K170394 · Achimhai Medical Corporation · DZE · Oct 6, 2017 · Dental
Device Facts
Record ID
K170394
Device Name
Kisses Mini
Applicant
Achimhai Medical Corporation
Product Code
DZE · Dental
Decision Date
Oct 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Kisses Mini is indicated for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Kisses Mini is for single and two stage surgical procedures. It is for delayed loading. The intended use of the 3.2 mm Internal Hex fixtures is limited to the replacement of maxillary lateral incisors and mandibular incisors. The intended use of the single-piece post type fixture is limited to replacement of mandibular central and lateral incisors. The ball-type single-piece implant is intended for stabilization and retention of overdentures.
Device Story
Kisses Mini is a dental implant system comprising fixtures (2-piece, post, and ball types), abutments, and screws. Fixtures are surgically placed into the jawbone to replace lost tooth roots; they feature self-tapping threads and SLA surface treatment. Abutments (duplex, simplex, angled, temporary) couple to fixtures to support prosthetic restorations. Ball-type implants use O-ring attachments for overdenture retention. Used in clinical dental settings by trained professionals for single or two-stage surgical procedures with delayed loading. The system provides structural support for artificial teeth, restoring chewing function and stabilizing dentures. Benefits include improved oral function and prosthetic retention for edentulous patients.
Clinical Evidence
Bench testing only. Includes sterilization validation (ISO 11137, ISO 17665), shelf-life validation, surface characterization (morphology, roughness, FT-IR), and mechanical performance testing (fatigue, compressive strength, rotational shear, screw loosening torque). Endotoxin testing per USP <85> (limit <0.5 EU/mL).
Technological Characteristics
Materials: Ti Gr.4 (fixtures), Ti 6Al 4V ELI Gr.23 (abutments/screws), Silicon (O-rings). Surface: SLA treatment. Sterilization: Gamma. Form factor: Internal hex, post, and ball-type fixtures; various abutment geometries. Connectivity: Mechanical coupling.
Indications for Use
Indicated for partially or fully edentulous patients (maxilla/mandible) requiring single or multiple-unit restorations (cemented/screw-retained/overdenture) or fixed bridgework support. Specific fixtures limited to maxillary lateral incisors and mandibular incisors. Ball-type implants indicated for overdenture stabilization.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 6, 2017
ACHIMHAI Medical Corporation % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 800 Roosevelt. Suite 417 Irvine, California 92620
Re: K170394
Trade/Device Name: Kisses Mini Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 4, 2017 Received: September 7, 2017
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K170394
Device Name Kisses Mini
Indications for Use (Describe)
The Kisses Mini is indicated for use in partially or fully edentulous maxilla, in support of single or multiple-unt restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Kisses Mini is for single and two stage surgical procedures. It is for delayed loading.
The intended use of the 3.2 mm Internal Hex fixtures is limited to the replacement of maxillary lateral incisors.
The intended use of the single-piece post type fixture to replacement of mandibular central and lateral incisors.
The ball-type single-piece implant is intended for stabilization and retention of overdentures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# (K170394)
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 10/06/2017
# 1. Submitter
ACHIMHAI Medical Corporation 28, Namyang-ro, 930beon-gil, Namyang-eup Hwaseong-si, Gyeonggido, Republic of Korea, 18255
# 2. U.S Agent/Contact Person
Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt Ste 417, Irvine CA 92620 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- Trade Name: Kisses Mini ●
- Common Name: Dental Implant System
- Classification Name: Endosseous Dental Implant System ●
- Product Code: DZE, NHA
- Classification regulation: 21CFR872.3640 ●
# 4. Predicate Device:
- . Primary Predicate Device: MiNi Internal Implant System (K150537) by MegaGen Implant Co., Ltd. S-MiNi Implant System (K112540) by Neobiotech Co., Ltd.
- Reference Predicate Device: ● Biogenesis Implant System – Kisses (K142813) by Achimhai Medical Corporation
# 5. Description:
KJ Mini offers three different types of implants: 2-Piece Type, Post Type, and Ball Type.
{5}------------------------------------------------
The 2-piece type fixture is a substructure of a dental implant system to replace a single tooth, and the lost root of edentulous patients. It consists of the hex part to be coupled to the superstructure, the single thread part to be fixed to the bone, and the cutting edge part with the self-tapping function. It offers single hex type and double hex type.
- 3.2mm Dia. x 8mm L ●
- 3.2mm Dia. x 9.5mm L ●
- 3.2mm Dia. x 11mm L
- 3.2mm Dia. x 12.5mm L
- 3.2mm Dia. x 14.5mm L
The post type fixture is a substructure of a dental implant system to replace a single tooth, and the lost root of edentulous patients.
