The OsteoMed Cannulated Screw System is indicated for bone fixation of hand and foot following trauma or osteotomy. Screws and washers are intended for single use only.
Device Story
OsteoMed Cannulated Screw System consists of metallic bone screws and washers used for fixation in hand and foot surgery following trauma or osteotomy. System includes various diameters (2.0mm to 8.0mm) and lengths; implants manufactured from Titanium alloy (ASTM F-136). Instruments (depth gauges, drivers, drills, countersinks, k-wires) facilitate screw placement. Device used by surgeons in clinical/OR settings. Output is mechanical stabilization of bone fragments. Benefits include restoration of skeletal alignment and support for bone healing. Subject device modifies predicate by redesigning 4.0mm screw for increased lag strength and adding 3.0mm diameter options.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing/verification comparison. Mechanical strength testing performed on 3.0mm and 4.0mm screws to ensure design features meet required criteria.
Technological Characteristics
Implants: Titanium alloy (ASTM F-136). Instruments: Stainless steel, anodized aluminum, medical grade polymer. Cannulated design for bone fixation. Sizes range 2.0mm-8.0mm diameter. Sterilization method not specified. No software or electronic components.
Indications for Use
Indicated for bone fixation of hand and foot following trauma or osteotomy in patients requiring surgical stabilization.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
OsteoMed Extended 2.0/2.4 Cannulated Screw System (K062863)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like figure with three heads, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2015
Osteomed Corp. Mrs. Piedad Peña Regulatory Affairs Specialist 3885 Arapaho Road Addison, Texas 75001
Re: K151021
Trade/Device Name: Osteomed Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: June 16, 2015 Received: June 17, 2015
Dear Mrs. Peña:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K151021 (page 1/1)
Device Name: OsteoMed Cannulated Screw System
Indications for Use:
The OsteoMed Cannulated Screw System is indicated for bone fixation of hand and foot following trauma or osteotomy. Screws and washers are intended for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in large, bold letters. The word "Osteo" is in blue, and the word "Med" is in gray. The logo is surrounded by a blue and gray oval shape. The logo is simple and modern, and it is likely used to represent a medical company.
### Special 510(k) Summary
| Name of Submitter: | OsteoMed<br>3885 Arapaho Road<br>Addison, Texas 75001<br>Phone: (972) 677-4600<br>Fax: (972) 677-4601 |
|--------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Piedad Peña |
| Date Prepared: | July 8, 2015 |
| Device Proprietary Name: | OsteoMed Cannulated Screw System |
| Device Common Name: | Cannulated bone screws, washers and associated instruments |
| Classification Name: | 21 CFR 888.3040: Smooth or threaded metallic bone fixation<br>fastener |
| Product Code: | HWC |
Predicate Device: OsteoMed Extended 2.0/2.4 Cannulated Screw System, K062863 Classification Name: Smooth or threaded metallic fixation fastener (21CFR 888.3040, Product Code HWC) Device Class: II Reference Device: OsteoMed Super Screw System, K954330 Classification Name: Smooth or threaded metallic fixation fastener (21CFR 888.3040, Product Code HWC) Device Class: II
Summary:
#### Device Description:
The OSTEOMED Cannulated Screw System is comprised of screws and washers used for bone fixation of the hand and foot following trauma or osteotomy. The System features cannulated screws in the following dimensions:
2.0mm screw diameter - 6mm to 42mm screw length;
2.4mm screw diameter - 6mm to 52mm screw length;
3.0mm screw diameter - 10mm to 40mm screw length;
3.5mm screw diameter - 12mm to 52mm screw length;
4.0mm screw diameter - 12mm to 52mm screw length;
5.5mm screw diameter - 12mm to 52mm screw length;
6.5mm screw diameter - 20mm to 120mm screw length;
8.0mm screw diameter - 30mm to 140mm screw length;
Image /page/3/Picture/16 description: The image contains the Colson logo. The logo consists of the word "Colson" in a bold, sans-serif font, with a vertical line extending above the "l". To the right of the logo is the text "A COLSON A", also in a sans-serif font, but smaller and less bold than the main logo. The background of the image is plain white.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OsteoMed. The logo is in an oval shape with a blue line at the top and a gray line at the bottom. The word "OsteoMed" is written in blue and gray inside the oval.
The implants (screws and washers) of the OsteoMed Cannulated Screw System are made from Titanium alloy (ASTM F-136). Modifications to screws and washer of the subject system include redesign of the 4.0mm cannulated screw to increase the strength of the lag portion and addition of a 3.0mm diameter cannulated screw and washer.
The system instruments include depth gauges, screw drivers, drills, countersinks, k-wires, and preparation instruments to facilitate the placement of screws. The instrumentation is made from various grades of stainless steel, anodized aluminum, and/medical grade polymer.
#### Indications for Use:
The OsteoMed Cannulated Screw System is indicated for bone fixation of hand and foot following trauma or osteotomy. Screws and washers are intended for single use only.
#### Technological Characteristics:
The OsteoMed Cannulated Screw System is available in a variety of sizes and is recommended for bone fixation in the hand and foot.
Cannulated Screw System implants (screws and washers) are manufactured from Titanium alloy (ASTM F-136), the same materials used in the manufacture of the predicate devices. This material is biocompatible.
#### Performance/Clinical Data:
The OsteoMed Cannulated Screw System was compared to the reference device, OsteoMed Super Screw System (K954330), and predicate device, OsteoMed Extended 2.0/2.4 Cannulated Screw System (K062863). The 3.0mm and 4.0mm cannulated screws underwent verification evaluation to ensure that the design features met the required mechanical strength criteria for their intended use. The intended use of the OsteoMed Cannulated Screw System is the same as OsteoMed Extended 2.0/2.4 Cannulated Screw System (K062863).
Performance equivalence was shown through the verification comparison to the predicate device and reference device.
Clinical testing is not required to support substantial equivalence.
#### Substantial Equivalence:
The basis of substantial equivalence for this device, OsteoMed Cannulated Screw System, is based on similarities in intended use, material, function, performance, design, technology, sterilization, and operational principles to the predicate device, OsteoMed Extended 2.0/2.4 Cannulated Screw System (K062863). A performance comparison was performed and verified that the cannualated screws met required mechanical strength criteria for their intended use compared to the legally marketed predicate device and reference device listed in this summary. OsteoMed believes that the non-clinical tests demonstrate that the device is as safe, and effective as the predicate device.
OsteoMed 3885 Arapaho Road Addison. Texas 75001 (973) 677-4600 FAX: (800) 390-2620 Customer Service: (800) 456-7779
Image /page/4/Picture/15 description: The image shows the logo for A Colson Associate. The logo consists of the word "Colson" inside of a circle. To the right of the circle is the text "A COLSON ASSOCIATE".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.