The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.
Device Story
Radical 7 Pulse CO-Oximeter utilizes Masimo Rainbow SET technology to monitor arterial oxygen saturation, pulse rate, carboxyhemoglobin, methemoglobin, total hemoglobin, and respiration rate. Device processes optical signals (red/infrared light absorption) and vibratory signals (respiratory sounds) to derive physiological parameters. Used in hospitals, mobile, and home environments by clinicians. Features touchscreen display, wireless radio, and output interfaces (SatShare, RS-232, Nurse Call) for integration with multi-parameter monitors. Provides real-time data to assist clinicians in patient assessment and decision-making; benefits include continuous, non-invasive monitoring of critical blood constituents and respiratory status.
Clinical Evidence
Clinical studies conducted on healthy adult volunteers (SpO2, SpCO, SpMet, SpHb) and 16 neonatal NICU patients (SpO2, SpMet). SpHb validated on surgical patients with varying skin pigmentation. Results demonstrated accuracy (rms) within specified ranges; no device-related adverse events reported. Studies performed per ISO 9919:2005.
Technological Characteristics
Pulse oximetry via red/infrared light absorption; respiratory rate via vibratory signal analysis. Includes wireless radio (802.11 a/b/g) and Bluetooth. Power: 100-240 VAC or rechargeable battery. Output: touchscreen, SatShare, RS-232, wireless. Class II device; no patient-contacting materials requiring biocompatibility testing.
Indications for Use
Indicated for continuous non-invasive monitoring of SpO2, pulse rate, SpCO, SpMet, SpHb, and RRa in adult, pediatric, and neonatal patients. Suitable for well or poorly perfused patients in hospital, mobile, and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET® Radical 7R Pulse CO-Oximeters and Accessories (K100428)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summarv
MAR 1 7 2011
| Submitted by: | Masimo Corporation<br>40 Parker, Irvine, CA 92618<br>Phone: (949) 297-7000; FAX: (949) 297-7592 |
|---------------------------------|-------------------------------------------------------------------------------------------------|
| Contact | Shelly Harris, Manager of Regulatory Affairs |
| Date Summary Prepared | January 3, 2010. |
| Establishment Registration No. | 2031172 |
| Trade Name | Masimo Radical 7 Pulse CO-Oximeter and Accessories |
| Common Name | Pulse Oximeter and Sensor |
| Regulation Number/Name/ Class | 21 CFR 870.2700/ Oximeter/ Class II |
| Product Code | DQA, BZQ, DPZ, JKS |
| Substantially Equivalent Device | Masimo Rainbow SET® Radical 7R Pulse CO-Oximeters and Accessories<br>510(k) Number - K100428 |
#### Description of the Device
The Radical 7 Pulse CO-Oximeter and accessories (Radical 7) include the Masimo Rainbow SET technology. The Radical 7 Pulse CO-Oximeter provides nonitoring of arterial oxygen saturation (SpO3), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiration rate (RRa). Other information displayed by the Radical 7 Pulse CO-Oximeter includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Hematocrit (SpHct), Signal Identification Quality (SIQa), Respiration Indicator (R), alarm status, alarm silence, battery life, sensor status, trends, and pleth waveform. The Radical 7 Pulse CO-Oximeter also has a touchscreen and output interfaces, which include: SatShare connection to multi-parameter monitors, Nurse Call analog output, and RS-232 serial output, and wireless radio.
#### Intended Use/Indications for Use
The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masino Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.
#### Principles of Operation
SpO2 and Pulse Rate: Pulse oximetry is governed by the principles that blood contents differ in their absorption of red and infrared light and blood pulsation.
SpCO, SpMet, and SpHb: The Radical 7 includes the Masimo Rainbow SET technology board, which calculates functional oxygen saturation (SpO₂), fractional concentration of carboxyhemoglobin (SpCO), fractional concentration of methemoglobin (SpMet), total hemoglobin concentration (SpHb) and pulse rate.
Respiratory or Respiration Rate (RRa) General Description: The Masimo Rainbow SET technology also provides respiratory or respiration rate measurements, based on vibratory signals from respiratory sounds.
