The IQvitals Zone is intended to be used by clinicians and medically qualified personnel for measuring and monitoring: · Noninvasive blood pressure for adult and pediatric patients (3 years and above) - · Pulse rate for adult and pediatric patients - · Noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2) for adult and pediatric patients - · Body temperature measured at Temporal Artery for adult and pediatric patients The most likely location for IQvitals Zone device to be used is for monitoring patient in general medical locations, hospitals and alternative care environments.
Device Story
IQvitals Zone is a multi-parameter patient monitor for clinical environments; measures noninvasive blood pressure (NIBP), pulse rate, SpO2, and temporal artery temperature. Inputs: oscillometric NIBP cuff, SpO2 sensor (Masimo or Nellcor), and Exergen temporal thermometer probe. Device uses a main controller board to process signals from modules via UART; displays results on integrated touch screen. Connects to computers via USB or Bluetooth Low Energy for data transmission. Operated by clinicians/medically qualified personnel in hospitals or alternative care settings. Output provides real-time vital signs to assist clinicians in patient assessment and monitoring. Benefits include simultaneous measurement and centralized viewing of patient vitals.
Clinical Evidence
Clinical validation of the NIBP module was conducted per ANSI/AAMI/ISO 81060-2:2013. Other parameters supported by bench testing and compliance with standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-30, IEC 60601-2-49, ISO 80601-2-56, and ISO 80601-2-61.
Technological Characteristics
Class II device; 21 CFR 870.2300. Components: Midmark NIBP module, Masimo/Nellcor SpO2 modules, Exergen temporal thermometer. Connectivity: USB 2.0, Bluetooth Low Energy. Power: 100-240 VAC or rechargeable lithium-ion battery. Form factor: Touch screen display, mobile cart or wall mount. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-30, IEC 60601-2-49, ISO 80601-2-56, ISO 80601-2-61.
Indications for Use
Indicated for adult and pediatric patients (3 years and older) in clinical settings for noninvasive blood pressure, pulse rate, SpO2, and temporal artery temperature monitoring.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2016
Midmark Corporation Ms. Maria Moreno Manager, Ouality Management System 690 Knox Street, Suite 100 Torrance, California 90502
Re: K161909 Trade/Device Name: IQvitals Zone: Models: 1-200-0310, 1-200-0320, 1-200-0330, 1-200-0340, 1-200-0350, 1-200-0360 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DOA, FLL Dated: September 30, 2016 Received: October 11, 2016
Dear Ms. Maria Moreno,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mada Yellen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K161909
### Device Name
IQvitals Zone Models: 1-200-0310, 1-200-0320, 1-200-0330, 1-200-0340, 1-200-0350, 1-200-0360
Indications for Use (Describe)
The IQvitals Zone is intended to be used by clinicians and medically qualified personnel for measuring and monitoring:
· Noninvasive blood pressure for adult and pediatric patients (3 years and above)
- · Pulse rate for adult and pediatric patients
- · Noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2) for adult and pediatric patients
- · Body temperature measured at Temporal Artery for adult and pediatric patients
The most likely location for IQvitals Zone device to be used is for monitoring patient in general medical locations, hospitals and alternative care environments.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5
# 510(k) Summary
#### 1. Sponsor
Midmark Corporation 690 Knox Street, Suite 100 Torrance, CA 90502 Phone: (800) 624-8950 Fax: (310) 516-6517 Contact: Mr. Asad Abu-Tarif; Vice President (R&D) Email: aabutarif@midmark.com Alternate Contact: Ms. Maria Moreno; Manager (Quality) Email: mmoreno@midmark.com
Summary Preparation Date: July 7, 2016
#### 2. Device
| Trade Name | IQvitals Zone |
|-------------------|---------------------------------------------------------------------------|
| Models | 1-200-0310, 1-200-0320, 1-200-0330, 1-200-0340,<br>1-200-0350, 1-200-0360 |
| Common Name | Patient Physiological Monitor |
| Classification | Class II |
| Product Code | MWI |
| Regulation Number | 21 CFR 870.2300 |
| Review Panel | Cardiovascular |
#### 3. Predicate Device
- Midmark Diagnostics IQmark Vital Signs Monitor (K072516)
- · Vital Signs Monitor VSM 6000 Series; (K101445)
- · Nellcor OxiMax Pulse Oximetry System with N-600x Pulse Oximeter and OxiMax Sensors and Cables (K060576)
- · Masimo Radical 7 Pulse CO-Oximeter and Accessories (K110028)
- · Exergen TemporalScanner Thermometer (K011291)
- 4. Device Description
{4}------------------------------------------------
The IQvitals Zone is designed to be used for measuring and monitoring systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. All functions of the device are performed via the touch screen display, except the power on/off function, which is a separate button on the back of the device.
