The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cementless application while the glenoid baseplate (metal-back or porous) is intended for cementless application with the addition of bone screws for fixation. All the implants of shoulder prosthesis are used for primary or revision surgeries. The glenoid bases (metal-back or porous) are intended for cementless application with the addition of cortical or cancellous bone screws. The porous glenoid base must be used for a total anatomical or reverse prosthesis.
Device Story
The ARROW Reverse Porous Glenoid is a shoulder prosthesis component designed for total reverse shoulder arthroplasty. It articulates with the ARROW Reverse shoulder system. The device is implanted by a surgeon using dedicated instrumentation. It is intended for cementless fixation to the glenoid bone, supplemented by cortical or cancellous bone screws. By replacing the damaged joint surfaces in patients with severe arthropathy and deficient rotator cuffs, the device restores shoulder function, provided the patient has a functional deltoid muscle. It is used in clinical settings for primary or revision shoulder surgeries.
Clinical Evidence
Bench testing only. Performance verified via testing to ASTM F1829-16, ASTM F2028-14, and ASTM F1378-12 standards. LAL testing confirmed endotoxin levels meet the 20EU/device limit.
Technological Characteristics
Constructed from titanium alloy. Designed for cementless fixation with bone screws. Dimensions available in a range of sizes. Tested per ASTM F1829-16, ASTM F2028-14, and ASTM F1378-12. No software or electronic components.
Indications for Use
Indicated for patients with severe shoulder arthropathy and grossly deficient rotator cuff, or failed shoulder joint replacement with grossly deficient rotator cuff. Requires functional deltoid muscle and adequate glenoid bone stock. Used for primary or revision surgeries.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 23, 2018
Fournitures Hospitalieres Industrie Patricia Donnard Regulatory Affairs Manager ZI de Kernevez - 6 rue Nobel Quimper 29000 FRANCE
Re: K171789
Trade/Device Name: ARROW® Reverse Porous Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWT Dated: December 14, 2017 Received: December 18, 2017
Dear Patricia Donnard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-<br>0120<br>Expiration Date: January 31, 2017<br>See PRA Statement on last page |
|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
510(k) Number (if known) K171789
Device Name
ARROW® Reverse porous alenoid
Indications for Use (Describe)
#### REVERSE PROSTHESIS (METAL-BACK OR POROUS GLENOID IMPLANT)
The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cementless application while the glenoid baseplate (metal-back or porous) is intended for cementless application with the addition of bone screws for fixation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork (PRA) Reduction Act Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1 FDAPSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance to the requirements of 21 CFR 807.92.
Date: January 17, 2018
# The assigned 510(k) number is: K171789
# Applicant
Fournitures Hospitalières Industrie 6 Rue Nobel, Z.I. de Kernevez 29000 QUIMPER - FRANCE Tel: (+33) 2.98.55.68.95 Fax: (+33) 2.98.53.42.13
## Company contact
Patricia DONNARD, Regulatory Affairs Manager p.donnard@fh-industrie.com Tel: (+33) 2.98.55.68.95
Naoual RAHIMI, Experienced RA Officer - Overseas Export Specialist n.rahimi@fh-industrie.com Tel: (+33) 2.98.55.68.95
# Product
ARROW® Reverse porous glenoid Trade name:
- Common name: Shoulder prosthesis
Classification: Shoulder joint metal/polymer semi-constrained cemented prosthesis. Product code: PHX, KWS, KWT Regulation number: 21 CFR 888.3660 Class: Il
#### Information on predicate devices to which substantial equivalence is claimed
| Manufacturer: | Fournitures Hospitalières Industrie |
|-------------------------------------|--------------------------------------------------------------------------------------------------------|
| Device Trade Name: | ARROW Reverse Shoulder System |
| 510(k): | K112193 |
| Manufacturer:<br>Device Trade Name: | Fournitures Hospitalières Industrie<br>ARROW Reverse Shoulder long keel and short keel<br>glenoid base |
| 510(k): | K142778 |
| Manufacturer: | Fournitures Hospitalières Industrie |
| Device Trade Name: | ARROW Anatomical porous glenoid base |
| 510(k): | K162068 |
{4}------------------------------------------------
| Manufacturer: | TORNIER SAS |
|--------------------|-------------------------------------------|
| Device Trade Name: | AEQUALIS Reversed Shoulder Prosthesis |
| 510(k): | K132285 |
| Manufacturer: | Zimmer |
| Device Trade Name: | TRABECULAR metal, Reverse Shoulder System |
| 510(k): | K052906 |
## Reference Devices:
| Manufacturer: | Fournitures Hospitalières Industrie |
|--------------------|--------------------------------------|
| Device Trade Name: | ARROW Anatomical porous glenoid base |
| 510(k): | K162068 |
| Manufacturer: | Fournitures Hospitalières Industrie |
| Device Trade Name: | CoLS Fixation System |
| 510(k): | K120740 |
| Manufacturer: | Fournitures Hospitalières Industrie |
| Device Trade Name: | CALCANAIL |
| 510(k): | K150463 and K150471 |
| Manufacturer: | Fournitures Hospitalières Industrie |
| Device Trade Name: | ARROW Anatomic shoulder system |
| 510(k): | K093599 |
# Device description
The ARROW® Reverse porous glenoid base is used in total reverse shoulder prosthesis and is designed to articulate with the ARROW® Reverse shoulder system (cleared in K112193). The porous glenoid base is used with bone screws for fixation (cleared in K112193).
The ARROW® Reverse porous glenoid is intended to be implanted using the dedicated instrumentation supplied by the manufacturer. This instrument set is common for all the configurations of prosthesis: simple humeral prosthesis, total anatomical prosthesis and reverse prosthesis.
#### Indications for use / Intended use
- Indications for use
REVERSE PROSTHESIS (METAL-BACK OR POROUS GLENOID IMPLANT) The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cemented or cementless application while the glenoid baseplate (metal-back or porous) is
intended for cementless application with the addition of bone screws for fixation.
{5}------------------------------------------------
- . Intended use
All the implants of shoulder prosthesis are used for primary or revision surgeries.
The glenoid bases (metal-back or porous) are intended for cementless application with the addition of cortical or cancellous bone screws. The porous glenoid base must be used for a total anatomical or reverse prosthesis.
## Comparison of technological characteristics
The ARROW Reverse porous glenoid and the above selected predicate devices have the same intended use and substantial similar indications for use and share the following similarities:
- they are made out of the same material (titanium alloy), י
- they are available in similar ranges of sizes,
- they bear design features similarities.
## Performances
The ARROW Reverse porous glenoid was tested according to the standards ASTM F1829-16, ASTM F2028-14 and ASTM F1378-12. After the tests were completed, it was determined that the ARROW Reverse porous glenoid performances were substantially equivalent to those of the selected predicate devices.
Risks to health have been addressed through the specified materials, processing controls, quality assurance and compliance to the Medical Device Good Manufacturing Practices Requlations.
The LAL testing meets the specified 20EU/device limit.
#### Substantial equivalence
The substantial equivalence of our products, when compared to the selected predicate devices, has been established following the commercial documents, 510(k) submission's information as well as conformance to standards in force.
The analysis of these technical data allows us to submit the ARROW Reverse porous glenoid, as being substantially equivalent to the already cleared predicate devices selected to draw a comparison.
#### Conclusion
Following the examination of all the above mentioned information, we believe that the ARROW Reverse porous glenoid is substantially equivalent to the selected predicate devices in terms of intended use, ranges of sizes, materials, performances, safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.