The SimpleLine II Abutment system is intended for use as an aid in prosthetic rehabilitation.
Device Story
SimpleLine II Abutment System consists of various dental abutment components (cover screws, healing abutments, solid/dual/angled/temporary/casting abutments) used in prosthetic rehabilitation. These components connect to endosseous dental implants to support artificial teeth. Used in clinical dental settings by licensed professionals; components are surgically placed or attached to implants to restore chewing function. The system provides the mechanical interface between the implant and the final dental prosthesis. Benefits include restoration of oral function and aesthetics for patients with missing teeth.
Clinical Evidence
No clinical data. Evidence consists of bench testing including biocompatibility (ISO 10993/ISO 7405: cytotoxicity, pyrogenicity, sensitization, intracutaneous reactivity, systemic toxicity) and mechanical performance testing (ISO static compressive torque force, fatigue, and adaptation accuracy).
Technological Characteristics
Materials: Ti-6Al-4V ELI alloy (ASTM F136), Pure Titanium Gr4 (ASTM F67), Ti-Gr2 (ASTM F67), Gold alloy, Co-Cr-Mo (ASTM F799). Components include various abutment types (solid, dual, angled, casting, temporary). Angled abutments available in 3.8-6.5mm diameters with 15° and 25° angles. Mechanical interface for endosseous dental implants.
Indications for Use
Indicated for use as an aid in prosthetic rehabilitation in patients requiring dental implant-supported restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
JAN 2 6 2012
## Dentium
## 510(k) Summary
### 2011.12.02
#### 1. Company and Correspondent making the submission
| Company | Dentium Co., Ltd. |
|---------|--------------------------------------------------------------------|
| Address | 27-5 Iui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-766, Korea |
| Contact | Jonghyun, Choi / R&D |
| E-Mail | eastman@dentium.com |
| Phone | +82 31 207 2200 |
| Fax | +82 31 207 3883 |
#### 2. Device Name
| Proprietary name: | SimpleLine II Abutment System |
|----------------------|------------------------------------------------------------|
| Common name: | Abutment |
| Classification name: | Abutment, Dental, Endosseous implant<br>NHA. 21CFR872.3630 |
#### 3. Predicated Device
Implantium Prosthetics, Dentium Co., Ltd. (K070228, K052957) Implantium, Dentium Co., Ltd. ( K041368)
#### 4. Device Description
ﭘﺮ ﺗ
SimpleLine II Abutment system is intended for use as an aid in prosthetic rehabilitation. It consists of Cover screw, Healing Abutment, Solid Abutment, Dual Abutment, Abutment Screw, SCA Abutment, Dual Milling Abutment, Angled Abutment (15° and 25°), Direct casting Abutment-Gold, Metal-Casting Abutment-Co-Cr, Temporary Abutment-Ti and Screw Abutment. All of abutment designs are based on existing Implantium that are cleared for marketing in the United States. The materials of the SimpleLine II Abutment System is Ti-6Al-4V ELI alloy(ASTM F136), Pure Titanium Gr4(ASTM F67). Gold alloy, Co-Cr-Mo (ASTM F799), Ti-Gr2(ASTM F67).
The Angled Abutments are available in diameters of 3.8, 4.3, 4.5, 5.5 and 6.5mm with 15° and 25° angles to the axis of the implant.
{1}------------------------------------------------
# entium
| Trade name | SimpleLine II<br>Abutment | Implantium<br>Prosthetics | Implantium<br>Prosthetics | Implantium |
|------------------------|----------------------------------------------------------------|---------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Dentium Co., Ltd. | Dentium Co., Ltd. | Dentium Co., Ltd. | Dentium Co., Ltd. |
| 510(k)<br>Number | New device | K052957 | K070228 | K041368 |
| Materials | Titanium, Titanium<br>Alloy, Co-Cr, Gold<br>Alloy | Titanium, Titanium<br>Alloy, Gold Alloy | Titanium, Titanium<br>Alloy | Titanium, Titanium<br>Alloy |
| Form | Abutment, Dental,<br>Endosseous<br>implants | Abutment, Dental,<br>Endosseous<br>implants | Abutment, Dental,<br>Endosseous<br>implants | Implant, endosseous,<br>root-form |
| Indications<br>for use | Intended for use as an<br>aid in prosthetic<br>rehabilitation. | SAME | SAME | Implantium is intended<br>to be surgically placed<br>in the bone of the upper<br>or lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as artificial teeth, and to<br>restore the patient's<br>chewing function. |
| Use | Prescription | Prescription | Prescription | Prescription |
#### 5. Substantial Equivalence:
The comparison between SimpleLine II Abutment and other predicated devices is claimed to be substantially equivalence in terms of indication for use, materials, product form, technology, and performance specifications.
The difference between SimpleLine II Abutment and predicate devices are the product shape and slight mechanical and physical characteristics. However, the slight differences do not affect to the application of the device. Therefore, we state that SimpleLine II Abutment system is substantially equivalent to the predicate devices.
#### 6. Indication for Use
The SimpleLine II Abutment system is intended for use as an aid in prosthetic rehabilitation.
### 7. Performance Testing
Biocompatibility testing on the proposed SimpleLine II Abutment has been completed. Requirements for biological evaluation of the proposed device were based on the 10(k) Summary Page 2 of 3
{2}------------------------------------------------
K112045
## entium
ISO7405(2008), "Dentistry-Evaluation of biocompatibility of medical devices used in dentistry." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed:
-ISO Cytotoxicity
-ISO Pyrogenicity
-ISO Sensitization
-ISO Intracutaneous reactivity
The proposed SimpleLine II Abutment was evaluated using the following performance bench testing to confirm the performance characteristics:
ISO Static compressive -Torque Force
-ISO Fatigue -Adaptation Accuracy
ISO Systemic toxicity
#### 8. Review
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the SlimpleLine II Abutment system met the established specifications necessary for consistent performance according to its intended use.
SimpleLine II Abutment system has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.
#### 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Dentium Co., Ltd. concludes that the SimpleLine II Abutment System is substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 2 6 2012
Mr. Jonghyun Choi Assistant Manager Dentium Company, Limited 27-5 lui-dong, Yeongtong-gu, Suwon-si, Gyeonggi-do Republic of KOREA 443-270
Re: K112045
Trade/Device Name: SimpleLine II Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: NHA Dated: January 13, 2012 Received: January 17, 2012
Dear Mr. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Choi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Susan Reonne
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Dentium
### Indications for Use
510(k) Number: K112045
Device Name:
SimpleLine II Abutment system
Indications for Use:
The SimpleLine II Abutment System is intended for use as an aid in prosthetic rehabilitation.
V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Russo
..
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112045
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.