THE ARROW ANATOMICAL POROUS GLENOID, DEPENDING ON THE COMPONENTS USED, IS DESIGNED FOR: - CENTRED GLENOHUMERAL OSTEOARTHRITIS - - RHEUMATOID POLYARTHRITIS - - POST-TRAUMATIC SEQUELA WITH GLENOID INJURY - -FRACTURES OF THE PROXIMAL HUMERUS WITH GLENOID INJURY - REVISION FOR GLENOID LOOSENING - - GLENOID BONE LOSS. WHERE BONE GRAFT IS NEEDED - - A FUNCTIONAL ROTATOR CUFF IS NECESSARY TO USE THIS DEVICE THE POROUS GLENOID BASE IS INTENDED FOR CEMENTLESS APPLICATION WITH THE ADDITION OF BONE SCREWS FOR FIXATION
Device Story
Shoulder prosthesis system; includes glenoid insert and porous glenoid base; used in total anatomical shoulder replacement. Implanted via dedicated instrumentation; base requires cementless application with bone screws for fixation. Articulates with ARROW anatomical shoulder system. Used by orthopedic surgeons in clinical settings for primary or revision surgeries. Provides mechanical joint replacement to restore shoulder function in patients with specific degenerative or traumatic conditions; requires functional rotator cuff for efficacy.
Clinical Evidence
Bench testing only. Performance evaluated according to ASTM F 1829 and ASTM F 2028 standards. Testing confirmed substantial equivalence to predicate devices. Product non-pyrogenicity established.
Technological Characteristics
Materials: titanium alloy (porous glenoid base) and polyethylene (glenoid insert). Design: nonconstrained/semi-constrained porous-coated uncemented prosthesis. Fixation: cementless with bone screws. Standards: ASTM F 1829, ASTM F 2028.
Indications for Use
Indicated for patients with centered glenohumeral osteoarthritis, rheumatoid polyarthritis, post-traumatic sequela with glenoid injury, proximal humerus fractures with glenoid injury, glenoid loosening, or glenoid bone loss requiring bone graft. Requires functional rotator cuff.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2017
Fournitures Hospitalieres Industrie Patricia Donnard Regulatory Affairs Manager Z.I. de Kernevez - 6 Rue Nobel Ouimper, 29000 FR
Re: K162068 Trade/Device Name: Arrow Anatomical Porous Glenoid Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented Prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWT, KWS Dated: March 7, 2017 Received: March 9, 2017
Dear Ms Donnard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
K162068
Device Name
#### ARROW® ANATOMICAL POROUS GLENOID
Indications for Use (Describe)
THE ARROW ANATOMICAL POROUS GLENOID, DEPENDING ON THE COMPONENTS USED, IS DESIGNED FOR:
- CENTRED GLENOHUMERAL OSTEOARTHRITIS -
- RHEUMATOID POLYARTHRITIS -
- POST-TRAUMATIC SEQUELA WITH GLENOID INJURY -
- -FRACTURES OF THE PROXIMAL HUMERUS WITH GLENOID INJURY
- REVISION FOR GLENOID LOOSENING -
- GLENOID BONE LOSS. WHERE BONE GRAFT IS NEEDED -
- A FUNCTIONAL ROTATOR CUFF IS NECESSARY TO USE THIS DEVICE
THE POROUS GLENOID BASE IS INTENDED FOR CEMENTLESS APPLICATION WITH THE ADDITION OF BONE SCREWS FOR FIXATION
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary of safety and effectiveness
This summary of 510(k) safety and effectiveness information is submitted in accordance to the requirements of SDMA 1990 and 21 CFR 807.92.
Date prepared: July 20th 2016
The assigned 510(k) number is: K162068
#### 1. Applicant
Fournitures Hospitalières industrie 6 Rue Nobel, Z.I. de Kernévez 29000 QUIMPER - FRANCE Tel: (+33) 2.98.55.68.95 Fax: (+33) 2.98.53.42.13
#### 2. Company contact
Philippe VEISTROFFER, General Manager Tel: (+33) 2.98.55.68.95
#### 3. Product
- Trade name: ARROW® anatomical porous glenoid
- Common name: Shoulder prosthesis
Classification:
Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis. Product code: MBF Regulation: 21 CFR 888.3670 Class: II
Shoulder joint metal/polymer non-constrained cemented prosthesis Product code: KWT Regulation: 21 CER 888.3650 Class: II
Shoulder joint metal/polymer semi-constrained cemented prosthesis Product code: KWS Regulation: 21 CFR 888.3660 Class: II
{4}------------------------------------------------
# 4. Predicate/ Leqally marketed devices
The ARROW anatomical porous glenoid, is substantially equivalent to the following legally marketed devices:
| Manufacturer: | Fournitures Hospitalières Industrie |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | ARROW Anatomical Shoulder System |
| 510 (K): | K093599 |
| Manufacturer: | Fournitures Hospitalières Industrie |
| Device Trade Name: | ARROW Reverse Shoulder System |
| 510 (K): | K112193 |
| Manufacturer:<br>Device Trade Name:<br>510 (K): | Fournitures Hospitalières Industrie<br>ARROW Reverse Shoulder long keel and short keel glenoid base<br>K142778 |
| Manufacturer: | Biomet Orthopedics |
| Device Trade Name: | Bio-Modular Shoulder System |
| 510 (K): | K030710 |
## 5. Device description
The ARROW anatomical porous glenoid consists of a glenoid insert and a porous glenoid base. The porous glenoid base is used with bone screws for fixation (cleared in K112193). The glenoid inserts and the porous glenoid bases are used in total anatomical prosthesis and are designed to articulate with the ARROW anatomical shoulder system (cleared in K093599).
