The CARESCAPE Central Station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data within a hospital or clinical environment. The CARESCAPE Central Station is intended to collect, display and print information from a network, including patient demographics, physiological parameters and waveforms, alarm annunciation and/or other non-medical information from monitors and telemetry systems. Physiological parameters and waveforms include electrocardiograph (ECG), pulse oximetry (SPO2), invasive blood pressures (IBP), non-invasive blood pressure (NIBP), respiration (RR), ventilator (VNT), carbon dioxide (CO2), oxygen (O2), mass spectrometry (Gas), temperature (Temp) and bispectral index (BIS), Beat to beat patient information for parameters and waveforms from the bedside and telemetry systems can be displayed. Patient monitor and telemetry system settings can be adjusted. Parameter values derived from patient data can be calculated, displayed, and printed. The CARESCAPE Central Station supports the ability to access information from GE products and hospital intranet in a web browser format. Additionally, CARESCAPE Central Station supports the ability to access patient information collected from the CARESCAPE network and stored on a network server.
Device Story
CARESCAPE Central Station v2 (CSCS) acts as a centralized viewing, reviewing, and monitoring station for patient data acquired from bedside monitors and telemetry transmitters via a network. Input includes patient demographics, physiological parameters (ECG, SPO2, IBP, NIBP, RR, VNT, CO2, O2, Gas, Temp, BIS), and waveforms. The device processes and displays this data, provides alarm annunciation, and allows remote configuration of monitor settings (e.g., alarm limits, ECG leads). Used in clinical settings by healthcare practitioners; output is viewed on a display or printed via laser printer/digital writer. Benefits include centralized patient oversight, historical trend review (up to 144 hours of full disclosure), and efficient alarm management. The system supports web browser access to hospital intranet and GE product information.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including risk analysis, design reviews, unit-level module verification, system integration testing, performance verification, safety testing, and simulated use validation.
Technological Characteristics
PC-based platform; Windows 7 embedded OS. Hardware includes upgraded processor, increased RAM, and SSD flash media. Connectivity via network (1000Mbps support). Supports widescreen displays. Alarm settings comply with IEC 60601-2-27:2011. Printing via PCL 6 compatible printers. No specific algorithm architecture (e.g., ML/DL) disclosed; device performs data collection, display, and calculation of physiological parameters.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in hospital or clinical environments requiring centralized monitoring of physiological data, waveforms, and alarm annunciation under the supervision of a licensed healthcare practitioner.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2016
GE Healthcare Mr. Robert Casarsa Regulatory Affairs Leader 8200 West Tower Ave. Milwaukee, Wisconsin 53223
Re: K162012
> Trade/Device Name: CARESCAPE Central Station V2 Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II Product Code: DXJ Dated: July 20, 2016 Received: July 21, 2016
Dear Mr. Robert Casarsa,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Manda Jellison
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K162012
Device Name CARESCAPE Central Station v2 (CSCS)
#### Indications for Use (Describe)
The CARESCAPE Central Station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data within a hospital or clinical environment.
The CARESCAPE Central Station is intended to collect, display and print information from a network, including patient demographics, physiological parameters and waveforms, alarm annunciation and/or other non-medical information from monitors and telemetry systems. Physiological parameters and waveforms include electrocardiograph (ECG), pulse oximetry (SPO2), invasive blood pressures (IBP), non-invasive blood pressure (NIBP), respiration (RR), ventilator (VNT), carbon dioxide (CO2), oxygen (O2), mass spectrometry (Gas), temperature (Temp) and bispectral index (BIS), Beat to beat patient information for parameters and waveforms from the bedside and telemetry systems can be displayed. Patient monitor and telemetry system settings can be adjusted. Parameter values derived from patient data can be calculated, displayed, and printed.
The CARESCAPE Central Station supports the ability to access information from GE products and hospital intranet in a web browser format. Additionally, CARESCAPE Central Station supports the ability to access patient information collected from the CARESCAPE network and stored on a network server.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif;">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are in a stylized font, and there are decorative swirls around the circle. The logo is simple and recognizable, and it is associated with a well-known company.
