The MiNi Internal Implant System is intended for two-stage surgical procedures in the following situations and with the following clinical protocols: - The intended use for the 3.0 mm diameter MiNi implant is limited to the replacement of maxillary lateral incisors and mandibular incisors. - Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge. - It is intended for delayed loading.
Device Story
MiNi Internal Implant System consists of CP Ti Grade 4 endosseous dental fixtures, abutments, and surgical instruments. Used in two-stage dental implant surgery to provide screw or cement-retained restorations. Fixtures feature S.L.A. (Sand-blasted, Large grit, Acid-etched) surface treatment. System includes various abutments (e.g., healing, angled, milling, temporary, impression copings). Clinicians perform surgical placement in dental clinics. Output is a stable root-form foundation for prosthetic teeth, restoring chewing function and aesthetics for patients with missing teeth.
Clinical Evidence
Bench testing only. Fatigue testing performed per ISO 14801:2007. Sterilization validation performed per ISO 11137 and ISO 17665 series. Biocompatibility, surface analysis, and residue analysis were not conducted, relying on equivalence in material, surface treatment, and manufacturing processes to the reference predicate (K140091).
Technological Characteristics
Materials: CP Ti Grade 4 (ASTM F67). Surface: S.L.A. (Sand-blasted, Large grit, Acid-etched). Fixture diameters: 3.0mm, 3.4mm. Lengths: 8.0mm to 14.5mm. Connection: Internal Hex. Sterilization: Gamma and steam sterilization.
Indications for Use
Indicated for two-stage surgical procedures for replacement of maxillary lateral incisors and mandibular incisors (3.0 mm diameter implants). Suitable for immediate placement in extraction sites or healed alveolar ridges. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with a symbol resembling a caduceus above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2015
MegaGen Implant Co., Ltd c/o Ms. Megan Holden Kodent, Inc. 325 N. Puente St., Unit B Brea, California 92821
Re: K150537
Trade/Device Name: MiNi Internal Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 16, 2015 Received: June 22, 2015
Dear Ms. Holden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for Use
510(K) Number (if known): K150537
Device Name: MiNi Internal Implant System
# Indications for Use:
The MiNi Internal Implant System is intended for two-stage surgical procedures in the following situations and with the following clinical protocols:
- The intended use for the 3.0 mm diameter MiNi implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. -
X Prescription Use ____________ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 510(k) Summary
#### (K150537)
This 510(k) Summary is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Submitter:
Dong Guk Ha MegaGen Implant Co., Ltd. 472, Hanjanggun-ro, Jain-myeon, Gyeongsan-si Gyeongsangbuk-do, South Korea Phone: 82-53-857-5770 82-53-857-5432 Fax:
# Contact /US agent:
Megan Holden KoDent, Inc. 325 N. Puente St. Unit B Brea, CA 92821 Phone: 714-525-0114 Fax: 714-525-0116
#### Device Information:
Device Name: MiNi Internal Implant System Classification Name: Endosseous Dental Implant Classification: Class II Product Code: DZE Subsequent Product Code: NHA Regulation number: 21 CFR 872.3640 Date Prepared: 7/17/2015
## General Description
MiNi Internal Implant System is composed of MiNi Internal Fixtures, abutments, and surgical instruments. The dental fixtures are made of CP Ti Grade 4 designed for use in dental implant surgery. This system is used as two stage surgery, root-form dental implants, associated with abutment systems, which provide the clinician with the screw and cement retained restoration for multi-mount options.
The fixtures are available in diameters of 3.0mm, 3.4mm and lengths of 8.0mm, 9.5mm, 11.0mm, 12.5mm, and 14.5mm. The fixtures, abutments, and surgical instruments are produced and packaged separately. The Surface treatment for this device is S.L.A (Sand-blasted, Large grit, Acid-etched).
This system includes various abutments, such as Abutment Screw, Cover Screw, Healing Abutment, EZ Post, Angled Abutment, Milling Abutment, Solid Abutment, Temporary Abutment, Impression Coping, Plastic Impression Coping, Guide Pin, Lab Analog, and Meg-Rhein Abutment. These abutments are compatible with MiNi Internal Fixtures.
