K101445 · Welch Allyn, Inc. · MWI · Jun 8, 2010 · Cardiovascular
Device Facts
Record ID
K101445
Device Name
VITAL SIGNS MONITOR-VSM 6000 SERIES
Applicant
Welch Allyn, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Jun 8, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The VSM 6000 series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric, and adult patients. The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. This product is available for sale only upon the order of a physician or licensed health care professional.
Device Story
VSM 6000 series is a modular, scalable vital signs monitor for spot checks and continuous monitoring. Inputs: oscillometric NIBP (cuff), SpO2 (light absorption sensor), and thermistor-based temperature probes. Device processes signals via internal modules to display vitals on a touchscreen LCD. Features include a silicone light bar for alarm conditions and USB connectivity for data transfer and nurse call systems. Operated by clinicians/medically qualified personnel in general medical/surgical floors and alternate care environments. Output provides real-time physiological data to assist clinical decision-making. Benefits include standardized, reliable patient monitoring across diverse clinical settings.
Clinical Evidence
No clinical studies were utilized for the purpose of obtaining safety and/or efficacy data. Bench testing only.
Technological Characteristics
Enclosure: engineering plastics with internal steel members. Sensing: oscillometric NIBP, thermistor temperature probes, light absorption SpO2. Power: AC and internal Lithium Ion battery. Connectivity: USB, nurse call relay. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, IEC 60601-2-30, ISO 9919, IEC 60601-2-49.
Indications for Use
Indicated for neonatal, pediatric, and adult patients requiring monitoring of noninvasive blood pressure, pulse rate, SpO2, and body temperature. Contraindicated for patients on heart/lung machines, patients being transported outside healthcare facilities, use near MRI machines, hyperbaric chambers, flammable anesthetics, or electro-cauterization devices.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
# WelchAllyn
Abbreviated 510(k) Premarket Notification VSM 6000 Series
| 5. 510(k) Summary | |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | February 15, 2010 |
| Submitted By: | Welch Allyn, Inc.<br>4341 State Street Road<br>P.O. Box 220<br>Skaneateles Falls, NY 13153-0220<br>Phone: (315) 685-3694<br>Fax: (315) 685-2532<br>Contact: Huy Doan, Regulatory Affairs Manager |
| Common Name: | Cardiovascular Monitoring Devices |
| Trade Name: | Vital Signs Monitor - VSM 6000 Series |
| Classification: | 21 CFR 870.2300; Monitor, Physiological, Patient (Without Arrhythmia<br>Detection or Alarms) |
| Predicate Device: | The predicate devices for these vital signs monitors are:<br>Welch Allyn VSM- Vital Signs Monitor (VSM 300) 53000 series cleared<br>through FDA 510(k) no. K063419 Welch Allyn Non-invasive Blood Pressure (NIBP) Device cleared<br>through FDA 510(k) no. K093907 Welch Allyn Spot Ultra Vital Signs Device cleared through FDA 510(k)<br>no. K040490 |
| Description: | The Welch Allyn VSM 6000 series is designed to provide a scalable,<br>modular system that could be configured to address the needs for vitals<br>signs spot check and monitoring, by using interchangeable components<br><br>The VSM 6000 series of monitors is intended to be used by clinicians and<br>medically qualified personnel for monitoring of noninvasive blood pressure,<br>pulse rate, noninvasive functional oxygen saturation of arteriolar<br>hemoglobin (SpO2) and body temperature of neonatal, pediatric and adult<br>patients.<br><br>The overall device has an enclosure constructed of engineering plastics with<br>internal steel members for strengthening. A silicone light bar is prominent<br>in a carry handle on the top of the device and illuminates for different alarm<br>conditions. A power button is located on the side of the device. A LCD with<br>a touch screen is prominent on the front of the device and provides the<br>primary interface for the user to interact with the device. Internal and<br>external communications is primarily by USB. External host USB<br>connections for accessories are tool accessible. A user accessible client<br>USB connection for data transfer is on the side of the device. A connection<br>to an internal relay for use with nurse call systems is provided on the side of<br>the device. The device contains an internal AC power supply for operating<br>the device and charging the internal Lithium Ion battery. |
{1}------------------------------------------------
## Welch Allyn
Abbreviated 510(k) Premarket Notification VSM 6000 Series
Noninvasive blood pressure is determined using the oscillometric method and is acquired during cuff inflation or cuff deflation. The blood pressure measurement capability is contained in an enclosed sensor module internal to the device that provides a hose connection out the side of the device. This device implements a previously cleared Welch Allyn NIBP module and cuffs.
