The Biogenesis Implant System –Kisses is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Biogenesis Implant System - Kisses is for single and two stage surgical procedures. It is for delayed loading.
Device Story
Biogenesis Implant System - Kisses consists of titanium dental implants and associated abutments (Duplex, Simplex, Solid Screw, Ball, Temporary) used to replace missing teeth in edentulous or partially edentulous patients. Implants are surgically placed into the mandible or maxilla by a dental professional. The system features an internal hex connection, self-tapping threads, and SLA surface treatment. Components are used to support prosthetic restorations. The device provides a stable foundation for artificial teeth, restoring chewing function and aesthetics. Implants are provided sterile via gamma radiation; abutments are non-sterile. Clinical benefit is the restoration of oral function through fixed or removable dental prostheses.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included sterilization validation (ISO 17665-1/2, ANSI/AANI ST79), shelf-life testing (ASTM F1980), and biocompatibility testing (ISO 10993-1) in accordance with FDA guidance for endosseous dental implants.
Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. Suitable for single or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
TS Fixture System by Ossetem Implant Co., Ltd. (K121995)
Implantium Abutments by Dentium Co., Ltd. (K052823)
EZ Plus Implant System by Megagen Co., Ltd. (K070562)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
Biogenesis Co., Ltd. c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 East Chapman Avenue Suite 110 Fullerton, California 92831
Re: K142813
Trade/Device Name: Biogenesis™ Implant System - Kisses Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: June 26, 2015 Received: July 01, 2015
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K142813
Device Name Biogenesis Implant System -Kisses
#### Indications for Use (Describe)
The Biogenesis Implant System –Kisses is in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
The Biogenesis Implant System - Kisses is for single and two stage surgical procedures. It is for delayed loading.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
■ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(k) Summary (K142813)
This summary of 510(K) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 07/30/2015
## 1. Applicant / Submitter
Biogenesis Co., Ltd. Mecha Zone #1109, SK Techno Park, 50. Wanam-Ro. Seongsan-Gu. Changwon-Si, Gyeongsangnam-Do, Republic of Korea, 642-971 Tel: +82-55-607-0345 Fax: +82-55-607-0347
## 2. Submission Correspondent
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- Trade Name: Biogenesis™ Implant System Kisses
- · Common Name: Dental Implant
- · Classification Name: Endosseous Dental Implant
- · Product Code: DZE, NHA
- Classification regulation: 21CFR872.3640
## 4. Predicate Device:
## - Primary Predicate Device:
TS Fixture System by Ossetem Implant Co., Ltd. (K121995)
## · Reference Predicate Devices:
Dentium Co., Ltd Implantium by Dentium Co., Ltd. (K041368) Implantium Abutments by Dentium Co., Ltd. (K052823) EZ Plus Implant System by Megagen Co., Ltd. (K070562)
{4}------------------------------------------------
# 5. Description:
The Biogenesis 110 Implant System - Kisses is a dental implant system made of titanium intended to be surgically placed in the bone of the upper or lower jaw arches. This product is a substructure of a dental implant system to replace a single tooth, partial tooth and the lost root of edentulous patients. It consists of the hex part to be coupled to the superstructure, the single thread part to be fixed to the bone, and the cutting edge part with the self-tapping function.
The Biogenesis™ Implant System offers bone level implants in the size range of 3.8 – 5.5 mm diameter with 7 – 14.5mm length.
- Diameter: 3.8mm / 4.2mm / 5.1mm / 5.5mm x Length: 7mm / 8mm / 9.5mm / 11mm / 12.5mm / 14.5mm
The Biogenesis TM Implant System also offers the following components.
- · Duplex Abutment
- Duplex Milling Abutment
- Simplex Abutment
- Temporary Abutment
- · Solid Screw Abutment
- · Ball Abutment
- Ball Cap
The implants are intended for use with straight implant only straight implants are included in the submission.
# 6. Indication for use:
The Biogenesis Implant System - Kisses is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Biogenesis Implant System is for single and two stage surgical procedures. It is intended for delayed loading.
