K171580 · Ge Medical Systems (China) Co., Ltd. · MHX · Nov 1, 2017 · Cardiovascular
Device Facts
Record ID
K171580
Device Name
Monitor B125, Monitor B105
Applicant
Ge Medical Systems (China) Co., Ltd.
Product Code
MHX · Cardiovascular
Decision Date
Nov 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B125/B105 is not intended for use during MRI. The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network. The Monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/ Myocardial/Core/Surface temperature, impedance respiration, respiration rate, CO2. The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia. VT>2. Ventricular Bradycardia. Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.
Device Story
Portable multi-parameter patient monitor; inputs include ECG, invasive/non-invasive blood pressure, SpO2, temperature, respiration, and CO2. Uses hemodynamic and E-MiniC modules for signal acquisition. Features 10/12-inch LCD touch screen; supports WLAN and network interfacing. Operated by clinicians in hospital environments. Provides real-time physiological data, ST-segment analysis, and arrhythmia detection. Output displayed on-screen; used by healthcare providers to assess patient status and trigger clinical interventions. Benefits include continuous monitoring of vital signs and automated detection of cardiac events to improve patient safety.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm system testing (IEC 60601-1-8), and performance testing for specific parameters (ECG, NIBP, IBP, gas, temperature, SpO2). Software verification and validation were conducted.
Technological Characteristics
Multi-parameter monitor with 10/12-inch LCD touch screen. Parameters: ECG, RESP, NIBP, IBP, TEMP, SpO2, CO2. Connectivity: WLAN, DVI-D, USB. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61. Software: Major level of concern.
Indications for Use
Indicated for monitoring, recording, and alarming of physiological parameters in adult, pediatric, and neonatal patients in hospital settings and during intra-hospital transport. Not for use in MRI environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1. 2017
GE Medical Systems (China) Co., Ltd. % Mr. Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223
Re: K171580
Trade/Device Name: Monitor B125/B105 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, DSK, GWQ, FLL, NHO, NHP, NHQ Dated: May 30, 2017 Received: May 31, 2017
Dear Mr. Robert Casarsa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Robert Casarsa
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Wilhelmmen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171580
Device Name Monitor B125/B105
### Indications for Use (Describe)
The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B125/B105 is not intended for use during MRI.
The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.
The Monitor B125/B105 monitors and displays : ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Tympani/Rectal/ Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration rate, CO2.
The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Couplet, Bigeminy, Trigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary
K171580
#### I. SUBMITTER
GE MEDICAL SYSTEMS (CHINA) CO., LTD. No. 19 Changjiang road National Hi-Tech Dev. Zone Wuxi, Jiangsu, China 214028 Sun YanLi Regulatory Affairs Leader
Primary Contact Person: Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. Telephone: 414-362-3063 Fax at 414-362-2585 E-mail: Robert.casarsa@ge.com
Secondary Contact Person: Douglas Kentz Regulatory Affairs Director GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 Phone: 414 362-2038 Fax: 414-262-2585 E-mail: Douglas.kentz@ge.com
#### II. DEVICE
Name of Device: Monitor B125/B105 Common or Usual Name: Multi-parameter Patient Monitor Classification Name: 21 CFR 870.1025 monitor, physiological, patient (with arrhythmia detection or alarms) 21 CFR 868.2375 monitor, breathing frequency 21 CFR 868.1700 analyzers, gas, nitrous-oxide, gaseous phase (anesthetic conc.) 21 CFR 868.1620 analyzers, gas, halothane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzers, gas, enflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1400 analyzers, gas, carbon-dioxide, gaseous-phase 21 CFR 868.1720 analyzers, gas, oxygen, gaseous-phase 21 CFR 870.1130 system, measurement, blood-pressure, non-invasive 21 CFR 870.2700 oximeters 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm)
{4}------------------------------------------------
- 21 CFR 870.2770 plethysmographs, impedance 21 CFR 870.1110 computers, blood-pressure 21 CFR 882.1400 full-montage standard electroencephalograph 21 CFR 880.2910 thermometers, electronic, clinical 21 CFR 868.1500 analyzers, gas, desflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, sevoflurane, gaseous-phase (anesthetic concentration) 21 CFR 868.1500 analyzer, gas, isoflurane, gaseous-phase (anesthetic concentration)
Regulatory Class: II Product Code: MHX, BZQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, DSK, GWQ, FLL, NHO, NHP, NHQ
#### III. PREDICATE DEVICE
K151063 Monitor B40 and K102239 CARESCAPE Monitor B650 These predicates have not been subject to a design-related recall. No reference devices were used in this submission.