Noninvasive functional oxygen saturation of arteriolar hemoglobin is determined through the absorption rate of transmitted or reflected light from a sensor applied to different body sites. This capability is contained in an enclosed sensor module that provides a sensor connection to device. This device implements sensor technology from previously cleared Nellcor and Masimo devices including circuit boards and sensors for different body sites.
Body temperature is determined by a thermister based temperature probes that can be used in oral, axillary, rectal body sites. Measurement is made through a predictive measurement as the probe temperature increases towards equilibrium with the body site and by streaming temperature data from the probe to the display. Contact of the stainless steel probe is separated from the patient by use of disposable plastic probe cover. This capability is contained in a sensor module internal to the device that provides a temperature probe connection out the front of the device. This device implements technology and probes from a previously cleared Welch Allyn device.
Intended Use: The VSM 6000 series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric, and adult patients. The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. This product is available for sale only upon the order of a physician or licensed health care professional.
### Contraindication for Use:
This system is not intended to be used:
- · on patients connected to heart/lung machines
- · on patients being transported outside a healthcare facility
- · near an MRI machine
- · in a hyperbaric chamber
- · near flammable anesthetics
- · near electro-cauterization devices
#### Technological Characteristics:
The Welch Allyn VSM 6000 Series Device utilizes an Oscillometric Blood Pressure Measurement (BP) algorithm, Thermometer and SpO2 algorithm . that is equivalent to the algorithm used in the Welch Allyn NIBP, Spot
{2}------------------------------------------------
## Welch Allyn
Abbreviated 510(k) Premarket Notification VSM 6000 Series 101445
>3/3
Ultra Vital Signs and Welch Allyn VSM 53000 (VSM 300 series). The Welch Allyn VSM 6000 Series Device is powered from AC power and Battery power. It uses the same operating principle and incorporates the same basic material as the previously cleared predicate devices. The subject device has the same technological characteristics as the predicate devices.
### Non-clinical performance data:
The subject device was also tested to evaluate its safety and effectiveness based on the following standards:
- · IEC 60601-1:1988+Al: 1991+A2:1995 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- · IEC 60601-1-2: 2001 + A1:2004 Medical Electrical Equipment Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Test
- · IEC 60601-1-4: 2000 General Requirement for Safety: Collateral Standard: Programmable Electrical Medical Systems
- · IEC 60601-1-8: 2003 General requirements for safety Collateral Standard: Alarm Systems - Requirements, tests and guidances - General requirements and guidelines for alarm systems in medical equipment
- · IEC 60601-2-30: 1999: Manual, electronic or automated sphygmomanometers
- · ISO 9919: 2005: Medical electrical equipment Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use
- · IEC 60601-2-49: 2001 Medical electrical equipment Part 2-49: Particular requirements for the safety of multifunction patient monitoring equipment.
### Clinical performance data:
No clinical studies were utilized for the purpose of obtaining safety and/or efficacy data.
- Conclusions: The intended use and the technological characteristics of the Welch Allyn VSM 6000 device are the same as the predicate devices and the product is as safe and effective as the predicate. Therefore Welch Allyn believes the Welch Allyn VSM 6000 series device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is simple, using only black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN - 8 2010
Welch Allyn, Inc. c/o Mr. Daniel Lehtonen Sr. Staff Engineer - Medical Devices Intertek Testing Services 2307 E Aurora Rd., Unit B7 Twinsburg, OH 44087
Re: K101445
Trade/Device Name: Vital Signs Monitor - VSM 6000 Series Regulatory Number: 21 CFR 870.2300 Regulation Name: Patient Physiological Monitor (without arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: MWI Dated: May 21, 2010 Received: May 24, 2010
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Daniel Lehtonen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W. Atto.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### K101445
### Welch Allyn
Abbreviated 510(k) Premarket Notification VSM 6000 Series
4. Statement of Indications For Use
510(k) Number (if known):
Device Name:
Vital Signs Monitor 6000 Series
Indications For Use:
The VSM 6000 series of monitors is intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric, and adult patients.
The most likely locations for patients to be monitored are general medical and surgical floors, general hospital, and alternate care environments. This product is available for sale only upon the order of a physician or licensed health care professional.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
W.W.A.C.
(Division Sign-Off)
Division of Cardiovascular Devices
**510(k) Number** K101445
Section 4 - Page 1 of 1
MPD FCD-0025, Rev. 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.