# 7. Basis for Substantial Equivalence
The Biogenesis Implant System – Kisses is substantially equivalent to previously marketed devices as presented in the comparison tables below.
The subject device is similar to the predicate devices based on the intended use, the principle of operations, the materials, the surface treatment, the size range and the technological characteristics. The verbiage of the Indications for Use of the subject device is slightly different than that of the declared predicates; however, these slight differences in wording does not change the intended use of the subject device has
{5}------------------------------------------------
compared to the declared predicate. In addition, the external design of the subject device is slightly different from the predicate devices; however, the performance testing provided in this 510k submission supports that the subject device is substantially equivalent to the predicate devices.
# Comparison Chart
#### 1) Fixture
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | N / A | K121995 | K041368 |
| Device Name | Biogenesis Implant System -<br>Kisses | TS Fixture System | Dentium Co., Ltd Implantium |
| Manufacturer | Biogenesis | Osstem | Dentium |
| Type | Internal hex | Internal hex | Internal hex |
| Design & Size Range | Diameter: 3.8 – 5.5 mm | Diameter: 3.5 – 5.0 mm | Diameter: 3.4 – 5.8 mm |
| | Length: 7 – 14.5 mm | Length: 7-15 mm | Length: 7-15 mm |
| Indications for Use | The Biogenesis Implant<br>System - Kisses is indicated for<br>use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including; cemented retained,<br>screw retained, or overdenture<br>restorations, and terminal or<br>intermediate abutment support<br>for fixed bridgework.<br>The Biogenesis Implant System<br>- Kisses is for single and two<br>stage surgical procedures. It is<br>for delayed loading | The TS fixture system is<br>indicated for use in partially or<br>fully edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including; cemented retained,<br>screw retained, or overdenture<br>restorations, and final or<br>temporary abutment support for<br>fixed bridgework. It is intended<br>for delayed loading.<br>Ts fixture system is compatible<br>with abutment in the et/ss<br>implant system | The dentium co., ltd.<br>Implantium is intended to be<br>surgically placed in the bone<br>of the upper or lower jaw<br>arches to provide support for<br>prosthetic devices, such as<br>artificial teeth, and to restore<br>the patient's chewing function.<br>This may be accomplished by<br>either a two-stage surgical<br>procedure or a single surgical<br>procedure. If a single surgical<br>procedure is used, single or<br>multiple implants may be<br>inserted (type i, or iii bone)<br>provided good initial stability<br>(> 40 ncm) is achieved. Not<br>intended for immediate<br>loading. |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | SLA Treatment | SLA Treatment | SLA Treatment |
| Sterile | Yes | Yes | Yes |
| Sterilization Method | Gamma | Gamma | Gamma |
| 510(K) Number | Subject Device | Reference Predicate Device | Reference Predicate Device |
| N / A | Biogenesis Implant System<br>- Kisses | K052823 | K070562 |
| Device Name | Biogenesis Implant System<br>- Kisses | Implantium Abutments | EZ Plus Implant System |
| Manufacturer | Biogenesis | Dentium | Megagen |
| Type | Internal | Internal | internal |
| Indications for Use | The Biogenesis Implant<br>System - Kisses is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate abutment<br>support for fixed<br>bridgework.