- 2.8mm Dia. x 8mm L ●
- 2.8mm Dia. x 9.5mm L ●
- 2.8mm Dia. x 11mm L ●
- 2.8mm Dia. x 12.5mm L ●
- 2.8mm Dia. x 14.5mm L
- 3mm Dia. x 8mm L
- 3mm Dia. x 9.5mm L
- 3mm Dia. x 11mm L
- 3mm Dia. x 12.5mm L
- 3mm Dia. x 14.5mm L .
The ball type fixture is a substructure of a dental implant system to replace a single tooth, and the lost root of edentulous patients.
- 2.8mm Dia. x 8mm L ●
- 2.8mm Dia. x 9.5mm L ●
- 2.8mm Dia. x 11mm L ●
- 2.8mm Dia. x 12.5mm L ●
- 2.8mm Dia. x 14.5mm L
- . 3mm Dia. x 8mm L
- . 3mm Dia. x 9.5mm L
- 3mm Dia. x 11mm L ●
- 3mm Dia. x 12.5mm L .
- 3mm Dia. x 14.5mm L ●
## 6. Indication for use:
The Kisses Mini is indicated for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Kisses Mini is for single and two stage surgical procedures. It is for delayed loading.
The intended use of the 3.2 mm Internal Hex fixtures is limited to the replacement of
{6}------------------------------------------------
maxillary lateral incisors and mandibular incisors.
The intended use of the single-piece post type fixture is limited to replacement of mandibular central and lateral incisors.
The ball-type single-piece implant is intended for stabilization and retention of overdentures.
## 7. Basis for Substantial Equivalence
## 7.1. Fixture
## Internal Hex Type
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | - | K150537 | K142813 |
| Device Name | Kisses Mini | MiNi Internal Implant System | Biogenesis Implant System -<br>Kisses |
| Manufacturer | Achimhai Medical Corporation | MegaGen Implant Co., Ltd. | Achimhai Medical Corporation<br>(Note: Our company name<br>has changed from Biogenesis<br>to Achimhai.) |
| Type | Internal hex | Internal hex | Internal hex |
| Design & Size Range | Diameter: 3.2 mm<br>Length: 8 - 14.5 mm | Diameter: 3.0 – 3.4 mm<br>Length: 8-14.5 mm | Diameter: 3.5 - 5.5 mm<br>Length: 7-15 mm |
| Indications for Use | The Kisses Mini is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxilla, in support of single or<br>multiple-unit restorations<br>including; cemented retained,<br>screw retained, or<br>overdenture restorations, and<br>terminal or intermediate<br>abutment support for fixed<br>bridgework. The Kisses Mini<br>is for single and two stage<br>surgical procedures. It is for<br>delayed loading.<br>The intended use of the 3.2<br>mm Internal Hex fixtures is<br>limited to the replacement of<br>maxillary lateral incisors and<br>mandibular incisors.<br>The intended use of the<br>single-piece post type fixture<br>is limited to replacement of<br>mandibular central and lateral<br>incisors. | The MiNi Internal Implant<br>System is intended for two<br>stage surgical procedures in<br>the following situations and<br>with the following clinical<br>protocols:<br>• The intended use for the 3.0<br>mm diameter MiNi implant is<br>limited to the replacement of<br>maxillary lateral incisors and<br>mandibular incisors.<br>• Immediate placement in<br>extraction sites and in<br>situations with a partially or<br>completely healed alveolar<br>ridge.<br>• It is intended for delayed<br>loading. | The Biogenesis Implant<br>System -Kisses is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of single<br>or multiple-unit restorations<br>including; cemented retained,<br>screw retained, or<br>overdenture restorations, and<br>terminal or intermediate<br>abutment support for fixed<br>bridgework.<br>The Biogenesis Implant<br>System - Kisses is for single<br>and two stage surgical<br>procedures. It is for delayed<br>loading. |
{7}------------------------------------------------
| | The ball-type single-piece<br>implant is intended for<br>stabilization and retention of<br>overdentures. | | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>Operation | The device is a substructure<br>of a dental implant system to<br>replace a single tooth and the<br>lost root of edentulous<br>patients. It consists of the hex<br>part to be coupled to the<br>superstructure, the single<br>thread part to be fixed to the<br>bone, and the cutting edge<br>part with the self-tapping<br>function. | The device is a substructure<br>of a dental implant system to<br>replace a single tooth and the<br>lost root of edentulous<br>patients. It consists of the hex<br>part to be coupled to the<br>superstructure, the single<br>thread part to be fixed to the<br>bone, and the cutting edge<br>part with the self-tapping<br>function. | The device is a substructure<br>of a dental implant system to<br>replace a single tooth and the<br>lost root of edentulous<br>patients. It consists of the hex<br>part to be coupled to the<br>superstructure, the single<br>thread part to be fixed to the<br>bone, and the cutting edge<br>part with the self-tapping<br>function. |
| Material Composition | Ti Gr.4 | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | SLA Treatment | SLA Treatment | SLA Treatment |
| Sterile | Yes | Yes | Yes |
| Sterilization Method | Gamma | Gamma | Gamma |
Kisses Mini Internal hex type fixture is substantially equivalent to the predicate devices in terms of intended use and technical characteristics. They have similar design that all the devices are two piece type fixtures with thread part to be fixed to the bone. They are made of the same material and all have cutting edge part for self-tapping function. The size range of the primary predicate device encompasses the size range of the subject device.