{1}------------------------------------------------
# 510(k) Summary
#### Comparison to Predicate Device
The Radical 7 in this filing is the same as the predicate in the K100428 filing, with the addition of wireless radio for remote monitoring, a Bluetooth radio module, and update to the user interface controls. Below are the specifications for the Radical 7.
| FEATURES | SPECIFICATIONS |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Range | SpO2: 0-100%; PR: 25-240 bpm; SpCO: 0-99%; SpMet: 0-99.9%; SpHb: 0-25 g/dL;<br>RRa: 0-70 breaths per minute; SpOC: 0-35 ml/dl, SpHct: 0-75%; PI: 0.02-20%; PVI: 0-100% |
| Accuracy: SpO2 | No Motion (adults/pediatrics/infants): 60-80 ± 3%; 70-100 ± 2%, ± 3%(neonates)<br>Motion (adults/pediatrics/infants/neonates): 70-100 ± 3%<br>Low Perfusion (adults/pediatrics/infants/neonates): 70-100 ± 2% |
| Accuracy: PR | No Motion/ Low Perfusion(adults/pediatrics/ infants/neonates): 25-240 ± 3 bpm<br>Motion (adults/pediatrics/infants/neonates): 25-240 ± 5 bpm |
| Accuracy: SpCO | 1-40 + 3%, adults/pediatrics/infants |
| Accuracy: SpMet | 1-15 ± 1%, adults/pediatrics/infants/neonates |
| Accuracy: SpHb | 8-17 +1 g/dl (arterial or venous), adults/pediatrics |
| Accuracy: RRa | 4-70 + 1 breath per minute, adults |
| Resolution | SpO2: 1%; PR: 1 bpm; SpCO: 1%; SpMet: 0.1%; SpHb: 0.1 g/dl; RRa: 1 breath per minute |
| Measurements | Low Signal IQ, PI, SpOC, SpHct, PVI, SIQa, RI |
| Power | AC Input Range: 100-240 VAC, 47-63 Hz; Rechargeable batteries |
| Environmental | Op temp: 32 to 122°F; Storage temp: -40 to 158°F; RH: 10-95% non-condensing |
| SpO2 Averaging | Averaging: 2, 4, 6, 8, 10, 12 and 16 seconds; FastSat |
| Mode/ Sensitivity | Sensitivity: APOD, Normal, Maximum |
| Alarm | Volume (pulse/tone), out of limit (high/low), sensor condition, system failure, low battery |
| Display/<br>Indicators | SpO2, PR, SPCO, SpMet, SpHb, SpHbv, RRa, SpOC, SpHct, PI, PVI, Pleth waveform, SIQ<br>SIQa, RI, sensitivity, sensor status/time/messages, alarm status, battery status |
| Output Interface | Trends, Satshare to Multiparameter monitors (SpO2 only), RS-232 (PC/printer), Vuelink,<br>Spacelabs Flexport, RadNet, PSN, wireless radio |
| Compliance | EMC: EN 60601-1-2, Class B; Electrical Safety: IEC 60601-1/UL 60601-1, Class 1 (AC<br>Power), internally powered, patient cable-Type BF applied part, Satshare cable-type CF<br>applied part; 802.11 a/b/g |
| Operation Mode | Continuous |
#### Clinical Summary
Clinical Studies: SpO2, SpCO and SpMet accuracy was determined by testing healthy adult volunteers in the range of 60-100% SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-Oximeter. SpHb accuracy has been validated on healthy adult volunteers and on surgical patients with light to dark skin pigmentation in the range of 8-17 g/dl SpHb against a laboratory CO-oximeter. The variation in these studies equals one standard deviation which encompasses 68% of the population. SpO₂ and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb with a resultant accuracy of 2 9% SpO2 and 0.9% SpMet.
Clinical Results: No device-related adverse events. The clinical studies were performed in accordance with ISO 9919:2005. The studies resulted in accuracies (rms) of equal to or less than the respective accuracies as stated above in the Radical 7 specifications.
#### Non-Clinical Summary
The Radical 7 complies with the voluntary standards as detailed in this submission. Laboratory testing for rne radical 7 compleo with the voluntary overify that the Radical 7 met all design specifications and was circularitially equivalent to the predicate device. This device was not tested for biocompatibility because like the predicate device, it does not contain any patient contacting materials.
{2}------------------------------------------------
# 510(k) Summary
### Conclusions
·
·
. .
!
The information in this 510(k) submission demonstrates that the Masimo Radical 7 Pulse CO-Oximeter is substantially equivalent to the predicate device, with respect to safety, effectiveness, and performance.
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
Ms. Shelly Harris Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618
MAR 1 7 2011
Re: K110028
Trade/Device Name: Radical 7 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, BZQ, DPZ, JKS Dated: February 14, 2011 Received: February 15, 2011
Dear Ms. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Harris
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Radical 7 Pulse CO-Oximeter and Accessories Device Name:
Indications For Use:
The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous noninvasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpQ₂) and pulse rate to multiparameter devices for the display of those devices.
Prescription Use (Per 21 CFR 801.109 Subpart D) AND/OR
Over-The-Counter Use (Per 21 CFR 801.109 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesional Devices Division of Anostrios of Antson of Antection Controllers
510(k) Number: K1100281
Special 510(k), Radical 7 01/03/2011, Page 4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.