{4}------------------------------------------------
#### Indication for use
The MiNi Internal Implant System is intended for two-stage surgical procedures in the following situations and with the following clinical protocols:
- -The intended use for the 3.0 mm diameter MiNi implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. -
## Predicate devices
- Primary Predicate Device : OsseoSpeed™ Plus (K120414) .
- Reference Predicate Device : XPEED AnyRidge Internal Implant System (K140091) .
## Substantial Equivalence Comparison
The MiNi Internal Implant System is similar designs and dimensions, and has the same material, intended use, and technological characteristics as the identified primary predicate device (K120414). The MiNi Internal Implant System is similar in fundamental scientific technology, and has the same material, surface treatment to the reference predicate device (K140091). When compared with predicate device, no new questions of substantial equivalence have been raised for the MiNi Internal Implant System.
| | Subject Device | Primary Predicate | Reference Predicate | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device | Device | | |
| 510(k) Number | K150537 | K120414 | K140091 | |
| Device Name | MiNi Internal<br>Implant System | OsseoSpeed™ Plus | XPEED AnyRidge<br>Internal Implant<br>System | |
| Manufacturer | Megagen Implant Co.,<br>Ltd. | Astra Tech AB | Megagen Implant Co.,<br>Ltd. | |
| Indications for Use | The MiNi Internal Implant<br>System is intended for two-<br>stage surgical procedures in<br>the following situations and<br>with the following clinical<br>protocols:<br>• The intended use for the<br>3.0 mm diameter MiNi<br>implant is limited to the<br>replacement of maxillary<br>lateral incisors and<br>mandibular incisors.<br>• Immediate placement in<br>extraction sites and in | Implants:<br>The Astra Tech Dental<br>Implants are intended for<br>both one- and two-stage<br>surgical procedures in the<br>following situations and<br>with the following clinical<br>protocols:<br>• replacing single and<br>multiple missing teeth in the<br>mandible and maxilla,<br>• immediate placement in<br>extraction sites and in | The XPEED AnyRidge<br>Internal Implant System is<br>intended to be surgically<br>placed in the maxillary or<br>mandibular molar areas for<br>the purpose providing<br>prosthetic support for dental<br>restorations (Crown, bridges,<br>and overdentures) in<br>partially or fully edentulous<br>individuals. It is used to<br>restore a patient's chewing<br>function. Smaller implants<br>(less than Ø6.0 mm) are<br>dedicated for immediate | |
| | | | | |
| Indications for Use<br>(continued) | situations with a partially or<br>completely healed alveolar<br>ridge.<br>• It is intended for delayed<br>loading. | completely healed alveolar<br>ridge,<br>• especially indicated for use<br>in soft bone applications<br>where implants with other<br>implant surface treatments<br>may be less effective,<br>• immediate loading in all<br>indications, except in single<br>tooth situations on implants<br>shorter than 8 mm or in soft | loading when good primary<br>stability is achieved and with<br>appropriate occlusal loading.<br>Larger implants are<br>dedicated for the molar<br>region and are indicated for<br>delayed loading. | |
| | | bone (type IV) where<br>implant stability may be<br>difficult to obtain and<br>immediate loading may not<br>be appropriate.<br>The intended use for | | |
| | | OsseoSpeed™ Plus 3.0S is<br>limited to replacement of<br>maxillary lateral incisors and<br>mandibular incisors.<br>Abutments: | | |
| | | Astra Tech Implant System<br>Plus abutments are intended<br>to be used in conjunction<br>with Astra Tech Implant<br>System Plus in fully<br>edentulous or partially<br>edentulous maxillary and/or<br>mandibular arches to<br>provide support for crowns, | | |
| | | bridges or overdentures.<br>Atlantis Abutments:<br>The Atlantis™ Abutment is<br>intended for use with an<br>endosseous implant to | | |
| | | support a prosthetic device<br>in a partially or completely<br>edentulous patient. It is<br>intended for use to support<br>single and multiple tooth<br>prostheses, in the mandible | | |
| | | or maxilla. The prosthesis<br>can be cemented, screw<br>retained or friction fit to the<br>abutment. The abutment<br>screw is intended to secure<br>the abutment to the | | |