There are slight differences in design such as thread sizes but it is very minor not affecting the substantial equivalence. Also, the Kisses Mini indications for use differ from the predicate devices in some languages for specific types and size, however, these differences do not raise new concerns because limiting the use of each type is a more conservative surgical approach.
### Post Type
| | Subject Device | Primary Predicate Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | N / A | K112540 |
| Device Name | Kisses Mini | S-MiNi Implant System |
| Manufacturer | Achimhai Medical Corporation | Neobiotech Co., Ltd. |
| Type | Post | Post |
| Design & Size Range | Image: A dental implant.<br>Diameter: 2.8 - 3.0 mm<br>Length: 8 - 14.5 mm<br>Cuff: 2.5 – 4.0 mm<br>Post: 8.0 mm | Image: A dental implant.<br>Diameter: 2.0 – 3.5 mm<br>Length: 7-15 mm<br>Cuff: 2.0 mm<br>Post: 10.0 mm |
| Intended Use | The Kisses Mini is indicated for use in partially<br>or fully edentulous mandibles and maxilla, in<br>support of single or multiple-unit restorations<br>including; cemented retained, screw retained,<br>or overdenture restorations, and terminal or | The S-MiNi Implant System is intended to use<br>in the treatment of missing mandibular central<br>and lateral incisors to support prosthetic<br>device, such as artificial teeth, in order to<br>restore chewing function partially edentulous. |
510(k) Submission
{8}------------------------------------------------
| intermediate abutment support for fixed<br>bridgework. The Kisses Mini is for single and<br>two stage surgical procedures. It is for<br>delayed loading.<br>The intended use of the 3.2 mm Internal Hex<br>fixtures is limited to the replacement of<br>maxillary lateral incisors and mandibular<br>incisors.<br>The intended use of the single-piece post type<br>fixture is limited to replacement of mandibular<br>central and lateral incisors.<br>The ball-type single-piece implant is intended<br>for stabilization and retention of overdentures. | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>Operation | The device is a substructure of a dental<br>implant system to replace a single tooth and<br>the lost root of edentulous patients. It consists<br>of the single thread part and the post part; the<br>single thread part to be fixed to the bone, and<br>the cutting edge part with the self-tapping<br>function. | The device is a substructure of a dental<br>implant system to replace a single tooth and<br>the lost root of edentulous patients. It consists<br>of the single thread part and the post part; the<br>single thread part to be fixed to the bone, and<br>the cutting edge part with the self-tapping<br>function. |
| Material Composition | Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | SLA Treatment | RBM Treatment |
| Sterile | Yes | Yes |
| Sterilization Method | Gamma | Gamma |
Kisses Mini Post type Implant is substantially equivalent to the predicate devices in terms of intended use and technical characteristics. The design is very similar that the both devices are one body type which a fixture and an abutment are combined. They are material and all have cutting edge part for self-tapping function. The size range of the predicate device encompasses the size range of the subject device.
The difference is in cuff size that the subject device cuff is longer than the predicate device. However, this difference does not affect substantial equivalence.
Another difference is in surface treatment that the subject device employs SLA and the predicate device employs RBM. For the subject deice, surface morphology, surface roughness Analysis, and FT-IR Analysis were performed to evaluate its characteristics, and the test results support it does not affect substantial equivalence.
Also, the Kisses Mini indications for use differ from the predicate devices in some languages for specific types and size, however, these differences do not raise new concerns because limiting the use of each type is a more conservative surgical approach.
