The Trident Extremity Fixation System consisting of plates and screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bone of the feet, hand, ankle and toes. Specific examples include: - Arthrodesis of the first metatarsalcuneiform joint (Lapidus fusion procedure) - Arthrodesis of the first metatarsalphalangeal joint (MTP fusion procedure) - Primary and revision MTP fusion due to hallux rigidus and/or hallux valgus corrections - Revision of failed first MTP arthroplasty implant 2nd, 3rd and 4th metatarsal weil osteotomy correction indications - Metatarsal osteotomies to correct hallux valgus bunion indications - Fractures of the fifth metatarsal bone in the foot - Carpal bone fusions and fractures Flatfoot Osteotomies: - Lateral column lengthening (Evans osteotomy) - Plantar flexion opening wedge osteotomy of the medial cuneiform (Cotton osteotomy) - Medial displacement calcaneal osteotomy - Midfoot/Hindfoot Arthrodesis/Fusions - Lisfranc arthrodesis and/or stabilization - 1st, 2nd, 3rd and 4th Tarsometatarsal (TMT) fusions - Navicular-Cuneiform (NC) joint fusions - Talo-Navicular (TN) joint fusions - Calcaneo-Cuboid (CC) joint fusions - Medial column fusions (NC and 1st TMT) The Trident Extremity Fixation System screws and plates are indicated for use with the Trident Plates and Screws of the same material. The locking and non-locking solid cortical screws, cannulated screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture fixation, appropriate for the size of the device.
Device Story
Trident Extremity Fixation System comprises titanium plates and screws (locking, non-locking, solid cortical, cannulated, headless) for orthopedic bone fixation; used by surgeons in clinical settings for fracture stabilization, osteotomy, and arthrodesis; system includes reusable instrumentation for implant delivery; provides mechanical stability to bone segments to facilitate healing and reconstruction; device size range (3.0-7.3mm diameter; 10-100mm length) allows selection based on anatomical site and patient needs.
Clinical Evidence
Bench testing only. Testing included static 4-point bending (ASTM F382-14) for plates; axial pullout, axial pushout, torque to failure (ASTM F543), and static/dynamic 3-point bending (ASTM F1264) for screws.
Indicated for stabilization of fresh fractures, revision procedures, joint fusion, and bone reconstruction in feet, hands, ankles, and toes. Includes specific procedures: Lapidus fusion, MTP fusion, hallux valgus/rigidus correction, Weil osteotomy, 5th metatarsal fractures, carpal fusions, flatfoot osteotomies (Evans, Cotton, calcaneal), and midfoot/hindfoot arthrodesis (Lisfranc, TMT, NC, TN, CC joints).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 8, 2018
Trident Orthopedics % Al Memmolo Official Correspondent 6648 Surf Crest Street Carlsbad, California 92011
### Re: K171690
Trade/Device Name: Trident Extremity Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 28, 2018 Received: March 30, 2018
Dear Al Memmolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known) K171690
Device Name
The Trident Extremity Fixation System
#### Indications for Use (Describe)
The Trident Extremity Fixation System consisting of plates and screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bone of the feet, hand, ankle and toes. Specific examples include:
- Arthrodesis of the first metatarsalcuneiform joint (Lapidus fusion procedure)
- Arthrodesis of the first metatarsalphalangeal joint (MTP fusion procedure)
- Primary and revision MTP fusion due to hallux rigidus and/or hallux valgus corrections
- Revision of failed first MTP arthroplasty implant 2nd, 3rd and 4th metatarsal weil osteotomy correction indications
- Metatarsal osteotomies to correct hallux valgus bunion indications
- Fractures of the fifth metatarsal bone in the foot
- Carpal bone fusions and fractures
Flatfoot Osteotomies:
- Lateral column lengthening (Evans osteotomy)
- Plantar flexion opening wedge osteotomy of the medial cuneiform (Cotton osteotomy)
- Medial displacement calcaneal osteotomy
- Midfoot/Hindfoot Arthrodesis/Fusions
- Lisfranc arthrodesis and/or stabilization
- 1st, 2nd, 3rd and 4th Tarsometatarsal (TMT) fusions
- Navicular-Cuneiform (NC) joint fusions
- Talo-Navicular (TN) joint fusions
- Calcaneo-Cuboid (CC) joint fusions
- Medial column fusions (NC and 1st TMT)
The Trident Extremity Fixation System screws and plates are indicated for use with the Trident Plates and Screws of the same material. The locking and non-locking solid cortical screws, cannulated screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary (as required by 21 CFR 807.92)
| Date Prepared | May 8, 2018 |
|----------------|------------------------------------------------|
| Manufacturer | Trident Orthopedics |
| Address | 212 Copperwood Court<br>Millersville, MD 21108 |
| Contact Person | Al Austin |
| Address | 212 Copperwood Court<br>Millersville, MD 21108 |
