K153263 · Toshibamedical Systems Corporation · JAK · Apr 1, 2016 · Radiology
Device Facts
Record ID
K153263
Device Name
Aquilion Lightning, TSX-035A/4 and /5, V7.0
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Apr 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion Lightning has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Device Story
16-row CT system; acquires cross-sectional X-ray images of whole body/head. Transforms X-ray attenuation data into volumetric image sets via reconstruction algorithms. Operated by trained physicians/technicians in clinical settings. Includes noise reduction (AIDR 3D, QDS) and metal artifact reduction (SEMAR). Output viewed on workstation for diagnostic interpretation; aids clinical decision-making by providing anatomical visualization. Benefits include diagnostic imaging capabilities for various patient conditions.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation testing confirmed performance specifications for modified X-ray generator, rotation speed, patient couch, and software applications.
Technological Characteristics
16-row CT system; 800 channels; 780mm gantry aperture; 0.5-0.75s rotation speed; max 50.4kW generator output. Materials/design per ISO 13485, IEC 60601-1, IEC 60601-2-44, NEMA XR-25/26. Includes QDS, AIDR 3D, and SEMAR processing. Connectivity via NEMA PS 3.1-3.18 (DICOM).
Indications for Use
Indicated for acquisition and display of cross-sectional whole-body volumes, including head, for use by trained physicians in specialized studies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 1, 2016
Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K153263
Trade/Device Name: Aquilion Lightning, TSX-035A/4 and /5, V7.0 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 2, 2016 Received: March 3, 2016
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K153263
Device Name
Aquilion Lightning, TSX-035A/4 and /5, V7.0
#### Indications for Use (Describe)
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.
The Aquilion Lightning has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### TOSHIBA AMERICA MEDICAL SYSTEMS, INC
2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
#### 510(k) SUMMARY
#### 1. SUBMITTER'S NAME:
Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
#### 2. OFFICIAL CORRESPONDENT:
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
# 3. ESTABLISHMENT REGISTRATION:
9614698
#### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
#### 5. Date Prepared:
March 2, 2016
# 6. TRADE NAME(S): Aquilion Lightning, TSX-035A/4 and /5, V7.0
- 7. COMMON NAME: System, X-ray, Computed Tomography
- 8. DEVICE CLASSIFICATION (Regulatory Class, CFR Reference, Name): Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
# 9. PRODUCT CODE / DESCRIPTION:
JAK / Computed Tomography X-Ray System
#### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
{4}------------------------------------------------
### 11. PREDICATE DEVICE:
| Product | Marketed | Regulation | Regulation | Product Code | 510(k) | Clearance |
|--------------------------------------|------------------------------------------|--------------------|----------------------------------------|---------------------------------------------------|---------|--------------------|
| | by | Number | Name | | Number | Date |
| Aquilion<br>Lightning,<br>TSX-035A/2 | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K151833 | October 9,<br>2015 |
### 12. REASON FOR SUBMISSION:
Modification to an existing device.
# 13. DEVICE DESCRIPTION:
The Aquilion Lightning, TSX-035A/4 and /5, v7.0 is a 16-row CT System that is intended to acquire and display cross-sectional volumes of the whole body, including the head. This system is based upon the technology and materials of previously marketed Toshiba CT systems.
# 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.
The Aquilion Lightning has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software, by a trained and qualified physician.
# 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion Lightning, TSX-035A/4 and /5, v7.0, is substantially equivalent to the Aquilion Lightning, TSX-035A/2, that received premarket clearance under K151833 and is marketed by Toshiba America Medical Systems. The Indications for Use for the subject device are identical to those of the predicate device. See below for a brief comparison of the technological characteristics between the subject and the predicate devices:
| Item | Aquilion Lightning, TSX-035A/4 and /5 | Aquilion Lightning, TSX-035A/2 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| 510(k) Number | This submission | K151833 |
| Gantry Rotation<br>Speed | 0.75 sec/rotation<br>Options:<br>• 0.6 sec (Available when 0.6s Fast Scan Kit,<br>CGS-61A, is installed)<br>• 0.5 sec (Available when 0.5s Fast Scan Kit,<br>CGS-67A, is installed) | 0.75 sec/rotation<br>Option:<br>• 0.6 sec (Available when 0.6s<br>Fast Scan Kit, CGS-61A, is<br>installed) |
| Detector | 800 channels × 16 rows | 800 channels × 16 rows |
| FOV (field of view) | 180/240/320/390/500 mm | 180/240/320/390/500 mm |
{5}------------------------------------------------
| X-ray rated output | Max. 36 kW<br>Option:<br>• Max. 50.4kW when X-ray Power-Up Kit,<br>CSGS-016A is installed | | | | Max. 36 kW | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------|------------|------------|--------------------------------|------------|
| X-ray tube voltage<br>X-ray tube current | 80/100/120/135 kV<br>10-300 mA<br>Option:<br>• 10-420 mA when X-ray Power-Up Kit,<br>CSGS-016A is installed | | | | 80/100/120/135 kV<br>10-300 mA | |
| X-ray fan angle | 43.9° | | | | 43.9° | |
| Patient Couch | Long Type | Long Type | Short Type | Short Type | Long Type | Short Type |
| • Couch-top stroke | 2190mm | 2390mm | 1890mm | 1890mm | 2190mm | 1890mm |
| • Scan permissible range<br>(Conventional) | 1830mm | 2030mm | 1530mm | 1530mm | 1830mm | 1530mm |
| • Scan permissible range (Helical) | 1780mm | 1980mm | 1480mm | 1480mm | 1780mm | 1480mm |
| • Max. guaranteed weight | 205kg | 300 kg | 205kg | 300 kg | 205kg | 205kg |
| Noise reduction processing | QDS, AIDR 3D , AIDR3D Enhanced | | | | QDS, AIDR 3D , AIDR3D Enhanced | |
| Metal Artifact Reduction | SEMAR (Helical Scan) | | | | SEMAR (Helical Scan) | |
| Gantry opening size (diameter) | 780 mm | | | | 780 mm | |
| Tilt angle | ±30° | | | | ±30° | |
| Installation area<br>(Short patient couch) | Minimum 17.2m²<br>(14.4 m² ) | | | | Minimum 17.1m²<br>(14.3 m² ) | |
{6}------------------------------------------------
| CT Cardiac Function Analysis Software | Previously cleared under K023760 |
|---------------------------------------|----------------------------------|
| Sure Plaque | Previously cleared under K043111 |
| ECG-Gating Reconstruction System | Previously cleared under K991766 |
| Scan Protocol Management | Previously cleared under K142465 |
| 4D Airways Analysis | Previously cleared under K143294 |
Previously cleared software options being implemented to the modified device:
# 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
# 17. TESTING
This submission includes summary tables detailing the risk analysis and verification/validation testing conducted through bench testing which demonstrates that the requirements for the modifications made to the system have been met. Evaluation of the modified device included, but was not limited to, confirmation that the associated performance specifications related to the increased X-ray generator output, increase of the rotation speed, implementation of a new patient couch and application software previously cleared on other Toshiba CT systems, were met.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the subject device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
# 18. CONCLUSION
The Aquilion Lightning, TSX-035A/4 and /5, v7.0, performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon bench testing, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.