K142465 · Toshibamedical Systems Corporation · JAK · Mar 10, 2015 · Radiology
Device Facts
Record ID
K142465
Device Name
Aquilion ONE Vision, v7.0
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Mar 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Image noise reduction
—
Substantially equivalent to or better than predicate device
Substantially equivalent to or better than predicate device
—
—
Bench testing using phantoms
—
4D Orthopedic Analysis
—
Comparable to manual measurements and/or segmentations
Comparable to manual measurements and/or segmentations
—
—
Clinical evaluations
—
4D Cerebral Artery Morphological Analysis
—
Comparable to manual measurements and/or segmentations
Comparable to manual measurements and/or segmentations
—
—
Clinical evaluations
—
Indications for Use
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software, of the whole organ by a trained and qualified physician.
Device Story
Whole-body CT scanner; captures cross-sectional volume data sets via X-ray. Operates in clinical settings; used by trained physicians. Input: X-ray projections; Output: reconstructed cross-sectional volume images. Features include AIDR 3D Enhanced (noise reduction), Single Energy Metal Artifact Reduction (SEMAR) for helical/volume scans, SurekV (automated kV selection), and SURE Exposure 3D (dose modulation). New clinical applications include 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis. Physician reviews images to perform specialized organ studies; aids in clinical diagnosis and treatment planning. Benefits include improved image quality, reduced noise, and optimized radiation dose.
Clinical Evidence
Clinical evaluations performed for 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis applications. Results confirmed performance comparable to manual measurements and segmentations. Bench testing utilized phantoms to validate AIDR 3D Enhanced, SurekV, and SEMAR features against predicates.
Technological Characteristics
Whole-body CT X-ray system. Features: AIDR 3D Enhanced noise reduction, SURE Exposure 3D, SurekV, SEMAR. Conforms to IEC 60601-1, -1-1, -1-2, -1-3, -1-4, -1-6, -1-9, -2-28, -2-32, -2-44, IEC 60825-1, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26. Radiation safety per 21 CFR 1010/1020.
Indications for Use
Indicated for patients requiring cross-sectional volume imaging of the head and whole organs (e.g., brain, heart, pancreas). Used by trained physicians for specialized diagnostic studies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Aquilion ONE Vision, TSX-301C/1 and 301C/2, v6.00 (K132222)
Aquilion ONE Vision, TSX-301C/3, 301C/4 and 301C/5, v6.00 (K133497)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2015
Toshiba Medical Systems Corporation % Mr. Orlando Tadeo, Jr. Manager, Regulatory Affairs 2441 Michelle Drive TUSTIN CA 92780
Re: K142465
Trade/Device Name: Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 9, 2015 Received: February 11, 2015
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142465
Device Name
Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0
Indications for Use (Describe)
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software, of the whole organ by a trained and qualified physician.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
TOSHIBA AMERICA MEDICAL SYSTEMS, INC.
2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
#### 510(k) SUMMARY
- 1. SUBMITTER'S NAME:
Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
# 2. OFFICIAL CORRESPONDENT:
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
- 3. ESTABLISHMENT REGISTRATION: 9614698
#### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
- 5. Date Prepared:
February 9, 2015
- 6. TRADE NAME(S): Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0
#### 7. COMMON NAME:
System, X-ray, Computed Tomography
#### 8. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
## 9. PRODUCT CODE / DESCRIPTION:
90JAK / Computed Tomography X-ray System
#### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
{4}------------------------------------------------
## 11. PREDICATE DEVICES:
| Product | Marketed by | 510(k) Number | Clearance Date |
|------------------------------------------------------------------|---------------------------------|---------------|-------------------|
| Aquilion ONE Vision, TSX-<br>301C/1 and 301C/2, v6.00 | Toshiba America Medical Systems | K132222 | November 7, 2013 |
| Aquilion ONE Vision, TSX-<br>301C/3, 301C/4 and 301C/5,<br>v6.00 | Toshiba America Medical Systems | K133497 | February 10, 2014 |
## 12. REASON FOR SUBMISSION:
Modification of a cleared device
## 13. DEVICE DESCRIPTION:
The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 are whole body CT scanners that capture cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. These systems are based upon the technology and materials of previously marketed Toshiba CT systems.
#### 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician.
## 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 is substantially equivalent to Aquilion ONE Vision, TSX-301C/1 and 301C/2, v6.00 and Aquilion ONE Vision, TSX-301C/3, 301C/4 and 301C/5, v6.00, which were cleared via Pre-Market Notification 510(k), K132222 and K133497, respectively. The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0, incorporates modifications to the cleared device which include a new AIDR 3D selection (AIDR 3D Enhanced), the introduction of two new clinical applications, the availability of Single Energy Metal Artifact Reduction in Helical scan mode, the addition of SurekV with SURE Exposure 3D and scan protocol management. The method of operation and manufacturing process of the CT system remain unchanged from the cleared device.
| Item | Aquilion ONE Vision,<br>TSX-301C/1, /2, /3,<br>/4, /5, /6, /7, /8, v7.0 | Aquilion ONE<br>Vision<br>TSX-301C/1, /2<br>v6.00 | Aquilion ONE Vision<br>TSX-301C/3, /4, /5,<br>v6.00 | Comments |
|-------------------------------|-------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|-----------------------------------|
| 510(k) Number | This submission | K132222 | K133497 | |
| Noise reduction<br>processing | AIDR 3D<br>AIDR 3D Enhanced | AIDR 3D | AIDR 3D | New selection<br>added (Enhanced) |
| Metal artifact<br>reduction | Volume Scan<br>Helical Scan | Volume Scan | Volume Scan | Helical scan<br>added |
A complete comparison table is included in this submission. See below for a brief summary of changes:
{5}------------------------------------------------
| ltem | Aquilion ONE Vision,<br>TSX-301C/1, /2, /3,<br>/4, /5, /6, /7, /8, v7.0 | Aquilion ONE<br>Vision<br>TSX-301C/1, /2<br>v6.00 | Aquilion ONE Vision<br>TSX-301C/3, /4, /5,<br>v6.00 | Comments |
|-------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|--------------------------------|
| Dose modulation | SURE Exposure 3D<br>Sure kV | SURE Exposure 3D | SURE Exposure 3D | Feature<br>enhancement |
| 4D Orthopedic<br>Analysis | Optional | N/A | N/A | New application |
| 4D Cerebral Artery<br>Morphological<br>Analysis | Optional | N/A | N/A | New application |
| Detector | Up to 10% noise<br>reduction * | N/A | Up to 10% noise<br>reduction * | * Applicable<br>configurations |
## 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-4, IEC60601-1-6, IEC60601-1-9, IEC60601-2-28, IEC60601-2-32, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
## 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the requirements for the modifications made to the system have been met. The modified system was evaluated, utilizing phantoms, to demonstrate that the image quality metric of AIDR 3D Enhanced versus current AIDR 3D is substantially equivalent to or better than the predicate device with regard to spatial resolution, CT number and contrast-to-noise ratio, noise properties and standard deviation of noise. Bench studies were also conducted to confirm that the SurekV feature selects the appropriate kV, to confirm that SEMAR functions similarly, with regard to image quality, when used in Volume and in Helical scan modes and that applicable configurations utilizing the "UREVision detector demonstrate improved noise reduction.
Additionally, clinical evaluations were conducted to demonstrate that the new applications 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis, perform as intended. The results confirmed that the applications were comparable to manual measurements and/or segmentations.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
{6}------------------------------------------------
## 18. CONCLUSION
The modifications incorporated into the Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 do not change the indications for use or the intended use of the device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.