K151235 · Medline Industries, Inc. · HRS · Aug 6, 2015 · Orthopedic
Device Facts
Record ID
K151235
Device Name
Medline Foot Plates and Screws
Applicant
Medline Industries, Inc.
Product Code
HRS · Orthopedic
Decision Date
Aug 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medline Foot Plates and Screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bones of the feet and toes. Specific examples include: - Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion) - Arthrodesis of the first metatarsophalangeal joint (MTP) including: - Primary MTP Fusion due to hallux ridgidus and/or hallux valgus - Revision MTP Fusion - Revision of failed first MTP Arthroplasty implant Flatfoot Osteotomies: - Lateral Column Lengthening (Evans Osteotomy) - Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy) - Medial Displacement Calcaneal Osteotomy (MDCO) Midfoot / Hindfoot Fusions: - LisFranc Arthrodesis and/or Stabilization - 1st(Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions - Intercuneiform Fusions - Navicular-Cuneiform (NC) Fusion - Talo-Navicular (TN) Fusion - Calcaneo-Cuboid (CC) Fusion - Medial Column Fusions (NC and 1st TMT) The Medline Locking and Non-Locking Cortical Screws are indicated for use with the Medline Foot Plates of the same base material. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Device Story
System consists of titanium alloy plates and 2.7mm/3.5mm locking/non-locking cortical screws; used for bone fixation, fusion, and reconstruction in feet and toes. Plates feature polyaxial locking holes and compression slots. Reusable instrumentation (drill guides, tissue protectors, drill bits) included. Operated by surgeons in clinical/OR settings. Device provides mechanical stabilization of bone segments to facilitate healing/fusion. Benefits include structural support for osteotomies and fracture repair.
Clinical Evidence
Bench testing only. Performance evaluated via static 4-point bending (ASTM F382-99), static axial pullout (ASTM F543-13), and static torque testing (ASTM F543-13). Results confirmed mechanical performance equivalent to predicate devices.
Indicated for patients requiring stabilization of fresh fractures, revision procedures, joint fusion, or bone reconstruction in feet and toes, including hallux rigidus, hallux valgus, flatfoot, and midfoot/hindfoot fusions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other. The faces are rendered in a simple, abstract style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2015
Medline Industries, Incorporated Ms. Jennifer Mason Senior Regulatory Affairs Specialist 1 Medline Place Mundelein, Illinois 60060
Re: K151235
Trade/Device Name: Medline Foot Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 8, 2015 Received: May 11, 2015
Dear Ms. Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151235
Device Name
Medline Foot Plates and Screws
Indications for Use (Describe)
The Medline Foot Plates and Screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bones of the feet and toes. Specific examples include:
- · Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion)
- · Arthrodesis of the first metatarsophalangeal joint (MTP) including:
- Primary MTP Fusion due to hallux ridgidus and/or hallux valgus
- Revision MTP Fusion
- · Revision of failed first MTP Arthroplasty implant
Flatfoot Osteotomies:
- · Lateral Column Lengthening (Evans Osteotomy)
- · Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy)
- · Medial Displacement Calcaneal Osteotomy (MDCO)
Midfoot / Hindfoot Fusions:
- · LisFranc Arthrodesis and/or Stabilization
- · 1st(Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions
- · Intercuneiform Fusions
- · Navicular-Cuneiform (NC) Fusion
- · Talo-Navicular (TN) Fusion
- · Calcaneo-Cuboid (CC) Fusion
- · Medial Column Fusions (NC and 1st TMT)
The Medline Locking and Non-Locking Cortical Screws are indicated for use with the Medline Foot Plates of the same base material. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
K151235 Page 1 of 3
Image /page/3/Picture/1 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, stacked vertically, with a white starburst-like symbol to the right of the text. The background is a solid dark blue color. The logo is simple and clean, with a focus on the company name and a memorable symbol.
ledline Industries, Inc. One Medline Place Mundelein, IL 60060
# SECTION 5 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
## Submitter / 510(k) Sponsor
Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060
Registration Number: 1417592
## Contact Person
Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason@medline.com
Summary Preparation Date July 31, 2015
Type of 510(k) Submission Traditional
## Device Name / Classification
Name of Device: Medline Foot Plates and Screws Proprietary Name: Medline Foot Plates and Screws Common Name: Plate, Fixation, Bone Classification Number: 21 CFR 888.3030 - Class II Classification Name: Single/multiple component metallic bone fixation appliances and accessories Product Code: HRS: Plate, Fixation, Bone HWC: Screw, Fixation Bone Classification Panel: Orthopedics
## Predicate Device
WRIGHT MEDICAL ORTHOLOC™ 3Di System K121651 (Primary) – ORTHOLOC™ 3Di Midfoot/Flatfoot System (Wright Midfoot/Flatfoot System)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a dark blue square. To the right of the word is a white star-like symbol with four points. The logo is positioned on the left side of the image.
