DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES
K052582 · GE Healthcare · CCK · Oct 14, 2005 · Anesthesiology
Device Facts
Record ID
K052582
Device Name
DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES
Applicant
GE Healthcare
Product Code
CCK · Anesthesiology
Decision Date
Oct 14, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Datex-Ohmeda S/5 Single-width airway module, E-miniC is intended to be used with Datex-Ohmeda S/5 modular monitors: S/5 Anesthesia Monitor, S/5 Compact Anesthesia Monitor, S/5 Critical Care Monitor and S/5 Critical Care Monitor for monitoring CO2 and respiration rate of all hospital patients.
Device Story
Side-stream gas analyzer module for Datex-Ohmeda S/5 modular monitors; measures CO2 and respiration rate. Inputs: airway gas samples via sampling line/airway adapter. Principle: infrared absorption at 4.2-4.3 mm using narrow-band filters. Output: numeric CO2 values, waveforms, and trends displayed on monitor. Used in clinical settings (anesthesia/critical care) by qualified medical personnel. Provides real-time breath-by-breath monitoring; generates audible/visual alarms for patient safety. Benefits: enables continuous respiratory status assessment to inform clinical decision-making.
Clinical Evidence
Bench testing only. Verification and validation performed against standards including IEC 60601-1, IEC 60601-1-2, ISO 9918:1993/EN 864:1996, and ASTM F-1456:2001. No clinical data presented.
Technological Characteristics
Side-stream infrared gas analyzer; measures CO2 absorption at 4.2-4.3 mm. Plug-in module for S/5 modular monitors. Accessories include airway gas sampling lines, mini D-fend water traps, and airway adapters. Complies with IEC 60601-1, ISO 9918, and ASTM F-1456. Connectivity via proprietary monitor interface.
Indications for Use
Indicated for monitoring CO2 and respiration rate in hospital patients weighing >5kg (11 lbs). For use by qualified medical personnel only.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
S/5 Compact Critical Care Monitor (CCCM) (K022740)
Submission Summary (Full Text)
{0}------------------------------------------------
K052582
OCI 1 4 2005
Page 1 of 3
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/50M Single-width Airway Module, E-MINIC and accessories
# GENERAL COMPANY INFORMATION as required by 807.92(a){1}
#### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
#### NAME OF CONTACT:
Mr. Joel Kent
DATE: September 16, 2005
#### DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5TM Single-width Airway Module, E-MINIC and accessories
## COMMON NAME:
Carbon dioxide gas monitor
#### CLASSIFICATION NAME:
## The following Class II classification appears applicable:
| Product Code | Classification Name | CFR Section |
|--------------|----------------------------------------------|-------------|
| CCK | Analyzer, Gas, Carbon-Dioxide, Gaseous-phase | 868.1400 |
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5™ Single-width airway module, E-miniC is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda Single-width airway module, M-miniC (K023454).
{1}------------------------------------------------
#### DEVICE DESCRIPTION as required by 807.92(a){4)
The Datex-Ohmeda E-miniC module is a single-width plug-in parameter module for a Datex-Ohmeda modular monitoring system. The Datex-Ohmeda S/5 E-mini(' module can be used with the following Datex-Ohmeda modular monitors:
- S/50M Anesthesia Monitor (AM) with main software L-ANE02(A)..00 (K021279) or . newer version
- S/5™ Compact Anesthesia Monitor (CAM) with main software L-CANE02(A). 00 . (K022485) or newer version.
- S/5TM Critical Care Monitor (CCM) with main software L-ICU02(A)..00 (K021376) or . newer version
- S/5TM Compact Critical Care Monitor (CCCM) with main software L-CICU02(A). 00 . (K022740) or newer version.
The E-miniC module uses mainly the same accessories as the predicate device, M-miniC (K023454). The main accessories include airway gas sampling lines, mini D-fend water traps and airway adapters. Some supplies have been discontinued, and the Exhaust line for gas return has been redesigned for a new gas return connector. The E-miniC module is a side stream gas analyzer for monitoring the Carbon dioxide (CO2) inhaled and exhaled by the patient by measuring absorbtion of CO2 at 4.2-4.3 mm using narrow band infrared filters/sensors. The EminiC also allows monitoring of the respiration rate. The CO2 is measured and displayed breath by breath. The module is first plugged into the frame of the Monitor. The sampling line is attached to the module connector. The monitor is switched on and the gas sampling line is attached to the airway adapter. The airway adapter is attached between ventilator Y-piece and Heat and moisture exchanger (HME) of the patient's intubation tube. The monitor displays measurements from the E-miniC module in the form of numeric values, CO2 waveform and trends. The waveform size, color and sweep speed can be adjusted. The monitor also generates audible and visual alarms for this module and indicates the priorities and sources of alarms, according to the user interface for alarms in Datex-Ohmeda S/5 patient monitors.