| Telephone | 410-903-9038 |
| Fax | 410-544-3547 |
| email | telosusa@aol.com |
| Trade Name | Trident Extremity Fixation System |
|---------------------|--------------------------------------------------------------------------------|
| Common Name | Bone Screw |
| Panel Code | Orthopaedics/87 |
| Classification Name | Single/multiple component metallic bone fixation appliances and<br>accessories |
| Class | Class II |
| Regulation Number | 21 CFR 888.3030 |
| Product Code | HRS: Plate, Fixation, Bone<br>HWC: Screw, Fixation, Bone |
| Predicate Devices | 510(k) # | Manufacturer |
|------------------------------------------------------------------------------------|---------------------|-----------------|
| Primary Predicate:<br>Medline Foot Plates and Screws | K151235 | Medline |
| Secondary Predicate:<br>Vilex Cannulated Bone Screw/DuVal<br>Cannulated Bone Screw | K991151,<br>K991197 | Vilex, Inc. |
| Secondary Predicate:<br>Osteomed Cannulated Screw System | K151021 | Osteomed, Corp. |
| Description | The Subject device is manufactured from titanium conforming to ASTM F-136. The system includes plates in various sizes as well as locking and non-locking screws to be used with the plates. The bones screws are available in diameter sizes from 3.0 mm to 7.3 mm diameter and lengths from 10 mm to 100 mm. The system also includes reusable instrumentation for delivery of the implantable components. The Subject device has a similar range of sizes as its predicate. |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications and<br>Intended Use | The Trident Extremity Fixation System consisting of plates and screws are<br>intended for use in stabilization and fixation of fresh fractures, revision<br>procedures, joint fusion and reconstruction of bone of the feet, hand, ankle<br>and toes. Specific examples include: |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | - Arthrodesis of the first metatarsalcuneiform joint (Lapidus fusion<br>procedure)<br>- Arthrodesis of the first metatarsalphalangeal joint (MTP fusion procedure) |
| | - Primary and revision MTP fusion due to hallux rigidus and/or hallux<br>valgus corrections |
{5}------------------------------------------------
## K171690
| - Revision of failed first MTP arthroplasty implant |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2nd, 3rd and 4th metatarsal weil osteotomy correction indications |
| Metatarsal osteotomies to correct hallux valgus bunion indications |
| Fractures of the fifth metatarsal bone in the foot |
| Carpal bone fusions and fractures |
| Flatfoot Osteotomies: |
| - Lateral column lengthening (Evans osteotomy) |
| - Plantar flexion opening wedge osteotomy of the medial cuneiform (Cotton<br>osteotomy) |
| - Medial displacement calcaneal osteotomy |
| Midfoot/Hindfoot Arthrodesis/Fusions |
| - Lisfranc arthrodesis and/or stabilization |
| - 1st, 2nd, 3rd and 4th Tarsometatarsal (TMT) fusions |
| - Navicular-Cuneiform (NC) joint fusions |
| - Talo-Navicular (TN) joint fusions |
| - Calcaneo-Cuboid (CC) joint fusions |
| - Medial column fusions (NC and 1st TMT) |
| The Trident Extremity Fixation System screws and plates are indicated for<br>use with the Trident Plates and Screws of the same material. The locking<br>and non-locking solid cortical screws, cannulated screws and headless<br>screws are also indicated for bone reconstruction, osteotomy, arthrodesis,<br>joint fusion, fracture repair and fracture fixation, appropriate for the size of<br>the device. |
| Technological<br>Characteristics<br>and Substantial<br>Equivalence | Documentation and performance testing was provided to demonstrate that<br>the Subject device is substantially equivalent to the Predicate Medline<br>Foot Plates and Screws (K151235) and the Predicate Vilex Cannulated<br>Bone Screw/DuVal Cannulated Bone Screw (K991151, K991197). The<br>Subject device is substantially equivalent to the predicate devices in<br>intended use, indications for use, materials, technological characteristics,<br>and labeling. |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data | Bench testing confirmed that the Subject device performed as intended<br>and is at least equivalent to the predicate devices. Static 4-Point bending<br>(ASTM F382-14) was conducted on the plates. Axial Pullout, Axial<br>Pushout, Torque to Failure (ASTM F543) Static Three-Point Bend, and<br>Dynamic Three-Point Bend (ASTM F1264) were performed on the screws. |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the intended use indication for cemented or hybrid<br>fixation, the subject device is substantially equivalent to the predicate<br>devices. |
| Conclusion | Based on the intended use, indications for use, technological<br>characteristics, materials, and comparison to predicate devices, the<br>Subject device has been shown to be substantially equivalent to legally<br>marketed predicate devices. |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.