#### DEVICE DESCRIPTION IV.
The Monitor B125/B105 is a multi-parameter patient monitor that is developed based on the predicate Monitor B40 (K151063) platform. The Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650(K102239)), WLAN (FCC ID: OU5B1X501) and touch screen.
As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 is a multiparameter patient monitor, utilizes 12inches /10inches LCD display with an integrated keypad and a pre-configuration (hemodynamic module) which provide basic parameters: ECG, RESP, NIBP, IBP, TEMP, SPO2.
As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has optional CO2 parameter provided by the identical E-MiniC module (K052582).
As with the predicate Monitor B40(K151063), the proposed Monitor B125/B105 has a mounting plate on the bottom of the monitor. The monitor can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.
{5}------------------------------------------------
#### V. INDICATONS FOR USE
The Monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner.
The Monitor B125/B105 is not intended for use during MRI.
The Monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.
The Monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/ Myocardial/Core/Surface temperature, impedance respiration, respiration rate, CO2.
The Monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia. VT>2. Ventricular Bradycardia. Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC) and Ventricular fibrillation.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The proposed Monitor B125/B105 is a multi-parameter patient monitor that is developed based on the predicate Monitor B40 (K151063) platform. The proposed Monitor B125/B105 provides additional support for ECG full arrhythmia (has been claimed in CARESCAPE Monitor B650 (K102239)), WLAN (FCC ID: OU5B1X501) and touch screen
The proposed Monitor B125/B105 uses the identical hemodynamic module (Hemo module) as the predicate Monitor B40 (K151063) in the monitoring of ECG, Resp, NIBP, IBP, SpO2 and Temp parameters data.
The proposed Monitor B125/B105 uses identical E-MiniC module (K052582) as the predicate Monitor B40 (K151063).
The fundamental technology of the proposed Monitor B125/B105 is the same as the predicate devices.
{6}------------------------------------------------
The proposed Monitors B125/B105 is as safe and effective as the predicate devices.
The following table includes comparisons of the main features of the device, and includes the features that are different from the predicate.
| WLAN | Support wireless communication.<br>WLAN module is FCC qualified, FCC ID: OU5B1X501) |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Touch screen | Support touch screen per customer needs |
| ECG arrhythmia<br>analysis | Enabled ECG full arrhythmia analysis<br>The proposed device uses the same ECG algorithm (EK-Pro<br>V12) as predicate device B40 (K151063)<br>The proposed device enabled full arrhythmia analysis which<br>disabled in the predicate device B40 (K151063).<br>Full arrhythmia analyses implemented with EK -ProV12 ECG<br>algorithm was cleared in B650 (K102239) |
| CARESCAPE<br>Central Station<br>V2 (K162012)<br>compatibility | Added the compatibility with CARESCAPE Central Station<br>V2 (K162012)<br>CARESCAPE Central Station V2 (K162012) has the same<br>communication protocol as CARESCAPE central station V1<br>(K133882) which has been cleared in predicate device B40<br>(K151063) |
| Improved<br>operating<br>workflow | Add standby and night mode Up to 168 hours trending storage and 200 snapshot checks Add event view software feature to combine alarm history and snapshot |
| Peripheral I/O<br>interface | Add DVI-D connector to support slave display. Add USB connector to support USB disk for service purpose |
| Masimo SpO2<br>accessories<br>change | Add Masimo RD sensors and related cables, the sensors are<br>compatible with MS-2011SB board (K053269)<br>Two reusable sensors (K051212)<br>MASIMO-RD-4050 |
| Four adhesive sensors (K042346): | |
| MASIMO-RD-4000 | |
| MASIMO-RD-4001 | |
| MASIMO-RD-4003 | |
| MASIMO-RD-4004 | |
| Two RD cables (K042536) | |
| MASIMO-RD-4085 | |
| MASIMO-RD-4084 | |
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
No biocompatibility testing was performed as the Monitor B125/B105 is not in contact with the patient. All supplies and accessories for use with the proposed Monitor B125/B105 that contact the patient have been previously cleared by FDA.