The IQvitals Zone can be wirelessly connected with low energy Bluetooth to mobile computers or be connected with a USB cable to computers and has the ability to send the measurement results to the computers.
The device has a rechargeable lithium ion battery and two mounting options: a mobile cart and a wall mount. All vitals parameters can be simultaneously measured and easily viewed on the touch screen display or the connected computer. Temperature is measured at Temporal Artery and the Temperature Probe is connected serially to the IQvitals Zone.
| Device Model<br>Configurations | Features | | | | |
|--------------------------------|--------------------------------------------------|----------------------------------------------|---------------------------|--------------------------|-------------------------|
| | Non-invasive<br>Blood Pressure<br>Monitor Module | Exergen<br>Temporal<br>Thermometer<br>Module | Nellcor<br>SpO₂<br>Module | Masimo<br>SpO₂<br>Module | Alarms<br>&<br>Wireless |
| 1-200-0310 | X | X | --- | --- | --- |
| 1-200-0320 | X | X | X | --- | --- |
| 1-200-0330 | X | X | --- | X | --- |
| 1-200-0340 | X | X | --- | --- | X |
| 1-200-0350 | X | X | X | --- | X |
| 1-200-0360 | X | X | --- | X | X |
The IQvitals Zone is offered in following models:
#### 5. Indications for Use
The IQvitals Zone is intended to be used by clinicians and medically qualified personnel for measuring and monitoring:
- Noninvasive blood pressure for adult and pediatric patients (3 years and above)
- · Pulse rate for adult and pediatric patients
- Noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2) for . adult and pediatric patients
- · Body temperature measured at Temporal Artery for adult and pediatric patients
The most likely location for IQvitals Zone device to be used is for monitoring patient in general medical locations, hospitals and alternative care environments.
{5}------------------------------------------------
#### 6. Technological Characteristics and Substantial Equivalence
The proposed 'IQvitals Zone' is substantially equivalent to the referenced predicate devices in regard to the intended use, fundamental scientific technology, design, and technological characteristics, as outlined in the following table:
| Characteristic | Subject Device | Predicate Devices | | |
|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | IQvitals Zone Wireless | Midmark Diagnostics<br>IQmark Vital Signs<br>Monitor (K072516) | Vital Signs Monitor<br>VSM 6000 Series;<br>(K101445) | |
| Submitter | Midmark Corporation | Midmark Diagnostics<br>Group | Welch Allyn, Inc. | |
| Classification | Device Class: 2<br>Product Code: MWI<br>21 CFR 870.2300 | Device Class: 2<br>Product Code: MWI<br>21 CFR 870.2300 | Device Class: 2<br>Product Code: MWI<br>21 CFR 870.2300 | |
| Indications for<br>Use | The IQVitals Zone is<br>intended to be used by<br>clinicians and medically<br>qualified personnel for<br>measuring and monitoring:<br>• Noninvasive blood<br>pressure for adult and<br>pediatric patients (3<br>years and above)<br>• Pulse rate for adult and<br>pediatric patients<br>• Noninvasive functional<br>oxygen saturation of<br>arteriolar hemoglobin<br>(SpO2) for adult and<br>pediatric patients<br>• Body temperature<br>measured at Temporal<br>Artery for adult and<br>pediatric patients<br>The most likely location for<br>IQvitals Zone device to be<br>used is for monitoring<br>patient in general medical<br>locations, hospitals and<br>alternative care<br>environments. | The Midmark Diagnostics<br>IQmark Vital Signs Monitor<br>is intended to be used by<br>clinicians and medically<br>qualified personnel for<br>monitoring adult, pediatric<br>and neonatal patients for<br>noninvasive blood<br>pressure, pulse rate<br>noninvasive functional<br>oxygen saturation of<br>arteriolar hemoglobin (Sp0<br>2), and body temperature.