<br>The Biogenesis Implant<br>System - Kisses is for single<br>and two stage surgical<br>procedures. It is for delayed<br>loading | The implantium abutments<br>are intended to be used with<br>the im-plantium root-form<br>endosseous dental implant to<br>aid in prosthetic rehabilitation<br>including overdenture<br>retention. After the root-form<br>endosseous dental implant is<br>surgically placed, the<br>endosseous dental implant<br>abutment device is attached<br>to it. | The ez plus implant systems<br>are intended to be surgically<br>placed in the upper or lower<br>jaw to support prosthetic<br>devices, such as artifical teeth<br>and to restore a patient's<br>chewing function. This may be<br>accomplished using either a<br>two stage surgical procedure<br>or a single stage surgical<br>procedure. Large angle<br>abutments (e.g. 25 degree)<br>on small diameter implants of<br>the ez plus internal<br>connection system are<br>intended for the anterior<br>region of the mouth and not<br>intended for use in the<br>posterior region of the mouth<br>due to limited strenth of the<br>implant |
| | Abutment 1- Duplex Abutment | | |
| <Duplex Abutment><br>Design & Size Range | Hex & Non-Hex<br>Diameter: 4.0mm - 6.5mm<br>Post Height : 4.0mm -<br>7.0mm<br>Gingival Height: 1mm -<br>7.0mm | Hex & Non-Hex<br>Diameter: 4.5mm - 6.5mm<br>Post Height : 5.5mm<br>Gingival Height: 1mm -<br>5.5mm | Trip & Non-Trip<br>Diameter: 4mm - 6mm<br>Post Height : 5mm<br>Gingival Height: 1mm - 5mm |
| Intended Use | Cement retained restoration | Cement retained restoration | Cement retained restoration |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | Anodizing coloring - Gold<br>color (Entire Body) | TiN coating - Gold color<br>(Partly) | Anodizing coloring - Gold<br>color(Entire Body) |
| Sterile | No | No | No |
| | Abutment 2- Duplex Milling Abutment | | |
| <Duplex Milling<br>Abutment><br>Design & Size Range | | | |
{6}------------------------------------------------
#### 2) Abutment
{7}------------------------------------------------
| | Hex & Non-Hex<br>Diameter: 4.0mm - 6.5mm<br>Gingival Height: 1.5mm –<br>3.0mm | Hex & Non-Hex<br>Diameter: 4.0mm - 6.5mm<br>Gingival Height: 1.5mm –<br>3.0mm | Trip & Non-Trip<br>Diameter: 5mm - 6mm<br>Gingival Height: 1mm – 5mm |
|---------------------------------------------|----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Intended Use | Cement retained restoration | Cement retained restoration | Cement retained restoration |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | Anodizing coloring - Gold<br>color(Entire Body) | TiN coating - Gold color<br>(Partly) | Anodizing coloring - Gold<br>color(Entire Body) |
| Sterile | No | No | No |
| Abutment 3- Simplex Abutment | | | |
| <Simplex Abutment><br>Design & Size Range | Diameter: 4.0mm - 6.5mm<br>Post Height : 4.0mm -<br>5.5mm<br>Gingival Height: 1mm –<br>7.0mm | Diameter: 4.5mm - 6.5mm<br>Post Height : 5.5mm<br>Gingival Height: 1mm -<br>5.5mm | Diameter: 4mm - 6mm<br>Post Height : 3.5 / 5mm<br>Gingival Height: 1mm - 5mm |
| Intended Use | Cement retained restoration | Cement retained restoration | Cement retained restoration |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | Anodizing coloring - Gold<br>color(Entire Body) | TiN coating - Gold color<br>(Partly) | Anodizing coloring - Gold<br>color(Entire Body) |
| Sterile | No | No | No |
| Abutment 4- Solid Screw Abutment | | | |
| <Solid Screw><br>Design & Size Range | Diameter: 4.8mm<br>Post Height : 4.0mm –<br>7.0mm<br>Gingival Height: 1.0mm –<br>7.0mm | Diameter: 4.5mm - 6.5mm<br>Post Height : 5.5mm<br>Gingival Height: 1mm -<br>5.5mm | Diameter: 4mm - 6mm<br>Post Height : 3.5 / 5mm<br>Gingival Height: 1mm -<br>5.0mm |
| Intended Use | Cement retained restoration | Cement retained restoration | Cement retained restoration |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | Anodizing coloring - Gold<br>color(Entire Body) | TiN coating - Gold color<br>(Partly) | Anodizing coloring - Gold<br>color(Entire Body) |
| Sterile | No | No | No |
| Abutment 5- Temporary Abutment | | | |
| <Temporary Abutment><br>Design & Size Range | Hex & Non-Hex | | Hex & Non-Hex |
{8}------------------------------------------------
| | Diameter: 4.0 / 4.5mm<br>Gingival Height: 1mm | Hex & Non-Hex<br>Diameter: 4.5mm<br>Gingival Height: 1mm | Diameter: 4mm / 4.5mm<br>Gingival Height: 1.5mm |
|--------------------------------------|------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------|