ledline Industries. Inc One Medline Place Mundelein, IL 60060
## Device Description
The Medline Foot Plates and Screws are manufactured from Titanium Alloy. The System includes plates offered in various styles, sizes and options; each contoured for specific anatomy and designed for specific procedures, and 2.7mm and 3.5mm diameter locking and non-locking cortical screws to be used with the polyaxial locking holes and compression slots included in the plates. The non-locking cortical screws can also be used for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture fixation, appropriate for the size of the device. The system also includes reusable instrumentation necessary to implant the plates and screws, e.g. drill guides, tissue protectors, and drill bits. The Medline Foot Plates and Screws are within the currently marketed sizes and indications of the identified predicate devices.
# Indications for Use
The Medline Foot Plates and Screws are intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bones of the feet and toes. Specific examples include:
- · Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion)
- · Arthrodesis of the first metatarsophalangeal joint (MTP) including:
- Primary MTP Fusion due to hallux ridgidus and/or hallux valgus
- Revision MTP Fusion
- · Revision of failed first MTP Arthroplasty implant
Flatfoot Osteotomies:
- · Lateral Column Lengthening (Evans Osteotomy)
- · Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy)
- · Medial Displacement Calcaneal Osteotomy (MDCO)
Midfoot / Hindfoot Fusions:
- LisFranc Arthrodesis and/or Stabilization
- · 1st(Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions
- · Intercuneiform Fusions
- Navicular-Cuneiform (NC) Fusion
- Talo-Navicular (TN) Fusion
- Calcaneo-Cuboid (CC) Fusion
- · Medial Column Fusions (NC and 1st TMT)
The Medline Locking and Non-Locking Cortical Screws are indicated for use with the Medline Foot Plates of the same base material. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the device,
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is the logo for Medline. The logo is a blue square with the word "MEDLINE" in white, sans-serif font. A white, stylized starburst is behind the word "MEDLINE". To the right of the blue square is a gray rectangle.
edline Industries. Inc ne Medline Place Mundelein II 60060
## Summary of Technological Characteristics
The proposed device is substantially equivalent to the predicate devices. The Wright Medical ORTHOLOC™ 3Di System K121651, ORTHOLOC 3Di Hallux System K120359, ORTHOLOC™ 3Di Ankle Plating System and the Ortholoc Bone Screws K120802 were selected as the predicates based on the same intended use and same materials.
# Summary of Non-Clinical Testing
| Test Type | Testing Standards | Performance Results |
|--------------------------------------------|-------------------|-------------------------|
| Static 4-Point Bending Testing<br>(Plates) | ASTM F382 | Equivalent to Predicate |
| Static Pullout Testing (Screws) | ASTM F543 | Equivalent to Predicate |
| Static Torque Testing (Screws) | ASTM F543 | Equivalent to Predicate |
## Summary of Performance Testing
## Static 4-Point Bend Testing
Static 4-point bend testing was conducted per ASTM F382-99 (2008). The purpose of this test was to ensure that the average load to failure of the proposed Medline Foot Plates was equivalent to the average load to failure bending stiffness of the predicate Wright Medical ORTHOLOC™ Plates.
## Static Axial Pullout Testing
Static axial pullout testing was conducted per ASTM F543-13. The purpose of this test was to ensure that the axial pullout strength of the proposed Medline Locking Screws and Non-Locking Cortical Screws were equivalent to the pullout strength of the predicate Wright Medical ORTHOLOC™ Locking and Bone Screws.
## Static Torque Testing
Static torque testing was conducted per ASTM F543-13. The purpose of this test was to ensure that the torsional strength of the proposed Medline Locking Screws and Non-Locking Cortical Screws were equivalent to the torsional strength of the predicate Wright Medical ORTHOLOC™ Locking and Bone Screws.
# Conclusion
In accordance with 21 CFR Part 807. and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Foot Plates and Screws are safe, effective and substantially equivalent to the predicate, Wright Medical ORTHOLOC 3Di System (K 121651) as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.