### Ball Type
| | Subject Device | Primary Predicate Device |
|---------------|------------------------------|--------------------------|
| 510(K) Number | N / A | K112540 |
| Device Name | Kisses Mini | S-MiNi Implant System |
| Manufacturer | Achimhai Medical Corporation | Neobiotech Co., Ltd. |
| Type | Ball | Ball |
{9}------------------------------------------------
| Design & Size Range | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image: Implant | Image: Implant |
| Diameter: 2.8 – 3.0 mm<br>Length: 8 - 14.5 mm<br>Cuff: 2.0-4.0 mm | Diameter: 2.0 – 3.5 mm<br>Length: 7 - 15 mm<br>Cuff: 3.0-4.0 mm |
| Intended Use | |
| The Kisses Mini is indicated for use in<br>partially or fully edentulous mandibles and<br>maxilla, in support of single or multiple-unit<br>restorations including; cemented retained,<br>screw retained, or overdenture restorations,<br>and terminal or intermediate abutment<br>support for fixed bridgework. The Kisses Mini<br>is for single and two stage surgical<br>procedures. It is for delayed loading.<br><br>The intended use of the 3.2 mm Internal Hex<br>fixtures is limited to the replacement of<br>maxillary lateral incisors and mandibular<br>incisors.<br><br>The intended use of the single-piece post<br>type fixture is limited to replacement of<br>mandibular central and lateral incisors.<br><br>The ball-type single-piece implant is intended<br>for stabilization and retention of overdentures. | The S-MiNi Implant System is intended to be<br>place in the bone of the upper or lower jaw<br>arches to provide support the prosthetic<br>devices to restore the patient's chewing<br>function, including the denture stabilization.<br>S-MiNi Implant System is intended for single<br>use only |
| Principle of Operation | |
| The device is a substructure of a dental<br>implant system to replace a single tooth and<br>the lost root of edentulous patients. It consists<br>of the single thread part and the ball part; the<br>single thread part to be fixed to the bone, the<br>cutting edge part with the self-tapping<br>function, and the ball part for detachable<br>denture. | The device is a substructure of a dental<br>implant system to replace a single tooth and<br>the lost root of edentulous patients. It consists<br>of the single thread part and the ball part; the<br>single thread part to be fixed to the bone, the<br>cutting edge part with the self-tapping<br>function, and the ball part for detachable<br>denture. |
| Material Composition | |
| Ti Gr.4 | Ti Gr.4 |
| Surface Treatment | |
| SLA Treatment | RBM Treatment |
| Sterile | |
| Yes | Yes |
| Sterilization Method | |
| Gamma | Gamma |
Kisses Mini Ball type Implant is substantially equivalent to the predicate devices in terms of intended use and technical characteristics. The design is also similar that the both devices have a ball-type structure for the connection of the O-ring attachment. They are made of the same material and all have cutting edge part for self-tapping function. The size range of the predicate device encompasses the size range of the subject device.
There is slight difference in design such as cuff design but it is very minor not affecting substantial equivalence.
Another difference is in surface treatment that the subject device employs SLA and the predicate device employs RBM. For the subject deice, surface morphology, surface roughness Analysis, and FT-IR Analysis were performed to evaluate its characteristics, and the test results support it does not affect substantial equivalence.
{10}------------------------------------------------
Also, the Kisses Mini indications for use differ from the predicate devices in some languages for specific types and size, however, these differences do not raise new concerns because limiting the use of each type is a more conservative surgical approach.
## 7.2. Abutment
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | | K150537 | K142813 |
| Device Name | Kisses Mini | MiNi Internal Implant System | Biogenesis Implant System -<br>Kisses |
| Manufacturer | Achimhai Medical<br>Corporation | MegaGen Implant Co., Ltd. | Achimhai Medical Corporation<br>(Note: Our company name<br>has changed from Biogenesis<br>to Achimhai.) |
| Type | Internal | Internal | Internal hex |
| Indications for Use | The Kisses Mini is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxilla, in support of single<br>or multiple-unit restorations<br>including; cemented<br>retained, screw retained, or<br>overdenture restorations,<br>and terminal or intermediate<br>abutment support for fixed<br>bridgework. The Kisses Mini<br>is for single and two stage<br>surgical procedures. It is for<br>delayed loading.<br>The intended use of the 3.2<br>mm Internal Hex fixtures is<br>limited to the replacement<br>of maxillary lateral incisors<br>and mandibular incisors.<br>The intended use of the<br>single-piece post type<br>fixture is limited to<br>re…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.