<br>In adults and pediatric<br>patients, temperature is<br>monitored orally, rectally,<br>or at axillary sites. In<br>neonates (to 1 month)<br>temperature is monitored<br>at axillary sites only.<br>The most likely locations<br>for patients to be<br>monitored are general<br>medical locations,<br>hospitals, and alternative<br>care environments. | The VSM 6000 series of<br>monitors is intended to be<br>used by clinicians and<br>medically qualified<br>personnel for monitoring of<br>noninvasive blood<br>pressure, pulse rate,<br>noninvasive functional<br>oxygen saturation of<br>arteriolar hemoglobin<br>(SpO2), and body<br>temperature in normal and<br>axillary modes of neonatal,<br>pediatric, and adult<br>patients.<br>The most likely locations<br>for patients to be<br>monitored are general<br>medical and surgical<br>floors, general hospital,<br>and alternate care<br>environments. This<br>product is available for<br>sale only upon the order of<br>a physician or licensed<br>health care professional. | |
| Prescription /<br>Over-The-<br>Counter (OTC)<br>Use | Prescription | Prescription | Prescription | |
| Characteristic | Subject Device | Predicate Devices | | |
| | IQvitals Zone Wireless | Midmark Diagnostics<br>IQmark Vital Signs<br>Monitor (K072516) | Vital Signs Monitor<br>VSM 6000 Series;<br>(K101445) | |
| Device Design<br>(components<br>and parts) | The IQvitals Zone includes<br>Midmark NIBP module,<br>Masimo or Nellcor OEM<br>SPO2 module, and<br>Exergen temporal<br>thermometer.<br><br>The Midmark NIBP module<br>uses oscillometric blood<br>pressure estimation<br>algorithm to estimate<br>patient's systolic, diastolic<br>and mean arterial blood<br>pressure. The SPO2<br>module is Masimo Radical<br>7 or Nellcor NELL-1. The<br>Main Board is the center<br>controller for the IQvitals<br>Zone. Internally it<br>communicates with<br>Wireless Board, NIBP<br>Board, SPO2 module,<br>Display Board, with UART<br>(Universal Asynchronous<br>Receiver/Transmitter)<br>interface. It also provides<br>RS232 interfaces for<br>Exergen temporal<br>thermometer and<br>compatible scales like<br>Midmark's IQscale or<br>Fairbanks scale.<br><br>The Alarm Module is | The IQmark Vital Signs<br>Monitor includes CAS Med<br>OEM NIBP module,<br>Masimo OEM SPO2<br>module, and Midmark<br>predictive temperature<br>module.<br><br>The CAS Med NIBP<br>modules uses oscillometric<br>blood pressure estimation<br>algorithm to estimate<br>patient's systolic, diastolic<br>and mean arterial blood<br>pressure. The SPO2<br>module is Masimo's MS-<br>11. The Midmark<br>thermometer is a<br>predictive temperature<br>module. All modules are<br>interfaced to the main<br>board through UART ports.<br><br>The IQmark Vital Signs<br>Monitor can communicate<br>with the laptop/PC through<br>a USB 1.1 port and /or an<br>Inferred wireless port.<br>The device uses numerical<br>LED display and buttons to<br>control the device<br>operation. | The Welch Allyn VSM<br>6000 Series Device<br>incorporates Oscillometric<br>Blood Pressure<br>Measurement (BP)<br>module, thermometer and<br>SPO2 modules<br><br>A power button is located<br>on the side of the device.<br>A LCD with a touch screen<br>is prominent on the front of<br>the device and provides<br>the primary interface for<br>the user to interact with<br>the device. Internal and<br>external communications<br>is primarily by USB. A user<br>accessible client USB<br>connection for data<br>transfer is on the side of<br>the device. | |
| | controlled directly by the<br>IO ports on the Main<br>Board. | | | |
