CT CARDIAC FUNCTION ANALYSIS SOFTWARE, MODELS CSCF-001A AND CSCF-001B
K023760 · Toshiba America Medical Systems, In.C · JAK · Nov 22, 2002 · Radiology
Device Facts
Record ID
K023760
Device Name
CT CARDIAC FUNCTION ANALYSIS SOFTWARE, MODELS CSCF-001A AND CSCF-001B
Applicant
Toshiba America Medical Systems, In.C
Product Code
JAK · Radiology
Decision Date
Nov 22, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Cardiac Function Analysis (CFA) is a software package that can be used in conjunction with CT cardiac images to semi-automatically calculate and display various Left Ventricular and Right Ventricular functional parameters such as End Systolic and End Diastolic volumes, stroke volume, LV ejection to include peak filling and ejection rates, Myocardial Mass calculations, regional wall motion display and analysis. When interpreted by a trained physician, the software aids in the assessment and in determination of cardiovascular disease diagnosis and management.
Device Story
Cardiac Function Analysis (CFA) software; post-processes ECG-gated cardiac CT images on CT system display console. Inputs: ECG-gated CT image data. Processing: Multi-planar reconstruction; manual or semi-automatic contour detection of ventricular heartwalls. Outputs: Ventricular wall motion, wall thickness, myocardium mass, ventricle volume, ejection fraction, peak filling/ejection rates; user-generated reports. Used in clinical settings by physicians to aid cardiovascular disease diagnosis and risk management. Benefits: Provides quantitative functional data to support clinical decision-making.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use as a post-processing software package.
Technological Characteristics
Post-processing software for CT scanner; operates on CT display console. Performs multi-planar reconstruction and semi-automatic contour detection. Conforms to IEC-60601 medical device safety standards and 21 CFR 1020.30/1020.33 diagnostic X-ray equipment standards.
Indications for Use
Indicated for use by trained physicians to assess cardiovascular disease and manage patient care using CT cardiac images to calculate and display ventricular functional parameters, including volumes, stroke volume, ejection fraction, myocardial mass, and wall motion.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
NOV 2 2 2002
# 510(k) Summary
| | K023760 |
|---------------------------------------|-----------------------------------------------------------------------------------|
| Date: | October 31, 2002 |
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
| Submitter's Address: | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068 |
| Submitter's Contact: | Paul Biggins, Regulatory Affairs Manager (714) 730-5000 |
| Establishment Registration<br>Number: | 2020563 |
| Device Proprietary Name: | CT Cardiac Function Analysis Software<br>CSCF-001A, CSCF-001B |
| Common Name: | Scanner, Computed Tomography, X-Ray<br>[Fed. Reg. No. 892.1750, Pro. Code: 90JAK] |
| Regulatory Class: | II (per 21 CFR 892.1750) |
| Performance Standard: | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard |
| Predicate Device(s): | General Electric CardIQ Function [K013422] |
| Reason For Submission | new device (software package) |
#### Description of this Device:
The Cardiac Function Analysis software is used to post-process cardiac CT images on the display console of the CT system. Using these post-processed images various measurements can be initiated by the user to obtain clinically relevant information pertaining to the heart. Additionally this information can be used in the detection of cardiovascular disease diagnosis and risk management. Support functions are:
- Generation of Multi-Planar Reconstructed images based upon ECG gated data. ●
- Manual or semi-automatic contour detection of the ventricular heartwalls. ●
- Measure and display ventricular wall motions, wall thickness, myocardium mass, ventricle volume and ejection fraction.
- . Various data outputs can be selected by the user to generate reports of the above measurements.
{1}------------------------------------------------
### Summary of Intended Uses:
The Cardiac Function Analysis (CFA) is a software package that can be used in conjunction with CT cardiac images to semi-automatically calculate and display various Left Ventricular and Right Ventricular functional parameters such as End Systolic and End Diastolic volumes, stroke volume, LV ejection to include peak filling and ejection rates, Myocardial Mass calculations, regional wall motion display and analysis. When interpreted by a trained physician, the software aids in the assessment and in determination of cardiovascular disease diagnosis and management.
#### Technological Characteristics:
This package is similar in uses and applications as those of the predicate devices. The main difference is in the method used to obtain the final results. Both this and the predicate devices are used as post-processing software to images collected by the parent CT Scanner.
#### Safety and Effectiveness Concerns:
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020.30 and 1020.33, that apply to this upgrade, will be met and reported via a supplement to the initial report for the predicate device. Additionally this system is in conformance with the applicable parts of the IEC-60601 -Medical Device Safety standards.
## Substantial Equivalence:
CT Cardiac Function Analysis Software (CFA), Model Numbers CSCF-001A, CSCF-001B is substantially equivalent to the predicate device. This package, and the predicate device, are both post-processing analysis software packages that allow the user to obtain clinically relevant anatomical and functional information from CT images of the cardiac region.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human profile.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 22 2002
Toshiba America Medical Systems, Inc. % Mr. Mark Job 510(K) Program Manager TÜV Product Service 1775 Old Highway 8, Suite 104 NEW BRIGHTON MN 55112-1891 Re: K023760 Trade/Device Name: Model CGCF-001A/CSCF-001B
Cardiac Function Analysis Software Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK
Dated: November 7, 2002 Received: November 8, 2002
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Verny C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: Cardiac Function Analysis: Model Number CSCF-001A/CSCF-001 B
Indications for Use:
The Cardiac Function Analysis (CFA) is a software package that can be used in conjunction with CT cardiac images to semi-automatically calculate and display various Left Ventricular and Right Ventricular functional parameters such as End Systolic and End Diastolic volumes, stroke volume, LV ejection fraction to include peak filling and ejection rates, Myocardial Mass calculations, regional wall motion display and analysis. When interpreted by a trained physician, the software aids in the assessment and in determination of cardiovascular disease diagnosis and management.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109) (Optional Format 1-2-96)
OR
B - 5
Over-The-Counter Use
David G. Lyon
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number_k_(
ﻠﺴﻠﺔ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.