#### INTENDED USE as required by 807.92(a)(5)
#### Intended Use:
The Datex-Ohmeda S/5 Single-width airway module, E-miniC is intended to be used with Datex-Ohmeda S/5 modular monitors: S/5 Anesthesia Monitor, S/5 Compact Anesthesia Monitor, S/5 Critical Care Monitor and S/5 Compact Critical Care Monitor for monitoring CO2 and respiration rate of all hospital patients.
## Indications for use:
The Datex-Ohmeda S/5 Single-width airway module, E-miniC and accessories is indicated for monitoring CO2 and respiration rate of all hospital patients. E-miniC is indicated for monitoring patients weighing more than 5kg (11 lbs.).
The device is indicated for use by qualified medical personnel only.
# SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5™ Single-width airway module, E-miniC is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda Single-width airway module, M-miniC (K023454).
{2}------------------------------------------------
The E-miniC module has the following similarities compared to the predicate M- miniC (K023454):
- identical intended use and indications for use .
- identical fundamental scientific technology .
- use the same operating principle .
- the same accessories (some supplies discontinued, one redesigned) .
- have the same user interface at the monitor and alarms (can be used with the same . monitor software)
- the Customer and parameter specifications are the same .
- have the same safety and effectiveness .
- are manufactured using the same processes .
The main differences between the new E- miniC and the predicate M-miniC (K023454) is primarily due to fact that the new E- miniC module has the following changes:
- New color, shape, and size and thus differing mechanics .
- The front panel and labeling have changed .
- E-miniC uses the improved CO2 measuring miniC unit of the previously cleared S/5 FM . Monitor (K043276)
Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of the Datex-Ohmeda S/5™ Single-width airway Module, E-miniC and accessories are substantially equivalent to the predicate Datex-Ohmeda M- miniC Module (K023454).
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5™ Single-width airway Module, E-miniC and accessories has been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications.
- COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning medical devices .
- FDA/DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993 .
- IEC 60601-1:1988 + Amdt. 1:1991 + Amdt. 2:1995 (Part 1: General requirements for safety) .
- EN 60601-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IEC60601-1:1988 + Amdt. . 1:1991 + Amdt. 2:1995)
- CAN/CSA C22.2 No. 601.1-M90 + S1:1994 (Canadian deviations to IEC 60601-1:1988 + Amdt. . 1:1991) + S2:1998 (=IEC Amdt 2:1995)
- UL 2601-1, October 24, 1997 (U.S. deviations to IEC 60601-1:1988 + Amdt. 1:1991+ Amd. . 2:1995)
- IEC 60601-1-2:2001 (Electromagnetic compatibility -- Requirements and tests) .
- FDA/ODE Guidance for Content of Premarket Submission for Software Contained in Medical . Devices. (May, 11, 2005)
- ISO 9918:1993 / EN 864:1996 Medical electrical equipment-Capnometers for use with humans-. Particular requirements
- ASTM F-1456: 2001 Standard Specification for Capnometers .
## CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5™ Single-width airway module, E-miniC as compared to the legally marketed (predicate) Datex-Ohmeda Single-width airway module, M-miniC (K023454).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" in a circular arrangement.
OCT 1 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joel Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusettes 02492
Re: K052582
Trade/Device Name: DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Monitor Regulatory Class: II Product Code: CCK Dated: September 16, 2005 Received: September 20, 205
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutte y Michael Omd.
Chris Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Datex-Ohmeda S/5 Single-width airway module, E-miniC and accessories.
Indications for use:
The Datex-Ohmeda S/5 Single-width airway module, E-miniC and accessories is indicated for monitoring CO2 and respiration rate of all hospital patients. E-miniC is indicated for monitoring patients weighing more than 5kg (11 lbs.).
The device is indicated for use by qualified medical personnel only.
Prescription Use __ X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suette Y. Michael Davis.
(Division Sign-Division of Anesthesiology, General Hospital, Infection Control, Dental Devic
510(k) Number __
Page __ of _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.