| | The IQvitals Zone can<br>communicate with a<br>laptop/PC through a USB<br>2.0 compliant full speed<br>port and/or a low energy<br>Bluetooth wireless<br>interface. The device has a<br>LCD touch screen that<br>provides graphical user<br>interface to operate the<br>device. | | | |
| Characteristic | Subject Device<br>IQvitals Zone Wireless | Predicate Devices<br>Midmark Diagnostics<br>IQmark Vital Signs<br>Monitor (K072516) | Vital Signs Monitor<br>VSM 6000 Series;<br>(K101445) | |
| Accessories &<br>Peripheral<br>Devices | SpO2 Sensors and cables Temperature Probes Temperature Probe Covers Blood Pressure Hose Blood Pressure Cuffs AC Power Adapter Lithium ion battery USB Cable Mobile Cart for IQvitals Zone IQvitals® Zone™ Wall Mount Articulating Arm 15" Laptop Tray 12" Tablet Tray | SPO2 sensors and cables Temperature Probes Temperature Probe Covers Temperature calibration key Blood Pressure Hose Blood Pressure Cuffs AC Power Adapter Lead acid battery USB cable | SPO2 sensors and cables Temperature Probes Temperature Probe Covers Temperature calibration key Blood Pressure Hose Blood Pressure Cuffs Lithium ion battery | |
| Available<br>Module<br>Configurations | Noninvasive Blood<br>Pressure (NIBP)<br>Pulse Oximeter<br>Temperature | Noninvasive Blood<br>Pressure (NIBP)<br>Pulse Oximeter<br>Temperature | Noninvasive Blood<br>Pressure (NIBP)<br>Pulse Oximeter<br>Temperature | |
| Device<br>Operational<br>Modes | Adult, Pediatric | Adult, Neonatal | Adult, Pediatric, Neonatal | |
| Noninvasive<br>Blood Pressure<br>(NIBP) Monitor<br>Module<br>Specifications | | | | |
| Operating<br>Principle /<br>Measurement Method | Oscillometric | Oscillometric | Oscillometric | |
| NIBP<br>Modes and<br>Range | Adult Mode<br>Systolic: 50-260 mmHg<br>Mean: 30-230 mmHg<br>Diastolic: 20-210 mmHg<br><br>Pediatric Mode<br>Systolic: 50-260 mmHg<br>Mean: 30-230 mmHg<br>Diastolic: 20-210 mmHg | Adult Mode<br>Systolic: 30-255 mmHg<br>Mean: 20-235 mmHg<br>Diastolic: 15-220 mmHg<br><br>Neonate Mode:<br>Systolic: 30-135mmHg<br>Mean: 20-125mmHg<br>Diastolic: 15-110mmHg | Adult Mode<br>Systolic: 30-260 mmHg<br>Mean: 280 mmHg<br>Diastolic: 20-220 mmHg<br><br>Pediatric Mode<br>Systolic: 30-260 mmHg<br>Mean: 280 mmHg<br>Diastolic: 20-220 mmHg<br><br>Neonate Mode<br>Systolic: 20 to 120 mmHg<br>Mean: 130 mmHg<br>Diastolic: 10-110 mmHg | |
| Characteristic | Subject Device | Predicate Devices | Vital Signs Monitor<br>VSM 6000 Series;<br>(K101445) | |
| | IQvitals Zone Wireless | Midmark Diagnostics<br>IQmark Vital Signs<br>Monitor (K072516) | | |
| •<br>Measurement<br>Accuracy | +/- 5 mmHg<br>(standard deviation < 8 mmHg)<br>Meets ANSI.AAMI<br>SP10:2002 standard<br>accuracy range | +/- 5 mmHg<br>(standard deviation < 8 mmHg)<br>Meets ANSI.AAMI<br>SP10:2002 standard<br>accuracy range | ±5 mmHg<br>(standard deviation < 8 mmHg)<br>Meets ANSI.AAMI<br>SP10:2002 standard<br>accuracy range | |
| •<br>Measurement<br>Resolution | 1 mmHg | 1 mmHg | 1 mmHg | |
| •<br>Pulse Rate<br>Range | 40 – 200 BPM | 30 – 240 BPM | 30 – 200 BPM (adult, pediatric)<br>35 – 220 BPM (neonate) | |
| •<br>Pulse Rate<br>Accuracy | ±3 bpm or ±5%, whichever<br>is greater | ± 2 bpm or ± 2%,<br>whichever is greater | ±3 bpm or ±5%, whichever<br>is greater | |
| •<br>Pulse Rate<br>Resolution | 1 bpm | 1 bpm | 1 bpm | |
| •<br>Initial Cuff<br>Pressure | Automatic or User-<br>selectable | User-selectable | Automatic | |
| •<br>Maximum<br>Cuff<br>Pressure | $280 \pm 5$ mmHg | 290 mmHg (adult mode)<br>145 mmHg (neonatal<br>mode) | 280 mmHg (adult, pediatric)<br>130 mmHg (neonate) | |
| •<br>Overpressure<br>Cutoff | $300 \pm 30$ mmHg…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.