The Kisses Plus Implant System is indicated for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Kisses Plus Implant System is for single and two stage surgical procedures. It is for delayed loading.
Device Story
Dental implant system comprising fixtures and abutments; used for prosthetic rehabilitation in edentulous mandibles and maxillae. Fixtures feature internal hex connection, macro threads, and self-tapping cutting edges; available in various diameters (3.8-5.5mm) and lengths (7-14.5mm). Abutments (angled, duplex, milling, simplex, temporary, solid screw, ball) couple to fixtures to support restorations. Used by dental clinicians in clinical settings. Device provides structural foundation for artificial teeth; enables fixed or removable prosthetic support. Benefits include restoration of chewing function and oral aesthetics. System utilizes delayed loading protocol.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via fatigue testing (ISO 14801), compressive strength, rotational shear strength, screw loosening torque, surface morphology/roughness analysis, and sterilization validation (ISO 11137, ISO 17665). Biocompatibility leveraged from predicate K142813.
Indicated for partially or fully edentulous mandibles and maxillae in patients requiring single or multiple-unit restorations (cemented, screw-retained, or overdenture) or terminal/intermediate abutment support for fixed bridgework. Suitable for single or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
ACHIMHAI Medical Corporation % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 690 Roosevelt Irvine, California 92620
Re: K172630
Trade/Device Name: Kisses Plus Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: March 1, 2018 Received: March 9, 2018
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
April 3, 2018
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
### K172630
Device Name Kisses Plus Implant System
Indications for Use (Describe)
The Kisses Plus Implant System is indicated for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
The Kisses Plus Implant System is for single and two stage surgical procedures. It is for delayed loading.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(k) Summary (K172630)
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 04/03/2018
## 1. Submitter
ACHIMHAI Medical Corporation 28, Namyang-ro, 930beon-gil, Namyang-eup Hwaseong-si, Gyeonggido, Republic of Korea, 18255
## 2. U.S Agent/Contact Person
Priscilla Chung LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- Trade Name: Kisses Plus Implant System ●
- Common Name: Dental Implant System
- Classification Name: Endosseous Dental Implant
- Product Code: DZE, NHA ●
- Classification regulation: 21CFR872.3640 ●
### 4. Predicate Device:
#### ● Primary Predicate Device:
TS Fixture System (K121995) by OSSTEM Implant Co., Ltd. (for fixtures) SimpleLine II Abutment System (K112045) by Dentium (for abutments)
#### Reference Device: ●
Biogenesis Implant System - Kisses (K142813) by Achimhai Medical Corporation EZ Plus Implant System (K070562) by Megagen
{4}------------------------------------------------
## 5. Description:
Kisses Plus Implant System offers three different types of implants: Double Hex, and Kisses Single Hex. The Double Hex type has the doubled hex part to be coupled to the superstructure, whereas, the Single Hex type and Kisses Single Hex type has single hex part.
The Double Hex type fixture is a substructure of a dental implant system. It consists of the hex part to be coupled to the superstructure, the macro thread part to be fixed to the bone, and the cutting edge part with the self-tapping function. It offers double hex type.
- 3.8mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 4.2mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L .
- 4.6mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 5.1mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- . 5.5mm Dia. x 7mm L. 8mm L. 9.5mm L. 11mm L. 12.5mm L. 14.5mm L
The Single Hex type fixture is a substructure of a dental implant system. It consists of the hex part to be coupled to the superstructure, the macro thread part to be fixed to the bone, and the cutting edge part with the self-tapping function. It offers single hex type.
- 3.8mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L
- 4.2mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 4.6mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 5.1mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 5.5mm Dia. x 7mm L. 8mm L. 9.5mm L. 11mm L. 12.5mm L. 14.5mm L .
The Kisses Single Hex type fixture is a substructure of a dental implant system. It consists of the hex part to be coupled to the superstructure, the macro thread part to be fixed to the bone, and the cutting edge part with the self-tapping function. It offers single hex type.
- 3.8mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 4.2mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 4.6mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L ●
- 5.1mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L
- 5.5mm Dia. x 7mm L, 8mm L, 9.5mm L, 11mm L, 12.5mm L, 14.5mm L .
Kisses Plus Implant System also includes the following components.
- -Angled abutment
The abutments and the screws below are manufactured by our company were cleared under K142813 and can be used with the subject fixtures.
- Duplex Abutment -
- -Duplex Milling Abutment
- -Simplex Abutment
- Temporary Abutment -
{5}------------------------------------------------
- Solid Screw Abutment -
- -Ball Abutment
- Ball Cap -
- Cover Screw -
- Healing Abutment -
## 6. Indication for use:
The Kisses Plus Implant System is indicated for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
The Kisses Plus Implant System is for single and two stage surgical procedures. It is for delayed loading.