| | | endosseous implant.<br>The Atlantis™ Crown<br>Abutment in Zirconia is<br>intended for use with an<br>endosseous implant to<br>function as a substructure | | |
| | | that also serves as the final<br>restoration, in partially or<br>completely edentulous<br>patients. The prosthesis is<br>screw retained. The<br>abutment screw is intended | | |
| | | to secure the crown<br>abutment to the endosseous<br>implant. | | |
| | | | | |
| Design | MiNi Internal Implant System, abutments and accessories have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments<br>Internal Hex | | -Single-stage or two-stage<br>- Single-tooth and/or multiple tooth applications<br>- OsseoSpeedTM Plus 3.0S is limited to replacement of maxillary lateral incisors and mandibular incisor<br>Internal Double Hex | XPEED AnyRidge Internal Implant System, abutments and accessories have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments<br>Internal Hex |
| | Material | | CP Ti Grade 4<br>(ASTM F67) | CP Ti Grade 4<br>(ASTM F67) |
| Sterilization | | Gamma sterilization | Gamma sterilization | Gamma sterilization |
| Fixture Diameter | | Internal Type:<br>Ø3.0, 3.4mm | Internal Type:<br>Ø3.0, 3.5, 4.2, 4.8, 5.4mm | Internal type<br>4.0 ~ 8.4mm |
| Fixture Height | | Internal Type:<br>8.0, 9.5, 11.0, 12.5<br>14.5mm | Internal Type:<br>6.0 – 17.0mm<br>(6.0mm height excluded for Ø3.0, 4.2, 4.8mm and 17.0mm height excluded for Ø3.0, 5.4mm) | Internal type:<br>7.7 ~ 14.4mm |
| Abutment | Diameters | Ø3.0 ~ 3.5mm | Ø3.0 ~ 5.4mm | Ø4.0 ~ 10.0mm |
| | Materials | Ti-6Al-4V ELI | | Ti-6Al-4V ELI,<br>CP Ti Grade 4 |
| Angulations of Angled<br>abutments | | 15° | 15° ~ 30° | 15°, 25° |
| Product Code | | DZE, NHA | DZE, NHA | DZE, NHA |
| Surface Treatment | | Sand-blasted, Large<br>grit, Acid-etched<br>(S.L.A) | Resorbable Blasting<br>Media (R.B.M)<br>and fluoride coating | Sand-blasted, Large<br>grit, Acid-etched<br>(S.L.A) |
{5}------------------------------------------------
{6}------------------------------------------------
The differences in indications between the MiNi Internal Implant System and primary predicate device (K120414) do not raise new questions of substantial equivalence because they are selected as a subset of the primary predicate. The MiNi Internal Implant System removed the indications of one-stage surgery, immediate loading, and soft bone clinical use compared to the primary predicate (K120414). Thus, the proposed indications do not increase risk nor change the intended use of the device and are found to be substantially equivalent.
## Non-Clinical Test Data
Sterilization validating testing has been performed in accordance with ISO 11137-1, ISO 11137-2,
{7}------------------------------------------------
ISO 11137-3, ISO 11737-1, and ISO 11737-2 for gamma sterilization and ISO 17665-1 and ISO 17665-2 for steam sterilization. Test results have demonstrated that the SAL of 10th was achieved and all testing requirements were met.
The fatigue test was performed on the subject device in accordance with ISO 14801:2007 Dentistry-Implants-Dynamic fatigue test for endosseous dental implants. The worst case scenario was chosen based on the FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
The results of the non-clinical testing demonstrate that the results have met the criteria of the standards, and the subject device is substantially equivalent to the predicate device.
| Testing Item | |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conducted test<br>on the subject device | Fatigue Test<br>Sterility test |
| Not conducted test<br>on the subject device | Biocompatibility test:<br>Cytotoxicity<br>Sensitization<br>Irritation or Intracutaneous reactivity<br>Systemic toxicity<br>Sub-chronic Toxicity<br>Genotoxocity<br>Implantation<br>Surface Analysis<br>Residue Analysis |
Biocompatibility Test, surface analysis, and residue analysis have not been conducted because the subject device has the same material, surface treatment, and manufacturing process as the reference predicate device (K140091). There is no difference between the subject and predicate device. There is no additional testing conducted on the subject device.
#### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Megagen Implant Co., Ltd. concludes that the MiNi Internal Implant system is substantially equivalent to predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.