# Electrical safety and electromagnetic compatibility (EMC)
The Monitor B125/B105 is designed and tested for compliance with the following performance standards:
- . IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-8:2006 + A1:2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- . IEC 60601-2-27: 2011 + C1:2012 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
- IEC 80601-2-30:2009 + A1:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers
- IEC 60601-2-34:2011 Medical electrical equipment - Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
- . IEC 60601-2-49:2011 Medical electrical equipment – Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
{8}------------------------------------------------
- ISO 80601-2-55:2011 Medical electrical equipment -- Part 2-55: Particular ● requirements for the basic safety and essential performance of respiratory gas monitors
- ISO 80601-2-56:2009 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- . ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- IEC 62366-1:2015 Medical devices-Part 1: Application of usability engineering to ● medical devices
- IEC 60601-1-6:2010+A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability
# Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Major" level of concern, where a failure could result in death or serious injury, either to a patient or to a user of the device. The following bench tests were performed:
Chapter 2: **Vectors and Scalars**
# Software Verification include below Testing:
- Unit Testing: Code reviews and/or static code analysis are the methods are adopted . for the software unit verification to ensure the unit acceptance criteria, as defined in the SDLC procedure, are met.
- Software Integration Test: Performed to confirm that integrated (changed) SW units conform to the architecture and requirements, and other portions of the system have not been adversely affected.
- . Software Regression Test: The purpose of this test is to verify the side effect of code changes. Software leader and engineering team shall review the side effect of all SPRs and it shall be mentioned in the SPR test description or individual test procedure.
- System Testing: Performance testing included the system level testing and subsystem level testing. System level procedures were executed to confirmed that all product system requirements specified in the product specifications were met. Design Specifications specify design requirements for each subsystem of Patient Monitor B125/B105. Each Design Specification has one or more corresponding Verification Procedures that were used to verify that the design requirements have been met.
Software Validation is integrated with System Validation: System Validation will confirm that the product fulfills the user needs and intended uses under actual or simulated use conditions.
{9}------------------------------------------------
K171580 Page 7 Of 7
## Environmental Testing:
Perform Rapid Change of Temperature, Change of Temperature, Dry heat for non-heatdissipating specimen with gradual change of temperature, Damp heat, Cold for non-heat dissipating specimen with gradual change of temperature, Cold for heat-dissipating specimen with gradual change of temperature and steady state testing to ensure the unit preference meet the environmental requirements.
### Mechanical stress Testing:
Perform sinusoidal vibration, shock, free fall, altitude test and broadband random vibration testing to ensure the unit meet the Mechanical stress requirements.
### Package Testing
Perform package transport testing to ensure the packaged product withstand the distribution environment and to show the serviceability of the transport packages.
### Animal Study
i
No animal studies have been performed on the Monitor B125/B105.
### Clinical Studies
No clinical studies have been performed on the Monitor B125/B105.
#### VIII. CONCLUSIONS
GE Healthcare considers the proposed Monitor B125/B105 to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.