## 7. Basis for Substantial Equivalence
| | Subject Device | Primary Predicate Device | Reference Device |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | - | K121995 | K142813 |
| Device Name | Kisses Plus Implant System | TS Fixture System | Biogenesis Implant System -<br>Kisses |
| Manufacturer | Achimhai Medical Corporation | OSSTEM Implant Co., Ltd. | Achimhai Medical Corporation<br>(Note: Our company name<br>has changed from Biogenesis<br>to Achimhai.) |
| Type | Internal hex | Internal hex | Internal hex |
| Design & Size Range | (Double Hex<br>Single Hex<br>Kisses Single Hex)<br>Diameter: 3.8 – 5.5 mm<br>Length: 7-14.5 mm | Diameter: 3.5 – 6.8 mm<br>Length: 7-15 mm | Diameter: 3.8 – 5.5 mm<br>Length: 7-14.5 mm |
| Differences in design | Taper angle : 3°<br>Image: Taper angle 3 degrees | Taper angle : about 1°<br>Image: Taper angle about 1 degree | Taper angle : 3°<br>Image: Taper angle 3 degrees |
| | Cutting edge | Cutting edge | Cutting edge |
| | Image: Platform edge | Image: Platform edge | Image: Platform edge |
| | Image: Thread shape | Image: Thread shape | Image: Thread shape |
| Indications for Use | The Kisses Plus Implant<br>System is indicated for use in<br>partially or fully edentulous<br>mandibles and maxillae, in<br>support of single or multiple-<br>unit restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate abutment<br>support for fixed bridgework. | The TS Fixture System is<br>indicated for use in partially or<br>fully edentulous mandibles<br>and maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment support<br>for fixed bridgework. | The Biogenesis Implant<br>System -Kisses is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of single<br>or multiple-unit restorations<br>including; cemented retained,<br>screw retained, or<br>overdenture restorations, and<br>terminal or intermediate<br>abutment support for fixed<br>bridgework. |
| | The Kisses Plus Implant<br>System is for single and two<br>stage surgical procedures. It<br>is for delayed loading. | It is intended for delayed<br>loading. TS Fixture System is<br>compatible with abutment in<br>the ET/SS Implant System. | The Biogenesis Implant<br>System - Kisses is for single<br>and two stage surgical<br>procedures. It is for delayed<br>loading. |
| Principle of<br>Operation | The device is a substructure<br>of a dental implant system. It<br>consists of the hex part to be<br>coupled to the superstructure,<br>the single thread part to be<br>fixed to the bone, and the<br>cutting edge part with the self-<br>tapping function. | The device is a substructure<br>of a dental implant system. It<br>consists of the hex part to be<br>coupled to the superstructure,<br>the single thread part to be<br>fixed to the bone, and the<br>cutting edge part with the self-<br>tapping function. | The device is a substructure<br>of a dental implant system. It<br>consists of the hex part to be<br>coupled to the superstructure,<br>the single thread part to be<br>fixed to the bone, and the<br>cutting edge part with the self-<br>tapping function. |
| Material Composition | Ti Gr.4(ASTM F67) | Ti Gr.4(ASTM F67) | Ti Gr.4(ASTM F67) |
| Surface Treatment | SLA Treatment | SLA Treatment | SLA Treatment |
| Sterile | Yes | Yes | Yes |
| Sterilization Method | Gamma Radiation | Gamma Radiation | Gamma Radiation |
### 7.1. Fixture
{6}------------------------------------------------
### Substantial Equivalence Discussion
Kisses Plus fixture is substantially equivalent to the predicate devices in terms of indications for use and technical characteristics. They have similar design that all the devices are two piece type fixtures with thread part to be fixed to the bone. They are made of the same material and all have cutting edge part for self-tapping function. The size range of the primary predicate device encompasses the size range of the subject device. There are slight differences in design such as taper angle, cutting edge shape, platform edge shape, thread shape. The taper angle of the subject device is different from the primary predicate device, but the same as the reference device.
{7}------------------------------------------------
The cutting edge of subject device has spiral shape. On the other hand, that of primary predicate device has pocket shape. However, the function of the cutting edge is the same as that of forming the path through which the bone is scratched when the fixture is placed. The reference device also has spiral cutting edge.
The platform edge of the subject device has different shape from the reference device but is the same as the primary predicate device.
Per the thread design, all the devices have the same pitch (0.8mm) but the angle of the thread is slightly different. However this does not raise an issue in effectiveness with this small difference.