The ARROW anatomical porous glenoid is intended to be implanted using the dedicated instrumentation supplied by the manufacturer. This instrument set is common for all the configurations of prosthesis namely: simple humeral prosthesis, total anatomical prosthesis (cemented glenoid implant with 4 pegs), total anatomical prosthesis (porous glenoid implant), pending, and reverse prosthesis.
## 6. Indications for use / Intended use
- . Indications for use
## TOTAL ANATOMICAL PROSTHESIS (POROUS GLENOID IMPLANT)
- Centred glenohumeral osteoarthritis ।
- -Rheumatoid polvarthritis
- -Post-traumatic sequela with glenoid injury
- Fractures of the proximal humerus with glenoid injury -
- -Revision for glenoid loosening
- -Glenoid bone loss, where bone graft is needed
{5}------------------------------------------------
A functional rotator cuff is necessary to use this device
The porous glenoid base is intended for cementless application with the addition of bone screws for fixation.
The ARROW anatomical porous glenoid is a part of the full ARROW shoulder system, which includes other prosthesis configurations with other indications, as below:
#### SIMPLE HUMERAL PROSTHESIS
- Fracture dislocation or complex four part fracture of the proximal humerus -
- Humeral head necrosis without injury to the glenoid cavity. -
- Extensive humeral head cartilage damage without injury to the glenoid cavity -
- -Centred osteoarthritis with a glenoid cavity not allowing implantation of a glenoid implant.
- Rheumatoid polyarthritis with thin rotator cuff. -
- Off-centred osteoarthritis with irreparable cuff, and with maintained active elevation of at least 120°.
TOTAL ANATOMICAL PROSTHESIS (CEMENTED GLENOID IMPLANT WITH 4 PEGS)
- Centred glenohumeral osteoarthritis with functional rotator cuff -
- Rheumatoid polyarthritis with functional rotator cuff -
- Fracture sequela, functional rotator cuff with glenoid iniury. -
#### REVERSE PROSTHESIS
The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cemented or cementless application while the metal-back glenoid baseplate is intended for cementless application with the addition of bone screws for fixation.
- . Intended use
All the implants of shoulder prosthesis are used for primary or revision surgeries.
The glenoid base is intended for cementless application with the addition of cortical or cancellous bone screws. The porous glenoid base must be used only for a total anatomical prosthesis.
## 7. Comparison of technological characteristics
The ARROW anatomical porous glenoid and the above selected predicate devices have the same intended use and substantial similar indications for use and share the following similarities:
- they are made out of the same materials (titanium alloy for the porous glenoid base and polyethylene for the glenoid insert),
- they are available in similar ranges of sizes, -
- they bear design features similarities.
{6}------------------------------------------------
## 8. Performances
The ARROW anatomical porous glenoid was tested according to the ASTM F 1829 and ASTM F 2028 standards. After the tests were completed, it was determined that the ARROW anatomical porous glenoid performances were substantially equivalent to those of the selected predicate devices.
Risks to health have been addressed through the specified materials, processing controls, quality assurance and compliance to the Medical Device Good Manufacturing Practices Regulations. Testing has been performed to establish product non-pyrogenicity.
## 9. Substantial equivalence
The substantial equivalence of our products, when compared to the selected predicate devices, has been established following the commercial documents, 510(k) submission's information as well as conformance to standards in force.
The analysis of these technical data allows us to submit the ARROW anatomical porous glenoid, as being substantially equivalent to the already cleared predicate devices selected to draw a comparison.
## 10. Conclusion
Following the examination of all the above mention, we believe that the ARROW anatomical porous glenoid is substantially equivalent to the selected predicate devices in terms of intended use, ranges of sizes, materials, performances, safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.