## K162012 PAGE 1 OF 6
### 510(K) SUMMARY
In accordance with 21 CFR 807.92 the following summary of information is provided:
- Date: July 20, 2016 Submitter: GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223 Contact Person: Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. Email: robert.casarsa@ge.com Ph: (414) 362-3063 Cell: (262) 358-1082 Device Trade Name: CARESCAPE Central Station v2 (CSCS) Common/Usual Central Station Name Classification: 21 CFR 870.2450 Product Code: DXJ BZO 21 CFR 868.2375 monitor, breathing frequency Secondary Codes: 21 CFR 868.1700 analyzer, gas, nitrous-oxide, gaseous CBR phase (anesthetic conc.) CBS 21 CFR 868.1620 analyzer, gas, halothane, gaseousphase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, enflurane, gaseous-CBQ phase (anesthetic concentration) CCK 21 CFR 868.1400 analyzer, gas, carbon-dioxide, gaseous-phase
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the well-known multinational conglomerate.
# GE Healthcare
510(k) Premarket Notification Submission
| CCL | 21 CFR 868.1720 | analyzer, gas, oxygen, gaseous-phase |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| DXN | 21 CFR 870.1130 | system, measurement, blood-pressure, noninvasive |
| DOA | 21 CFR 870.2700 | oximeter |
| DPT | 21 CFR 870.2300 | monitor, cardiac (incl. cardiotachometer & rate alarm) |
| DSB | 21 CFR 870.2770 | plethysmograph, impedance |
| DSK | 21 CFR 870.1110 | computer, blood-pressure |
| GWQ | 21 CFR 882.1400 | full-montage standard electroencephalograph |
| FLL | 21 CFR 880.2910 | thermometer, electronic, clinical |
| NHO | 21 CFR 868.1500 | analyzer, gas, desflurane, gaseous-phase (anesthetic concentration) |
| NHP | 21 CFR 868.1500 | analyzer, gas, sevoflurane, gaseous-phase (anesthetic concentration) |
| NHQ | 21 CFR 868.1500 | analyzer, gas, isoflurane, gaseous-phase (anesthetic concentration) |
| BSE | 21 CFR 868.1640 | Helium gas analyzer |
| JEG | 21 CFR 868.1075 | Argon gas analyzer |
| CCI | 21 CFR 868.1690 | Nitrogen gas analyzer |
| Predicate Device(s): | CARESCAPE Central Station (K133882)<br>MUSE ST Guard (K842308)<br>12SL V22 – ACS (K092369)<br>MUSE CV (K110132) | |
| Device Description: | The CARESCAPE Central Station (CSCS) is based on a PC technology platform and is user friendly for easy operation using a simple logical screen menu. The interactive controls include the use of a computer mouse, keyboard and touch screen. Optional writers for the purpose of graphing waveforms and printing patient information include a 2-inch Direct Digital Writer and/or a laser printer. Internal speakers provide alarm audio indication.<br>The CSCS provides centralized monitoring of patients connected to GE Medical Systems Information Technologies, Inc.'s monitors and telemetry transmitters. It may be configured to display up to four real-time waveforms per patient for up to 16 patients and up to 9 waveforms for a single selected patient. Waveforms include ECG, SPO2, respiration ventilation flow and pressure, invasive blood pressure and CO2. | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized, cursive font. The letters are enclosed within a blue circle with a decorative, swirling border. The logo is simple, recognizable, and represents the General Electric brand.
The device acts as a (1) viewing station, (2) reviewing station and (3) live monitoring station by obtaining its information from acquisition devices off the CARESCAPE network.
Patients may be admitted to and discharged from monitors and telemetry devices from the central location. The central station is also the control and display device for telemetry monitoring.
The CSCS provides secondary annunciation of alarms from primary bedside monitoring devices and primary annunciation of alarms from wireless telemetry devices.
The display window for each patient shows waveforms and vital information including: patient name, bed number, arrhythmia and alarm visual indicators, system messages, audio pause indicator, audio alarm indicator, alarm message line, heart rate, PVC count, transmitter number, ECG lead label, pacemaker status, ST measurement, and graph status. Physiological parameter values and waveforms from the GE Medical Systems Information Technologies, Inc.s' monitors can be displayed and printed from the CSCS.