| 7.2. Abutments | | | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | Subject Device | Primary Predicate Device | Reference Device |
| | | K112045 | K070562 |
| Device Name | Kisses Plus Implant System | SimpleLine II Abutment System | EZ Plus Implant System |
| Manufacturer | Achimhai Medical Corporation | Dentium | Megagen |
| Type | Internal | Internal | Internal |
| Indications for Use | The Kisses Plus Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.<br>The Kisses Plus Implant System is for single and two stage surgical procedures. It is for delayed loading. | The SimpleLine II Abutment system is intended for use as an aid in prosthetic rehabilitation. | The EZ Plus Implant Systems are intended to be surgically placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surgical procedure. Large angle abutments (e.g. 25°) on small diameter implants of the EZ Plus internal connection system are intended for the anterior region of the mouth and not intended for use in the posterior region of the mouth due to limited strength of the implant. |
| Principle of Operation | The abutments are to be coupled to the fixtures to make temporary or final prosthesis. | The abutments are to be coupled to the fixtures to make temporary or final prosthesis. | The abutments are to be coupled to the fixtures to make temporary or final prosthesis. |
| Abutment 1- Angled Abutment | | | |
| <Angled Abutment> Design & Size Range | Image: Hex<br>Hex<br>Image: Non-Hex<br>Non-Hex<br>Angle : 15-25°<br>Diameter: 4.0-5.5mm<br>Post Height : 8.0mm<br>Gingival Height: 2.0-6.0mm | Image: Octa<br>Octa<br>Angle : 15-25°<br>Diameter: 3.8-6.5mm<br>Post Height : 9.0mm<br>Gingival Height: 1mm | Image: Trip<br>Trip<br>Image: Non-Trip<br>Non-Trip<br>Angle : 15-25°<br>Diameter: 4.0 - 6.0mm<br>Post Height : 7.0mm<br>Gingival Height: 2.0 - 5.0mm |
| Difference in design | Anti-rotation : Hex | Anti-rotation : Octa | Anti-rotation : Trip<br>Fixture side |
| Intended Use | Cement retained restoration | Cement retained restoration | Cement retained restoration |
| Material Composition | Ti Gr.4(ASTM F67) | Ti Gr.4(ASTM F67) | Ti Gr.4(ASTM F67) |
| Surface Treatment | Anodizing coloring - Gold<br>color (Entire Body) | No | Anodizing coloring — Gold<br>color (Entire Body) |
| Sterile | No | No | No |
### 7.2. Abutments
510(k) Submission.
{8}------------------------------------------------
### Substantial Equivalence Discussion
The subject Angled Abutment is substantially equivalent to the predicate devices in terms of indications for use and technical characteristics. They are material and have similar design. The size range of the predicate devices encompasses the size range of the subject device. Differences in design are anti-rotation feature. The Anti-rotation serves to prevent the rotation of the
prosthesis in a single case. Each device has its own anti-rotation feature design for the subject device. Octa design for the primary predicate device, and the Trip design for the reference device. Even though the design is different among the devices, this feature serves the same function. Also, the test result of the fatigue test supported substantial equivalence despite this difference.
### Substantial Equivalence Discussion
Kisses Plus Implant System is substantially equivalent to the predicate devices in terms of indications for use and technical characteristics. The indications for use statement of the subject device is exactly same as the reference device (k142813) made by our company even in wording. When comparing that of the primary predicate device, the wording, "intended for" has changed to "indicated for" but this change does not constitute a new claim. The subject device also a claim for "single and two stage surgical procedures" which the primary predicate device does not have, but the reference device has the same claim. For the abutments, the subject device mentions "terminal or intermediate abutment" in the indications for use which is different from the primary predicate device that uses the term "final or temporary abutment", however, the meaning of the verbiage is not altered. .
They are made of the same material and have similar design. The size range of the predicate device encompasses the size range of the subject device. There are slight differences in design, however, it is very minor not affecting substantial equivalence. We have performed the fatigue test, material/surface characterization, biocompatibility, sterilization validation, shelf life validation studies to make sure the difference in design does not raise an issue in safety and effectiveness and the test result of the test supported substantial equivalence.
Based on the information and test results provided in submission, we conclude that the subject device is substantially equivalent to the predicate devices.
{9}------------------------------------------------
## 8. Non-Clinical Testing
- Sterilization validating testing has been performed in accordance with ISO 11137 for ● gamma sterilization and ISO 17665-1 and ISO 17665-2 for steam sterilization.
- . Five year of shelf life has been validated through accelerating testing.
- . Surface Morphology, Surface Roughness Analysis, and FT-IR Analysis were performed to evaluate the fixture surface characteristics after SLA treatment.
- . Other Performance tests: Visual inspection, Dimension, Packing inspection, Sterility, Adaption accuracy, Marginal adaptation, Compressive strength, Fatigue (ISO 14801), Rotational sheer strength, Screw loosening torque tests
- . Biocompatibility: The fixtures and the abutments of the subject device are the same as the predicate device (BIOGENESIS IMPLANT SYSTEM, K142813) manufactured by our company in terms of raw materials, surface treatment, sterilization method, and the manufacturing processes. No biocompatibility testing was conducted as no new materials, manufacturing or processes were introduced.
### 9. Conclusion
The subject devices and the predicate device have the same indications for use and have the same technological characteristics.
Overall, the Kisses Plus Implant System has the following similarities to the predicate device:
- * have the same indications for use,
- * use the same operating principle,
- * incorporate the same design,
- * incorporate the same material and the sterilization method.
Based on the similarities, we conclude that the Kisses Plus Implant System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.