp cooling<br>was not routinely<br>available | Not reported | To compare overall survival<br>according to whether or not<br>scalp cooling was used during<br>neoadjuvant or adjuvant<br>chemotherapy for non-<br>metastatic breast cancer. | Median follow-up was 6.3<br>years for scalp-cooled group<br>and 8.0 years for non-scalp-<br>cooled group.<br>No difference in overall<br>mortality was observed<br>between scalp-cooled<br>patients and non-scalp-cooled<br>patients (adjusted hazard<br>ratio 0.89, 95% confidence<br>interval 0.68-1.17, p=0.40). | Among women undergoing<br>neoadjuvant or adjuvant<br>chemotherapy for non-<br>metastatic breast cancer,<br>scalp cooling used to<br>prevent CIA had no<br>negative effect on survival. | |
| Author | Type of Study/<br>Method | Patients | Chemotherapy Agents | Objective | Results<br>(Performance and Safety) | Conclusion | |
| Nangia J, et al.<br>(2017)<br><br>Clinicaltrials.gov<br>identifier<br>NCT01986140 | Randomized,<br>multi-center<br>(7)<br><br>A comfort<br>scale was<br>administered<br>after each<br>treatment in<br>scalp-cooling<br>group.<br><br>Alopecia<br>assessments<br>using Common<br>Terminology<br>Criteria for<br>Adverse<br>Events<br>(CTCAE) v4.0<br>at baseline<br>and after<br>chemotherapy<br>cycle by a<br>blinded<br>clinician,<br>patient's<br>clinician and<br>patient;<br>participants | n=182 women with<br>stage 1 or II breast<br>cancer undergoing<br>chemotherapy from<br>December 2013 to<br>September 2016<br><br>Randomized to<br>receive scalp<br>cooling (n=119) or<br>no cooling (n=63)<br><br>Each patient<br>underwent scalp<br>cooling for 30<br>minutes pre-<br>infusion, during<br>infusion and 90<br>minutes post-<br>infusion | Planning to receive at<br>least 4 cycles of taxane-<br>(n=91, 64%) and/or<br>anthracycline-based<br>chemotherapy for<br>curative intent (n=51,<br>36%) | To assess whether use of the<br>Orbis Paxman Hair Loss<br>Prevention System is safe and<br>effective in reducing CIA in<br>woman with breast cancer<br>undergoing neoadjuvant or<br>adjuvant chemotherapy. | Publication reports on interim<br>analysis (n=142 patients);<br>patients will be followed for 5<br>years for safety (time and site<br>of first recurrence) and overall<br>survival.<br><br>48 of 95 (50.5%) in cooling<br>group had successful hair<br>preservation (95% confidence<br>interval 40.7%-60.4%)<br>compared to 0 of 47 (0%) in<br>the control group (95%<br>confidence interval, 0%-7.6%).<br><br>Success rate difference was<br>50.5% (95% confidence<br>interval, 40.5%-60.6%).<br><br>The trial was stopped early for<br>superiority (p=0.0061).<br><br>No statistically significant<br>differences in changes in any<br>of the quality-of-life (QOL)<br>scales from baseline to<br>chemotherapy cycle 4 were<br>observed between the scalp<br>cooling and control groups.<br><br>54 adverse events (all grades<br>1 and 2) were reported in the | Among women with stage<br>I to II breast cancer<br>receiving chemotherapy<br>with a taxane,<br>anthracycline, or both,<br>those who underwent<br>scalp cooling were<br>significantly more likely to<br>have < 50% hair loss after<br>the 4th chemotherapy<br>cycle compared with those<br>who received no scalp<br>cooling. | |
| Author | Type of Study/<br>Method | Patients | Chemotherapy Agents | Objective | Results<br>(Performance and Safety) | Conclusion | |
| | were asked if<br>they needed<br>to use a wig<br>and/or a head<br>wrap | | | | | | |
| Massey C (2004)<br>Note: Data is<br>included in<br>Paxman UK<br>Clinical Study of<br>Efficacy | Open, non-<br>randomized,<br>observational,<br>multi-center<br>(8)<br>Alopecia was<br>assessed using<br>the World<br>Health<br>Organization<br>(WHO) grading<br>system;<br>patient<br>acceptability<br>assessed by<br>questionnaire;<br>results<br>compiled by<br>Scalp Cooling<br>Assessment<br>Groups | n=94 breast cancer<br>patients who<br>underwent<br>treatment 1997-<br>2000 | 5-fluorouracil,<br>epirubicin and<br>cyclophosphamide<br>(FEC) regimen | To assess the efficacy of scalp<br>cooling to reduce alopecia for<br>women undergoing treatment<br>for breast cancer using the<br>Paxman Scalp Cooler.<br>To assess patient views on the<br>comfort and acceptability of<br>scalp cooling using the<br>Paxman Scalp Cooler. | Use of the Paxman Scalp<br>Cooler was judged a success<br>for 89% of all patients using<br>the WHO grading system for<br>alopecia and for 87% of<br>patients being specifically<br>administered the commonly<br>used 5-fluorouracil, epirubicin<br>and cyclophosphamide (FEC)<br>regimen.<br>When asked about degrees of<br>comfort during the scalp-<br>cooling process, 85% of<br>patients described it as very<br>comfortable, reasonably<br>comfortable or comfortable,<br>with only 15% of patients<br>reporting a description of<br>uncomfortable or very<br>uncomfortable. | Scalp cooling using the<br>Paxman Scalp Cooler was<br>found to be an effective<br>technique with minimal<br>side-effects for patients<br>treated with commonly<br>prescribed alopecia-<br>inducing chemotherapy<br>drugs. | |
| Bini M, et al. | Observational | N=47 breast cancer | 70% of patients | To verify effectiveness of | Median number of the | The Paxman scalp cooler | |
| Author | Type of Study/<br>Method | Patients | Chemotherapy Agents…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.