Non-real time patient information available for reviewing and printing includes: Graphic Trends, Tabular Numeric Vital SignsTrends, Event HistoryReview, Full Disclosure, Calipers, and ST Review. Data can be printed to a networked laser printer. In the case of Event Review, data can also be printed to a PDF file.
The CARESCAPE Central Station also provides remote control of patient monitor and telemetry device configuration settings that includes:
- Admitted patient demographics like name and medical record number;
- Alarm Settings like high/low limit values and alarm priority levels:
- Printing settings like selection of which waveforms to print on graphs and the printed output destinations;
- ECG settings like primary lead selection, ST analysis on/off and pace maker detection on/off;
Initiate and terminate combination monitoring where a bedside patient monitor accepts ECG data from a telemetry transmitter:
- Non-ECG parameter settings like respiration lead selection
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, decorative elements, also in blue, giving the logo a classic and recognizable appearance.
and NBP cuff size selection.
The Full Disclosure option provides up to 144 hours of beat-tobeat patient information from the bedside or telemetry system for parameters and waveforms. Full Disclosure also stores resting ECGs from 16 patients once per minute for 144 hours and up to 2000 alarm histories with waveform snippets for each patient. This information can be displayed at the CSCS in detailed and summary mode formats.
The CARESCAPE Central Station is intended for use under the Intended Use: direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data within a hospital or clinical environment.
> The CARESCAPE Central Station is intended to collect, display and print information from a network, including patient demographics, physiological parameters and waveforms, alarm annunciation and/or other non-medical information from monitors and telemetry systems. Physiological parameters and waveforms include electrocardiograph (ECG), pulse oximetry (SPO2), invasive blood pressures (IBP), non-invasive blood pressure (NIBP), respiration (RR), ventilator (VNT), carbon dioxide (CO2), oxygen (O2), mass spectrometry (Gas), temperature (Temp) and bispectral index (BIS). Beat to beat patient information for parameters and waveforms from the bedside and telemetry systems can be displayed. Patient monitor and telemetry system settings can be adjusted. Parameter values derived from patient data can be calculated, displayed, and printed.
The CARESCAPE Central Station supports the ability to access information from GE products and hospital intranet in a web browser format. Additionally, CARESCAPE Central Station supports the ability to access patient information collected from the CARESCAPE network and stored on a network server.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has a wavy or swirling pattern around the letters, giving it a dynamic and recognizable appearance.
The CARESCAPE Central Station employs the same functional Technology: scientific technology as its predicate devices. The device is a software driven device running on a PC platform. The
> CARESCAPE Central Station V2 utilizes new hardware platforms and embedded operating system.
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The CARESCAPE Central Station and its applications were tested to, and comply with, applicable voluntary standards. The CARESCAPE Central Station was tested to assure that the device meets its design specifications. Testing included all new or modified features.
The following quality assurance measures were applied to the development and testing of the of the system:
Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) Summary of Clinical Tests:
The subject of this premarket submission, CARESCAPE Central Station, did not require clinical studies to support substantial equivalence.
Comparison: The following contains the differences of the CARESCAPE Central Station V2 when compared to the predicate CARESCAPE Central Station V1:
### Hardware
Upgraded Platform; faster processor, increased RAM, increased SSD Flash Media, removed external RS-232 Serial port, removed DVI and VGA interfaces, removed unused network port, and increased MC support to 1000Mbps.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular frame that has decorative, swirling elements around the perimeter. The color of the logo is a light blue.
### GE Healthcare
510(k) Premarket Notification Submission
- 2 configurations available, Desktop and Integrated hardware options
Supports widescreen displays
Eliminated cooling fan for low power CPU
#### Software
Upgraded platform from XP based OS to Windows 7 based.
#### Historical Trending
- Added ability for additional trends where the bedside is already monitoring the now supported additional parameters
#### Alarming
Settings are password protected
Alarm levels now conform to IEC 60601-2-27:2011 Clause 208.6.6.2.105, Clause 208.6.1.2, and Clause 208.6.11.2.2
### Printing
- . Compatible with PCL 6 Compatible HP Universal Printers to support additional printers.
- GE Healthcare considers the CARESCAPE Central Station to Conclusion: be as safe and as effective, and its performance substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.