| Intended Use | To manufacture temporary prostheses | To manufacture temporary prostheses | To manufacture temporary prostheses |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.3 |
| Surface Treatment | No | No | No |
| Sterile | No | No | No |
| Material Composition | Delrin 500P NC010 | Delrin 500P NC010 | Delrin 500P NC010 |
| Surface Treatment | No | No | No |
| Sterile | No | No | No |
| Abutment 6- Ball abutment | | | |
| <Solid Screw><br>Design & Size Range | Diameter: 3.0, 3.5mm<br>Gingival Height: 0.5mm – 7mm | Diameter: 3.3mm –3.5mm<br>Gingival Height: 0mm - 5mm | |
| Intended Use | Cement retained restoration | Cement retained restoration | |
| Material Composition | Ti 6Al 4V ELI, Gr.23 | Ti Gr.4 | |
| Surface Treatment | Anodizing coloring – Gold<br>color(Entire Body) | No | |
| Sterile | No | No | |
#### 3) Screws
| | Subject Device | Reference Predicate Device |
|-----------------------------------------|----------------------------------------|--------------------------------------------------|
| 510(K) Number | N / A | K052823 |
| Device Name | Biogenesis Implant System - Kisses | Implantium Abutments |
| Manufacturer | Biogenesis | Dentium |
| Abutment Screw | | |
| <Abutment Screw><br>Design & Size Range | Image: Abutment Screw Subject Device | Image: Abutment Screw Reference Predicate Device |
| Intended Use | To connect the abutment to the fixture | To connect the abutment to the fixture |
| Material Composition | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 |
| Surface Treatment | No | No |
| Sterile | No | No |
| Cover Screw | | |
| <Cover Screw><br>Design & Size Range | Image: Cover Screw Subject Device | Image: Cover Screw Reference Predicate Device |
{9}------------------------------------------------
| Intended Use | To provide sealing effect for fixture | To provide sealing effect for fixture |
|----------------------------------------|-----------------------------------------------------|-----------------------------------------------------|
| Material Composition | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 |
| Surface Treatment | Anodizing coloring(Entire Body) | Anodizing coloring(Entire Body) |
| Sterile | Yes | Yes |
| Sterilization Method | Gamma | Gamma |
| Healing Screw | | |
| <Healing Screw><br>Design & Size Range | Image: Healing Screw | Image: Healing Screw |
| Intended Use | To help the soft tissue of gum naturally<br>formed. | To help the soft tissue of gum naturally<br>formed. |
| Material Composition | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 |
| Surface Treatment | No | No |
| Sterile | Yes | Yes |
| Sterilization Method | Gamma | Gamma |
# 8. Non-Clinical Testing
Various performance testing has been performed on the Biogenesis Implant System-Kisses and the test results met the pre-set criteria. Testing included:
- · Sterilization validating testing according to ISO 17665-1/2 and ANSI/AANI ST79
- · Shelf life testing according to ASTM F1980
- · Biocompatibility testing per ISO 10993-1
- · Conformance to FDA Guidance Document for Endosseous Dental Implants and Abutments
There might be some differences in sterilization parameters, shelf life and manufacturing processes between the subject device and the predicate devices, however, the test results supported that the subject device is substantially equivalent to the predicate devices.
# 9. Conclusion
The subject device and the predicate devices have the same intended use and have similar technological characteristics.
Overall, the Biogenesis Implant System -Kisses has the following similarities to the predicate devices:
- has the same intended use,
- · uses the same operating principle,
- · incorporates the same basic design,
- incorporates the same material and the surface treatment.
{10}------------------------------------------------
Based on the similarities, we conclude that the Biogenesis Implant System –Kisses is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.